ASSET DATA
Asset Condemnation Approval/Rejection
Sr# | BER | RBER | View RBER Document | RBER Remark | RBER Upload Date | Asset# / Barcode# | District | Manufacturer | Department | Facility Name | Asset | Criticality | Remark | Approve | Reject |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
1 | Pending | Approved | View | verified | 14-06-2024 | 0642031 | Idukki | Respironics | DIALYSIS UNIT | 24X7 PHC ARAKULAM | Nebuliser | NON CRITICAL |
|
||
2 | Pending | Approved | View | Equipment completed 6 year 10 months repairing cost of equipment is 77% of asset value, both criteria for rber met | 14-10-2024 | 1040260 | Malappuram | APC | LABORATORY | 24X7 PHC ATHANIKKAL | UPS | NON CRITICAL |
|
||
3 | Pending | Approved | View | Equipment completed 9 year, End of life letter from oem received | 02-12-2023 | 1340681 | Kannur | Robonik | LABORATORY | 24X7 PHC CHERUTHAZHAM | Biochemistry Analyzer | CRITICAL |
|
||
4 | Pending | Approved | View | Equipment completed 15+ year, repairing cost of equipment is 61% of asset value, both criteria for rber met | 17-07-2024 | 1340709 | Kannur | KELVINATOR | pharmacy | 24X7 PHC CHERUTHAZHAM | Pharmacy Refrigerator | NON CRITICAL |
|
||
5 | Pending | Approved | View | VERIFIED | 31-12-2024 | 0440334 | Alappuzha | Agappe Diagnostics LTD. | LABORATORY | 24X7 PHC KALAVOOR | Biochemistry Analyzer | CRITICAL |
|
||
6 | Pending | Approved | View | Equipment completed 12+ year reapiring cost of equipment is 66%of asset value, both criteria met | 22-11-2023 | 1345310 | Kannur | Lifeplus | DRESSING ROOM | 24X7 PHC KELAKAM | Sterlizer | NON CRITICAL |
|
||
7 | Pending | Approved | View | VERIFIED | 27-11-2024 | 0740204 | Ernakulam | Godrej | LABORATORY | 24X7 PHC KODANAD | Pharmacy Refrigerator | NON CRITICAL |
|
||
8 | Pending | Approved | View | VERIFIED | 31-07-2024 | 0740190 | Ernakulam | REMI ELEKTROTECHNIK LTD. | LABORATORY | 24X7 PHC KODANAD | Centrifuge | NON CRITICAL |
|
||
9 | Pending | Approved | View | Equipment completed 8 year 2 months Oem robonick informed Robonick prietest touch is an end of life declared model, as per tender clause 5.3.14.2 we recommend equipment for condemination | 17-12-2024 | 1342109 | Kannur | Robonik | LABORATORY | 24X7 PHC KOODALI | Biochemistry Analyzer | CRITICAL |
|
||
10 | Pending | Approved | View | Equipment completed 9 year 4 months oem informed end of life for robonick prietest touch semiauto analyser, as per tender clause 5.3.14.2 we recommend equipment for condemination | 23-11-2024 | 1331002 | Kannur | Robonik | LABORATORY | 24X7 PHC KOOTTUMUGHAM | Semi-Auto Biochemistry Analyzer | NON CRITICAL |
|
||
11 | Pending | Approved | View | VERIFIED | 08-08-2024 | 0540869 | Kottayam | LIFE LINE | OP | 24X7 PHC MANIMALA | Nebuliser | NON CRITICAL |
|
||
12 | Pending | Approved | View | Totally Rusted- Verified- Recommended RBER | 22-12-2023 | 0541748 | Kottayam | nill | WARD | 24X7 PHC MARANGATTUPALLY | Sterlizer | NON CRITICAL |
|
||
13 | Pending | Approved | View | Equipment completed 8+ year Oem informed Robonick prietest touch is an obselate model , as per tender clause 5.3.14.2. we recommend equipment for condemination | 12-07-2024 | 1341414 | Kannur | Robonik | LABORATORY | 24X7 PHC MATOOL | Biochemistry Analyzer | CRITICAL |
|
||
14 | Pending | Approved | View | VERIFIED | 24-07-2024 | 0740791 | Ernakulam | Robonik | LABORATORY | 24X7 PHC PANANGAD | Biochemistry Analyzer | CRITICAL |
|
||
15 | Pending | Approved | View | Equipment completed 6+ year , repairing cost of equipment is 83% of asset value, both criteria for rber met | 22-08-2024 | 0942163 | Palakkad | Medicare Products INC | OP | 24X7 PHC PERINGOTTUKURISSI | B P Apparatus | NON CRITICAL |
|
||
16 | Pending | Approved | View | VERIFIED | 29-11-2024 | 0640227 | Idukki | Rotek Instruments | LABORATORY | 24X7 PHC RAJAKUMARY | Centrifuge | NON CRITICAL |
|
||
17 | Pending | Approved | View | VERIFIED | 12-07-2024 | 0540476 | Kottayam | READY MIST | OP | 24X7 PHC THRIKODITHANAM | Nebuliser | NON CRITICAL |
|
||
18 | Pending | Approved | View | Equipment completed 8 year 8 months, eol declared model robonick prietest touch as per tender clause 5.3.14.2 | 06-02-2024 | 1340949 | Kannur | Robonik | LABORATORY | 24X7 PHC UDAYAGIRI | Semi-Auto Biochemistry Analyzer | NON CRITICAL |
|
||
19 | Pending | Approved | View | Equipment outlived 9 year 5 month , oem informed robonick prietest touch semi auto analyser is an obselate model, as per tender clause 5.3.14.2 we recommend equipment for condemination | 12-12-2024 | 1040972 | Malappuram | Robonik | LABORATORY | 24X7 PHC VALANCHERY | Biochemistry Analyzer | CRITICAL |
|
||
20 | Pending | Approved | View | VERIFIED | 06-08-2024 | 0141438 | Thiruvananthapuram | SAMSUNG | OP | 24X7 PHC VAMANAPURAM | Pharmacy Refrigerator | NON CRITICAL |
|
||
21 | Pending | Approved | View | Equipment completed 9 year 2 months with an asset cost of rupees 1353, Quotation not submitted since spares are not available in the market | 18-09-2024 | 1420319 | KASARGODE | NIDEK | WARD | THQH BEDADKA | Nebuliser | NON CRITICAL |
|
||
22 | Pending | Approved | View | Equipment completed 10+ year , repairing cost of equipment is more than 100%,Both criteria for rber met | 26-07-2024 | 0931009 | Palakkad | B.BRAUN | DIALYSIS UNIT | CHC CHALISSERY | Dialysis Machine | CRITICAL |
|
||
23 | Pending | Approved | View | Equipment outlived 23+ year,repairing cost of equipment is more than 100%, Both criteria for RBER Met | 28-06-2024 | 0930933 | Palakkad | nill | CH WARD | CHC CHERPULASSERY | Suction Apparatus | NON CRITICAL |
|
||
24 | Pending | Approved | View | Equipment completed 8+ year, repairing cost of equipment is more than 100%, both criteria for rber met | 02-12-2023 | 1430055 | KASARGODE | MINDRAY | LABORATORY | CHC CHERUVATHUR | Haematology Analyser | CRITICAL |
|
||
25 | Pending | Approved | View | Documents verified- Quotation above 100%- Year met- Recommended for RBER | 17-10-2023 | 0431232 | Alappuzha | PROKAN | LABORATORY | CHC CHUNAKKARA | Haematology Analyser | CRITICAL |
|
||
26 | Pending | Approved | View | VERIFIED | 17-07-2024 | 0530416 | Kottayam | Godrej | OP | CHC EDAYAZHAM | Pharmacy Refrigerator | NON CRITICAL |
|
||
27 | Pending | Approved | View | VERIFIED | 24-07-2024 | 0732859 | Ernakulam | LABOMED | LABORATORY | CHC EZHIKKARA | Laboratory Microscope | NON CRITICAL |
|
||
28 | Pending | Approved | View | VERIFIED | 07-11-2024 | 0331344 | Pathanamthitta | KELVINATOR | OP | CHC EZHUMATTOOR | Pharmacy Refrigerator | NON CRITICAL |
|
||
29 | Pending | Approved | View | Equipment completed 7 year, repairing cost of equipmnent is more than 100%,Both criteria for Rber Met | 06-03-2024 | 0930748 | Palakkad | PROKAN | LABORATORY | CHC Kadampazhipuram | Haematology Analyser | CRITICAL |
|
||
30 | Pending | Approved | View | Equipment completed 11 year 4 months, oem declared end of life for the said equipment as per tender clause 5.3.14.2 we recommend equipment for condemination | 19-11-2024 | 1030234 | Malappuram | Robonik | LABORATORY | CHC KALIKAVU | Biochemistry Analyzer | CRITICAL |
|
||
31 | Pending | Approved | View | VERIFIED | 19-10-2024 | 0130446 | Thiruvananthapuram | Kadavil Electro Mechanical Industries | LABORATORY | CHC KALLARA | Centrifuge | NON CRITICAL |
|
||
32 | Pending | Approved | View | VERIFIED | 06-09-2024 | 0130483 | Thiruvananthapuram | APC | LABORATORY | CHC KANYAKULANGARA | UPS | NON CRITICAL |
|
||
33 | Pending | Approved | View | Equipment completed 14+ year Oem informed Robonick prietest touch in obselate model as per tender clause 5.3.14.2 we recommend equipment for condemination | 17-07-2024 | 1330033 | Kannur | Robonik | LABORATORY | CHC KARIVELLUR | Biochemistry Analyzer | CRITICAL |
|
||
34 | Pending | Approved | View | Equipment completed 17+year, Repairing cost of equipment is more than 1005, both criteria for rber met | 22-08-2024 | 1330016 | Kannur | Changsha Jinde Technology Co,Ltd | OP | CHC KARIVELLUR | X Ray Film Viewer | NON CRITICAL |
|
||
35 | Pending | Approved | View | verified | 29-01-2024 | 0530343 | Kottayam | NISCO | OP | CHC KOODALLOOR | B P Apparatus | NON CRITICAL |
|
||
36 | Pending | Approved | View | Equipment completed 7 year 7 month with an asset cost of rupees 850, quotation not submitted since spares are not available in the market | 24-04-2024 | 0930044 | Palakkad | Rossmax | OP | CHC KOZHINJAMPARA | Weighing Machine | NON CRITICAL |
|
||
37 | Pending | Approved | View | EQUIPMENT COMPLETED 10 YEAR 5 MONTH EOL DECLARED MODEL AS PER TENDER CLAUSE 5.3.14.2 RECOMMEND EQUIPMENT FOR CONDEMINATION | 19-02-2024 | 1430390 | KASARGODE | Robonik | LABORATORY | CHC KUMBLA | Semi-Auto Biochemistry Analyzer | NON CRITICAL |
|
||
38 | Pending | Approved | View | Equipment completed 8+ year , repairing cost of equipment is 92% of asset value, both criteria for rber met | 09-08-2024 | 1430392 | KASARGODE | PROKAN | LABORATORY | CHC KUMBLA | Haematology Analyser | CRITICAL |
|
||
39 | Pending | Approved | View | VERIFIED | 02-11-2024 | 0430084 | Alappuzha | LIFELINE | CASUALTY | CHC KURATHIKAD | B P Apparatus | NON CRITICAL |
|
||
40 | Pending | Approved | View | Equipment completed 20+ year, repairing cost of equipment is 68% of asset value, both criteria for rber met as per tender clause 5.3.14.1 we recommend equipment for condemination | 19-09-2024 | 0831556 | Thrissur | LABOMED | LABORATORY | CHC MADAVANA | Laboratory Microscope | NON CRITICAL |
|
||
41 | Pending | Approved | View | Equipment completed 12 year 5 months, repairing cost of equipment is 72% of assetvalue, both criteria for rber met as per tender clause 5.3.14.1 we recommend equipment for condemination | 07-01-2024 | 1420352 | KASARGODE | MINDRAY | LABORATORY | THQH MANGALPADY | Haematology Analyser | CRITICAL |
|
||
42 | Pending | Approved | View | Equipment completed 7 year 9 months repairing cost of equipment is 79% of asset value, both criteria for rber met | 20-11-2024 | 1420353 | KASARGODE | PROKAN | LABORATORY | THQH MANGALPADY | Haematology Analyser | CRITICAL |
|
||
43 | Pending | Approved | View | Equipment completed 8 years, Robonick prietest touch is an obselate model as per tender clause 5.3.14.2 we recommend equipment for condemination | 29-04-2024 | 1430346 | KASARGODE | Robonik | LABORATORY | CHC MANJESWAR | Semi-Auto Biochemistry Analyzer | NON CRITICAL |
|
||
44 | Pending | Approved | View | Equipment completed 14+ year, repairing cost of equipment is 69% of asset value,both criteria for RBER Met | 17-07-2024 | 1031316 | Malappuram | SYSTRONICS | LABORATORY | CHC MANKADA | Colorimeter | NON CRITICAL |
|
||
45 | Pending | Approved | View | Equipment installed on 11-10-2013 and aged upto 11+ years, authorised supplier sree gokulam healthcare infomed thet machiene is very old and spare parts are not available so we recommend equipment for condemination | 05-11-2024 | 1330261 | Kannur | EDAN | OP | CHC MATTANNUR | ECG Machine | CRITICAL |
|
||
46 | Pending | Approved | View | Equipment outlived 12 year, repairing cost of equipment is more than 100% with respect to asset value, Both criteria for RBR met | 04-11-2023 | 1330311 | Kannur | REMI ELEKTROTECHNIK LTD. | LABORATORY | CHC MATTANNUR | Centrifuge | NON CRITICAL |
|
||
47 | Pending | Approved | View | Equipment completed 8 year, repairing cost of equipment is 92% of asset value, both criteria for RBER met | 15-05-2024 | 1331064 | Kannur | Agappe Diagnostics LTD. | LABORATORY | CHC MAYYIL | Biochemistry Analyzer | CRITICAL |
|
||
48 | Pending | Approved | View | Equipment completed 7 year, repairing cost is more than 100% of asset value, both criteria for rber met | 23-05-2024 | 1331125 | Kannur | Reico | OP | CHC MAYYIL | Sterlizer | NON CRITICAL |
|
||
49 | Pending | Approved | View | Equipment completed 12 year, repairing cost of equipment is 67% of asset value, both criteria for rber met | 26-11-2024 | 1031643 | Malappuram | CONTEC MEDICAL SYSTEMS CO.LTD. | OP | CHC MELATTUR | ECG Machine | CRITICAL |
|
||
50 | Pending | Approved | View | Equipment outlived 11+ year, Oem beston industries informed model discontinued and spares are not available, as per tender clause 5.3.14.2 we recommend equipment for condemination | 19-06-2024 | 0832395 | Thrissur | Beston Industries | LABORATORY | CHC MULLASSERY | Hot Air Oven | NON CRITICAL |
|
||
51 | Pending | Approved | View | VERIFIED | 06-08-2024 | 0630657 | Idukki | LABOMED | LABORATORY | CHC MUTTOM | Laboratory Microscope | NON CRITICAL |
|
||
52 | Pending | Approved | View | Equipment completed 11+ year, oem confirmed that model is obselate and spares are not available | 10-07-2021 | 1231138 | Wayanad | TRIVITRON | ART RADIO THERAPY | CHC NALLOORNAD | Infusion Pump | CRITICAL |
|
||
53 | Pending | Approved | View | Equipment completed 5 +year, quotation not submitted since spares are not available in the market | 30-01-2024 | 1030698 | Malappuram | Comfo Care | nursing station | CHC NEDUVA | Nebuliser | NON CRITICAL |
|
||
54 | Pending | Approved | View | Equipment completed 11+ year repairing cost of equipmnt is 66% of asset value,both criteria for rber met | 31-07-2024 | 1030749 | Malappuram | REMI ELEKTROTECHNIK LTD. | LABORATORY | CHC NEDUVA | Centrifuge | NON CRITICAL |
|
||
55 | Pending | Approved | View | Equipment completed11+ year with an asset cost of rupees 4900, Repairing cost of equipment is 87% of asset value, both criteria for rber met | 24-08-2024 | 1030761 | Malappuram | Labtronics | LABORATORY | CHC NEDUVA | Colorimeter | NON CRITICAL |
|
||
56 | Pending | Approved | View | Equipment completd 10 year 1 month , repairing cost of equipment is 79% of asset value, both criteria for rber met | 20-12-2024 | 0930280 | Palakkad | MINDRAY | LABORATORY | CHC NEMMARA | Haematology Analyser | CRITICAL |
|
||
57 | Pending | Approved | View | Equipment outlived 10+year repairing cost of equipment is more than 100%,Both criteria for RBER Met | 20-05-2024 | 1130018 | Kozhikode | BPL | OP | CHC OLAVANNA | ECG Machine | CRITICAL |
|
||
58 | Pending | Approved | View | Equipment completed 12+ year, repairing cost of equipment is 93% of asset value, both criteria for rber met | 29-01-2024 | 1130777 | Kozhikode | REMI ELEKTROTECHNIK LTD. | LABORATORY | CHC Orkkatteri | Centrifuge | NON CRITICAL |
|
||
59 | Pending | Approved | View | Equipment completed 10 year 8 months , repairing cost of equipment is more than 100% of asset value, both criteria for rber met | 28-09-2024 | 1420464 | KASARGODE | MEDICA | LABORATORY | THQH PANATHADY | Electrolyte Analyser | CRITICAL |
|
||
60 | Pending | Approved | View | Equipment completed 6+ year with an asset cost of rupees 1500,Quotation not submitted since spares are not available in the market | 23-08-2024 | 1420549 | KASARGODE | Infi | WARD | THQH PANATHADY | B P Apparatus | NON CRITICAL |
|
||
61 | Pending | Approved | View | Equipment completed 11 year, Authorised dealer informed that Edan se 1200 ecg machiene is very old and spare parts are not available, so we recommend equipment for condemination | 30-11-2024 | 1330853 | Kannur | EDAN | OP | CHC PAPINISERY | ECG Machine | CRITICAL |
|
||
62 | Pending | Approved | View | Equipment completed 14+ year,Authorised dealer informed that Ecg machiene EDAN SE1200 is an obselate model, as per tender clause 5.3.14.2 we recommend equipment for condemination06 | 06-04-2024 | 0831807 | Thrissur | EDAN | OP | CHC PERINJANAM | ECG Machine | CRITICAL |
|
||
63 | Pending | Approved | View | VERIFIED | 26-10-2024 | 0630361 | Idukki | MISPA | LABORATORY | CHC PURAPUZHA | Urine Analyzer | NON CRITICAL |
|
||
64 | Pending | Approved | View | VERIFIED | 30-07-2024 | 0530897 | Kottayam | Robonik | LABORATORY | CHC SACHIVOTHAMAPURAM | Biochemistry Analyzer | CRITICAL |
|
||
65 | Pending | Approved | View | VERIFIED | 19-07-2024 | 0530856 | Kottayam | EZ Life | OP | CHC SACHIVOTHAMAPURAM | Nebuliser | NON CRITICAL |
|
||
66 | Pending | Approved | View | Equipment completed 9+ year with an asset cost of rupees 1353,Quotation not submitted since spares are not available in the market | 09-08-2024 | 1130328 | Kozhikode | Comfo Care | nursing station | CHC Thiruvangoor | Nebuliser | NON CRITICAL |
|
||
67 | Pending | Approved | View | VERIFIED | 15-05-2024 | 0530201 | Kottayam | NA | OP | CHC THOTTACKAD | Suction Apparatus | NON CRITICAL |
|
||
68 | Pending | Approved | View | VERIFIED | 25-11-2024 | 0530231 | Kottayam | Robonik | LABORATORY | CHC THOTTACKAD | Biochemistry Analyzer | CRITICAL |
|
||
69 | Pending | Approved | View | Equipment completed 13+ year, Oem declared that equipment has served its best life till date and on examination found equipment is in non repairable condition | 15-06-2024 | 1130999 | Kozhikode | REICO | OP | CHC Ulliyeri | Sterlizer | NON CRITICAL |
|
||
70 | Pending | Approved | View | VERIFIED | 10-07-2024 | 0130338 | Thiruvananthapuram | REMI ELEKTROTECHNIK LTD. | LABORATORY | CHC VAKKOM | Centrifuge | NON CRITICAL |
|
||
71 | Pending | Approved | View | Equipment completed 11+ year, OEM informed the model is obselate and spare parts are not available, as per tender clause 5.3.14.2 we recommend equipment for condemination | 25-06-2024 | 1130857 | Kozhikode | PHOTO ELECTRIC INSTRU PVT.LTD | LABORATORY | CHC VALAYAM | Colorimeter | NON CRITICAL |
|
||
72 | Pending | Approved | View | VERIFIED | 10-06-2024 | 0231075 | Kollam | PROKAN | LABORATORY | CHC VELINALLOOR | Haematology Analyser | CRITICAL |
|
||
73 | Pending | Approved | View | VERIFIED | 12-11-2024 | 0430470 | Alappuzha | NA | immunization | CHC VELIYANADU | Sterlizer | NON CRITICAL |
|
||
74 | Pending | Approved | View | VERIFIED | 27-06-2024 | 0711371 | Ernakulam | HERCULES | DIALYSIS UNIT | DH ALUVA | Weighing Machine | NON CRITICAL |
|
||
75 | Pending | Approved | View | VERIFIED | 29-11-2024 | 0712037 | Ernakulam | Labline Instruments | BLOOD BANK | DH ALUVA | Blood Bank Refrigerator | NON CRITICAL |
|
||
76 | Pending | Approved | View | Vendor not Existing- images and Documents verified | 16-12-2023 | 0610106 | Idukki | JOVE | BLOOD BANK | DH IDUKKI | Blood Collection Monitor | NON CRITICAL |
|
||
77 | Pending | Approved | View | Images and Documents Verified- Year met- Below 10K asset- Recommended for RBEr | 19-12-2023 | 0610631 | Idukki | ANITA INDUSTRIES | ISOLATION WARD | DH IDUKKI | B P Apparatus | NON CRITICAL |
|
||
78 | Pending | Approved | View | VERIFIED | 23-09-2024 | 0610567 | Idukki | EDAN | LABOUR ROOM | DH IDUKKI | NST Machine (CTG) | NON CRITICAL |
|
||
79 | Pending | Approved | View | Equipment outlived 10+ year , As per Oem model is obselate and spares are not available, as per tender clause 5.3.14.2 we recommend equipment for condmination | 11-03-2024 | 1410212 | KASARGODE | HICARE | BLOOD BANK | DH KANHANGAD | Blood Collection Monitor | NON CRITICAL |
|
||
80 | Pending | Approved | View | Equipment completed 14+ year , repairing cost of equipment is 84% of asset value, both criteria for rber met | 23-08-2024 | 1410348 | KASARGODE | Air Liquide Medical Systems PVT.LTD. | ICU | DH KANHANGAD | Ventilator | CRITICAL |
|
||
81 | Pending | Approved | View | Equipment completed 13+ year, repairing cost of equipment is more than 100%, both criteria for rber met | 23-08-2024 | 1410398 | KASARGODE | MINDRAY | MM WARD | DH KANHANGAD | Multipara Monitor | CRITICAL |
|
||
82 | Pending | Approved | View | Equipment outlived 12 years , Oem states that HeartstartXl is an obselate model As per tender clause 5.3.14.2 we recommend equipment for condemination | 27-04-2024 | 1410471 | KASARGODE | PHILIPS | OT | DH KANHANGAD | Defibrillator | NON CRITICAL |
|
||
83 | Pending | Approved | View | Equipment completed 12 year,repairing cost of equipment is 84%,both criteria for rber met | 09-02-2024 | 1410437 | KASARGODE | Allengers | OT | DH KANHANGAD | C-ARM | NON CRITICAL |
|
||
84 | Pending | Approved | View | Equipment completed 8 year 11 month, Oem GE declared GE-lOGIQ PC Is obselate model and end of life letter issued, as per tender clause 5.3.14.2 we declared equipment for condemination | 16-03-2024 | 1311085 | Kannur | GE Elpro Medical Systems | CATH LAB | DH KANNUR | Ultrasound Machine | CRITICAL |
|
||
85 | Pending | Approved | View | Equipment completed 13+ year , repairing cost of equipment is more than 100%, both criteria for rber met | 11-12-2024 | 1310082 | Kannur | REMI ELEKTROTECHNIK LTD. | BLOOD BANK | DH KANNUR | Centrifuge | NON CRITICAL |
|
||
86 | Pending | Approved | View | Equipment completed 10+ year, repairing cost of equipment is more than 100% both criteria for rber met | 22-08-2024 | 1310046 | Kannur | MAQUET | DIALYSIS UNIT | DH KANNUR | Ventilator | CRITICAL |
|
||
87 | Pending | Approved | View | Equipment outlived 14 years, Equipment is an obselate model and end of letter provided by oem,As per tender clause 5.3.14.2 we recommend equipment for condemination | 15-05-2024 | 1310612 | Kannur | stryker instrument | ot2 | DH KANNUR | Powered Instrument Drive | NON CRITICAL |
|
||
88 | Pending | Approved | View | EQUIPMENT COMPLETED 17+ YEAR, REPAIRING COST OF EQUIPMENT IS 65% OF ASSET VALUE, BOTH CRITERIA FOR RBER MET | 05-09-2024 | 1310714 | Kannur | BPL | Ot 1 | DH KANNUR | Defibrillator | NON CRITICAL |
|
||
89 | Pending | Approved | View | Equipment outlived 22+ year, oem informed that equipment is in non repairable condition, as has served its best till date,end of life declared,so we recommentd to condem equipement as per tender clause 5.3.14.2 | 23-08-2024 | 1310351 | Kannur | Reico | CSSD | DH KANNUR | Sterlizer | NON CRITICAL |
|
||
90 | Pending | Approved | View | Equipment completed 12 year, repairing cost of equipment is 60% of asset value, both criteria met for RBER | 06-11-2023 | 1310517 | Kannur | KELVINATOR | DENTAL DEPARTMENT | DH KANNUR | Pharmacy Refrigerator | NON CRITICAL |
|
||
91 | Pending | Approved | View | checked equipment found mainboard,powersupply board and keypad unit defective.Equipment completed 8 year 6 months , oem informed that this machiene already passed more than 8 years,so instead of purchasing mainboard of unit, requesting to purchase new cautery machiene,becuase the cost of mainboard is almost equal to the price of new cautery machiene, after the end of working life of cautery machiene leads to deterioration of the internal electronic spare parts which may create difficulty during the surgical procedure , so as per tender clause5.3.14.2 we recommending equipment for condemination | 10-12-2024 | 1310920 | Kannur | XcelLance Medical Technologies PVT.LTD. | OT | DH KANNUR | Cautery | NON CRITICAL |
|
||
92 | Pending | Approved | View | VERIFIED | 06-05-2024 | 0210027 | Kollam | Allengers | X RAY | DH KOLLAM | X Ray Machine | CRITICAL |
|
||
93 | Pending | Approved | View | VERIFIED | 13-11-2024 | 0210114 | Kollam | LG | MICU | DH KOLLAM | Pharmacy Refrigerator | NON CRITICAL |
|
||
94 | Pending | Approved | View | VERIFIED | 26-06-2024 | 0210279 | Kollam | Biorad | BLOOD BANK | DH KOLLAM | Elisa Washer | NON CRITICAL |
|
||
95 | Pending | Approved | View | VERIFIED EOL | 18-05-2024 | 0311211 | Pathanamthitta | Carl Zeiss | EYE OP | DH KOZHENCHERRY | Automatic Field Analyzer | NON CRITICAL |
|
||
96 | Pending | Approved | View | Equipment completed 12 year 8 months, oem informed philips heartstart xl defibrilator is an obselate model as per tender clause 5.3.14.2 we recommend equipment for condemination | 18-10-2024 | 1011406 | Malappuram | PHILIPS | MICU | DH NILAMBUR | Defibrillator | NON CRITICAL |
|
||
97 | Pending | Approved | View | Equipment outlived 9+ year, repairing cost of equipment is more than 100%. both criteria met for RBER. | 03-11-2023 | 1011708 | Malappuram | ADONIS MEDICAL SYSTEM PVT.LTD. | RADIOLOGY | DH NILAMBUR | X Ray Machine | CRITICAL |
|
||
98 | Pending | Approved | View | Equipment completed 12 year 5 months repairing cost of equipment is more than 100% of asset value, both criteria for rber met | 30-11-2024 | 0911202 | Palakkad | Erba | LABORATORY | DH PALAKKAD | Fully Automated Random Access Clinical Chemistry Analyzer | NON CRITICAL |
|
||
99 | Pending | Approved | View | Equioment completed 13 years, oem declared end of life for robonick preitest touch semi auto analyser | 30-11-2024 | 0911201 | Palakkad | Robonik | LABORATORY | DH PALAKKAD | Biochemistry Analyzer | CRITICAL |
|
||
100 | Pending | Approved | View | Equipment outlived 14+ year oem recommended equipment consumed its shelf life and the repair of same is not cost effective, so we recommend equipment for condemination as per tender clause 5.3.14.2 | 09-07-2024 | 0911197 | Palakkad | SYSMEX | LABORATORY | DH PALAKKAD | Haematology Analyser | CRITICAL |
|
||
101 | Pending | Approved | View | Equipment completed 17+ year,The actual book value of equipment is 9600 rupees, were asset value is 26494 rupees,The repairing cost of equipment is more than 100% with respect to book value | 04-07-2024 | 0911233 | Palakkad | ELECTROLUX | LABORATORY | DH PALAKKAD | Pharmacy Refrigerator | NON CRITICAL |
|
||
102 | Pending | Approved | View | Equipment completed 17+year,Repairing cost of equipment is 65% of asset value, both criteria for Rber met | 04-07-2024 | 0911254 | Palakkad | Godrej | LABORATORY | DH PALAKKAD | Pharmacy Refrigerator | NON CRITICAL |
|
||
103 | Pending | Approved | View | Equipment completed 29+ year, Repairing cost of equipment is more than 69%, Both criteria for rber met | 09-08-2024 | 0911085 | Palakkad | Godrej | LABORATORY | DH PALAKKAD | Pharmacy Refrigerator | NON CRITICAL |
|
||
104 | Pending | Approved | View | Equipment outlived 11+years,repairing cost of equipment is more than 100%,bothcriteria for Rber met | 20-04-2024 | 0911160 | Palakkad | B.BRAUN | DIALYSIS UNIT | DH PALAKKAD | Dialysis Machine | CRITICAL |
|
||
105 | Pending | Approved | View | Equipment completed 10+ year, oem informed some of the spare parts are not available currently due to aging of equipment, so oem schiller advised to upgrade or condemn the unit | 16-11-2024 | 0911365 | Palakkad | SCHILLER | CATH ICU | DH PALAKKAD | Syringe Pump | CRITICAL |
|
||
106 | Pending | Approved | View | Equipment completed 6+ year with an asset cost of rupees 1353, Quotation not submitted since spares are not available in the market | 19-08-2024 | 0913391 | Palakkad | Safe An B Touch | PSYCHIATRY WARD | DH PALAKKAD | Nebuliser | NON CRITICAL |
|
||
107 | Pending | Approved | View | Equipment completed 20+ year, repairing cost of equipment is 72%,Both criteria for RBER Met | 10-06-2024 | 0911094 | Palakkad | Whirlpool | ICTC | DH PALAKKAD | Pharmacy Refrigerator | NON CRITICAL |
|
||
108 | Pending | Approved | View | Equipment completed 18+ year repairing cost of equipment is 64% of asset value, both criteria for rber met | 10-07-2024 | 0911090 | Palakkad | ELECTROLUX | ICTC LAB | DH PALAKKAD | Pharmacy Refrigerator | NON CRITICAL |
|
||
109 | Pending | Approved | View | Equipment completed 9+ year, Authorised dealer informed that Multiparamonitor philips,Gold way UT6000 is an obselate model, as per tender clause 5.3.14.2 we recommend equipment for condemination | 26-07-2024 | 0911339 | Palakkad | Shenzhen Goldway Industrial Inc. | CATH ICU | DH PALAKKAD | Multipara Monitor | CRITICAL |
|
||
110 | Pending | Approved | View | Equipment completed 11 years 7 months repairing cost of equipment is 66% of asset value, both criteria for rber met | 26-12-2024 | 1010034 | Malappuram | REMI ELEKTROTECHNIK LTD. | LABORATORY | DH PERINTHALMANNA | Centrifuge | NON CRITICAL |
|
||
111 | Pending | Approved | View | Equipment outlived 19+ year,Orginal equipment manufacture provided End of life letter and recommend for decomission of equipment considering age of the equipment, increasing maintanence cost, and new technology standards. | 20-11-2023 | 1010315 | Malappuram | Terumo penpol | BLOOD BANK | DH PERINTHALMANNA | Blood Collection Monitor | NON CRITICAL |
|
||
112 | Pending | Approved | View | VERIFIED | 17-08-2024 | 0611167 | Idukki | PHILIPS | OT | DH THODUPUZHA | Multipara Monitor | CRITICAL |
|
||
113 | Pending | Approved | View | VERIFIED | 15-02-2024 | 0610763 | Idukki | LIFE LINE | DIALYSIS UNIT | DH THODUPUZHA | B P Apparatus | NON CRITICAL |
|
||
114 | Pending | Approved | View | VERIFIED | 24-05-2024 | 0610796 | Idukki | Whirlpool | LABORATORY | DH THODUPUZHA | Pharmacy Refrigerator | NON CRITICAL |
|
||
115 | Pending | Approved | View | VERIFIED | 10-12-2024 | 0610800 | Idukki | Biorad | LABORATORY | DH THODUPUZHA | Elisa Reader | NON CRITICAL |
|
||
116 | Pending | Approved | View | Equipment completed 14 year 8 months repairing cost of equipment is more than 100% , both criteria for rber met , as per tender clause 5.3.14.1 we recommend equipment for condemintion | 24-10-2024 | 1010540 | Malappuram | Allengers | RADIOLOGY | DH TIRUR | X Ray Machine | CRITICAL |
|
||
117 | Pending | Approved | View | Equipment completed 10+ year oem philips declared end of life for Heartstart XL Defibrilators, as per tender clause 5.3.14.2 we recommend equipment for condemination | 12-12-2024 | 1010654 | Malappuram | PHILIPS | DIALYSIS UNIT | DH TIRUR | Defibrillator | NON CRITICAL |
|
||
118 | Pending | Approved | View | Equipment completed 8 year,repairing cost of equipment is 80% of asset value,both critera met for rber | 28-01-2024 | 1010870 | Malappuram | HAMILTON | ICU | DH TIRUR | Ventilator | CRITICAL |
|
||
119 | Pending | Approved | View | Equipment completed 12+ year, repairing cost of equipment is more than 100% both criteria for rber met | 19-04-2024 | 1113054 | Kozhikode | Rotek Instruments | LABORATORY | DH VADAKARA | Centrifuge | NON CRITICAL |
|
||
120 | Pending | Approved | View | Equipment outlived 10+ year, repairing cost of equipment is 87%,Both criteria for RBER met | 23-04-2024 | 1113082 | Kozhikode | FRESENIOUS MEDICAL CARE | DIALYSIS UNIT | GH KOZHIKODE | Dialysis Machine | CRITICAL |
|
||
121 | Pending | Approved | View | Equipment completed 8 year 11 months supplier informed getidy -jy23 is an obselate model,so we recommend equipment for condmeination as per tender clause 5.3.14.2 | 02-09-2024 | 0812270 | Thrissur | LY PLANS SYSTEMS | DENTAL DEPARTMENT | DH WADAKKANCHERRY | Sterlizer | NON CRITICAL |
|
||
122 | Pending | Approved | View | Equipment outlived 16+ year, Oem Delta medical coperation states that Delcatt 250B Model is very old and they have stop manufacturing , so as per tender clause 5.3.14.2 we recommend equipment for condemination | 09-01-2024 | 0812315 | Thrissur | DELTA | OT | DH WADAKKANCHERRY | Cautery | NON CRITICAL |
|
||
123 | Pending | Approved | View | Equipment completed 12 year 5 month inour effort to source spare parts for replacement with orginal manufacture ascon healthcare it has come to our attention that ascon healthcare has ceased its operation and permanently stopped its operations and no manufacturing exisitng in adherance to tender clause 5.3.14.2 that manufacture not exisiting we recommend equipment for condemination | 02-09-2024 | 0812317 | Thrissur | ASCON | OT | DH WADAKKANCHERRY | Cautery | NON CRITICAL |
|
||
124 | Pending | Approved | View | Equipment completed 5 year 7 months with an asset cost of rupees 1353, Quotation not submitted since spares are not available in the market | 18-09-2024 | 0812139 | Thrissur | EASY BREATH | CASUALTY | DH WADAKKANCHERRY | Nebuliser | NON CRITICAL |
|
||
125 | Pending | Approved | View | VERIFIED | 28-11-2024 | 0113714 | Thiruvananthapuram | LABOMED | LABORATORY | DISTRICT MODEL HOSPITAL PEROORKADA | Laboratory Microscope | NON CRITICAL |
|
||
126 | Pending | Approved | View | Equipment completed 12 year,End of life declared model as per tenderclause5.3.14.2 criteria for rber met | 26-03-2024 | 1012135 | Malappuram | Robonik | LABORATORY | DISTRICT TB CENTRE MANJERI | Biochemistry Analyzer | CRITICAL |
|
||
127 | Pending | Approved | View | RESOLVED | 25-07-2024 | 0111569 | Thiruvananthapuram | Robonik | LABORATORY | DISTRICT TB CENTRE THIRUVANANTHAPURAM | Semi-Auto Biochemistry Analyzer | NON CRITICAL |
|
||
128 | Pending | Approved | View | VERIFIED | 25-07-2024 | 0111571 | Thiruvananthapuram | Kadavil Electro Mechanical Industries | LABORATORY | DISTRICT TB CENTRE THIRUVANANTHAPURAM | Centrifuge | NON CRITICAL |
|
||
129 | Pending | Approved | View | VERIFIED | 21-04-2024 | 0111556 | Thiruvananthapuram | CELFROST | LABORATORY | DISTRICT TB CENTRE THIRUVANANTHAPURAM | Deep Freezer | NON CRITICAL |
|
||
130 | Pending | Approved | View | VERIFIED | 18-04-2024 | 0111591 | Thiruvananthapuram | LG | IRL | State TB Centre | Pharmacy Refrigerator | NON CRITICAL |
|
||
131 | Pending | Approved | View | VERIFIED | 11-11-2024 | 0312111 | Pathanamthitta | Nihon kohden | OT | GH ADOOR | Defibrillator | NON CRITICAL |
|
||
132 | Pending | Approved | View | VERIFIED | 07-09-2024 | 0410041 | Alappuzha | BPL | ECG ROOM | GH ALAPPUZHA | ECG Machine | CRITICAL |
|
||
133 | Pending | Approved | View | VERIFIED | 19-08-2024 | 0410342 | Alappuzha | NATIONAL INDUSTRIAL CORP. | PMR | GH ALAPPUZHA | Infrared Lamp | NON CRITICAL |
|
||
134 | Pending | Approved | View | VERIFIED | 19-08-2024 | 0410358 | Alappuzha | NATIONAL INDUSTRIAL CORP. | PMR | GH ALAPPUZHA | Infrared Lamp | NON CRITICAL |
|
||
135 | Pending | Approved | View | VERIFIED | 19-08-2024 | 0410361 | Alappuzha | NATIONAL INDUSTRIAL CORP. | PMR | GH ALAPPUZHA | Infrared Lamp | NON CRITICAL |
|
||
136 | Pending | Approved | View | VERIFIED | 09-03-2024 | 0410532 | Alappuzha | Appasamy Associates PVT.LTD. | EYE OT | GH ALAPPUZHA | Phaco Emulsification Unit | NON CRITICAL |
|
||
137 | Pending | Approved | View | VERIFIED | 09-08-2024 | 0410525 | Alappuzha | Appasamy Associates PVT.LTD. | EYE OT | GH ALAPPUZHA | Vitrectomy set | NON CRITICAL |
|
||
138 | Pending | Approved | View | VERIFIED | 13-06-2024 | 0410087 | Alappuzha | LG | CASUALTY | GH ALAPPUZHA | Pharmacy Refrigerator | NON CRITICAL |
|
||
139 | Pending | Approved | View | VERIFIED | 16-04-2024 | 0710116 | Ernakulam | B.BRAUN | DIALYSIS UNIT | GH ERNAKULAM | Dialysis Machine | CRITICAL |
|
||
140 | Pending | Approved | View | Equipment outlived 10+ year repairing cost of equipment is more than 100%,Both criteria for RBER met | 27-04-2024 | 1410898 | KASARGODE | B.BRAUN | DIALYSIS UNIT | GH KASARAGODE | Dialysis Machine | CRITICAL |
|
||
141 | Pending | Approved | View | Equipment outlived 13+year,the spare & components necessary to put the instruments back in working condition are no longer avaiable with oem Remi,model discontinued and letter from oem attached as per tender clause 5.3.14.2 | 22-01-2024 | 1410848 | KASARGODE | REMI ELEKTROTECHNIK LTD. | BLOOD BANK | GH KASARAGODE | Blood Bank Refrigerator | NON CRITICAL |
|
||
142 | Pending | Approved | View | Equipment completed 13 years, reparing cost of equipment is 82% of asset value, both criteria met for rber | 12-02-2024 | 1410966 | KASARGODE | Air Liquide Medical Systems PVT.LTD. | ICU | GH KASARAGODE | Ventilator | CRITICAL |
|
||
143 | Pending | Approved | View | Equipment outlived 13+year, Oem philips india ltd has declared Heartstart XL defibrilator as an end of support device, so as pewr tender clause 5.3.14.2 | 27-10-2023 | 1410927 | KASARGODE | PHILIPS | ICU | GH KASARAGODE | Defibrillator | NON CRITICAL |
|
||
144 | Pending | Approved | View | Equipmemt installed on 15-12-2017, outlived 6 year, repairing cost of equipment is 93% of asset value, both criteria for RBER met | 18-04-2024 | 1411052 | KASARGODE | Seefar | NICU | GH KASARAGODE | Phototherapy | CRITICAL |
|
||
145 | Pending | Approved | View | VERIFIED | 04-07-2024 | 0510706 | Kottayam | AKAS MEDICAL | EYE OT | GH KOTTAYAM | Pulse Oximeter | NON CRITICAL |
|
||
146 | Pending | Approved | View | VERIFIED | 27-10-2024 | 0510200 | Kottayam | PHilips Medical systems | OT | GH KOTTAYAM | Multipara Monitor | CRITICAL |
|
||
147 | Pending | Approved | View | Spare not available with Vendor- Cyrix Explanation letter attached- Verified- Recommended RBER | 12-12-2023 | 0510202 | Kottayam | ALAN ELECTRONIC SYSTEM PVT.LTD. | OT | GH KOTTAYAM | Cautery | NON CRITICAL |
|
||
148 | Pending | Approved | View | VERIFIED | 28-11-2024 | 0510392 | Kottayam | BPL | CASUALTY | GH KOTTAYAM | Defibrillator | NON CRITICAL |
|
||
149 | Pending | Approved | View | VERIFIED | 18-03-2024 | 0510869 | Kottayam | MICO SURICAL EQUIPMENTS PVT LTD | WARD 6 | GH KOTTAYAM | B P Apparatus | NON CRITICAL |
|
||
150 | Pending | Approved | View | Verified- Recommended RBER- Machine very Old | 15-02-2024 | 0510712 | Kottayam | videocon | EYE OT | GH KOTTAYAM | Pharmacy Refrigerator | NON CRITICAL |
|
||
151 | Pending | Approved | View | Quotation and service report attached- Recommended RBER | 28-10-2023 | 0112852 | Thiruvananthapuram | B.BRAUN | DIALYSIS UNIT | GH NEYYATTINKARA | Dialysis Machine | CRITICAL |
|
||
152 | Pending | Approved | View | VERIFIED | 04-04-2024 | 0115441 | Thiruvananthapuram | SONY | OT | GH NEYYATTINKARA | Monitor (LCD TV) | NON CRITICAL |
|
||
153 | Pending | Approved | View | VERIFIED | 09-05-2024 | 0511672 | Kottayam | Tokyo BOEKI medical system | LABORATORY | GH PALA | Biochemistry Analyzer | CRITICAL |
|
||
154 | Pending | Approved | View | VERIFIED | 29-07-2024 | 0511767 | Kottayam | Local purchase | LABOUR ROOM | GH PALA | OT Light | CRITICAL |
|
||
155 | Pending | Approved | View | VERIFIED | 06-07-2024 | 0511205 | Kottayam | HOSPITECH | OT | GH PALA | OT Light | CRITICAL |
|
||
156 | Pending | Approved | View | VERIFIED | 26-07-2024 | 0511204 | Kottayam | HOSPITECH | OT | GH PALA | OT Light | CRITICAL |
|
||
157 | Pending | Approved | View | Mercury Spillage- Verified- Recommended RBER | 22-12-2023 | 0511295 | Kottayam | ANITA INDUSTRIES | CASUALTY | GH PALA | B P Apparatus | NON CRITICAL |
|
||
158 | Pending | Approved | View | VERIFIED | 06-12-2024 | 0511673 | Kottayam | PROKAN | LABORATORY | GH PALA | Haematology Analyser | CRITICAL |
|
||
159 | Pending | Approved | View | VERIFIED | 04-01-2025 | 0310167 | Pathanamthitta | REMI ELEKTROTECHNIK LTD. | BLOOD BANK | GH PATHANMTHITTA | Blood Bank Refrigerator | NON CRITICAL |
|
||
160 | Pending | Approved | View | VERIFIED | 21-08-2024 | 0310131 | Pathanamthitta | B.BRAUN | DIALYSIS UNIT | GH PATHANMTHITTA | Dialysis Machine | CRITICAL |
|
||
161 | Pending | Approved | View | VERIFIED | 21-08-2024 | 0310132 | Pathanamthitta | B.BRAUN | DIALYSIS UNIT | GH PATHANMTHITTA | Dialysis Machine | CRITICAL |
|
||
162 | Pending | Approved | View | VERIFIED | 06-11-2024 | 0310459 | Pathanamthitta | Getidy | EYE OT | GH PATHANMTHITTA | Sterlizer | NON CRITICAL |
|
||
163 | Pending | Approved | View | VERIFIED | 03-12-2024 | 0310002 | Pathanamthitta | videocon | CASUALTY | GH PATHANMTHITTA | Pharmacy Refrigerator | NON CRITICAL |
|
||
164 | Pending | Approved | View | Equipment outlived 10 year 11 months,repairing cost of equipment is 75% of asset value,both critera for rber met | 02-02-2024 | 1311179 | Kannur | Biorad | BLOOD BANK | GH THALASSERY | Elisa Washer | NON CRITICAL |
|
||
165 | Pending | Approved | View | Equipment completed 8+ year, repairing cost of equipment is 89% of asset value, both criteria for rber met | 04-09-2024 | 1311416 | Kannur | Eltek | LABORATORY | GH THALASSERY | Centrifuge | NON CRITICAL |
|
||
166 | Pending | Approved | View | Equipment completed 25+ year, repairing cost of equipment is 66% of asset value, both criteria for rber met | 11-09-2024 | 1311422 | Kannur | LABLINE INSTRUMENTS | LABORATORY | GH THALASSERY | Incubator | NON CRITICAL |
|
||
167 | Pending | Approved | View | Equipment completed 11+ year, Authorised dealer informed that philips UT 6000 series is an obselate model, as per tender clause 5.3.14.2 we recommend equipment for condemination | 27-06-2024 | 1312001 | Kannur | PHILIPS | Labour OT | GH THALASSERY | Multipara Monitor | CRITICAL |
|
||
168 | Pending | Approved | View | Equipment completed 11+ year repairing cost of equipment is 76% of asset value, both criteria for rber met | 11-09-2024 | 1311317 | Kannur | BMS | Physiotherapy | GH THALASSERY | Digital Stimulator (IFT+MST+TENS) | NON CRITICAL |
|
||
169 | Pending | Approved | View | Equipment completed 8 year, Automatic feild analyser Zeiss-Humphery 7401 is an obselate model;As per tender clause 5.3.14.2 | 28-03-2024 | 1311473 | Kannur | ZEISS | Ophthalmology Op | GH THALASSERY | Automatic Field Analyzer | NON CRITICAL |
|
||
170 | Pending | Approved | View | Equipment completed 8+ year, Oem confirmed the international shelf life of the unit is 7+ year, after the end of working life of cautery machiene leads to deteriotation of the internal electronic spare parts which may create dififcult during surgical procedure, | 17-08-2024 | 1311847 | Kannur | XcelLance Medical Technologies PVT.LTD. | OT | GH THALASSERY | Cautery | NON CRITICAL |
|
||
171 | Pending | Approved | View | Rber | 15-03-2023 | 1311433 | Kannur | videocon | LABORATORY | GH THALASSERY | Pharmacy Refrigerator | NON CRITICAL |
|
||
172 | Pending | Approved | View | Equipment completed 25+ year Reparing cost of equipment is 68% of asset value both criteria for rber met | 26-05-2024 | 1311161 | Kannur | BPL | BLOOD BANK | GH THALASSERY | Pharmacy Refrigerator | NON CRITICAL |
|
||
173 | Pending | Approved | View | Equipment completed 14+ year,repairing cost of equipment is more than 100%,Both criteria for rber met as per tender clause 5.3.14.1 we recommend equipment for condemination | 24-08-2024 | 1311221 | Kannur | BPL | Sbicu | GH THALASSERY | Multipara Monitor | CRITICAL |
|
||
174 | Pending | Approved | View | Equipment completed 6 year 11 months, the ventilator was donated to the hospital and also siemens -sv300 model ventilator is an obselate model and source of necessary spare parts are not available, as per tender clause 5.3.14.2 we recommend equipment for condemination | 16-10-2024 | 1311204 | Kannur | SIEMENS | Sbicu | GH THALASSERY | Ventilator | CRITICAL |
|
||
175 | Pending | Approved | View | Equipment covered 5 years, repairing cost is 75% , with respect to asset value, both criteria for RBER Met | 12-10-2023 | 1311736 | Kannur | SUPREME ENTERPRISES | FM WARD | GH THALASSERY | Suction Apparatus | NON CRITICAL |
|
||
176 | Pending | Approved | View | Equipment completed 7 year 7 months repairing cost of equipment is 85% of asset value, both criteria for rber met as per tender clause 5.3.14.2 we recommend equipment for condemination | 24-09-2024 | 1311107 | Kannur | NIPRO | DIALYSIS UNIT | GH THALASSERY | Dialysis Machine | CRITICAL |
|
||
177 | Pending | Approved | View | VERIFIED | 12-06-2024 | 0111861 | Thiruvananthapuram | PROMIS | OT | GH THIRUVANANTHAPURAM | Laparoscopic Insufflator | NON CRITICAL |
|
||
178 | Pending | Approved | View | VERIFIED | 04-04-2024 | 0111959 | Thiruvananthapuram | ASCON | OT | GH THIRUVANANTHAPURAM | Operating Microscope | NON CRITICAL |
|
||
179 | Pending | Approved | View | VERIFIED | 04-04-2024 | 0112023 | Thiruvananthapuram | Appasamy Associates PVT.LTD. | OT | GH THIRUVANANTHAPURAM | Phaco Emulsification Unit | NON CRITICAL |
|
||
180 | Pending | Approved | View | Verified- OEM Letter attached- Recommended RBER | 29-12-2023 | 0111870 | Thiruvananthapuram | KARL STORZ | OT | GH THIRUVANANTHAPURAM | Laser lithotripter with flexible ureteroscope | NON CRITICAL |
|
||
181 | Pending | Approved | View | Documents verified- Both criteria met- Recommended RBER | 28-11-2023 | 0111053 | Thiruvananthapuram | GE Elpro Medical Systems | COVID ICU | GH THIRUVANANTHAPURAM | Multipara Monitor | CRITICAL |
|
||
182 | Pending | Approved | View | VERIFIED | 25-07-2024 | 0110667 | Thiruvananthapuram | TRISCO | OP | GH THIRUVANANTHAPURAM | Dental Chair | NON CRITICAL |
|
||
183 | Pending | Approved | View | VERIFIED | 28-11-2024 | 0110223 | Thiruvananthapuram | TRUCHECK | LABORATORY | GH THIRUVANANTHAPURAM | Nebuliser | NON CRITICAL |
|
||
184 | Pending | Approved | View | VERIFIED | 14-02-2024 | 0110234 | Thiruvananthapuram | L&T | MALE WARD | GH THIRUVANANTHAPURAM | Pulse Oximeter | NON CRITICAL |
|
||
185 | Pending | Approved | View | VERIFIED | 23-08-2024 | 0110413 | Thiruvananthapuram | Agappe Diagnostics LTD. | LABORATORY | GH THIRUVANANTHAPURAM | Specific Protein Analyser | NON CRITICAL |
|
||
186 | Pending | Approved | View | Equipment completed 12+ year , repairing cost of equipment is more than 100%, both criteria for rber met | 16-12-2023 | 0810849 | Thrissur | Erba | LABORATORY | GH THRISSUR | Fully Automated Random Access Clinical Chemistry Analyzer | NON CRITICAL |
|
||
187 | Pending | Approved | View | Equipment completed 10+ year ,Oem declared that the phaco emulsification unit, appasamy galaxy pro is outdated and model obselate, as per tender clause 5.3.14.2 we recommend equipment for condemination | 25-06-2024 | 0810727 | Thrissur | Appasamy Associates PVT.LTD. | OT | GH THRISSUR | Phaco Emulsification Unit | NON CRITICAL |
|
||
188 | Pending | Approved | View | Upon thorough inspection of the Boyle’s Apparatus (Make: Hospitech, Model: Orracle SS501, Serial Number: HT10B1365), installed on 28/04/2010 and aged over 14 years, we found critical damages, including a non-functional mixer control unit, damaged N2 and O2 valves, deteriorated PU tube connections, faulty diaphragm valve set, a malfunctioning pressure regulator, and broken wheels. With the original equipment manufacturer, Hospitech Mumbai, no longer operational and spare parts unavailable, it is impractical to repair the apparatus. As per Tender Clause 5.3.14.2, we recommend the equipment for condemnation due to its age, obsolete technology, safety concerns, non-compliance with current medical standards, and the high cost of repair relative to replacement. Therefore, we request the tender inviting authority to consider its condemnation and appropriate disposal | 21-10-2024 | 0810273 | Thrissur | HOSPITECH | FP OT | GH THRISSUR | Anesthasia Machine | CRITICAL |
|
||
189 | Pending | Approved | View | Equipment aged upto 12+ year,Repairing cost of equipment is more than 100%,Both criteria for RBER met | 17-04-2024 | 0810776 | Thrissur | Anand | OT | GH THRISSUR | Suction Apparatus | NON CRITICAL |
|
||
190 | Pending | Approved | View | Equipment completed 19 year 2 months, the equipment was obsolete model as per oem so as per tender clause 5.3.14.2 we recommend equipment for condmeination | 16-07-2024 | 0810725 | Thrissur | VERSAMED | OT | GH THRISSUR | Ventilator | CRITICAL |
|
||
191 | Pending | Approved | View | Equipment outlived 10+ year,repairing cost of equipment is more than 100%,Both criteria for RBER Met | 02-02-2024 | 0810468 | Thrissur | B.BRAUN | DIALYSIS UNIT | GH THRISSUR | Dialysis Machine | CRITICAL |
|
||
192 | Pending | Approved | View | Equipment completed 10+ year, repairing cost of equipement is more than 100%,Both criteria for rber met | 09-07-2024 | 0810471 | Thrissur | B.BRAUN | DIALYSIS UNIT | GH THRISSUR | Dialysis Machine | CRITICAL |
|
||
193 | Pending | Approved | View | Equipment completed 11 year 1 months, repairing cost of equipment is more than 100%, both criteria for rber met | 04-11-2024 | 0810467 | Thrissur | B.BRAUN | DIALYSIS UNIT | GH THRISSUR | Dialysis Machine | CRITICAL |
|
||
194 | Pending | Approved | View | Equipment completed 10+year repairing cost of equipment is more than 100%,Both criteria for rber met | 09-07-2024 | 0810427 | Thrissur | B.BRAUN | Remove | GH THRISSUR | Dialysis Machine | CRITICAL |
|
||
195 | Pending | Approved | View | Equipment installed on 08-04-2013 and aged upto 10 year, Aspen jsb,1200 is an End of life declared model, so we recommend equipment for condemination as per tender clause 5.3.14..2 | 04-03-2024 | 0810484 | Thrissur | ASPEN | MICU | GH THRISSUR | Infusion Pump | CRITICAL |
|
||
196 | Pending | Approved | View | Equipment completed 11+ year, Oem declared aspen jsb1200 is an end of life declared model so as per tender clause 5.3.14.2 we recommend equipment for condemination | 26-04-2024 | 0810405 | Thrissur | ASPEN | MICU | GH THRISSUR | Infusion Pump | CRITICAL |
|
||
197 | Pending | Approved | View | Equipment installed on 11-03-2013 and aged upto 10 years. Aspen jsb 1200 is an obselate model infusion pump, as per tender clause 5.3.14.2 we recommend equipment for condemination | 04-03-2024 | 0810493 | Thrissur | ASPEN | MICU | GH THRISSUR | Infusion Pump | CRITICAL |
|
||
198 | Pending | Approved | View | Equipment installed on 12-03-2013 and aged upto 10 year,Aspen jsb1200 is an obselate model, as per tender clause 5.3.14.2 we recommend equipment for condemination | 04-03-2024 | 0810486 | Thrissur | ASPEN | MICU | GH THRISSUR | Infusion Pump | CRITICAL |
|
||
199 | Pending | Approved | View | Equipment completed 11 year 6 months repairing cost of equipement is 98% of asset value, both criteria for rber met | 30-09-2024 | 0810906 | Thrissur | Sonomed | EYE OP | GH THRISSUR | Ocular Ultrasound | NON CRITICAL |
|
||
200 | Pending | Approved | View | Equipment completed 9 year, repairing cost of equipment is more than 100% of asset value, both criteria for rber met | 04-10-2024 | 0810914 | Thrissur | Alps | AUDIOLOGY | GH THRISSUR | Audiometer | NON CRITICAL |
|
||
201 | Pending | Approved | View | Equipment outlived 16+ year, repairing cost of equipment is 74% with respect to asset value, Both criteria met for RBER as per tender clause 5.3.14.2 | 13-12-2023 | 1111809 | Kozhikode | Mc Lord's | WARD 4 | GMHC KOZHIKODE | Suction Apparatus | NON CRITICAL |
|
||
202 | Pending | Approved | View | VERIFIED | 01-07-2024 | 0114490 | Thiruvananthapuram | SYSMEX | LABORATORY | GMHC PEROORKADA THIRVANANTHAPURAM | Haematology Analyser | CRITICAL |
|
||
203 | Pending | Approved | View | VERIFIED | 22-10-2024 | 0143072 | Thiruvananthapuram | Robonik | LABORATORY | PHC ANAKUDY | Semi-Auto Biochemistry Analyzer | NON CRITICAL |
|
||
204 | Pending | Approved | View | VERIFIED | 30-03-2024 | 0341153 | Pathanamthitta | CONTEC MEDICAL SYSTEMS CO.LTD. | PP UNIT | PHC ANICADU | ECG Machine | CRITICAL |
|
||
205 | Pending | Approved | View | VERIFIED | 21-10-2024 | 0541652 | Kottayam | Local purchase | OP | PHC ATHIRAMPUZHA | OT Light | CRITICAL |
|
||
206 | Pending | Approved | View | VERIFIED | 10-11-2024 | 0141614 | Thiruvananthapuram | Robonik | LABORATORY | PHC AZHOOR | Mitek | NON CRITICAL |
|
||
207 | Pending | Approved | View | Equipment completed 10 years,End of life declared model as per tenderclause 5.3.14.2 both criteria for rber met | 25-03-2024 | 1345236 | Kannur | Robonik | LABORATORY | PHC CHITTARIPARAMBA | Semi-Auto Biochemistry Analyzer | NON CRITICAL |
|
||
208 | Pending | Approved | View | Equipment completed 9+ year, repairing cost of equipment is 71% of asset value, both criteria for rber met | 05-09-2024 | 1345246 | Kannur | Rotek Instruments | LABORATORY | PHC CHITTARIPARAMBA | Centrifuge | NON CRITICAL |
|
||
209 | Pending | Approved | View | VERIFIED | 29-11-2024 | 0641210 | Idukki | Robonik | LABORATORY | PHC DEVIYAR COLONY | Semi-Auto Biochemistry Analyzer | NON CRITICAL |
|
||
210 | Pending | Approved | View | Equipment completed 10 years, oem informed robonick prietest touch is an end of life declared model, as per tender clause 5.3.14.2 we recommend equipment for condemination | 25-09-2024 | 1344494 | Kannur | Robonik | LABORATORY | PHC EDAKKAD | Biochemistry Analyzer | CRITICAL |
|
||
211 | Pending | Approved | View | VERIFIED | 19-07-2024 | 0141767 | Thiruvananthapuram | Godrej | OP | PHC EDAVA | Pharmacy Refrigerator | NON CRITICAL |
|
||
212 | Pending | Approved | View | VERIFIED | 17-05-2024 | 0340270 | Pathanamthitta | Robonik | LABORATORY | PHC EZHAMKULAM | Semi-Auto Biochemistry Analyzer | NON CRITICAL |
|
||
213 | Pending | Approved | View | Equipment completed 8 years, Oem declared that Robonick prietest touch is an obselate model as per tender clause 5.3.14.2 we recommend equipment for condemination | 27-02-2024 | 1342587 | Kannur | Robonik | LABORATORY | PHC EZHOME | Biochemistry Analyzer | CRITICAL |
|
||
214 | Pending | Approved | View | Equipment completred 18 year 8 months, repairing cost of equipment is 61% of asset value, both criteria for rber met as per tender clause 5.3.14.1 we recommend equipment for condemination | 23-09-2024 | 1344359 | Kannur | Godrej | pharmacy | PHC KANICHAR | Pharmacy Refrigerator | NON CRITICAL |
|
||
215 | Pending | Approved | View | Equipment outlived 10+ year, Repairing cost of equipment is more than 100%, Both criteria for RBER Met | 15-11-2023 | 1342623 | Kannur | HEMANT SURGICAL INDUSTRIES LTD. | OP | PHC KANNAPURAM | Nebuliser | NON CRITICAL |
|
||
216 | Pending | Approved | View | Equipment completed 7+ year, repairing cost of equipment is more than 100%,Both criteria for RBER Met | 25-04-2024 | 1042894 | Malappuram | APC | LABORATORY | PHC KEEZHUPARAMBA | UPS | NON CRITICAL |
|
||
217 | Pending | Approved | View | Equipment outlived 6+ year with an asset cost of 850 rupees, Quotation not submitted since spares are not avialable in the market | 28-06-2024 | 0943556 | Palakkad | AIMS | OP | PHC KIZHAKKENCHERY | Weighing Machine | NON CRITICAL |
|
||
218 | Pending | Approved | View | Equipment completed 8 year 6 months, oem declared end of life for said machiene, as per tender clause 5.3.14..2 we recommend equipment for condemination | 30-09-2024 | 1342439 | Kannur | Robonik | LABORATORY | PHC KOLACHERY | Biochemistry Analyzer | CRITICAL |
|
||
219 | Pending | Approved | View | VERIFIED | 01-10-2024 | 0144926 | Thiruvananthapuram | Local purchase | LABORATORY | PHC KULATHOOR | Colorimeter | NON CRITICAL |
|
||
220 | Pending | Approved | View | Equipment completed 14+ year with an asset cost of rupees 1353.Quotation not submitted since spares are not available in the market | 06-08-2024 | 0940985 | Palakkad | Infi-neb | OP | PHC MARUTHARODE | Nebuliser | NON CRITICAL |
|
||
221 | Pending | Approved | View | Equipment completed 7 year 8 months, repairing cost of equipment is 83% of asset value, both criteria for rber met, as per tender clause 5.3.14.1 we recommend equipment for condemination | 26-12-2024 | 0940168 | Palakkad | ANITA INDUSTRIES | OP | PHC MARUTHARODE | B P Apparatus | NON CRITICAL |
|
||
222 | Pending | Approved | View | Equipment completed 6 year 11 month , in ticket master asset value is 49491 rupees,as per stockbook value the equipment cost only 7670 rupees, considering the orginal book value, repairing cost of equipment is 78% of asset value, both criteria for rber met | 18-11-2024 | 1440420 | KASARGODE | Sreeraj surgical | OP | PHC MOUCODE | Sterlizer | NON CRITICAL |
|
||
223 | Pending | Approved | View | Equipment completed 8+ year, repairing cost of equipment is more than 100% both criteria for RBER Met | 03-04-2024 | 1440108 | KASARGODE | BIO PLUS | OP | PHC NARKILAKKAD | B P Apparatus | NON CRITICAL |
|
||
224 | Pending | Approved | View | Equipment completed 10+ year oem declared end of life for said machiene robonick prietest touch semi auto analyser, as per tender clause 5.3.14.2 we recommend equipment for condemination | 30-09-2024 | 0842348 | Thrissur | Robonik | LABORATORY | PHC NATTIKA | Biochemistry Analyzer | CRITICAL |
|
||
225 | Pending | Approved | View | VERIFIED | 15-06-2024 | 0542132 | Kottayam | ANITA INDUSTRIES | SUB CENTRE | PHC ONAMTHURUTH | B P Apparatus | NON CRITICAL |
|
||
226 | Pending | Approved | View | Equipment completed 9 year repairing cost of equipment is more than 100% of asset value, both criteria for rber met | 25-09-2024 | 0941046 | Palakkad | LABOMED | LABORATORY | PHC OZHALAPATHY | Laboratory Microscope | NON CRITICAL |
|
||
227 | Pending | Approved | View | Equipment completed 5+ year with an asset cost of rupees 3800, Quotation not submitted since spares are not available in the market | 12-04-2024 | 0941078 | Palakkad | OMRON | OP | PHC OZHALAPATHY | Weighing Machine | NON CRITICAL |
|
||
228 | Pending | Approved | View | EQUIPMENT COMPLETED 14+ YEAR ,HOSPITAL HAS DECIDED TO CONDEM THE EQUIPMENT, CONDEMINATION REPORT FROM HOSPITAL ATTACHED | 06-02-2024 | 1041019 | Malappuram | LABOMED | LABORATORY | PHC OZHOOR | Laboratory Microscope | NON CRITICAL |
|
||
229 | Pending | Approved | View | VERIFIED | 17-05-2024 | 0740139 | Ernakulam | Dr Morepen | OP | PHC PAIPRA | B P Apparatus | NON CRITICAL |
|
||
230 | Pending | Approved | View | VERIFIED | 06-04-2024 | 0543330 | Kottayam | NIL | OP | PHC PALLICKATHODU | Sterlizer | NON CRITICAL |
|
||
231 | Pending | Approved | View | VERIFIED | 06-12-2024 | 0542929 | Kottayam | Robonik | LABORATORY | PHC PARATHODU | Semi-Auto Biochemistry Analyzer | NON CRITICAL |
|
||
232 | Pending | Approved | View | Equipment completed 8+ year,oem informed robonick preitest touch semi auto analyser is an obselate model,as per tenderclause 5.3.14.2 we recommend equipment for condemination | 04-07-2024 | 0943347 | Palakkad | Robonik | LABORATORY | PHC Pattithara | Semi-Auto Biochemistry Analyzer | NON CRITICAL |
|
||
233 | Pending | Approved | View | Equipment completed 7+ years, repairing cost of equipment is 91% of asset value, both criteria for rber met , as per tender clause 5.3.14.1 we recommend equipment for condemination | 20-09-2024 | 1040500 | Malappuram | PROKAN | LABORATORY | PHC Ponmundam | Haematology Analyser | CRITICAL |
|
||
234 | Pending | Approved | View | Equipment completed 9 year 9 month , oem issued end of life letter for robonick preitest touch semi auto analyser,as per tender clause 5.3.14.2 we recommend equipment for condemination | 07-12-2024 | 0844945 | Thrissur | Robonik | LABORATORY | PHC POOKODE | Biochemistry Analyzer | CRITICAL |
|
||
235 | Pending | Approved | View | Fully Rusted and Dilapidated condition- Images attached- Verified- Recommended for RBER | 09-11-2023 | 0340980 | Pathanamthitta | LG | observation room | PHC PRAMADAM | Pharmacy Refrigerator | NON CRITICAL |
|
||
236 | Pending | Approved | View | Equipment completed 8+ year, End of life letter from OEM received | 23-12-2023 | 1342816 | Kannur | Robonik | LABORATORY | PHC RAMANTHALI | Biochemistry Analyzer | CRITICAL |
|
||
237 | Pending | Approved | View | Very Old machine- Asset value Below 10K- Images attached- Verified- Recommended for RBER | 13-11-2023 | 0543507 | Kottayam | Whirlpool | OP | PHC THALANAD | Pharmacy Refrigerator | NON CRITICAL |
|
||
238 | Pending | Approved | View | Document verified- RBER Approved | 13-11-2023 | 0543495 | Kottayam | Diamond | OP | PHC THALANAD | B P Apparatus | NON CRITICAL |
|
||
239 | Pending | Approved | View | Verified- Recommended for RBER | 04-01-2024 | 0543432 | Kottayam | Infi-neb | OP | PHC THALAYAZAHAM | Nebuliser | NON CRITICAL |
|
||
240 | Pending | Approved | View | Equipment outlived 16+ year,repairing cost of equipment is more than 100%,both criteria for RBER Met | 05-01-2024 | 1241103 | Wayanad | ELECTROLUX | LABORATORY | PHC THONDERNADU | Pharmacy Refrigerator | NON CRITICAL |
|
||
241 | Pending | Approved | View | Equipment completed 9 year 8 months oem informed robonick prietest touch is an end of life notified | 16-11-2024 | 1440446 | KASARGODE | Robonik | LABORATORY | PHC UDUMA | Semi-Auto Biochemistry Analyzer | NON CRITICAL |
|
||
242 | Pending | Approved | View | VERIFIED | 27-08-2024 | 0640063 | Idukki | LIFE LINE | PP UNIT | PHC UDUMPANCHOLA | B P Apparatus | NON CRITICAL |
|
||
243 | Pending | Approved | View | Equipment completed 8+year, End of life letter provide by oem robonick as per tenderclause 5.3.14.2 | 24-01-2024 | 1041250 | Malappuram | Robonik | LABORATORY | PHC VATTAMKULAM | Biochemistry Analyzer | CRITICAL |
|
||
244 | Pending | Approved | View | Equipment completed 10 year,Eol letter provided by oem attached, as per tender clause 5.3.14.2 | 04-01-2024 | 1241206 | Wayanad | Robonik | LABORATORY | PHC Vellamunda | Biochemistry Analyzer | CRITICAL |
|
||
245 | Pending | Approved | View | verified | 10-06-2024 | 0142245 | Thiruvananthapuram | Robonik | LABORATORY | PHC VEMBAYAM | Semi-Auto Biochemistry Analyzer | NON CRITICAL |
|
||
246 | Pending | Approved | View | Equipment completed 9+ year, Oem informed Robonick preitest touch semi auto analyser has covered its maximum life, end of life declared,as per tender clause 5.3.14.2 we recommend equipment for condemination | 22-04-2024 | 0843670 | Thrissur | Robonik | LABORATORY | PHC VENKIDANGU | Biochemistry Analyzer | CRITICAL |
|
||
247 | Pending | Approved | View | VERIFIED | 22-08-2024 | 0241427 | Kollam | Robonik | LABORATORY | PHC VILAKKUDY | Semi-Auto Biochemistry Analyzer | NON CRITICAL |
|
||
248 | Pending | Approved | View | RBER | 29-03-2023 | 0843770 | Thrissur | Robonik | LABORATORY | PHC VILVATTOM | Biochemistry Analyzer | CRITICAL |
|
||
249 | Pending | Approved | View | Equipment completed 10 years repiring cost of equipment is 80% of asset value, both criteria for rber met | 16-11-2024 | 0940029 | Palakkad | REMI ELEKTROTECHNIK LTD. | LABORATORY | PHC Kalladikodu | Centrifuge | NON CRITICAL |
|
||
250 | Pending | Approved | View | Equipment completed 12 year,repairing cost of equipment is more than 100% both criteria for rber met | 14-12-2023 | 1311996 | Kannur | CORNLEY | LABORATORY | RPH LAB THALAP KANNUR | Electrolyte Analyser | CRITICAL |
|
||
251 | Pending | Approved | View | Equipment completd 10 year 11 months,repairing cost of equipment is 71% of asset value, both criteria for rber met, as per tender clause 5.3.14.1 we recommend equipment for condemination | 20-12-2024 | 1311938 | Kannur | KELVINATOR | LABORATORY | RPH LAB THALAP KANNUR | Pharmacy Refrigerator | NON CRITICAL |
|
||
252 | Pending | Approved | View | VERIFIED | 23-02-2024 | 0411931 | Alappuzha | Robonik | LABORATORY | TB CENTER ALAPPUZHA | Biochemistry Analyzer | CRITICAL |
|
||
253 | Pending | Approved | View | VERIFIED | 29-05-2024 | 0211566 | Kollam | Rotek Instruments | LABORATORY | TB CENTRE KOLLAM | WATER BATH | NON CRITICAL |
|
||
254 | Pending | Approved | View | Equipment completed 11 year 11 months oem issued end of life for semiauto analyser robonick prietest touch so as per tender clause 5.3.14.2 we recommend equipment for condemination | 22-11-2024 | 1012251 | Malappuram | Robonik | LABORATORY | TB CLINIC PONNANI | Semi-Auto Biochemistry Analyzer | NON CRITICAL |
|
||
255 | Pending | Approved | View | Equipment completed 9 year with an asset cost of rupees 1353.Quotation not submitted since spares are not available in the market | 17-02-2024 | 0920917 | Palakkad | NA | Labour Ward | THQH ALATHUR | Nebuliser | NON CRITICAL |
|
||
256 | Pending | Approved | View | Checked equipment found equipment outlived 9+ year, repairing cost of equipment is 88% of asset value, both criteria for rber met | 27-05-2024 | 0920786 | Palakkad | BIOMEDIX | EYE OP | THQH ALATHUR | Ocular Ultrasound | NON CRITICAL |
|
||
257 | Pending | Approved | View | VERIFIED | 07-12-2024 | 0120085 | Thiruvananthapuram | ADONIS MEDICAL SYSTEM PVT.LTD. | X RAY | THQH ATTINGAL | X Ray Machine | CRITICAL |
|
||
258 | Pending | Approved | View | VERIFIED | 20-07-2024 | 0120831 | Thiruvananthapuram | Steri | OT | THQH ATTINGAL | Sterlizer | NON CRITICAL |
|
||
259 | Pending | Approved | View | Equipment completed 7+ year repairing cost of equipment is more than 100% both criteria for RBER Met | 08-07-2024 | 0820100 | Thrissur | SENSA CORE | LABORATORY | THQH CHALAKUDY | Electrolyte Analyser | CRITICAL |
|
||
260 | Pending | Approved | View | Equipment completed 7 year 11 months with An asset cost of rupees 1353, Quotation not submitted since spares are not available in the market | 27-11-2024 | 0820691 | Thrissur | ACCUNEB JL | LABOUR ROOM | THQH CHALAKUDY | Nebuliser | NON CRITICAL |
|
||
261 | Pending | Approved | View | Equipment outlived 8+ year repairing cost of equipment is more than 100%,both criteria for rber met | 10-06-2024 | 0820723 | Thrissur | Transasia Biomedical LTD. | LABORATORY | THQH CHAVAKAD | ESR Analyser | NON CRITICAL |
|
||
262 | Pending | Approved | View | Equipment completed 12+ year,Oem robonick informed that prietest touch semi auto have covered its maximum life and end of life declared as per tender clause 5.3.14.2 we recommend equipment for condemination | 15-02-2024 | 0820706 | Thrissur | Robonik | LABORATORY | THQH CHAVAKAD | Biochemistry Analyzer | CRITICAL |
|
||
263 | Pending | Approved | View | VERIFIED | 06-07-2023 | 0420249 | Alappuzha | ASPEN | WARD 1 | THQH CHERTHALA | Infusion Pump | CRITICAL |
|
||
264 | Pending | Approved | View | VERIFIED | 16-12-2024 | 0420241 | Alappuzha | manchal | OT | THQH CHERTHALA | Tourniquet | NON CRITICAL |
|
||
265 | Pending | Approved | View | VERIFIED | 07-05-2024 | 0420188 | Alappuzha | Robonik | LABORATORY | THQH CHERTHALA | Biochemistry Analyzer | CRITICAL |
|
||
266 | Pending | Approved | View | VERIFIED | 16-04-2024 | 0121066 | Thiruvananthapuram | BPL | OT | THQH CHIRAYINKEEZH | Multipara Monitor | CRITICAL |
|
||
267 | Pending | Approved | View | Equipment completed 10+ year, repaiirng cost of equipment is 87%, both criteria for rber met as per tender clause 5.3.14.1 we recommend equipment for condemination | 21-08-2024 | 0920077 | Palakkad | Supreme Enterprises | LABOUR ROOM | THQH CHITTUR | Suction Apparatus | NON CRITICAL |
|
||
268 | Pending | Approved | View | VERIFIED | 19-08-2023 | 0720734 | Ernakulam | SURGI-VAC | OT | THQH FORTKOCHI | Suction Apparatus | NON CRITICAL |
|
||
269 | Pending | Approved | View | verified | 20-02-2024 | 0420910 | Alappuzha | DRAGER | LABOUR ROOM | THQH HARIPPAD | Bilirubinometer | NON CRITICAL |
|
||
270 | Pending | Approved | View | VERIFIED | 25-10-2023 | 0420779 | Alappuzha | SKANRAY | pediatric | THQH HARIPPAD | Pulse Oximeter | NON CRITICAL |
|
||
271 | Pending | Approved | View | Equipment completed 13+ years, Oem provided End of life for the said machiene, as per tender clause 5.3.14.2 we recommend equipmemt for condemination | 04-07-2024 | 1320386 | Kannur | Robonik | LABORATORY | THQH IRITTY | Biochemistry Analyzer | CRITICAL |
|
||
272 | Pending | Approved | View | VERIFIED | 02-11-2024 | 0220140 | Kollam | GE Elpro Medical Systems | ultrasound Room | THQH KADAKKAL | Ultrasound Machine | CRITICAL |
|
||
273 | Pending | Approved | View | VERIFIED | 21-10-2024 | 0220213 | Kollam | Biorad | LABORATORY | THQH KARUNAGAPPALLY | Elisa Washer | NON CRITICAL |
|
||
274 | Pending | Approved | View | VERIFIED | 14-05-2024 | 0421651 | Alappuzha | Labline Instruments | LABORATORY | THQH KAYAMKULAM | Hot Air Oven | NON CRITICAL |
|
||
275 | Pending | Approved | View | VERIFIED | 21-12-2024 | 0421677 | Alappuzha | REMI ELEKTROTECHNIK LTD. | LABORATORY | THQH KAYAMKULAM | Centrifuge | NON CRITICAL |
|
||
276 | Pending | Approved | View | VERIFIED | 31-12-2024 | 0421572 | Alappuzha | Nice Neotech Medical Systems PVT.LTD | LABOUR ROOM | THQH KAYAMKULAM | Phototherapy | CRITICAL |
|
||
277 | Pending | Approved | View | VERIFIED | 05-04-2024 | 0421040 | Alappuzha | BPL | ECG ROOM | THQH KAYAMKULAM | ECG Machine | CRITICAL |
|
||
278 | Pending | Approved | View | Equipment completed 15+ year, Oem informed the model was discontinued and no spareparts and services are available, end of life letter issued as per tender clause 5.3.14.2 we recommend equipment for condemination | 07-03-2024 | 0821196 | Thrissur | Reico | POST OPERATIVE WARD | THQH KODUNGALLUR | Sterlizer | NON CRITICAL |
|
||
279 | Pending | Approved | View | Equipment completed 12 year 7 months with an asset cost of rupees 1353, Quotation not submitted since spares are not available in the market | 27-11-2024 | 1321562 | Kannur | noren | OP | THQH KOOTHUPARAMBA | Nebuliser | NON CRITICAL |
|
||
280 | Pending | Approved | View | Equipment completed 5+ year, with an asset cost of rupees 1323, repairing cost of equipment is more than 100% both criteria for rber met | 22-05-2024 | 1321613 | Kannur | DISHA | CASUALTY | THQH KOOTHUPARAMBA | B P Apparatus | NON CRITICAL |
|
||
281 | Pending | Approved | View | Equipment completed 8 year 10 months , repairing cost of equipment is 66% of asset value,both criteria for rber met | 21-02-2024 | 1321512 | Kannur | LABOMED | Blood storage unit | THQH KOOTHUPARAMBA | Laboratory Microscope | NON CRITICAL |
|
||
282 | Pending | Approved | View | VERIFIED | 03-10-2024 | 0722230 | Ernakulam | PHOENIX | NBSU | THQH KOTHAMANGALAM | Infant Radiant Warmer | CRITICAL |
|
||
283 | Pending | Approved | View | Verified- Recommended RBER | 03-10-2023 | 0722365 | Ernakulam | PHILIPS | OT | THQH KOTHAMANGALAM | OT Light | CRITICAL |
|
||
284 | Pending | Approved | View | VERIFIED | 04-11-2024 | 0220614 | Kollam | Allengers | X RAY | THQH KOTTARAKKARA | X Ray Machine | CRITICAL |
|
||
285 | Pending | Approved | View | Equipment outlived 9+ year, repairing cost of equipment is more than 100%,Both criteria for RBER Met | 24-06-2024 | 1121526 | Kozhikode | BPL | LABOUR ROOM | THQH KOYILANDY | Multipara Monitor | CRITICAL |
|
||
286 | Pending | Approved | View | Equipment completed 13 year 8 months, repairing cost of equipment is more than 100%, both criteria for rber met | 14-10-2024 | 1121140 | Kozhikode | Allengers | X RAY | THQH KOYILANDY | X Ray Machine | CRITICAL |
|
||
287 | Pending | Approved | View | Equipment completed 9 year 6months , repairing cost of equipment is more than 100% ,both criteria for rber met | 24-09-2024 | 0822564 | Thrissur | Erba | LABORATORY | THQH KUNNAMKULAM | Biochemistry Analyzer | CRITICAL |
|
||
288 | Pending | Approved | View | Equipment completed 19+ year repairing cost of equipment is 75% of asset value, both criteria for rber met | 02-09-2024 | 1120685 | Kozhikode | Kadavil Electro Mechanical Industries | LABORATORY | THQH KUTTIADI | Centrifuge | NON CRITICAL |
|
||
289 | Pending | Approved | View | Equipment completed 14 year 7 month, oem fuji film informed that Aloka- prosound is end of life declared model and spare parts were not available in the market. As per tender clause 5.3.14.2 we recommend equipment for condemination | 20-12-2024 | 1021715 | Malappuram | ALOKA | LABOUR ROOM | THQH MALAPPURAM | Ultrasound Machine | CRITICAL |
|
||
290 | Pending | Approved | View | Equipment completed 11 years, oem informed that model is obselated and spares are not availble as per tender clause 5.3.14.2 we recommend equipment for condemination | 26-09-2024 | 0920599 | Palakkad | JY23 | DENTAL DEPARTMENT | THQH MANNARKKAD | Sterlizer | NON CRITICAL |
|
||
291 | Pending | Approved | View | Equipment completed 12+ year, repairing cost of equipment is more than 100%, both criteria for rber met | 09-01-2024 | 0920467 | Palakkad | Allengers | OT | THQH MANNARKKAD | C-ARM | NON CRITICAL |
|
||
292 | Pending | Approved | View | Letter attached- Images and Documents Verified- Recommended for RBER | 20-10-2023 | 0620715 | Idukki | videocon | LABORATORY | THQH NEDUMKANDAM | Pharmacy Refrigerator | NON CRITICAL |
|
||
293 | Pending | Approved | View | VERIFIED | 21-05-2024 | 0620782 | Idukki | Beston Industries | LABORATORY | THQH NEDUMKANDAM | Nebuliser | NON CRITICAL |
|
||
294 | Pending | Approved | View | VERIFIED | 24-07-2024 | 0221271 | Kollam | Godrej | CASUALTY | THQH NEDUNGOLAM | Pharmacy Refrigerator | NON CRITICAL |
|
||
295 | Pending | Approved | View | VERIFIED | 30-01-2024 | 0721491 | Ernakulam | BPL | CASUALTY | THQH NORTH PARAVOOR | Oxygen Concentrator | NON CRITICAL |
|
||
296 | Pending | Approved | View | VERIFIED | 11-03-2024 | 0721445 | Ernakulam | Dr Morepen | ORTHO WARD | THQH NORTH PARAVOOR | Nebuliser | NON CRITICAL |
|
||
297 | Pending | Approved | View | VERIFIED | 27-03-2024 | 0721406 | Ernakulam | Arizant Healthcare | OT | THQH NORTH PARAVOOR | Fluid Warmer | NON CRITICAL |
|
||
298 | Pending | Approved | View | VERIFIED | 17-05-2024 | 0721556 | Ernakulam | MINDRAY | OT | THQH NORTH PARAVOOR | Multipara Monitor | CRITICAL |
|
||
299 | Pending | Approved | View | Equipment completed 8 year 11 months, repairing cost of equipment is 78% of asset value, both criteria for rber met | 06-11-2024 | 67093022 | Palakkad | B BRAWN | DIALYSIS UNIT | THQH OTTAPALAM | Dialysis Machine | CRITICAL |
|
||
300 | Pending | Approved | View | VERIFIED | 30-09-2024 | 0120460 | Thiruvananthapuram | CAVITRON | DENTAL DEPARTMENT | THQH PARASSALA | Dental Scalar | NON CRITICAL |
|
||
301 | Pending | Approved | View | Equipment completed 12 years,mindray dp9900 plus is an obselate model as per tender clause 5.3.14.2 Met rber criteria | 22-03-2024 | 0921465 | Palakkad | MINDRAY | Labour Ward | THQH PATTAMBI | Ultrasound Machine | CRITICAL |
|
||
302 | Pending | Approved | View | Equipment completed 8 year 6 months with an asset cost of rupees 1323, Quotation not submitted since spares are not available in the market | 25-10-2024 | 0921596 | Palakkad | Elkometer | CASUALTY | THQH PATTAMBI | B P Apparatus | NON CRITICAL |
|
||
303 | Pending | Approved | View | Equipment completed 13+ year, repairing cost of equipment is more than 100%, both criteria for rber met | 24-08-2024 | 1320776 | Kannur | BPL | OT | THQH PAYYANNUR | Multipara Monitor | CRITICAL |
|
||
304 | Pending | Approved | View | The equipment installed on 14/01/2010 has outlived 14+ years. The asset value of the equipment is 26,494 rupees, while the actual book value is 6,500 rupees. The repair cost of the equipment is more than 100% of the book value, and both criteria for RBER have been met considering the book value. | 09-03-2024 | 1321108 | Kannur | videocon | Labour Ward | THQH PAYYANNUR | Pharmacy Refrigerator | NON CRITICAL |
|
||
305 | Pending | Approved | View | Equipment completed 12 year, Robonick preitest touch is an Eol declared model, as per tender clause 5.3.14.2 we recommend equipment for condemination | 30-05-2024 | 1320022 | Kannur | Robonik | LABORATORY | THQH PERAVOOR | Biochemistry Analyzer | CRITICAL |
|
||
306 | Pending | Approved | View | Equipment completed 9 year 6 months, repairing cost of equipment is 80% of asset value, both criteria for rber met | 30-12-2024 | 1320019 | Kannur | MINDRAY | LABORATORY | THQH PERAVOOR | Haematology Analyser | CRITICAL |
|
||
307 | Pending | Approved | View | Equipment completed 14+ year,with an asset value of rupes 26494, the orginal book value of the equipment is 11400 rupees, the repairing cost of the equipment is 71% with respect to orginal book value | 27-03-2024 | 1320135 | Kannur | Godrej | ICTC LAB | THQH PERAVOOR | Pharmacy Refrigerator | NON CRITICAL |
|
||
308 | Pending | Approved | View | VERIFIED | 24-07-2024 | 0721842 | Ernakulam | PARAS ENTERPRISES | WARD 7 | THQH PERUMBAVOOR | Sterlizer | NON CRITICAL |
|
||
309 | Pending | Approved | View | VERIFIED | 17-12-2024 | 0421124 | Alappuzha | OMRON HEALTHCARE CO.LTD. | opd | THQH PULINKUNNU | Nebuliser | NON CRITICAL |
|
||
310 | Pending | Approved | View | VERIFIED | 19-03-2024 | 0222270 | Kollam | DIAMOND | LABOUR ROOM | THQH SASTHAMKOTTA | B P Apparatus | NON CRITICAL |
|
||
311 | Pending | Approved | View | Equipment outlived 10+ year, repairing cost of equipment is 79%, both criteria for rber met as per tender clause 5.3.14.1 we recommend equipment for condemination | 29-08-2024 | 1120223 | Kozhikode | INTERACOUSTICS | NICU | THQH THAMARASSERY | Oto Acoustic Emission Screener (OAE) | NON CRITICAL |
|
||
312 | Pending | Approved | View | Equipment completed 11 year 7 month,repairing cost of equipment is 91% both critera for rber met | 18-01-2024 | 1120071 | Kozhikode | MINDRAY | LABORATORY | THQH THAMARASSERY | Haematology Analyser | CRITICAL |
|
||
313 | Pending | Approved | View | Equipment completed 12+ year, repairing cost of equipment is more than 100% both criteria for rber met | 30-07-2024 | 1120099 | Kozhikode | MINDRAY | OT | THQH THAMARASSERY | Multipara Monitor | CRITICAL |
|
||
314 | Pending | Approved | View | Equipment completed 7 year 11 month, repairing cost of equipment is 66%, Both criteria for rber met | 22-02-2024 | 1120220 | Kozhikode | PRAGATHI MEDICAL SYSTEMS | NICU | THQH THAMARASSERY | Weighing Machine | NON CRITICAL |
|
||
315 | Pending | Approved | View | VERIFIED | 10-03-2024 | 0720183 | Ernakulam | ALMICRO | LABORATORY | THQH THRIPUNITHURA | Laboratory Microscope | NON CRITICAL |
|
||
316 | Pending | Approved | View | verified | 02-03-2024 | 0720054 | Ernakulam | MINDRAY | LABORATORY | THQH THRIPUNITHURA | Haematology Analyser | CRITICAL |
|
||
317 | Pending | Approved | View | VERIFIED | 14-06-2024 | 0421201 | Alappuzha | SPAN | LABORATORY | THQH THURAVOOR | Elisa Reader | NON CRITICAL |
|
||
318 | Pending | Approved | View | Equipment condemned from hospital as per order no G8421/2023 dtaed 01-01-2024 | 20-08-2024 | 1020810 | Malappuram | NEOTECH | LABOUR ROOM | THQH TIRURANGADI | Infant Radiant Warmer with Phototherapy Unit | CRITICAL |
|
||
319 | Pending | Approved | View | Equipment completed 10+ year repairing cost of equipment is 89% of asset value,both criteria for rber met | 20-08-2024 | 1021860 | Malappuram | Mindray | LABORATORY | THQH TIRURANGADI | Haematology Analyser | CRITICAL |
|
||
320 | Pending | Approved | View | Equipment completed 9+ year with an asset cost of rupees 1353, Quotation not submitted since spares are not available in the market | 01-09-2024 | 1420090 | KASARGODE | Easy Care | OP | THQH Trikkaripur | Nebuliser | NON CRITICAL |
|
||
321 | Pending | Approved | View | Verified- Recommended RBER | 18-12-2023 | 0121261 | Thiruvananthapuram | PHILIPS | OT | THQH VARKALA | Defibrillator | NON CRITICAL |
|
||
322 | Pending | Approved | View | Equipment outlived 12+ year, repairing cost of equipment is 99% of asset value, both criteria for RBER Met | 29-05-2024 | 1020644 | Malappuram | MINDRAY | LABORATORY | THQH WANDOOR | Haematology Analyser | CRITICAL |
|
||
323 | Pending | Approved | View | Equipment completed 10+year, repairing cost of equipment is more than 100%,both criteria for RBER met | 23-10-2023 | 0913284 | Palakkad | CORNLEY | LABORATORY | TRIBAL SPECIALITY HOSPITAL KOTTATHARA | Electrolyte Analyser | CRITICAL |
|
||
324 | Pending | Approved | View | Equipment outlived 5+ year, with an asset cost of rupees 1353,Quotation not submitted since spares are not available in the market | 03-07-2024 | 0913276 | Palakkad | Transasia Biomedical LTD. | LABORATORY | TRIBAL SPECIALITY HOSPITAL KOTTATHARA | ESR Analyser | NON CRITICAL |
|
||
325 | Pending | Approved | View | Equipment completed 10 year 7 months , repairing cost of equipment is more than 100%, both criteria for rber met | 21-12-2024 | 0913283 | Palakkad | Erba | LABORATORY | TRIBAL SPECIALITY HOSPITAL KOTTATHARA | Biochemistry Analyzer | CRITICAL |
|
||
326 | Pending | Approved | View | VERIFIED | 18-12-2023 | 0140728 | Thiruvananthapuram | DISHA | OP | UPHC CHAKKAI | B P Apparatus | NON CRITICAL |
|
||
327 | Pending | Approved | View | VERIFIED | 04-01-2024 | 0140722 | Thiruvananthapuram | REMI ELEKTROTECHNIK LTD. | LABORATORY | UPHC CHAKKAI | Centrifuge | NON CRITICAL |
|
||
328 | Pending | Approved | View | VERIFIED | 10-06-2024 | 0140989 | Thiruvananthapuram | LABLINE INSTRUMENTS | LABORATORY | UPHC CHALAI | Incubator | NON CRITICAL |
|
||
329 | Pending | Approved | View | Equipment completed 9 year 3 months, oem issued end of life letter, as per tender clause 5.3.14.2 we recommend equipment for condemination | 24-10-2024 | 1140467 | Kozhikode | Robonik | LABORATORY | UPHC CHELAVOOR | Semi-Auto Biochemistry Analyzer | NON CRITICAL |
|
||
330 | Pending | Approved | View | VERIFIED | 03-10-2024 | 0741930 | Ernakulam | REMI ELEKTROTECHNIK LTD. | LABORATORY | UPHC KALOOR | Centrifuge | NON CRITICAL |
|
||
331 | Pending | Approved | View | Equipment completed 8+ year, repaiirng cost of equipment is 86% of asset value, both criteria for rber met | 19-08-2024 | 1440898 | KASARGODE | BPL | OP | UPHC KANHANGAD | ECG Machine | CRITICAL |
|
||
332 | Pending | Approved | View | VERIFIED | 10-07-2024 | 0141223 | Thiruvananthapuram | Robonik | LABORATORY | UPHC KANNAMMOOLA | Biochemistry Analyzer | CRITICAL |
|
||
333 | Pending | Approved | View | Equipment completed 8 year 10 month End of life for the equipment was issued by oem robonick | 30-05-2024 | 1142023 | Kozhikode | Robonik | LABORATORY | UPHC KANNANCHERI | Biochemistry Analyzer | CRITICAL |
|
||
334 | Pending | Approved | View | VERIFIED | 22-05-2024 | 0240123 | Kollam | KELVINATOR | LABORATORY | UPHC KARUNAGAPALLY | Pharmacy Refrigerator | NON CRITICAL |
|
||
335 | Pending | Approved | View | Equipment completed 9 year 7 month, oem informed the said equipment was end of life notified so as per tender clause 5.3.14.2 we recommend equipment for condemination | 18-09-2024 | 1140114 | Kozhikode | Robonik | LABORATORY | UPHC KINASSERY | Biochemistry Analyzer | CRITICAL |
|
||
336 | Pending | Approved | View | Equipment completed 9 year 4 month repairing cost of equipment is 71% of asset value, both criteria for rber met as per tender clause 5.3.14.1 we recommend equipment for condemination | 09-12-2024 | 0943361 | Palakkad | REMI ELEKTROTECHNIK LTD. | LABORATORY | UPHC OTTAPALAM | Centrifuge | NON CRITICAL |
|
||
337 | Pending | Approved | View | Equipment completed 9 year 8 month with an asset cost of rupees 1353, Quotation not submitted since spares are not available in the market | 24-09-2023 | 0845041 | Thrissur | OMRON HEALTHCARE CO.LTD. | OP | UPHC V.R PURAM | Nebuliser | NON CRITICAL |
|
||
338 | Pending | Approved | View | VERIFIED | 07-08-2024 | 0540411 | Kottayam | Robonik | LABORATORY | UPHC VELOOR | Semi-Auto Biochemistry Analyzer | NON CRITICAL |
|
||
339 | Pending | Approved | View | Equipment completed 10 year 6 months, repairing cost of equipment is 97% of asset value, both criteria for rber met | 14-10-2024 | 1042976 | Malappuram | REMI ELEKTROTECHNIK LTD. | LABORATORY | UPHC VETTEKODE | Centrifuge | NON CRITICAL |
|
||
340 | Pending | Approved | View | verified | 02-11-2024 | 0412253 | Alappuzha | UNIVERSAL BIOCHEMICALS | BLOOD BANK | WOMEN & CHILDREN HOSPITAL ALAPPUZHA | Incubator | NON CRITICAL |
|
||
341 | Pending | Approved | View | VERIFIED | 13-06-2024 | 0412239 | Alappuzha | HEMOCUE | BLOOD BANK | WOMEN & CHILDREN HOSPITAL ALAPPUZHA | Haemoglobinometer | NON CRITICAL |
|
||
342 | Pending | Approved | View | Equipment completed 17 year 2 months, repairing cost of equipment is 83% of asset value, both criteria for rber met | 26-10-2024 | 1111330 | Kozhikode | termopenpol | BLOOD BANK | WOMEN & CHILDREN HOSPITAL KOZHIKODE | Blood Collection Monitor | NON CRITICAL |
|
||
343 | Pending | Approved | View | As per asset equipment installed on 01-04-2015 , but actual installation date as per stock is 06-03-2010, considering stock equipment outlived 14 year 8 months, repairing cost of equipment is 71% of asset value, both criteria for rber met, as per tender clause 5.3.14.1 we recommend equipment for condmeination | 13-11-2024 | 1111169 | Kozhikode | MINDRAY | LABORATORY | WOMEN & CHILDREN HOSPITAL KOZHIKODE | Haematology Analyser | CRITICAL |
|
||
344 | Pending | Approved | View | Equipment completed 7 year 4 month repairing cost of equipment is 83% of asset value, both criteria for rber met | 29-11-2023 | 1111318 | Kozhikode | Chromate | BLOOD BANK | WOMEN & CHILDREN HOSPITAL KOZHIKODE | Elisa Reader | NON CRITICAL |
|
||
345 | Pending | Approved | View | Equipment outlived 10+ year, Repairing cost of equipment is 71% of asset value, both criteria for rber met | 19-06-2024 | 0912654 | Palakkad | Transasia Biomedical LTD. | LABORATORY | WOMEN & CHILDREN HOSPITAL PALAKKAD | Fully Automated Random Access Clinical Chemistry Analyzer | NON CRITICAL |
|
||
346 | Pending | Approved | View | Equipment completed 10+ year, repairing cost of equipment is more than 100%,Both criteria for rber met | 07-08-2024 | 0912705 | Palakkad | INTERACOUSTICS | DEIC | WOMEN & CHILDREN HOSPITAL PALAKKAD | Oto Acoustic Emission Screener (OAE) | NON CRITICAL |
|
||
347 | Pending | Approved | View | Equipment completed 9 year 11 months repairing cost of equipment is 82% of asset value, both criteria for rber met | 0912721 | Palakkad | Appasamy Associates PVT.LTD. | DEIC | WOMEN & CHILDREN HOSPITAL PALAKKAD | Ophthalmoscope | NON CRITICAL |
|
|||
348 | Pending | Approved | View | Verified | 12-06-2024 | 0114531 | Thiruvananthapuram | ANITA INDUSTRIES | CASUALTY | WOMEN & CHILDREN HOSPITAL THYCAUD | B P Apparatus | NON CRITICAL |
|
||
349 | Pending | Approved | View | VERIFIED | 07-06-2024 | 0211111 | Kollam | PHOENIX | NICU | WOMEN & CHILDREN VICTORIA HOSPITAL KOLLAM | Infant Radiant Warmer | CRITICAL |
|
||
350 | Pending | Approved | View | Equipment completed 7 years, repairing cost of equipment is 82% both criteria for rber met | 11-01-2024 | 1331063 | Kannur | PROKAN | LABORATORY | CHC MAYYIL | Haematology Analyser | CRITICAL |
|
||
351 | Pending | Approved | View | Equipment completed 5 year 3 months, OEM informed spare parts are not available model obselate as per tender clause 5.3.14.2 | 28-12-2023 | 1311185 | Kannur | Biorad | BLOOD BANK | GH THALASSERY | Elisa Reader | NON CRITICAL |
|
||
352 | Pending | Approved | View | Equipment completed 7 year 6 months oem informed model was discontinued so spares parts are not available as per tender clause 5.3.14.2 we recommend equipment for condemination | 16-10-2024 | 1142919 | Kozhikode | LABY | LABORATORY | PHC MANGAD | Centrifuge | NON CRITICAL |
|
||
353 | Pending | Approved | View | Equipment completed 5 year 9 month, with an asset cost of rupees 1875, Quotation not submitted since spares are not available in the market | 03-12-2024 | 0840327 | Thrissur | OMRON HEALTHCARE CO.LTD. | OP | UPHC ANAPUZHA | Nebuliser | NON CRITICAL |
|
||
354 | Pending | Approved | View | Equipment completed 4+ year with an asset cost of rupees 2104.Quotation not submitted since spares are not available in the market | 22-02-2024 | 1440113 | KASARGODE | DIAMOND | OP | PHC NARKILAKKAD | B P Apparatus | NON CRITICAL |
|
||
355 | Pending | Approved | View | Equipment completed 5 years 2 months with an asset cost of rupees 850,Quotation not submitted since spares are not available in the market | 16-10-2024 | 1340075 | Kannur | SAMSO | SUB CENTRE | PHC MUNDERI | Weighing Machine | NON CRITICAL |
|
||
356 | Pending | Approved | View | Equipment completed 4+ year with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market | 02-08-2024 | 1342342 | Kannur | EKCO | OP | PHC PAMPURUTHY | B P Apparatus | NON CRITICAL |
|
||
357 | Pending | Approved | View | Equipment completed 8 year 5 months, repairing cost of equipment is 79% of asset value, both criteria for rber met, as per tender clause 5.3.14.1 we recommend equipment for condemination | 19-09-2024 | 1311339 | Kannur | Prognosys | X RAY | GH THALASSERY | X Ray Film Processor | NON CRITICAL |
|
||
358 | Pending | Approved | View | Equipment outlived 8+ year,with an asset cost of rupees 1875,Quotation not submitted since spares are not available in the market | 01-02-2024 | 1420433 | KASARGODE | Olex | WARD | THQH MANGALPADY | Nebuliser | NON CRITICAL |
|
||
359 | Pending | Approved | View | VERIFIED | 30-08-2024 | 0540870 | Kottayam | Lifecare | OP | 24X7 PHC MANIMALA | Nebuliser | NON CRITICAL |
|
||
360 | Pending | Approved | View | Equipment completed 7 year 4 month, repairing cost of equipment is more than 68% of asset value, both criteria for rber met | 21-08-2024 | 1440715 | KASARGODE | AGD Biomedical | LABORATORY | PHC PADNA | Biochemistry Analyzer | CRITICAL |
|
||
361 | Pending | Approved | View | Equipment completed 5 year 8 months with an asset cost of rupees 1500, Quotation not submitted since spares are not available in the market | 16-11-2024 | 0832698 | Thrissur | BIO PLUS | PP UNIT | CHC MULLASSERY | B P Apparatus | NON CRITICAL |
|
||
362 | Pending | Approved | View | VERIFIED | 16-08-2024 | 0440753 | Alappuzha | Rossmax | OP | 24X7 PHC PALLITHODU | B P Apparatus | NON CRITICAL |
|
||
363 | Pending | Approved | View | VERIFIED | 27-08-2024 | 0640077 | Idukki | LIFE LINE | SUB CENTRE | PHC UDUMPANCHOLA | B P Apparatus | NON CRITICAL |
|
||
364 | Pending | Approved | View | VERIFIED | 29-07-2024 | 0640067 | Idukki | LIFE LINE | SUB CENTRE | PHC UDUMPANCHOLA | B P Apparatus | NON CRITICAL |
|
||
365 | Pending | Approved | View | Equipment completed 4+ year with an asset cost of rupees 1500.Quotation not submitted since spares are not available in the market | 10-01-2024 | 0845467 | Thrissur | ELKO | SUB CENTRE | PHC PERUMBILAVU | B P Apparatus | NON CRITICAL |
|
||
366 | Pending | Approved | View | VERIFIED | 29-07-2024 | 0740390 | Ernakulam | Agappe Diagnostics LTD. | LABORATORY | 24X7 PHC RAYAMANGALAM | Specific Protein Analyser | NON CRITICAL |
|
||
367 | Pending | Approved | View | VERIFIED | 17-10-2024 | 0721915 | Ernakulam | SAFE POWER | DIALYSIS UNIT | THQH PERUMBAVOOR | UPS | NON CRITICAL |
|
||
368 | Pending | Approved | View | VERIFIED | 15-05-2024 | 0520027 | Kottayam | AccuSure | CASUALTY | THQH VAIKOM | B P Apparatus | NON CRITICAL |
|
||
369 | Pending | Approved | View | Equipment outlived 6+ year, reparing cost of equipment is more than 60%, both criteria for rber met | 12-04-2024 | 0942302 | Palakkad | Sidco | OP | PHC VANIYAMKULAM | Sterlizer | NON CRITICAL |
|
||
370 | Pending | Approved | View | Equipment completed 6+ year with an asset cost of rupees 850, Quotation not submitted since spares are not available in the market | 09-08-2024 | 0940859 | Palakkad | SAMSO | OP | PHC KUMARAMPUTHUR | Weighing Machine | NON CRITICAL |
|
||
371 | Pending | Approved | View | Equipment completed 4 year 7 months with an asset cost of rupees 715,Quotation not submitted since spares are not available in the market | 15-11-2024 | 0942063 | Palakkad | SAMSO | SUB CENTRE | PHC ADAKKAPUTHUR | Weighing Machine | NON CRITICAL |
|
||
372 | Pending | Approved | View | VERIFIED | 23-08-2024 | 0230514 | Kollam | ATLAS SURGICALS | AUTOCLAVE ROOM | CHC PATHANAPURAM | Sterlizer | NON CRITICAL |
|
||
373 | Pending | Approved | View | Equipment completed 8 year 6 months repairing cost of equipment is 64% of asset value, both criteria for rber met. | 07-12-2024 | 0943471 | Palakkad | NIL | OP | PHC Pattithara | Sterlizer | NON CRITICAL |
|
||
374 | Pending | Approved | View | VERIFIED | 26-10-2024 | 0642235 | Idukki | OMRON HEALTHCARE CO.LTD. | OP | PHC CHEMPAKAPPARA | B P Apparatus | NON CRITICAL |
|
||
375 | Pending | Approved | View | VERIFIED | 24-06-2024 | 0740046 | Ernakulam | Euro Diagnostic Systems PVT.LTD | LABORATORY | PHC KOTTAPPADY | Centrifuge | NON CRITICAL |
|
||
376 | Pending | Approved | View | Equipment completed 4 year 8 months, Quotation not submitted since spares are not available in the market | 17-10-2024 | 1344428 | Kannur | SAMSO | OP | PHC MOKERI | Weighing Machine | NON CRITICAL |
|
||
377 | Pending | Approved | View | Equipment completed 4 year 11 months with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market | 16-11-2024 | 0832699 | Thrissur | LIFELINE | PP UNIT | CHC MULLASSERY | B P Apparatus | NON CRITICAL |
|
||
378 | Pending | Approved | View | Equipment completed 5 year 2 months with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market | 05-11-2024 | 1340072 | Kannur | EKCO | SUB CENTRE | PHC MUNDERI | B P Apparatus | NON CRITICAL |
|
||
379 | Pending | Approved | View | VERIFIED | 03-07-2024 | 0531006 | Kottayam | LIFE CARE EQUIPMENTS | SUB CENTRE | CHC KADAPLAMATTOM | B P Apparatus | NON CRITICAL |
|
||
380 | Pending | Approved | View | Year not met- Special case- Images and Documents verified- Recommended for RBER | 11-10-2023 | 0543629 | Kottayam | LIFE CARE EQUIPMENTS | SUB CENTRE | PHC VIZHIKKATHODU | B P Apparatus | NON CRITICAL |
|
||
381 | Pending | Approved | View | VERIFIED | 10-07-2024 | 0241496 | Kollam | Robonik | LABORATORY | PHC K.S PURAM | Semi-Auto Biochemistry Analyzer | NON CRITICAL |
|
||
382 | Pending | Approved | View | Equipment completed 4+ year with an asset cost of rupees 2200,Quotation not submitted since spares are not availble in the market | 22-07-2024 | 1343110 | Kannur | EKCO | SUB CENTRE | PHC CHERUKUNNU(P) | B P Apparatus | NON CRITICAL |
|
||
383 | Pending | Approved | View | Equipment outlived 19+ year repairing cost of equipment is 83% of asset value, both criteria for rber met | 03-12-2024 | 1121881 | Kozhikode | Bioplus | CASUALTY | THQH BALUSSERY | B P Apparatus | NON CRITICAL |
|
||
384 | Pending | Approved | View | Equipment completed 6+ year with an asset cost of rupees 1353. Quotation not submitted since spares are not available in the market | 08-07-2024 | 1142364 | Kozhikode | LIFE LINE | OP | PHC ARIKKULAM | Nebuliser | NON CRITICAL |
|
||
385 | Pending | Approved | View | Equipment completed 4+ year with an asset cost of rupees 1875.Quotation not submitted since spares are not available in the market | 08-07-2024 | 1142351 | Kozhikode | Lifecare | OP | PHC ARIKKULAM | Nebuliser | NON CRITICAL |
|
||
386 | Pending | Approved | View | Equipment completed 04+ year with an asset cost of rupees 1875, Quotation not submitted since spares are not available in the market. | 08-07-2024 | 1142366 | Kozhikode | Lifecare | OP | PHC ARIKKULAM | Nebuliser | NON CRITICAL |
|
||
387 | Pending | Approved | View | Equipment completed 7+ year, repairing cos of equipment is 80% of asset value, both criteria for rber met | 30-06-2024 | 0830274 | Thrissur | NIPRO | DIALYSIS UNIT | CHC MALA | Dialysis Machine | CRITICAL |
|
||
388 | Pending | Approved | View | Equipment outlived 5+ year with an asset cost of rupees 715.Quotation not submitted since spares are not available in the market | 24-05-2024 | 1344504 | Kannur | SAMSO | OP | PHC EDAKKAD | Weighing Machine | NON CRITICAL |
|
||
389 | Pending | Approved | View | Equipment completed 4+ year with an asset cost of rupees 715 Quotation not submitted since spares are not available in the market | 16-07-2024 | 1344587 | Kannur | SAMSO | OP | PHC TRIPPANGOTTOOOR | Weighing Machine | NON CRITICAL |
|
||
390 | Pending | Approved | View | Equipment completed 7+ year, repairinmg cost of equipment is more than 100%, both criteria for rber met | 02-08-2024 | 0831253 | Thrissur | PROKAN | LABORATORY | CHC ALAPAD | Haematology Analyser | CRITICAL |
|
||
391 | Pending | Approved | View | Equipment completed 13 years, repairing cost of equipment is 69.5% both criteria for Rber met | 14-11-2023 | 1330525 | Kannur | LG | OP | CHC IRIKKUR | Pharmacy Refrigerator | NON CRITICAL |
|
||
392 | Pending | Approved | View | Equipment completed 3+ year, oem informed that oxymed Om -501 ,has been discontinued and repair parts are no longer available for these models, as per tender clause 5.3.14.2 we recommend equipment for condemination | 04-07-2024 | 1321039 | Kannur | MEDOXY | WARD | THQH IRITTY | Oxygen Concentrator | NON CRITICAL |
|
||
393 | Pending | Approved | View | Equipment completed 6+ year, repairing cost of equipment is 90% of asset value, both criteria for rber met, as per temder clause 5.3.14.2 we recommend equipment for condmination | 23-08-2024 | 1340479 | Kannur | Heidelco | OP | PHC CHIRAKKAL | ECG Machine | CRITICAL |
|
||
394 | Pending | Approved | View | Equipment installed on 26-08-2019 with an asset cost of rupees 715 has outlived 4 years, Quotation not submitted since spares are not available in the market | 24-04-2024 | 1341991 | Kannur | SAMSO | maincenter | PHC MALAPPATTAM | Weighing Machine | NON CRITICAL |
|
||
395 | Pending | Approved | View | Equipment outlived 5+ years Quotation not submitted since spares are not available in the market | 18-04-2024 | 1340813 | Kannur | Healcure | OP | PHC ALAKODE (T) | Weighing Machine | NON CRITICAL |
|
||
396 | Pending | Approved | View | Equipment outlived 4+ year with an asset cost of rupees 715, Quotation not submitted since spares are not avialble in the market | 23-08-2024 | 1340579 | Kannur | SAMSO | SUB CENTRE | PHC ERAMAM KUTTUR | Weighing Machine | NON CRITICAL |
|
||
397 | Pending | Approved | View | Equipment completed 5 years with an asset cost of rupees 715, quotation not submitted since spares are not available in the market | 04-11-2024 | 1342983 | Kannur | SAMSO | SUB CENTRE | 24X7 PHC PARASSINIKADAVU | Weighing Machine | NON CRITICAL |
|
||
398 | Pending | Approved | View | Equipment completed 6 year 11month with an asset cost of rupees 850.Quotation not submitted since spares are not available in the market | 31-12-2024 | 1340378 | Kannur | Healcure | OP | 24X7 PHC VALAPATTANAM | Weighing Machine | NON CRITICAL |
|
||
399 | Pending | Approved | View | Equipment completed 4 year 11 months with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market. | 22-08-2024 | 1342765 | Kannur | Med Vision | OP | PHC ETTIKULAM | B P Apparatus | NON CRITICAL |
|
||
400 | Pending | Approved | View | Equipment outlived 6 months,body fully rusted,weaked body structure posing hazard to patients and user.the extensive rusting causes suction apparatus integrity, moreover, motor,suction bottle and tubings damaged.all these imperfections is potentially harmefull for patients and operational efficency | 13-03-2024 | 1311868 | Kannur | Supreme Enterprises | PAY WARD | GH THALASSERY | Suction Apparatus | NON CRITICAL |
|
||
401 | Pending | Approved | View | Equipment outlived 5+year,with an asset cost of rupees 1353,Quotation not submitted since spares are not available in the market | 29-05-2024 | 1330854 | Kannur | READY MIST | OP | CHC PAPINISERY | Nebuliser | NON CRITICAL |
|
||
402 | Pending | Approved | View | Equipment outlived 6+ year,Repairing cost more than 100%,Both critria Met for RBER | 06-10-2023 | 1330876 | Kannur | OMRON | OP | CHC PAPINISERY | B P Apparatus | NON CRITICAL |
|
||
403 | Pending | Approved | View | Equipment completed 4 year, Repairing cost of equipment is 81% with respect to asset value, Both criteria for RBER Met | 18-10-2023 | 1344755 | Kannur | ELKO | SUB CENTRE | PHC CHELORA | B P Apparatus | NON CRITICAL |
|
||
404 | Pending | Approved | View | Equipment completed 5 year 6 month, repairing cost of equipment is 70%,both criteria for Rber met | 25-03-2024 | 1240111 | Wayanad | ECG Mac | OP | 24X7 PHC Appappara | ECG Machine | CRITICAL |
|
||
405 | Pending | Approved | View | Equipment completed 21 years, repairing cost of equipment is more than 100% of asset value, both criteria for rber met, as per tenderclause 5.3.14.1 we recommend equipment for condemination | 25-09-2024 | 0930194 | Palakkad | NIL | OP | CHC NANNIODE | Sterlizer | NON CRITICAL |
|
||
406 | Pending | Approved | View | VERIFIED | 09-08-2024 | 0410245 | Alappuzha | HEINE | EYE OP | GH ALAPPUZHA | Opthalmoscope | NON CRITICAL |
|
||
407 | Pending | Approved | View | VERIFIED | 27-09-2024 | 0440485 | Alappuzha | PROKAN | LABORATORY | UPHC MANGALAM | Haematology Analyser | CRITICAL |
|
||
408 | Pending | Approved | View | VERIFIED | 19-08-2024 | 0410364 | Alappuzha | MURPHY | PMR | GH ALAPPUZHA | Infrared Lamp | NON CRITICAL |
|
||
409 | Pending | Approved | View | VERIFIED | 10-05-2024 | 0520013 | Kottayam | AccuSure | Casuality | THQH VAIKOM | B P Apparatus | NON CRITICAL |
|
||
410 | Pending | Approved | View | Equipment completed 6 year 2 months, repairimng cost of equipment is more than 100%, Both criteria for rber met | 02-09-2024 | 0840075 | Thrissur | ECG Mac | OP | PHC DESHAMANGALAM | ECG Machine | CRITICAL |
|
||
411 | Pending | Approved | View | VERIFIED | 12-06-2024 | 0321671 | Pathanamthitta | ELKO SURGICALS AND AILLED PRODUCTS | CASUALTY | THQH MALLAPPALLY | Nebuliser | NON CRITICAL |
|
||
412 | Pending | Approved | View | Equipment completed 6 year 7 month repairing cost of equipment is 81% of asset value, both criteria for rber met | 04-01-2025 | 1342580 | Kannur | Rotek Instruments | LABORATORY | PHC EZHOME | Centrifuge | NON CRITICAL |
|
||
413 | Pending | Approved | View | Equipment completed 6 year 5 months with an asset cost of rupees 1353, Quotation not submitted since spares are not available in the market | 24-12-2024 | 1411106 | KASARGODE | LIFE LINE | MALE WARD | GH KASARAGODE | Nebuliser | NON CRITICAL |
|
||
414 | Pending | Approved | View | Equipment completed 7 year 2 months, with an asset cost of rupees 1353.Quotation not submitted since spares are not available in the market | 11-11-2024 | 1342577 | Kannur | NISCOMED | OP | PHC EZHOME | Nebuliser | NON CRITICAL |
|
||
415 | Pending | Approved | View | Equipment outlived 3 year 8 month, Quotation not submitted since spares are not available in the market | 20-01-2024 | 0942289 | Palakkad | MEDICARE | OP | PHC VANIYAMKULAM | B P Apparatus | NON CRITICAL |
|
||
416 | Pending | Approved | View | Equipment completed 4 year 3 month , spare quote not submitted since spare are not available in the market, | 13-12-2023 | 1341869 | Kannur | SAMSO | SUB CENTRE | PHC KADIRUR | Weighing Machine | NON CRITICAL |
|
||
417 | Pending | Approved | View | Equipment completed 5 year 1 month, repairing cost of equipment is 83% of asset value, both criteria for rber met | 31-12-2024 | 0930126 | Palakkad | DIAMOND | CASUALTY | CHC KUZALMANNAM | B P Apparatus | NON CRITICAL |
|
||
418 | Pending | Approved | View | VERIFIED | 16-08-2024 | 0410106 | Alappuzha | Agappe Diagnostics LTD. | LABORATORY | GH ALAPPUZHA | Specific Protein Analyser | NON CRITICAL |
|
||
419 | Pending | Approved | View | Equipment completed 9+ year,Repairing cost of equipment is 78% of asset value,both criteria for RBER Met | 08-05-2024 | 1331244 | Kannur | Whirlpool | pharmacy | CHC PINARAYI | Pharmacy Refrigerator | NON CRITICAL |
|
||
420 | Pending | Approved | View | Equipment outlived 4+ year with an asset cost of Rupees 1875.Quotation not submitted since spares are not available in the market | 19-07-2024 | 0921587 | Palakkad | NA | CASUALTY | THQH PATTAMBI | Nebuliser | NON CRITICAL |
|
||
421 | Pending | Approved | View | Equipment completed 10+ year repairing cost of equipment is 71% of asset value, both criteria for rber met | 08-07-2024 | 1040951 | Malappuram | Whirlpool | pharmacy | 24X7 PHC VALANCHERY | Pharmacy Refrigerator | NON CRITICAL |
|
||
422 | Pending | Approved | View | Equipment outlived 5+ year repairing cost of equipment is 79% of asset value, both criteria for rber met | 27-07-2024 | 1141311 | Kozhikode | Rotek Instruments | LABORATORY | PHC MANIYUR | Centrifuge | NON CRITICAL |
|
||
423 | Pending | Approved | View | Equipment completed 9 year repairing cost of equipment is 72% of asset value, both criteria for rber met | 08-08-2024 | 1010855 | Malappuram | SKANRAY | ICU | DH TIRUR | Multipara Monitor | CRITICAL |
|
||
424 | Pending | Approved | View | Equipment outlived 12 year and 7 months, found problem with main board and xray tube. the oem allengers has confirmed that the xray opel 2.5 is an obselate model, as per tender clause 5.3.14.2 we recommend equipment for condemination | 13-12-2024 | 0810464 | Thrissur | Welcare | DIALYSIS UNIT | GH THRISSUR | Multipara Monitor | CRITICAL |
|
||
425 | Pending | Approved | View | One more Month Required to met Criteria- Digital BP- Special case- | 18-10-2023 | 0542831 | Kottayam | DIAMOND | SUB CENTRE | PHC PALLICKATHODU | B P Apparatus | NON CRITICAL |
|
||
426 | Pending | Approved | View | Verified- Recommended RBER | 30-01-2024 | 0541913 | Kottayam | ANITA INDUSTRIES | OP | 24X7 PHC POONJAR | B P Apparatus | NON CRITICAL |
|
||
427 | Pending | Approved | View | Equipment aged upto 13+ years, Oem informed BPL, INTELIX Is an obselate model as per tender clause 5.3.14.2 we recommend equipment for condemination | 23-04-2024 | 1320835 | Kannur | BPL | X RAY | THQH PAYYANNUR | X Ray Machine Portable | CRITICAL |
|
||
428 | Pending | Approved | View | Equipment completed 3 year,Asset cost of equipment is rupees 2200, Quotation not submitted, since spares are not available in the market a | 20-02-2024 | 1340094 | Kannur | NA | OP | UPHC Porora | B P Apparatus | NON CRITICAL |
|
||
429 | Pending | Approved | View | Equipment completed 5 year, with an asset cost of rupees 2200, quotation not submitted since spares are not available in the market | 16-01-2024 | 0920802 | Palakkad | DIAMOND | MEDICAL WARD | THQH ALATHUR | B P Apparatus | NON CRITICAL |
|
||
430 | Pending | Approved | View | Equipment completed 5 year 6 months with an asset cost of rupees 1353,Quotation not submitted since spares are not available in the market | 20-09-2024 | 0920940 | Palakkad | OMRON HEALTHCARE CO.LTD. | CH WARD | THQH ALATHUR | Nebuliser | NON CRITICAL |
|
||
431 | Pending | Approved | View | VERIFIED | 23-07-2024 | 0440270 | Alappuzha | PROKAN | LABORATORY | UPHC CHERAVALLY | Haematology Analyser | CRITICAL |
|
||
432 | Pending | Approved | View | Equipment completed 6 year and 7 month repairing cost of the equipment is 79% of asset value, both criteria for rber met | 09-12-2024 | 1040447 | Malappuram | ECG Mac | ECG ROOM | PHC PULIKKAL | ECG Machine | CRITICAL |
|
||
433 | Pending | Approved | View | Equipment completed 3+ year with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market | 31-08-2024 | 1342576 | Kannur | OMRON | OP | PHC EZHOME | B P Apparatus | NON CRITICAL |
|
||
434 | Pending | Approved | View | Equipment completed 4 year 9 month, oem declared merilyzer celquant 3 aspire is discontinued , end of life declared model | 09-02-2024 | 1130114 | Kozhikode | Merilyzer CliniQuant | LABORATORY | CHC Kodanchery | Haematology Analyser | CRITICAL |
|
||
435 | Pending | Approved | View | VERIFIED | 15-06-2024 | 0540837 | Kottayam | APC | LABORATORY | 24X7 PHC MADAPPALLY | UPS | NON CRITICAL |
|
||
436 | Pending | Approved | View | Equipment outlived 6+ year, Manufacture cura is not existing as per tender clause 5.3.14.2 we recommend equipment for condemination | 15-05-2024 | 0921140 | Palakkad | Tuscano Equipments PVT.LTD. | DIALYSIS UNIT | THQH MANNARKKAD | Bicarbonate Mixer | NON CRITICAL |
|
||
437 | Pending | Approved | View | verified- Recommended RBER | 14-02-2024 | 0611314 | Idukki | EKCO | PAIN & PALLIATIVE | DH THODUPUZHA | B P Apparatus | NON CRITICAL |
|
||
438 | Pending | Approved | View | Documents and Images verified- Recommended for RBER | 16-10-2023 | 0541842 | Kottayam | SAMSO | OP | 24X7 PHC ERATTUPETTA | Weighing Machine | NON CRITICAL |
|
||
439 | Pending | Approved | View | VERIFIED | 01-07-2024 | 0510494 | Kottayam | Tanita ,BC 601 | NCD CLINIC | GH KOTTAYAM | Body Composition Analyser | NON CRITICAL |
|
||
440 | Pending | Approved | View | Equipment completed 7 year 1 month repairing cost of equipment is 86% of asset value, both criteria for rber met | 31-12-2024 | 1321711 | Kannur | PROKAN | LABORATORY | THQH THALIPARAMBA | Haematology Analyser | CRITICAL |
|
||
441 | Pending | Approved | View | VERIFIED | 21-06-2024 | 0420258 | Alappuzha | janak Healthcare Pvt Ltd | DIALYSIS UNIT | THQH CHERTHALA | ICU Cot | CRITICAL |
|
||
442 | Pending | Approved | View | Equipment completed 7+ yesar repairing cost of equipment is more than 100% both criteria for rber met | 16-03-2024 | 1420471 | KASARGODE | APC | LABORATORY | THQH PANATHADY | UPS | NON CRITICAL |
|
||
443 | Pending | Approved | View | Mercury BP- Documents verified- Recommended RBER | 23-11-2023 | 0120659 | Thiruvananthapuram | ANITA INDUSTRIES | CASUALTY | THQH PARASSALA | B P Apparatus | NON CRITICAL |
|
||
444 | Pending | Approved | View | VERIFIED | 06-09-2024 | 0131006 | Thiruvananthapuram | APC | LABORATORY | CHC VILAPPIL | UPS | NON CRITICAL |
|
||
445 | Pending | Approved | View | VERIFIED | 10-12-2023 | 0510210 | Kottayam | CHAOTUO | OT | GH KOTTAYAM | Sterlizer | NON CRITICAL |
|
||
446 | Pending | Approved | View | Equipment completed 3+ year with an asset cost of rupees 850, Quotation not submitted since spares are not available in the ,market | 25-05-2024 | 0845045 | Thrissur | S.CURE | OP | UPHC V.R PURAM | Weighing Machine | NON CRITICAL |
|
||
447 | Pending | Approved | View | VERIFIED | 15-06-2024 | 0510486 | Kottayam | Yuwell | NCD CLINIC | GH KOTTAYAM | B P Apparatus | NON CRITICAL |
|
||
448 | Pending | Approved | View | Equipment completed 3 year 11 month with an asset cost of rupees 1875, Quotation not submitted since spares are not available in the market | 04-09-2024 | 1310341 | Kannur | Easy Care | CH Ward | DH KANNUR | Nebuliser | NON CRITICAL |
|
||
449 | Pending | Approved | View | VERIFIED | 29-11-2024 | 0620567 | Idukki | MONARCH SURGICAL INSUSTRIES | CASUALTY | THQH KATTAPPANA | Suction Apparatus | NON CRITICAL |
|
||
450 | Pending | Approved | View | Equipment completed 6 year 6 months, repairing cost of equipment is 62% of asset value, both criteria for rber met | 30-12-2024 | 1410654 | KASARGODE | Masimo California | DEIC | DH KANHANGAD | Pulse Oximeter | NON CRITICAL |
|
||
451 | Pending | Approved | View | Equipment completed 4+ year with an asset cost of rupees 1875,Quotation not submitted since spares are not available in the market | 04-07-2024 | 0940997 | Palakkad | Lifecare | OP | PHC MARUTHARODE | Nebuliser | NON CRITICAL |
|
||
452 | Pending | Approved | View | Equipment completed 5+ year with an asset cost of rupees 1800,Quotation not submitted since spares are not available in the market | 27-05-2024 | 1345029 | Kannur | SAMSO | OP | PHC KODIYERI | Weighing Machine | NON CRITICAL |
|
||
453 | Pending | Approved | View | Equipment completed 5 year 6 month, Quotation not submitted since spares are not available in the market | 09-10-2024 | 1345028 | Kannur | SAMSO | OP | PHC KODIYERI | Weighing Machine | NON CRITICAL |
|
||
454 | Pending | Approved | View | Equipment outlived 5+ year,Oem informed through mail that company stopped the production and distribution of mixer and its spares, so spares are not available,As per tender clause 5.3.14.2 we recommend equipment for condemination | 29-05-2024 | 1011305 | Malappuram | Euro Diagnostic Systems PVT.LTD | LABORATORY | DH NILAMBUR | Haemo Mixer | NON CRITICAL |
|
||
455 | Pending | Approved | View | VERIFIED | 19-12-2024 | 0241396 | Kollam | Olex | OP | PHC VILAKKUDY | B P Apparatus | NON CRITICAL |
|
||
456 | Pending | Approved | View | Equipment completyed 5 year,Quotation not submitted since spares are not available in the market | 21-03-2024 | 1420524 | KASARGODE | Easy Care | WARD | THQH PANATHADY | Nebuliser | NON CRITICAL |
|
||
457 | Pending | Approved | View | Flood Damage- Documents verified- Recommended RBER | 30-10-2023 | 0141212 | Thiruvananthapuram | Lifecare | nursing station | UPHC KANNAMMOOLA | B P Apparatus | NON CRITICAL |
|
||
458 | Pending | Approved | View | Equipment completed 5+ year with an asset cost of rupees 850, Quotation not submitted since spares are not available in the market | 16-01-2024 | 0941592 | Palakkad | Unigrip | OP | PHC MUTHALAMADA | Weighing Machine | NON CRITICAL |
|
||
459 | Pending | Approved | View | VERIFIED | 27-06-2024 | 0442350 | Alappuzha | Erba | LABORATORY | PHC ARYAD | Biochemistry Analyzer | CRITICAL |
|
||
460 | Pending | Approved | View | Equipment completed 4 year 5 month with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market | 09-09-2024 | 1343427 | Kannur | OMRON | OP | 24X7 PHC KEEHPALLY | B P Apparatus | NON CRITICAL |
|
||
461 | Pending | Approved | View | Equipment completed 7+ year , repairing cost of equipment is more than 100%,Both criteria for rber met | 19-06-2024 | 0830473 | Thrissur | PROKAN | LABORATORY | CHC MATTATHUR | Haematology Analyser | CRITICAL |
|
||
462 | Pending | Approved | View | Equipment completed 5 year 1 month with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market | 25-10-2024 | 0943007 | Palakkad | EKCO | OP | PHC PUDUKODE | B P Apparatus | NON CRITICAL |
|
||
463 | Pending | Approved | View | VERIFIED | 06-12-2023 | 0114845 | Thiruvananthapuram | ELECTROLUX | BLOOD BANK | WOMEN & CHILDREN HOSPITAL THYCAUD | Pharmacy Refrigerator | NON CRITICAL |
|
||
464 | Pending | Approved | View | Equipment completed 4 year 11 months with an asset cost of rupees 1600, Quotation not submitted since spares are not available in the market | 06-09-2024 | 1341471 | Kannur | SAMSO | OP | 24X7 PHC MATOOL | Weighing Machine | NON CRITICAL |
|
||
465 | Pending | Approved | View | Equipment completed 5 year repairing cost of equipment is 60% of asset value, both criteria for rber met | 30-09-2024 | 1341535 | Kannur | HYKON | LABORATORY | PHC KANKOL ALAPADAMBA | UPS | NON CRITICAL |
|
||
466 | Pending | Approved | View | VERIFIED | 30-04-2024 | 0541338 | Kottayam | APC | LABORATORY | PHC VAZHAPPALLY | UPS | NON CRITICAL |
|
||
467 | Pending | Approved | View | Equipment outlived 6+ year with an asset cost of rupees 2200,Quotation not submitted since spares are not available in the market | 10-02-2024 | 0842466 | Thrissur | OMRON | OP | PHC ARIMBUR | B P Apparatus | NON CRITICAL |
|
||
468 | Pending | Approved | View | VERIFIED | 08-10-2024 | 0443353 | Alappuzha | Robonik | LABORATORY | PHC PATHIYOOR | Biochemistry Analyzer | CRITICAL |
|
||
469 | Pending | Approved | View | Equipment completed 7+ year, repairing cost of equipment is more than 80%, both criteria for rber met | 21-08-2024 | 0912502 | Palakkad | ANITA INDUSTRIES | CH Ward | WOMEN & CHILDREN HOSPITAL PALAKKAD | B P Apparatus | NON CRITICAL |
|
||
470 | Pending | Approved | View | Equipment completed 5+ year with an asset cost of rupees 2200,Quotation not submitted since spares are not available in the market | 21-03-2024 | 1420254 | KASARGODE | Dr Morepen | WARD | THQH NILESWAR | B P Apparatus | NON CRITICAL |
|
||
471 | Pending | Approved | View | VERIFIED | 06-09-2024 | 0112118 | Thiruvananthapuram | KELVINATOR | ICTC LAB | DH NEDUMANGAD | Pharmacy Refrigerator | NON CRITICAL |
|
||
472 | Pending | Approved | View | Equipment out lived 6+ year, Oem cura healthcare not exisiting and spares are not available,As per tenderclause 5.3.14.2 we recommend equipment for condemination | 23-12-2023 | 0812017 | Thrissur | Tuscano Equipments PVT.LTD. | DIALYSIS UNIT | DH WADAKKANCHERRY | Bicarbonate Mixer | NON CRITICAL |
|
||
473 | Pending | Approved | View | VERIFIED | 07-05-2024 | 0421300 | Alappuzha | MEDEC | casulty | THQH THURAVOOR | Suction Apparatus | NON CRITICAL |
|
||
474 | Pending | Approved | View | Equipment completed 5+ year with an asset cost of rupees 2200. Quotation not submitted since spares are not available in the market | 22-02-2024 | 0831104 | Thrissur | OMRON | CASUALTY | CHC VALAPAD | B P Apparatus | NON CRITICAL |
|
||
475 | Pending | Approved | View | Equipment completed 5 year 6 months repairing cost of equipment is 85% of asset value, both criteria for rber met | 04-07-2024 | 0822184 | Thrissur | CONTEC MEDICAL SYSTEMS CO.LTD. | OT | THQH PUTHUKKAD | Pulse Oximeter | NON CRITICAL |
|
||
476 | Pending | Approved | View | Equipment completed 6 year 2 months repairing cost of equipment is more than 100% both criteria for rber met | 05-11-2024 | 0920202 | Palakkad | DELTA | LABORATORY | THQH CHITTUR | UPS | NON CRITICAL |
|
||
477 | Pending | Approved | View | VERIFIED | 31-08-2024 | 0144737 | Thiruvananthapuram | RMS | observation room | PHC OTTOOR | ECG Machine | CRITICAL |
|
||
478 | Pending | Approved | View | VERIFIED EOL | 01-06-2024 | 0710021 | Ernakulam | LY PLANS SYSTEMS | RADIOLOGY | GH ERNAKULAM | OPG (Dental) | NON CRITICAL |
|
||
479 | Pending | Approved | View | EQUIPMENT COMPLETED 4 YEAR 5 MONTHS REPAIRING COST OF EQUIPMENT IS 62% OF ASSET VALUE, BOTH CRITERIA FOR RBER MET | 22-08-2024 | 1342699 | Kannur | IBALL | LABORATORY | PHC MORAZHA | UPS | NON CRITICAL |
|
||
480 | Pending | Approved | View | Equipment completed 5+ year repairing cost of equipment is 66%, both criteria for rber met | 19-08-2024 | 1340041 | Kannur | APC | LABORATORY | PHC MUNDERI | UPS | NON CRITICAL |
|
||
481 | Pending | Approved | View | VERIFIED | 24-07-2024 | 0142139 | Thiruvananthapuram | TRUCHEK | OP | UPHC MUTTADA | B P Apparatus | NON CRITICAL |
|
||
482 | Pending | Approved | View | Equipment completed 5+ year repairing cost of equipment is 76%,Both criteria for Rber met | 03-04-2024 | 0920395 | Palakkad | Hardik Medi Tech | CASUALTY | THQH MANNARKKAD | Spot Light | NON CRITICAL |
|
||
483 | Pending | Approved | View | Equipment completed 4+ year, upon inspection found significant structural damage including bulging and destroyed inner chamber and multiple cracks in the body, the equipment is not safe for using for sterilisation due to these critical defects, | 01-08-2024 | 0912374 | Palakkad | Esteem Industries Inc. | High Dependency Unit(HDU) | WOMEN & CHILDREN HOSPITAL PALAKKAD | Sterlizer | NON CRITICAL |
|
||
484 | Pending | Approved | View | Equipment completed 3 year with an asset cost of rupees 2200, in fully damaged condition becuase of main board and display with flex cable damaged, Quotation not submitted since spares are not available in the market | 23-02-2024 | 0940835 | Palakkad | OMRON HEALTHCARE CO.LTD. | OP | PHC KUMARAMPUTHUR | B P Apparatus | NON CRITICAL |
|
||
485 | Pending | Approved | View | Equipment installed on 23-03-2018, outlived 6 year with an asset cost of rupees 2200.Quotation not submitted since spares are not available in the market | 25-04-2024 | 0930554 | Palakkad | OMRON | OP | CHC KOPPAM | B P Apparatus | NON CRITICAL |
|
||
486 | Pending | Approved | View | Covid supply- No vendor details- Donated- Spare parts not available in market- Verified- Recommended RBER | 21-12-2023 | 0513233 | Kottayam | nill | MEDICAL WARD | GH KANJIRAPALLY | Oxygen Concentrator | NON CRITICAL |
|
||
487 | Pending | Approved | View | Equipment completed 4+ year with an asset cost of rupees 1875, Quotation not submitted since spares are not available in the market | 12-02-2024 | 0830991 | Thrissur | Lifecare | OP | CHC VATANAPPALLY | Nebuliser | NON CRITICAL |
|
||
488 | Pending | Approved | View | VERIFIED | 06-08-2024 | 0110375 | Thiruvananthapuram | Biotech | PMR | GH THIRUVANANTHAPURAM | Ultrasound Therapy Unit | NON CRITICAL |
|
||
489 | Pending | Approved | View | Equipment completed 6+year,repairing cost of equipment is 87% both criteria for Rber met | 02-11-2023 | 1011320 | Malappuram | Kadavil Electro Mechanical Industries | LABORATORY | DH NILAMBUR | Centrifuge | NON CRITICAL |
|
||
490 | Pending | Approved | View | verified | 02-02-2024 | 0541924 | Kottayam | Ready Matic | OP | 24X7 PHC POONJAR | Suction Apparatus | NON CRITICAL |
|
||
491 | Pending | Approved | View | VERIFIED | 12-12-2024 | 0111891 | Thiruvananthapuram | MDE | OT | GH THIRUVANANTHAPURAM | Tourniquet | NON CRITICAL |
|
||
492 | Pending | Approved | View | VERIFIED | 23-07-2024 | 0410658 | Alappuzha | Transasia Biomedical LTD. | LABORATORY | DH CHENGANNUR | ESR Analyser | NON CRITICAL |
|
||
493 | Pending | Approved | View | Equipment completed 5 + year, with an asset cost of rupees 2200, quotation not submitted since spares are not available in the market | 03-11-2023 | 0920915 | Palakkad | Lifecare | LABOUR ROOM | THQH ALATHUR | B P Apparatus | NON CRITICAL |
|
||
494 | Pending | Approved | View | VERIFIED | 22-06-2024 | 0140138 | Thiruvananthapuram | TRUCHECK | OP | UPHC VETTUKAD | B P Apparatus | NON CRITICAL |
|
||
495 | Pending | Approved | View | VERIFIED | 22-06-2024 | 0140136 | Thiruvananthapuram | Lifecare | OP | UPHC VETTUKAD | B P Apparatus | NON CRITICAL |
|
||
496 | Pending | Approved | View | Equipment competleted 5+ year, repairing cost of equipment is more than 71% of asset value,both criteria for rber met | 17-11-2023 | 1311919 | Kannur | Enertech Electronics Pvt.Ltd. | urology | GH THALASSERY | UPS | NON CRITICAL |
|
||
497 | Pending | Approved | View | Year not met- Special case- Document verified- Recommended RBER | 10-11-2023 | 0130153 | Thiruvananthapuram | Lifecare | OP | CHC ANCHUTHENGU | B P Apparatus | NON CRITICAL |
|
||
498 | Pending | Approved | View | VERIFIED | 19-08-2024 | 0410365 | Alappuzha | NISCO | PMR | GH ALAPPUZHA | Infrared Lamp | NON CRITICAL |
|
||
499 | Pending | Approved | View | Equipment completed 4 year 5 months, with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market | 19-10-2024 | 1440315 | KASARGODE | MediTech Industries | OP | PHC PANATHUR | B P Apparatus | NON CRITICAL |
|
||
500 | Pending | Approved | View | Equipment completed 4 year 10 months with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market | 30-11-2024 | 0943390 | Palakkad | Ready Matic | OP | PHC ANANGANADI | B P Apparatus | NON CRITICAL |
|
||
501 | Pending | Approved | View | Both criteria met- Recommended RBER | 25-10-2023 | 0141572 | Thiruvananthapuram | ECG Mac | ECG ROOM | PHC KEEZHATTINGAL | ECG Machine | CRITICAL |
|
||
502 | Pending | Approved | View | VERIFIED | 18-05-2024 | 0311357 | Pathanamthitta | CONTEC MEDICAL SYSTEMS CO.LTD. | ECG ROOM | DH KOZHENCHERRY | ECG Machine | CRITICAL |
|
||
503 | Pending | Approved | View | VERIFIED | 10-06-2024 | 0311431 | Pathanamthitta | B.BRAUN | DIALYSIS UNIT | DH KOZHENCHERRY | Dialysis Machine | CRITICAL |
|
||
504 | Pending | Approved | View | VERIFIED | 02-12-2024 | 0410865 | Alappuzha | BPL | MEDICAL WARD | DH CHENGANNUR | Oxygen Concentrator | NON CRITICAL |
|
||
505 | Pending | Approved | View | Equipment completed 9 year, repairing cost of equipment is 191% of asset value, both criteria for Rber Met | 03-04-2024 | 1110172 | Kozhikode | B.BRAUN | DIALYSIS UNIT | GH KOZHIKODE | Dialysis Machine | CRITICAL |
|
||
506 | Pending | Approved | View | Equipment completed 9+myear, repairing cost of equipment is more than 100%,Both criteria for rber met as per tender clause 5.3.14.1 we recommend equipment for condemination | 21-06-2024 | 1110182 | Kozhikode | B.BRAUN | DIALYSIS UNIT | GH KOZHIKODE | Dialysis Machine | CRITICAL |
|
||
507 | Pending | Approved | View | Equipment completed 9+ year, repairing cost is more than 100%,both criteria for rber met 0 | 07-05-2024 | 1110186 | Kozhikode | B.BRAUN | DIALYSIS UNIT | GH KOZHIKODE | Dialysis Machine | CRITICAL |
|
||
508 | Pending | Approved | View | VERIFIED | 02-04-2024 | 0110011 | Thiruvananthapuram | Fresenius Medical Care AG & Co. | DIALYSIS UNIT | GH THIRUVANANTHAPURAM | Dialysis Machine | CRITICAL |
|
||
509 | Pending | Approved | View | Eqruipment completed 3+ year with an asset cost of rupees 1875 ,Quotation not submitted since spares are not available in the market | 29-01-2024 | 1010567 | Malappuram | TRUCHECK | oncology | DH TIRUR | Nebuliser | NON CRITICAL |
|
||
510 | Pending | Approved | View | VERIFIED | 24-07-2024 | 0711233 | Ernakulam | TRUCHECK | SPECIAL WARD | DH ALUVA | Nebuliser | NON CRITICAL |
|
||
511 | Pending | Approved | View | Equipment outlived 10+ year oem informed that service cost will be higher and some of the spare parts not available currently due to aging, oem schiller recommended upgrading or condem unit | 21-11-2024 | 0911929 | Palakkad | SCHILLER | CATH ICU | DH PALAKKAD | Syringe Pump | CRITICAL |
|
||
512 | Pending | Approved | View | VERIFIED | 27-07-2024 | 0110038 | Thiruvananthapuram | Matrix | BLOOD BANK | GH THIRUVANANTHAPURAM | Card Centrifuge | NON CRITICAL |
|
||
513 | Pending | Approved | View | Equipment completed 12+ year , repairing cost of equipment is 86% of asset value, both criteria for rber met, as per tender clause 5.3.14.1 we recommend equipment for condemination | 21-08-2024 | 0911531 | Palakkad | MAQUET | MICU | DH PALAKKAD | Ventilator | CRITICAL |
|
||
514 | Pending | Approved | View | VERIFIED | 23-03-2024 | 0110059 | Thiruvananthapuram | L&T | MALE WARD | GH THIRUVANANTHAPURAM | Pulse Oximeter | NON CRITICAL |
|
||
515 | Pending | Approved | View | Equipment completed 8 year, repairing cost of equipment is more than 80%,Both criteria for rber met | 03-05-2024 | 1310731 | Kannur | Alps | AUDIOLOGY | DH KANNUR | Audiometer | NON CRITICAL |
|
||
516 | Pending | Approved | View | VERIFIED | 29-10-2023 | 0711498 | Ernakulam | LIFE PLUS | INJECTION ROOM | DH ALUVA | Fetal doppler | NON CRITICAL |
|
||
517 | Pending | Approved | View | VERIFIED | 02-03-2024 | 0411127 | Alappuzha | MEDITECH | MEDICAL WARD | DH MAVELIKARA | Nebuliser | NON CRITICAL |
|
||
518 | Pending | Approved | View | VERIFIED | 23-07-2024 | 0411120 | Alappuzha | Infi-neb | MEDICAL WARD | DH MAVELIKARA | Nebuliser | NON CRITICAL |
|
||
519 | Pending | Approved | View | VERIFIED | 19-08-2024 | 0610447 | Idukki | NIDEK | LABOUR ROOM | DH IDUKKI | Fetal doppler | NON CRITICAL |
|
||
520 | Pending | Approved | View | VERIFIED | 31-08-2024 | 0610448 | Idukki | NIDEK | LABOUR ROOM | DH IDUKKI | Fetal doppler | NON CRITICAL |
|
||
521 | Pending | Approved | View | Equipment completed 13 year 10 months, repairing cost of equipment is 88% of asset value, both criteria for rber met as per tender clause 5.3.14.1 we recommend equipment for condemination | 30-09-2024 | 0811491 | Thrissur | TRISCO | OT | GH IRINJALAKUDA | OT Table | CRITICAL |
|
||
522 | Pending | Approved | View | VERIFIED | 28-03-2024 | 0714035 | Ernakulam | PRAGATHI MEDICAL SYSTEMS | LABOUR ROOM | GH MUVATTUPUZHA | Weighing Machine | NON CRITICAL |
|
||
523 | Pending | Approved | View | VERIFIED | 23-07-2024 | 0411326 | Alappuzha | Faith Innovations | OT | DH MAVELIKARA | Fumigator | NON CRITICAL |
|
||
524 | Pending | Approved | View | Equipment completed 11+ year, Oem informed model is eol notified,as per tender clause 5.3.14.2 we recommend equipment for condemination | 08-07-2024 | 0911853 | Palakkad | Robonik | OP | DISTRICT TB CENTRE PALAKKAD | Biochemistry Analyzer | CRITICAL |
|
||
525 | Pending | Approved | View | VERIFIED | 18-03-2024 | 0411373 | Alappuzha | Supreme Enterprises | OT | DH MAVELIKARA | Suction Apparatus | NON CRITICAL |
|
||
526 | Pending | Approved | View | Equipment completed 12+ year repairing cost of equipment is 73% of asset value, both criteria for rber met | 09-08-2024 | 0911249 | Palakkad | TATA VOLTAS | LABORATORY | DH PALAKKAD | Pharmacy Refrigerator | NON CRITICAL |
|
||
527 | Pending | Approved | View | Equipment completed 3+ year with an asset cost of rupees 1875, Quotation not submitted since spares are not available in the market | 18-11-2024 | 1410573 | KASARGODE | Safe & Touch | FS WARD | DH KANHANGAD | Nebuliser | NON CRITICAL |
|
||
528 | Pending | Approved | View | Equipment outlived 4+ year with an asset cost of rupees 1350.Quotation not submitted since spares are not available in the market | 29-11-2023 | 0810204 | Thrissur | OMRON HEALTHCARE CO.LTD. | CASUALTY | GH THRISSUR | Nebuliser | NON CRITICAL |
|
||
529 | Pending | Approved | View | Unserviceable letter attached- Year met- Recommended for RBER | 11-10-2023 | 0310056 | Pathanamthitta | HYKON | LABORATORY | GH PATHANAMTHITTA | UPS | NON CRITICAL |
|
||
530 | Pending | Approved | View | VERIFIED | 17-12-2024 | 0610380 | Idukki | MEDEC | ENT WARD | DH IDUKKI | Suction Apparatus | NON CRITICAL |
|
||
531 | Pending | Approved | View | VERIFIED | 24-10-2024 | 0512112 | Kottayam | Beurer | LABOUR ROOM | GH CHANGANACHERRY | B P Apparatus | NON CRITICAL |
|
||
532 | Pending | Approved | View | Equipment completed 3+ year with an asset cost of rupees 1353,Quotation not submitted since spares are not available in the market | 08-08-2024 | 0911390 | Palakkad | MEDITECH | CASUALTY | DH PALAKKAD | Nebuliser | NON CRITICAL |
|
||
533 | Pending | Approved | View | VERIFIED | 19-03-2024 | 0310110 | Pathanamthitta | SOCOMEC | RADIOLOGY | GH PATHANAMTHITTA | UPS | NON CRITICAL |
|
||
534 | Pending | Approved | View | Equipment outlived 6+ year, repairing cost of equipment is 68% of asset value, both criteria for rber met | 27-12-2023 | 0810312 | Thrissur | NUMERIC | BLOOD BANK | GH THRISSUR | UPS | NON CRITICAL |
|
||
535 | Pending | Approved | View | Recommended for RBER.End of life by OEM. | 17-10-2023 | 0512142 | Kottayam | ALAN ELECTRONIC SYSTEM PVT.LTD. | main ot | GH CHANGANACHERRY | Cautery | NON CRITICAL |
|
||
536 | Pending | Approved | View | VERIFIED | 31-08-2024 | 0610566 | Idukki | BPL | LABOUR ROOM | DH IDUKKI | Fetal doppler | NON CRITICAL |
|
||
537 | Pending | Approved | View | Equipment completed 5 , found fail calibration and has exhibited disrepancies in volume dispensing which are beyond acceptable tolerance limits for accurate and reliable lab workm therefore we recommend this pipeete for condemination | 15-10-2024 | 0810343 | Thrissur | Matrix | BLOOD BANK | GH THRISSUR | Micro Pippette | NON CRITICAL |
|
||
538 | Pending | Approved | View | Equipment completed 4+ year with an asset cost of rupees 850,Quotation not submitted since spares are not available in the market | 26-06-2024 | 1310850 | Kannur | Dr.Morepen | G1 | DH KANNUR | Weighing Machine | NON CRITICAL |
|
||
539 | Pending | Approved | View | tEquipment condemned from hospital as per order no A6/7297/2023 Dated 26-02-2024 | 22-08-2024 | 0911432 | Palakkad | SAMSO | CASUALTY | DH PALAKKAD | Weighing Machine | NON CRITICAL |
|
||
540 | Pending | Approved | View | VERIFIED | 18-11-2024 | 0610604 | Idukki | Swan | NEW ICU | DH IDUKKI | B P Apparatus | NON CRITICAL |
|
||
541 | Pending | Approved | View | VERIFIED | 18-11-2024 | 0610605 | Idukki | ANITA INDUSTRIES | NEW ICU | DH IDUKKI | B P Apparatus | NON CRITICAL |
|
||
542 | Pending | Approved | View | Equipment completed 12+ year, Oem informed Infusion pump allied is an obselate model, as per tender clause 5.3.14.2 we recommend equipment for condemination | 26-04-2024 | 0810445 | Thrissur | Allied Photographics Limited | MICU | GH THRISSUR | Infusion Pump | CRITICAL |
|
||
543 | Pending | Approved | View | Totaly rusted- Verified- Recommended RBER | 17-01-2024 | 0610914 | Idukki | ELECTROLUX | MALE WARD | DH THODUPUZHA | Pharmacy Refrigerator | NON CRITICAL |
|
||
544 | Pending | Approved | View | Equipment outlived 5+ year,Repairing cost of equipment is more than 100% both criteria for rber met | 09-05-2024 | 1111152 | Kozhikode | Consul Neowatt | RADIOLOGY | WOMEN & CHILDREN HOSPITAL KOZHIKODE | UPS | NON CRITICAL |
|
||
545 | Pending | Approved | View | Equipment completed 7 year 10 month with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market. | 02-01-2025 | 1111214 | Kozhikode | EKCO | CASUALTY | WOMEN & CHILDREN HOSPITAL KOZHIKODE | B P Apparatus | NON CRITICAL |
|
||
546 | Pending | Approved | View | Mercury Spillage- Verified- Recommended RBER | 22-12-2023 | 0511297 | Kottayam | ANITA INDUSTRIES | CASUALTY | GH PALA | B P Apparatus | NON CRITICAL |
|
||
547 | Pending | Approved | View | VERIFIED | 29-06-2024 | 0712042 | Ernakulam | Local purchase | OT | DH ALUVA | Fumigator | NON CRITICAL |
|
||
548 | Pending | Approved | View | Verified- Recommended RBER | 02-01-2024 | 0511301 | Kottayam | DR.Gene | CASUALTY | GH PALA | Weighing Machine | NON CRITICAL |
|
||
549 | Pending | Approved | View | VERIFIED | 28-05-2024 | 0310451 | Pathanamthitta | GE Elpro Medical Systems | CATH LAB | GH PATHANAMTHITTA | Ultrasound Machine | CRITICAL |
|
||
550 | Pending | Approved | View | Equipment completed 3 year 5 months, Quotation not submitted since spares are not available in the market | 03-04-2024 | 0810921 | Thrissur | PHILIPS | OP | GH THRISSUR | Nebuliser | NON CRITICAL |
|
||
551 | Pending | Approved | View | Equipment completed 4 year 2 month with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market | 10-09-2024 | 0810963 | Thrissur | ELKO | WARD 7/8 | GH THRISSUR | B P Apparatus | NON CRITICAL |
|
||
552 | Pending | Approved | View | Equipment completed 4 year 10 months with an asset cost of rupees 1353,Quotation not submitted since spares are not available in the market | 18-09-2024 | 1111366 | Kozhikode | OMRON HEALTHCARE CO.LTD. | WARD 9/11 | WOMEN & CHILDREN HOSPITAL KOZHIKODE | Nebuliser | NON CRITICAL |
|
||
553 | Pending | Approved | View | Equipment covered 7+ year, quotation not submitted since spares are not available in the market | 19-01-2024 | 0912440 | Palakkad | Citizen Systems Japan CO.LTD | Antenatal Ward | WOMEN & CHILDREN HOSPITAL PALAKKAD | B P Apparatus | NON CRITICAL |
|
||
554 | Pending | Approved | View | Equipment completed 5+ year, repairing cost of equipment is 89% of asset value,both criteria for RBER Met. | 11-12-2023 | 0912442 | Palakkad | perfecxa | Antenatal Ward | WOMEN & CHILDREN HOSPITAL PALAKKAD | Fetal doppler | NON CRITICAL |
|
||
555 | Pending | Approved | View | Verified- Recommended RBER | 13-02-2024 | 0611250 | Idukki | ALPHANEB | ISOLATION WARD | DH THODUPUZHA | Nebuliser | NON CRITICAL |
|
||
556 | Pending | Approved | View | VERIFIED | 17-09-2024 | 0410043 | Alappuzha | BPL | ECG ROOM | GH ALAPPUZHA | ECG Machine | CRITICAL |
|
||
557 | Pending | Approved | View | Equipment completed 5+ year with an asset cost of rupees 1353, Quotation not submitted since spares are not available in the market | 11-09-2024 | 1411092 | KASARGODE | READY MIST | Paediatric ward | GH KASARAGODE | Nebuliser | NON CRITICAL |
|
||
558 | Pending | Approved | View | Equipment completed 9 years 7 month Quotation not submitted since spares are not available in the market | 18-10-2024 | 1310941 | Kannur | MEDITECH | Rotary pay ward | DH KANNUR | Nebuliser | NON CRITICAL |
|
||
559 | Pending | Approved | View | Equipment completed 25+ year,repairing cost of equipment is more than 100% both criteria for rber met | 14-06-2024 | 1010433 | Malappuram | HEINE | EYE OP | DH PERINTHALMANNA | Opthalmoscope | NON CRITICAL |
|
||
560 | Pending | Approved | View | VERIFIED | 17-07-2024 | 0720144 | Ernakulam | Datex Ohmeda Inc. | OT | THQH THRIPUNITHURA | Anesthasia Machine | CRITICAL |
|
||
561 | Pending | Approved | View | VERIFIED | 27-05-2024 | 0411756 | Alappuzha | BMS | PHYSIOTHERAPY | LS NOORANAD | Wax Bath | NON CRITICAL |
|
||
562 | Pending | Approved | View | Equipment completed 4 year with an asset cost of rupees 1353, quotation not submitted since spares are not available in the market | 22-03-2024 | 0912486 | Palakkad | Instapro | PICU | WOMEN & CHILDREN HOSPITAL PALAKKAD | Nebuliser | NON CRITICAL |
|
||
563 | Pending | Approved | View | Equipment completed 7+ year with an asset cost of rupees 1353.Quotation not submitted since spares are not available in the market | 17-12-2024 | 1411269 | KASARGODE | LIFE LINE | LABOUR ROOM | GH KASARAGODE | Nebuliser | NON CRITICAL |
|
||
564 | Pending | Approved | View | RESOLVED | 15-07-2024 | 0410212 | Alappuzha | Local purchase | SURGERY OP | GH ALAPPUZHA | Sterlizer | NON CRITICAL |
|
||
565 | Pending | Approved | View | Equipment surpassed its shelf life of 12 years and 6 months,no stock details available at hospital,attested document by store keeper attached, as per tender clause 5.3.14.2 which stipulates manufacture not existing the equipment should evaluate for condemination we exclate these as a special case to tender invite authority | 29-11-2023 | 1411353 | KASARGODE | MEMMERT GmbH | TB CENTER | DISTRICT TB CENTRE KASARAGOD | Hot Air Oven | NON CRITICAL |
|
||
566 | Pending | Approved | View | Equipment completed 9 year 4 month. Ultrasound machiene mindray DP9900PLUS is an obselate model and oem stopped production, as per tender clause 5.3.14.2 we recommend equipment for condemination | 16-12-2024 | 0912330 | Palakkad | MINDRAY | CASUALTY | WOMEN & CHILDREN HOSPITAL PALAKKAD | Ultrasound Machine | CRITICAL |
|
||
567 | Pending | Approved | View | VERIFIED | 07-10-2024 | 0513425 | Kottayam | nill | pediatric | GH KANJIRAPALLY | Weighing Machine | NON CRITICAL |
|
||
568 | Pending | Approved | View | VERIFIED | 24-04-2024 | 0721924 | Ernakulam | SAFE POWER | DIALYSIS UNIT | THQH PERUMBAVOOR | UPS | NON CRITICAL |
|
||
569 | Pending | Approved | View | VERIFIED | 30-10-2024 | 0310906 | Pathanamthitta | MONARCH SURGICAL INSUSTRIES | FEMALE WARD | GH ADOOR | Suction Apparatus | NON CRITICAL |
|
||
570 | Pending | Approved | View | VERIFIED | 19-08-2024 | 0410357 | Alappuzha | NISCO | PMR | GH ALAPPUZHA | Infrared Lamp | NON CRITICAL |
|
||
571 | Pending | Approved | View | VERIFIED | 19-08-2024 | 0410359 | Alappuzha | NISCO | PMR | GH ALAPPUZHA | Infrared Lamp | NON CRITICAL |
|
||
572 | Pending | Approved | View | VERIFIED | 10-06-2024 | 0110192 | Thiruvananthapuram | Energio lite | RADIOLOGY | GH THIRUVANANTHAPURAM | UPS | NON CRITICAL |
|
||
573 | Pending | Approved | View | Equipment completed 5+ year with an asset cost of rupees 1353, Quotation not submitted since spares are not available in the market | 08-08-2024 | 1420226 | KASARGODE | DeVilbiss | WARD | THQH NILESWAR | Nebuliser | NON CRITICAL |
|
||
574 | Pending | Approved | View | VERIFIED | 21-03-2024 | 0110269 | Thiruvananthapuram | OMRON HEALTHCARE CO.LTD. | MICU | GH THIRUVANANTHAPURAM | Nebuliser | NON CRITICAL |
|
||
575 | Pending | Approved | View | VERIFIED | 02-08-2024 | 0110298 | Thiruvananthapuram | Biotech | PMR | GH THIRUVANANTHAPURAM | Ultrasound Therapy Unit | NON CRITICAL |
|
||
576 | Pending | Approved | View | Equipment outlived 5+ year, Oem cura healthcare not existing, as per tender clause 5.3.14.2 we recommend equipment for condemination | 19-01-2024 | 0812018 | Thrissur | Cura | DIALYSIS UNIT | DH WADAKKANCHERRY | Dialyzer Reprocessing Unit | NON CRITICAL |
|
||
577 | Pending | Approved | View | Equipment completed 7+ year with an asset cost of rupees 1353,Quotation not submitted since spares are not available in the market | 20-07-2024 | 0812022 | Thrissur | DeVilbiss | khrw payward | DH WADAKKANCHERRY | Nebuliser | NON CRITICAL |
|
||
578 | Pending | Approved | View | Equipment completed 5 year 6 month, asset value of the equipment is 67815 rupees, where actual book value of the equipment is 47978, the repair cost based on actual book value is 69%, both criteria for rber met | 05-08-2024 | 1420333 | KASARGODE | Biotech | PHYSIOTHERAPY | THQH BEDADKA | Diathermy | NON CRITICAL |
|
||
579 | Pending | Approved | View | VERIFIED | 27-08-2024 | 0620030 | Idukki | INTERACOUSTICS | LABOUR ROOM | THQH ADIMALY | Oto Acoustic Emission Screener (OAE) | NON CRITICAL |
|
||
580 | Pending | Approved | View | VERIFIED | 05-06-2024 | 0721560 | Ernakulam | Datex Ohmeda Inc. | PP UNIT | THQH NORTH PARAVOOR | Anesthasia Machine | CRITICAL |
|
||
581 | Pending | Approved | View | VERIFIED | 24-07-2024 | 0721565 | Ernakulam | MONARCH SURGICAL INSUSTRIES | PP UNIT | THQH NORTH PARAVOOR | Suction Apparatus | NON CRITICAL |
|
||
582 | Pending | Approved | View | VERIFIED | 31-07-2024 | 0722220 | Ernakulam | perfecxa | Labour Ward | THQH KOTHAMANGALAM | Fetal doppler | NON CRITICAL |
|
||
583 | Pending | Approved | View | Equipment Outlived 13 years, repairing cost of equipment is 91.63%, Both criteria for Rber Met | 04-03-2023 | 1311338 | Kannur | ALLENGERS | X RAY | GH THALASSERY | X Ray Machine | CRITICAL |
|
||
584 | Pending | Approved | View | Equipment completed 7 years,Repairing cost of equipment is More than 100%,Both criteria for RBER Met | 23-05-2024 | 1311525 | Kannur | Confident | CASUALTY | GH THALASSERY | Spot Light | NON CRITICAL |
|
||
585 | Pending | Approved | View | Equipment completed 6+ year, repairing cost of equipment is 96.74% of asset value,both criteria for rber met | 09-02-2024 | 0912641 | Palakkad | Welch Allyn | CASUALTY | WOMEN & CHILDREN HOSPITAL PALAKKAD | Otoscope | NON CRITICAL |
|
||
586 | Pending | Approved | View | VERIFIED | 15-07-2024 | 0410537 | Alappuzha | Local purchase | EYE OT | GH ALAPPUZHA | OT Table | CRITICAL |
|
||
587 | Pending | Approved | View | Verified- Recommended RBER | 30-09-2023 | 0110569 | Thiruvananthapuram | Dr.Morepen | CARDIOLOGY OP | GH THIRUVANANTHAPURAM | B P Apparatus | NON CRITICAL |
|
||
588 | Pending | Approved | View | Equipment completed 8 year repairing cost of equipment is 79% of asset value, both criteria for rber met | 03-01-2024 | 0811908 | Thrissur | ZEISS | LABORATORY | TB CENTER THRISSUR | Laboratory Microscope | NON CRITICAL |
|
||
589 | Pending | Approved | View | Verified- OEM Letter- Recommended RBER | 30-12-2023 | 0110584 | Thiruvananthapuram | Appasamy Associates PVT.LTD. | EYE OP | GH THIRUVANANTHAPURAM | Slit Lamp | NON CRITICAL |
|
||
590 | Pending | Approved | View | VERIFIED | 07-08-2024 | 0110614 | Thiruvananthapuram | TRUCHECK | GENERAL WARD | GH THIRUVANANTHAPURAM | Nebuliser | NON CRITICAL |
|
||
591 | Pending | Approved | View | VERIFIED- RECOMMENDED RBER | 22-01-2024 | 0110616 | Thiruvananthapuram | Diamond | GENERAL WARD | GH THIRUVANANTHAPURAM | B P Apparatus | NON CRITICAL |
|
||
592 | Pending | Approved | View | Equipment completed 6+ year repairing cost of equipment is 79% of asset value, both criteria for rber met | 05-09-2024 | 0811909 | Thrissur | Masimo California | OP | DEIC-THRISSUR | Pulse Oximeter | NON CRITICAL |
|
||
593 | Pending | Approved | View | Found crack on the boiler- Service not possible- Image attached- Verified- Recommended RBER | 06-10-2023 | 0110682 | Thiruvananthapuram | Shiva | DENTAL DEPARTMENT | GH THIRUVANANTHAPURAM | Sterlizer | NON CRITICAL |
|
||
594 | Pending | Approved | View | VERIFIED | 16-10-2024 | 0110683 | Thiruvananthapuram | Goley | DENTAL DEPARTMENT | GH THIRUVANANTHAPURAM | Fumigator | NON CRITICAL |
|
||
595 | Pending | Approved | View | VERIFIED | 18-06-2024 | 0110692 | Thiruvananthapuram | BPL | FS WARD | GH THIRUVANANTHAPURAM | B P Apparatus | NON CRITICAL |
|
||
596 | Pending | Approved | View | VERIFIED | 23-08-2024 | 0110758 | Thiruvananthapuram | SUPREME ENTERPRISES | CATH LAB | GH THIRUVANANTHAPURAM | Suction Apparatus | NON CRITICAL |
|
||
597 | Pending | Approved | View | VERIFIED | 10-04-2024 | 0110800 | Thiruvananthapuram | Lifecare | GERIATRIC WARD | GH THIRUVANANTHAPURAM | Nebuliser | NON CRITICAL |
|
||
598 | Pending | Approved | View | Equipment completed 13+ year,repairing cost of equipment is 82% of asset value, both criteria for rber met | 12-03-2024 | 0812426 | Thrissur | RIESTER | EYE OP | DH WADAKKANCHERRY | Opthalmoscope | NON CRITICAL |
|
||
599 | Pending | Approved | View | VERIFIED | 23-08-2024 | 0110825 | Thiruvananthapuram | MONARCH SURGICAL INSUSTRIES | OT | GH THIRUVANANTHAPURAM | Suction Apparatus | NON CRITICAL |
|
||
600 | Pending | Approved | View | VERIFIED | 24-07-2024 | 0722532 | Ernakulam | Dr.Morepen | CASUALTY | THQH NARAKKAL | B P Apparatus | NON CRITICAL |
|
||
601 | Pending | Approved | View | VERIFIED | 16-04-2024 | 0722454 | Ernakulam | Easy Care | PAIN & PALLIATIVE | THQH KOTHAMANGALAM | Portable suction | NON CRITICAL |
|
||
602 | Pending | Approved | View | Equipment completed 14+ year , repairing cost of equipment is 64% of asset value, both criteria for rber met, as per tender clause 5.3.14.1, we recommend equipment for condemination | 19-09-2024 | 0812481 | Thrissur | Nice Neotech Medical Systems PVT.LTD | CH WARD | DH WADAKKANCHERRY | Infant Radiant Warmer | CRITICAL |
|
||
603 | Pending | Approved | View | VERIFIED | 20-03-2024 | 0110845 | Thiruvananthapuram | SUPREME ENTERPRISES | OT | GH THIRUVANANTHAPURAM | Suction Apparatus | NON CRITICAL |
|
||
604 | Pending | Approved | View | Equipment completed 2 year 8 months with an asst cost of rupees 1353, Quotation not submitted since spares are not available in the market | 10-09-2024 | 0812485 | Thrissur | Lifecare | MM WARD | DH WADAKKANCHERRY | Nebuliser | NON CRITICAL |
|
||
605 | Pending | Approved | View | Equipment outlived 9+years, repairing cost of equipment is 86% of asset value, both criteria for RBER met | 21-03-2024 | 0812450 | Thrissur | BIOMEDIX | EYE OP | DH WADAKKANCHERRY | Ocular Ultrasound | NON CRITICAL |
|
||
606 | Pending | Approved | View | VERIFIED | 17-10-2024 | 0722569 | Ernakulam | MEDITECH | OP | THQH NARAKKAL | Nebuliser | NON CRITICAL |
|
||
607 | Pending | Approved | View | VERIFIED | 27-08-2024 | 0640065 | Idukki | BOKANG INSTRUMENTS CO. LTD | SUB CENTRE | PHC UDUMPANCHOLA | B P Apparatus | NON CRITICAL |
|
||
608 | Pending | Approved | View | VERIFIED | 28-09-2024 | 0140088 | Thiruvananthapuram | EKCO | immunization | UPHC MAMBAZHAKKARA | B P Apparatus | NON CRITICAL |
|
||
609 | Pending | Approved | View | One more month required to meet Criteria- Documents verified- Recommended RBER | 31-10-2023 | 0140100 | Thiruvananthapuram | EKCO | immunization | UPHC MAMBAZHAKKARA | B P Apparatus | NON CRITICAL |
|
||
610 | Pending | Approved | View | Equipment completed 11+ year repairing cost of equipment is 79% of asset value, both criteria for rber met | 30-12-2023 | 0811896 | Thrissur | ZEISS | LABORATORY | TB CENTER THRISSUR | Laboratory Microscope | NON CRITICAL |
|
||
611 | Pending | Approved | View | VERIFIED | 24-06-2024 | 0740215 | Ernakulam | Dr Morepen | INJECTION ROOM | 24X7 PHC KODANAD | Nebuliser | NON CRITICAL |
|
||
612 | Pending | Approved | View | Equipment completed 2 year 5 months,with an asset cost of rupees 1600, Quotation not submitted since spares are not available in the market | 06-03-2024 | 0840035 | Thrissur | SAMSO | INJECTION ROOM | UFHC Porkkalengad | Weighing Machine | NON CRITICAL |
|
||
613 | Pending | Approved | View | VERIFIED | 18-06-2024 | 0740226 | Ernakulam | Medigold | DRESSING ROOM | 24X7 PHC KODANAD | Spot Light | NON CRITICAL |
|
||
614 | Pending | Approved | View | VERIFIED | 16-11-2024 | 0140369 | Thiruvananthapuram | ELKO | FWC | PHC KOTTUKAL | B P Apparatus | NON CRITICAL |
|
||
615 | Pending | Approved | View | VERIFIED | 18-07-2024 | 0140268 | Thiruvananthapuram | CE | OP | PHC KOTTUKAL | Weighing Machine | NON CRITICAL |
|
||
616 | Pending | Approved | View | Equipment completed 5 year 2 months with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market | 08-11-2024 | 1340180 | Kannur | EKCO | SUB CENTRE | PHC KALLIASSERY | B P Apparatus | NON CRITICAL |
|
||
617 | Pending | Approved | View | Equipment completed 10 year 2months with an asset cost of rupees 495,repairing cost of equipment is more than 100% of asset value, both criteria for rber met | 19-11-2024 | 1440287 | KASARGODE | NIL | PHYSIOTHERAPY | PHC CHITTARIKAL | Infrared Lamp | NON CRITICAL |
|
||
618 | Pending | Approved | View | VERIFIED | 31-07-2024 | 0740209 | Ernakulam | LABLINE INSTRUMENTS | LABORATORY | 24X7 PHC KODANAD | WATER BATH | NON CRITICAL |
|
||
619 | Pending | Approved | View | VERIFIED | 31-07-2024 | 0740210 | Ernakulam | Labtronics | LABORATORY | 24X7 PHC KODANAD | Colorimeter | NON CRITICAL |
|
||
620 | Pending | Approved | View | Equipment completed 5 year 2 months with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market | 05-11-2024 | 1340063 | Kannur | ANITA INDUSTRIES | SUB CENTRE | PHC MUNDERI | B P Apparatus | NON CRITICAL |
|
||
621 | Pending | Approved | View | verified | 15-11-2023 | 0722675 | Ernakulam | Beurer | CASUALTY | THQH ANGAMALI | B P Apparatus | NON CRITICAL |
|
||
622 | Pending | Approved | View | VERIFIED | 08-08-2024 | 0411777 | Alappuzha | MEDICARE | PHYSIOTHERAPY | LS NOORANAD | B P Apparatus | NON CRITICAL |
|
||
623 | Pending | Approved | View | verified | 27-03-2024 | 0740365 | Ernakulam | DR.MOREPEN | observation room | 24X7 PHC RAYAMANGALAM | B P Apparatus | NON CRITICAL |
|
||
624 | Pending | Approved | View | VERIFIED | 10-10-2024 | 0140452 | Thiruvananthapuram | Safe & Touch | observation room | PHC AMACHAL | Nebuliser | NON CRITICAL |
|
||
625 | Pending | Approved | View | VERIFIED | 30-12-2024 | 0140455 | Thiruvananthapuram | LIFE LINE | observation room | PHC AMACHAL | Nebuliser | NON CRITICAL |
|
||
626 | Pending | Approved | View | VERIFIED | 11-10-2024 | 0140461 | Thiruvananthapuram | SAMSO | observation room | PHC AMACHAL | Weighing Machine | NON CRITICAL |
|
||
627 | Pending | Approved | View | VERIFIED EOL | 27-05-2024 | 0140472 | Thiruvananthapuram | Robonik | LABORATORY | PHC AMACHAL | Semi-Auto Biochemistry Analyzer | NON CRITICAL |
|
||
628 | Pending | Approved | View | VERIFIED | 08-02-2024 | 0540071 | Kottayam | ASIA POWER | LABORATORY | PHC KALLARA | UPS | NON CRITICAL |
|
||
629 | Pending | Approved | View | Manufacture cura healthcare not existing, As per tender clause 5.3.14.2 which stipulates that manufacture not existing the equipment should be evaluated for condemination, we highlighting the tender inviting authority as these as special case | 06-11-2023 | 1340215 | Kannur | NIL | DIALYSIS UNIT | UDC URUVACHAL | Bicarbonate Mixer | NON CRITICAL |
|
||
630 | Pending | Approved | View | Equipment already condemn during KTPL Period on 10/11/2021,with barcode 70133068, with complaint on 148837 | 15-07-2024 | 1340217 | Kannur | NIL | DIALYSIS UNIT | UDC URUVACHAL | UPS | NON CRITICAL |
|
||
631 | Pending | Approved | View | Equipment out lived 6+ year, the oem cura healthcare closed its operations and completly closed down and manufacture not existing as per tender clause 5.3.14..2 we recommend equipment for condmeination | 15-07-2024 | 1340218 | Kannur | Tuscan | DIALYSIS UNIT | UDC URUVACHAL | Dialyzer Reprocessing Unit | NON CRITICAL |
|
||
632 | Pending | Approved | View | Equipment completed 12 year 8 months, oem philips declared end of life for philips, heartstartXL Defibrilator as per tender clause 5.3.14.2 we recommend equipment for condemination | 23-12-2024 | 1312031 | Kannur | PHILIPS | Labour OT | GH THALASSERY | Defibrillator | NON CRITICAL |
|
||
633 | Pending | Approved | View | VERIFIED | 21-10-2024 | 0540087 | Kottayam | Naulakha | OP | PHC KALLARA | B P Apparatus | NON CRITICAL |
|
||
634 | Pending | Approved | View | VERIFIED | 28-09-2024 | 0140616 | Thiruvananthapuram | Lifecare | STORE | PHC KALLIKKAD NEW | Nebuliser | NON CRITICAL |
|
||
635 | Pending | Approved | View | VERIFIED | 23-09-2024 | 0140615 | Thiruvananthapuram | MICROLIFE | STORE | PHC KALLIKKAD NEW | B P Apparatus | NON CRITICAL |
|
||
636 | Pending | Approved | View | VERIFIED | 28-09-2024 | 0140611 | Thiruvananthapuram | Lifecare | observation room | PHC KALLIKKAD NEW | Nebuliser | NON CRITICAL |
|
||
637 | Pending | Approved | View | Two more months required to meet Criteria- Special case- Images and Documents verified-Recommended RBER | 11-11-2023 | 0540108 | Kottayam | NIL | OP | PHC KALLARA | Sterlizer | NON CRITICAL |
|
||
638 | Pending | Approved | View | VERIFIED | 26-07-2024 | 0140776 | Thiruvananthapuram | videocon | PAIN & PALLIATIVE | PHC ANAD | Pharmacy Refrigerator | NON CRITICAL |
|
||
639 | Pending | Approved | View | Verified- Recommended RBER | 25-10-2023 | 0140841 | Thiruvananthapuram | REMI ELEKTROTECHNIK LTD. | LABORATORY | UPHC VATTIYOORKAVU | Centrifuge | NON CRITICAL |
|
||
640 | Pending | Approved | View | Equipemnt completed 11 + year, repairing cost of equipment is more than 100% both criteria for rber met as per tender clause 5.3.14.1 we recommend equipment for condemination | 24-10-2024 | 1420506 | KASARGODE | Beurer | PHYSIOTHERAPY | THQH PANATHADY | Infrared Lamp | NON CRITICAL |
|
||
641 | Pending | Approved | View | VERIFIED | 16-04-2024 | 0730025 | Ernakulam | TRISCO | ISOLATION WARD | CHC RAMAMANGALAM | Suction Apparatus | NON CRITICAL |
|
||
642 | Pending | Approved | View | VERIFIED | 30-10-2024 | 0440214 | Alappuzha | onlineups, bpe | LABORATORY | UPHC Nehrutrophy | UPS | NON CRITICAL |
|
||
643 | Pending | Approved | View | VERIFIED | 30-10-2024 | 0440218 | Alappuzha | NA | LABORATORY | UPHC Nehrutrophy | UPS | NON CRITICAL |
|
||
644 | Pending | Approved | View | verified | 18-01-2024 | 0340038 | Pathanamthitta | Not available | LABORATORY | PHC Chenneerkkara | UPS | NON CRITICAL |
|
||
645 | Pending | Approved | View | VERIFIED | 17-05-2024 | 0720974 | Ernakulam | Labtech Medico PVT.LTD. | LABORATORY | THQH PUTHENVELIKARA | Laboratory Microscope | NON CRITICAL |
|
||
646 | Pending | Approved | View | Equipment completed 4+year, Quotation not submitted since spares are not available in the market | 03-04-2024 | 1340654 | Kannur | Dr Morepen | observation room | PHC ERAMAM KUTTUR | Nebuliser | NON CRITICAL |
|
||
647 | Pending | Approved | View | Equipment installed on 28-08-2019 with an asset cost of rupees 5000.Quotation not submitted since spares are not available in the market. | 13-03-2024 | 1340558 | Kannur | SAMSO | SUB CENTRE | PHC CHIRAKKAL | Weighing Machine | NON CRITICAL |
|
||
648 | Pending | Approved | View | Equipment completed 12+ years,Oem carestream Dry view 5850 laser imager oem informed model is discontinued | 19-07-2024 | 1210127 | Wayanad | Carestream | RADIOLOGY | DH MANANTHAVADY | CR Printer | NON CRITICAL |
|
||
649 | Pending | Approved | View | VERIFIED | 10-06-2024 | 0110770 | Thiruvananthapuram | MONARCH SURGICAL INSUSTRIES | SICU | GH THIRUVANANTHAPURAM | Electrical Suction Unit | NON CRITICAL |
|
||
650 | Pending | Approved | View | Equipment completed 4 year 5 month oem informed model is obselate and endof life issued, As per tender clause 5.3.14.2 we recommend equipment for condemination | 11-12-2024 | 1341220 | Kannur | LABLINE INSTRUMENTS | LABORATORY | UPHC KARIYAD | Incubator | NON CRITICAL |
|
||
651 | Pending | Approved | View | VERIFIED | 09-10-2024 | 0140703 | Thiruvananthapuram | ABE | OP | UPHC PALKULANGARA | ENT Head Light | NON CRITICAL |
|
||
652 | Pending | Approved | View | Equipment completed 5+ year with an asset cost of rupees 1600,Quotation not submitted since spares are not available in the market | 28-11-2023 | 0840502 | Thrissur | SAMSO | OP | PHC PARIYARAM | Weighing Machine | NON CRITICAL |
|
||
653 | Pending | Approved | View | VERIFIED | 18-07-2024 | 0140883 | Thiruvananthapuram | Robonik | LABORATORY | UPHC KALIPPANKULAM | Biochemistry Analyzer | CRITICAL |
|
||
654 | Pending | Approved | View | VERIFIED | 24-04-2024 | 0740935 | Ernakulam | nill | observation room | UPHC KADAVANTHRA | Sterlizer | NON CRITICAL |
|
||
655 | Pending | Approved | View | VERIFIED | 14-08-2024 | 0130044 | Thiruvananthapuram | Infi-neb | OP | CHC IRANIMUTTOM | Nebuliser | NON CRITICAL |
|
||
656 | Pending | Approved | View | Equipment completed 5+ year with an asset cost of rupees 2200,Quotation not submitted since spares are not available in the market | 07-02-2024 | 0840563 | Thrissur | OMRON | OP | PHC NENMANIKARA | B P Apparatus | NON CRITICAL |
|
||
657 | Pending | Approved | View | Equipment completed 5 year 9 months, oem owgels oxymed model is obselated and spare are not available, as per tender clause 5.3.14.2 we recommend equipment for condemination | 23-10-2024 | 0840560 | Thrissur | Lifecare | OP | PHC NENMANIKARA | Nebuliser | NON CRITICAL |
|
||
658 | Pending | Approved | View | Equipment completed 5+ year with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market | 23-08-2024 | 1341133 | Kannur | OMRON | OP | 24X7 PHC PATTIAM | B P Apparatus | NON CRITICAL |
|
||
659 | Pending | Approved | View | Equipment completed 14+ year, repairing cost of equipment is more than 100%,both criteria for rber met | 17-07-2024 | 1341274 | Kannur | PARAS ENTERPRISES | OP | FHC Kottiyoor | Sterlizer | NON CRITICAL |
|
||
660 | Pending | Approved | View | Documents Verified- Criteria met- Recommended RBER | 16-11-2023 | 0130075 | Thiruvananthapuram | Labtronics | LABORATORY | CHC POONTHURA | Colorimeter | NON CRITICAL |
|
||
661 | Pending | Approved | View | Equipment completed 5 year 5 month,with an asset cost of rupees 1600.Quotation not submitted since spares are not available in the market | 18-12-2024 | 1140418 | Kozhikode | SAMSO | OP | GRD NGO Quatters | Weighing Machine | NON CRITICAL |
|
||
662 | Pending | Approved | View | Equipment completed 7 year 6 months, repairing cost of equipment is 64% of asset value, both criteria for rber met | 01-10-2024 | 1030007 | Malappuram | APC | LABORATORY | CHC Kottakkal | UPS | NON CRITICAL |
|
||
663 | Pending | Approved | View | VERIFIED | 09-10-2024 | 0640696 | Idukki | LIFE LINE | immunization | UPHC Kattappana | B P Apparatus | NON CRITICAL |
|
||
664 | Pending | Approved | View | VERIFIED | 19-07-2024 | 0640703 | Idukki | LIFE LINE | OP | UPHC Kattappana | B P Apparatus | NON CRITICAL |
|
||
665 | Pending | Approved | View | Equipment completed 7+ year, repairing cost of equipment is 70% of asset value, both criteria for rber met | 26-06-2024 | 1341348 | Kannur | BPE | LABORATORY | UPHC KOLASSERY | UPS | NON CRITICAL |
|
||
666 | Pending | Approved | View | Equipment completed 3 year 7 month , Quotation not submitted since spares are not avialble in the market | 30-09-2024 | 0840520 | Thrissur | nill | OP | PHC ANDATHODE | B P Apparatus | NON CRITICAL |
|
||
667 | Pending | Approved | View | Equipment completed 5 year 9 months the said model owgels oxymed 5lpm was an obselote model and spares were mot available | 23-10-2024 | 0840650 | Thrissur | nill | PAIN & PALLIATIVE | PHC ANDATHODE | Oxygen Concentrator | NON CRITICAL |
|
||
668 | Pending | Approved | View | Verified- Recommended for RBER | 16-12-2023 | 0540462 | Kottayam | SAMSO | OP | PHC THIRUVARPU | Weighing Machine | NON CRITICAL |
|
||
669 | Pending | Approved | View | Verified- Recommended RBER | 04-01-2024 | 0540460 | Kottayam | SAMSO | OP | PHC THIRUVARPU | Weighing Machine | NON CRITICAL |
|
||
670 | Pending | Approved | View | Equipment completed 5 year 11 months, oem informed Owgels,Oxymed -om501 is an obselate model oxygen concentrator, as per tender clause 5.3.14.2 we recommend equipment for condemination | 21-12-2024 | 0840659 | Thrissur | Oxymed | PAIN & PALLIATIVE | PHC ANDATHODE | Oxygen Concentrator | NON CRITICAL |
|
||
671 | Pending | Approved | View | Equipment completed 4+ year with an asset cost of rupees 1323, Quotation not submitted since spares are not available in market | 27-08-2024 | 0840653 | Thrissur | ELKO | PAIN & PALLIATIVE | PHC ANDATHODE | B P Apparatus | NON CRITICAL |
|
||
672 | Pending | Approved | View | Equipment completed 7 years 9 months repairing cost of equipment is 78% of asset value both criteria for rber met | 16-12-2024 | 1341629 | Kannur | Shiva | OP | PHC CHERUKUNNU | Sterlizer | NON CRITICAL |
|
||
673 | Pending | Approved | View | VERIFIED | 17-08-2024 | 0412116 | Alappuzha | Comfo Care | CASUALTY | WOMEN & CHILDREN HOSPITAL ALAPPUZHA | Nebuliser | NON CRITICAL |
|
||
674 | Pending | Approved | View | EQUIPMENT COMPLETED 5 YEAR 6 MONTHS WITH AN ASSET COST OF RUPEES 1875, QUOTATION NOT SUBMITTED SINCE SPARES ARE NOT AVAILABLE IN THE MARKET | 30-08-2024 | 1410759 | KASARGODE | noren | pediatric | DH KANHANGAD | Nebuliser | NON CRITICAL |
|
||
675 | Pending | Approved | View | EQUIPMENT COMPLETED 5+ YEAR WITH AN ASSET COST OF RUPEES 1353.QUOTATION NOT SUBMITTED SINCE SPARES ARE NOT AVAILABLE IN THE MARKET | 03-09-2024 | 0840870 | Thrissur | Lifecare | OP | PHC VELUR | Nebuliser | NON CRITICAL |
|
||
676 | Pending | Approved | View | Equipment outlived 5+ year with an asset cost of rupees 1353.Quotation not submitted since spares are not available in the market | 24-04-2024 | 0840876 | Thrissur | Lifecare | OP | PHC VELUR | Nebuliser | NON CRITICAL |
|
||
677 | Pending | Approved | View | VERIFIED | 23-09-2024 | 0111520 | Thiruvananthapuram | HYKON | IRL | State TB Centre | UPS | NON CRITICAL |
|
||
678 | Pending | Approved | View | VERIFIED | 04-07-2024 | 0411815 | Alappuzha | DRAGER | SNCU | SNCU VANDANAM | Bilirubinometer | NON CRITICAL |
|
||
679 | Pending | Approved | View | VERIFIED | 10-06-2024 | 0111446 | Thiruvananthapuram | Rotek Instruments | MOLECULAR DIAGONOSTIC SECTION | STATE PUBLIC HEALTH LABORATORY | Centrifuge | NON CRITICAL |
|
||
680 | Pending | Approved | View | VERIFIED | 31-05-2024 | 0311929 | Pathanamthitta | LABOPETTE | SEROLOGY LAB | PH LAB KOZHENCHERRY | Micro Pippette | NON CRITICAL |
|
||
681 | Pending | Approved | View | VERIFIED | 21-10-2024 | 0540808 | Kottayam | Yuwell | OP | 24X7 PHC MADAPPALLY | Oxygen Concentrator | NON CRITICAL |
|
||
682 | Pending | Approved | View | Equipment completed 3+ year with an asset cost of rupees 1600, Quotation not submitted since spares are not available in the market | 22-08-2024 | 1141649 | Kozhikode | SAMSO | OP | PHC NARIPATTA | Weighing Machine | NON CRITICAL |
|
||
683 | Pending | Approved | View | VERIFIED | 06-04-2024 | 0210624 | Kollam | Whirlpool | STORE | DH KOLLAM | Pharmacy Refrigerator | NON CRITICAL |
|
||
684 | Pending | Approved | View | VERIFIED | 31-08-2024 | 0111466 | Thiruvananthapuram | KELVINATOR | MOLECULAR DIAGONOSTIC SECTION | STATE PUBLIC HEALTH LABORATORY | Pharmacy Refrigerator | NON CRITICAL |
|
||
685 | Pending | Approved | View | VERIFIED | 25-04-2024 | 0722833 | Ernakulam | LONGFIAN | PAIN & PALLIATIVE | THQH ANGAMALI | Oxygen Concentrator | NON CRITICAL |
|
||
686 | Pending | Approved | View | VERIFIED | 10-06-2024 | 0111323 | Thiruvananthapuram | MICRO | MEDIA | STATE PUBLIC HEALTH LABORATORY | Ph Meter | NON CRITICAL |
|
||
687 | Pending | Approved | View | VERIFIED | 31-08-2024 | 0540733 | Kottayam | nill | OP | 24X7 PHC MUNDANKUNNU | Oxygen Concentrator | NON CRITICAL |
|
||
688 | Pending | Approved | View | Equipment completed 4+ year repairing cost of equipment is 90%,Both criteria for RBER Met. | 25-10-2023 | 1320079 | Kannur | EKCO | SUB CENTRE | THQH PERAVOOR | B P Apparatus | NON CRITICAL |
|
||
689 | Pending | Approved | View | VERIFIED | 26-10-2023 | 0730176 | Ernakulam | REMI ELEKTROTECHNIK LTD. | LABORATORY | CHC VENGOLA | Centrifuge | NON CRITICAL |
|
||
690 | Pending | Approved | View | Equipment outlived 3+ year with an asset cost of rupees 2200,Quotation not submitted since spares are not available in the market | 16-02-2024 | 0830059 | Thrissur | TRUCHECK | PAIN & PALLIATIVE | CHC KADAPURAM | B P Apparatus | NON CRITICAL |
|
||
691 | Pending | Approved | View | Equipment completed 4+ year with an asset cost of rupees 2200.Quotation not submitted since spares are not available in the market | 16-02-2024 | 0830061 | Thrissur | TRUCHECK | PAIN & PALLIATIVE | CHC KADAPURAM | B P Apparatus | NON CRITICAL |
|
||
692 | Pending | Approved | View | VERIFIED | 17-04-2024 | 0111677 | Thiruvananthapuram | LABOMED | MLT | STATE PUBLIC HEALTH LABORATORY | Laboratory Microscope | NON CRITICAL |
|
||
693 | Pending | Approved | View | Equipment completed 3+ year repairing cost of equipment is more than 100%, both criteria for rber met | 05-09-2024 | 1320342 | Kannur | BPL | LABOUR ROOM | THQH PERAVOOR | Fetal doppler | NON CRITICAL |
|
||
694 | Pending | Approved | View | VERIFIED | 04-04-2024 | 0730197 | Ernakulam | LIFE LINE | PALIATIVE | CHC VENGOLA | B P Apparatus | NON CRITICAL |
|
||
695 | Pending | Approved | View | VERIFIED | 02-11-2024 | 0412383 | Alappuzha | LIFE LINE | Antenatal Ward | WOMEN & CHILDREN HOSPITAL ALAPPUZHA | Fetal doppler | NON CRITICAL |
|
||
696 | Pending | Approved | View | VERIFIED | 13-05-2024 | 0111760 | Thiruvananthapuram | DIAMOND | WARD 2 | GH THIRUVANANTHAPURAM | B P Apparatus | NON CRITICAL |
|
||
697 | Pending | Approved | View | VERIFIED | 10-05-2024 | 0111777 | Thiruvananthapuram | Lifecare | FEVER WARD | GH THIRUVANANTHAPURAM | Nebuliser | NON CRITICAL |
|
||
698 | Pending | Approved | View | Recommended RBER | 22-01-2024 | 0111780 | Thiruvananthapuram | READY MIST | WARD 9 | GH THIRUVANANTHAPURAM | Nebuliser | NON CRITICAL |
|
||
699 | Pending | Approved | View | RECOMMENDED RBER | 17-01-2024 | 0111805 | Thiruvananthapuram | Lifecare | WARD 7 | GH THIRUVANANTHAPURAM | Suction Apparatus | NON CRITICAL |
|
||
700 | Pending | Approved | View | VERIFIED | 19-10-2024 | 0141051 | Thiruvananthapuram | Robonik | LABORATORY | UPHC THRIKKANNAPURAM | Biochemistry Analyzer | CRITICAL |
|
||
701 | Pending | Approved | View | Equipment completed 2 year 8 month with an asset cost of rupees 1353.Quotation not submitted since spares are not available in the market | 20-12-2024 | 1320753 | Kannur | Easy Care | FEMALE WARD | THQH PAYANGADI | Nebuliser | NON CRITICAL |
|
||
702 | Pending | Approved | View | VERIFIED | 21-05-2024 | 0222321 | Kollam | Lifecare | MM WARD | THQH SASTHAMKOTTA | Nebuliser | NON CRITICAL |
|
||
703 | Pending | Approved | View | Equipment completed 13+ year, Repairing cost of equipment is 66% of asset value both criteria for rber met | 19-06-2024 | 0830638 | Thrissur | LG | pharmacy | CHC CHERPU | Pharmacy Refrigerator | NON CRITICAL |
|
||
704 | Pending | Approved | View | VERIFIED | 26-10-2024 | 0641182 | Idukki | S Cure | OP | PHC DEVIYAR COLONY | B P Apparatus | NON CRITICAL |
|
||
705 | Pending | Approved | View | VERIFIED | 26-10-2024 | 0641183 | Idukki | Dr.Odin | OP | PHC DEVIYAR COLONY | B P Apparatus | NON CRITICAL |
|
||
706 | Pending | Approved | View | VERIFIED | 04-04-2024 | 0111869 | Thiruvananthapuram | MONARCH SURGICAL INSUSTRIES | OT | GH THIRUVANANTHAPURAM | Suction Apparatus | NON CRITICAL |
|
||
707 | Pending | Approved | View | VERIFIED | 04-04-2024 | 0111906 | Thiruvananthapuram | MONARCH SURGICAL INSUSTRIES | OT | GH THIRUVANANTHAPURAM | Suction Apparatus | NON CRITICAL |
|
||
708 | Pending | Approved | View | Equipment completed 4 year 3months,with an asset cost of rupees 4990.Quotation not submitted since spares are not available in the market | 29-08-2024 | 1320280 | Kannur | Sknol | FEMALE WARD | THQH PERAVOOR | Weighing Machine | NON CRITICAL |
|
||
709 | Pending | Approved | View | Equipment completed 4 year 5 month, Quotation not submitted since spares are not available in the market | 13-12-2023 | 1341575 | Kannur | SAMSO | SUBCENTER | PHC KANKOL ALAPADAMBA | Weighing Machine | NON CRITICAL |
|
||
710 | Pending | Approved | View | Equipment installed on 14-10-1998 and aged upto 25+ year, Repairing cost of equipment is more than 100% of asset value, Both criteria for RBER Met | 25-03-2024 | 1210566 | Wayanad | Wipro GE Medical Systems Ltd. | X RAY | DH MANANTHAVADY | X Ray Machine | CRITICAL |
|
||
711 | Pending | Approved | View | Equipment completed 3+ year with an asset cost of rupees 1600, Quotation not submitted since spares are not available in the market | 01-07-2024 | 1210571 | Wayanad | BIO PLUS | CH WARD | DH MANANTHAVADY | Portable suction | NON CRITICAL |
|
||
712 | Pending | Approved | View | Equipment completed 5 year 7 months with an asset cost of rupees 1353, Quotation not submitted since spares are not available in the market | 25-11-2024 | 1210625 | Wayanad | TRUCHECK | injuction room | DH MANANTHAVADY | Nebuliser | NON CRITICAL |
|
||
713 | Pending | Approved | View | VERIFIED | 07-12-2024 | 0112131 | Thiruvananthapuram | LABY | LABORATORY | DH NEDUMANGAD | Centrifuge | NON CRITICAL |
|
||
714 | Pending | Approved | View | VERIFIED | 29-11-2024 | 0640752 | Idukki | NIL | PP UNIT | 24X7 PHC KANCHIYAR | Weighing Machine | NON CRITICAL |
|
||
715 | Pending | Approved | View | Verified- Recommended RBER | 17-02-2024 | 0640755 | Idukki | NIL | DUTY ROOM OP | 24X7 PHC KANCHIYAR | B P Apparatus | NON CRITICAL |
|
||
716 | Pending | Approved | View | Equipment completed 2+ year,the supplier detail is unavailable from end user as the equipment was donated without manufacture details during covid pandemic | 22-07-2024 | 0830319 | Thrissur | NIL | FEMALE WARD | CHC MALA | Oxygen Concentrator | NON CRITICAL |
|
||
717 | Pending | Approved | View | Equipment outlived 10 year, end of life declared model, as per tender clause 5.3.14.2 we recommend equipment for condemination | 28-02-2024 | 1210668 | Wayanad | Appasamy Associates PVT.LTD. | EYE OP | DH MANANTHAVADY | OT Table | CRITICAL |
|
||
718 | Pending | Approved | View | VERIFIED | 24-09-2024 | 0112231 | Thiruvananthapuram | Getidy | DENTAL DEPARTMENT | DH NEDUMANGAD | Sterlizer | NON CRITICAL |
|
||
719 | Pending | Approved | View | VERIFIED | 08-08-2024 | 0112287 | Thiruvananthapuram | Biotech | PHYSIOTHERAPY | DH NEDUMANGAD | Ultrasound Therapy Unit | NON CRITICAL |
|
||
720 | Pending | Approved | View | VERIFIED | 24-09-2024 | 0112285 | Thiruvananthapuram | BMS | PHYSIOTHERAPY | DH NEDUMANGAD | TENS | NON CRITICAL |
|
||
721 | Pending | Approved | View | Equipment completed 3+ year, oem informed that oxymed owgels om 501 is discontinued model and spares are not available,in the market as per tender clause5.3.14.2 we recommend equipment for condemination | 04-07-2024 | 1321050 | Kannur | Oxymed | WARD | THQH IRITTY | Oxygen Concentrator | NON CRITICAL |
|
||
722 | Pending | Approved | View | Equipment completed 3 year and 3 month, quotation not submitted because spares are not available in the market | 29-12-2023 | 1321331 | Kannur | Lifecare | WARD | THQH IRITTY | Nebuliser | NON CRITICAL |
|
||
723 | Pending | Approved | View | Equipment outlived 8 years with an asset cost of rupees791.Calibration of said equipment has been failed due to damages in spares such as rotating key, numbers and rings, quotation of spares not submitted since spares are not available in the market | 18-01-2024 | 0830740 | Thrissur | DEEP VISION | LABORATORY | CHC VELLANIKKARA | Micro Pippette | NON CRITICAL |
|
||
724 | Pending | Approved | View | VERIFIED | 10-04-2024 | 0112397 | Thiruvananthapuram | PHILIPS | pediatric | DH NEDUMANGAD | Nebuliser | NON CRITICAL |
|
||
725 | Pending | Approved | View | Equipment completed 11+ year with an asset cost of rupees 1353, Quotation not submitted since spares are not available in the market | 24-06-2024 | 0830757 | Thrissur | Easy Care | CASUALTY | CHC PUTHENCHIRA | Nebuliser | NON CRITICAL |
|
||
726 | Pending | Approved | View | Equipment completed 10 year 1 months repairing cost of equipment is 71% of asset value,both criteria for rber met | 13-11-2024 | 0830466 | Thrissur | REMI ELEKTROTECHNIK LTD. | LABORATORY | CHC MATTATHUR | Centrifuge | NON CRITICAL |
|
||
727 | Pending | Approved | View | Equipment completed 4 year 11 months repiring cost of equipment is 64%,Both criteria for rber met | 31-08-2024 | 1321503 | Kannur | Blue Star | pharmacy | THQH KOOTHUPARAMBA | Stabilizer | NON CRITICAL |
|
||
728 | Pending | Approved | View | verified | 29-08-2024 | 0240195 | Kollam | PROKAN | LABORATORY | UPHC Uliyakovil | Haematology Analyser | CRITICAL |
|
||
729 | Pending | Approved | View | VERIFIED | 26-02-2024 | 0731113 | Ernakulam | OMRON | OP | CHC MALIPPURAM | B P Apparatus | NON CRITICAL |
|
||
730 | Pending | Approved | View | VERIFIED | 17-10-2024 | 0141155 | Thiruvananthapuram | Infi | nursing station | UPHC NANTHANCODE | B P Apparatus | NON CRITICAL |
|
||
731 | Pending | Approved | View | VERIFIED | 02-04-2024 | 0111929 | Thiruvananthapuram | ANITA INDUSTRIES | speciality op | GH THIRUVANANTHAPURAM | B P Apparatus | NON CRITICAL |
|
||
732 | Pending | Approved | View | VERIFIED | 12-07-2024 | 0220430 | Kollam | Lifecare | MM WARD | THQH KARUNAGAPPALLY | Nebuliser | NON CRITICAL |
|
||
733 | Pending | Approved | View | Equipment completed 7+ year, repairing cost of equipment is 68% of asset value, both criteria for rber met | 18-07-2024 | 0830815 | Thrissur | IBALL | LABORATORY | CHC PUTHENCHIRA | UPS | NON CRITICAL |
|
||
734 | Pending | Approved | View | Equipment outlived 5+ year with an asset cost of rupees 1875, Quotation not submitted since spares are not available in the market | 02-05-2024 | 1121279 | Kozhikode | Lifecare | CASUALTY | THQH NADAPURAM | Nebuliser | NON CRITICAL |
|
||
735 | Pending | Approved | View | VERIFIED | 05-04-2024 | 0120438 | Thiruvananthapuram | Local purchase | DENTAL DEPARTMENT | THQH PARASSALA | Sterlizer | NON CRITICAL |
|
||
736 | Pending | Approved | View | VERIFIED | 23-08-2024 | 0120445 | Thiruvananthapuram | TRISCO | DENTAL DEPARTMENT | THQH PARASSALA | Dental Chair | NON CRITICAL |
|
||
737 | Pending | Approved | View | VERIFIED | 05-07-2024 | 0220458 | Kollam | Hardik Medi Tech | OT | THQH KARUNAGAPPALLY | Spot Light | NON CRITICAL |
|
||
738 | Pending | Approved | View | VERIFIED | 10-07-2024 | 0120487 | Thiruvananthapuram | SUPREME ENTERPRISES | OT | THQH PARASSALA | Suction Apparatus | NON CRITICAL |
|
||
739 | Pending | Approved | View | VERIFIED | 17-04-2024 | 0220478 | Kollam | marathon | DENTAL DEPARTMENT | THQH KARUNAGAPPALLY | Micromotor | NON CRITICAL |
|
||
740 | Pending | Approved | View | VERIFIED | 10-08-2024 | 0120599 | Thiruvananthapuram | Local purchase | ENT OP | THQH PARASSALA | ENT Head Light | NON CRITICAL |
|
||
741 | Pending | Approved | View | VERIFIED | 10-08-2024 | 0120595 | Thiruvananthapuram | Local purchase | ENT OP | THQH PARASSALA | ENT Head Light | NON CRITICAL |
|
||
742 | Pending | Approved | View | Equipment completed 6 year 10 months, repairing cost of equipment is more than 100% both criteria for rber met | 04-01-2025 | 1120775 | Kozhikode | nill | OT | THQH KUTTIADI | UPS | NON CRITICAL |
|
||
743 | Pending | Approved | View | VERIFIED | 12-06-2024 | 0321540 | Pathanamthitta | PHILIPS | WARD | THQH MALLAPPALLY | Nebuliser | NON CRITICAL |
|
||
744 | Pending | Approved | View | VERIFIED | 31-07-2024 | 0741769 | Ernakulam | SAMSO | observation room | 24X7 PHC KADAVOOR | Weighing Machine | NON CRITICAL |
|
||
745 | Pending | Approved | View | Equipment completed 7+ year,supplier details are not available at hospital because equipment was donated item,( china make equipment) due to the lack of supplier and spare information we request tender inviting authority to consider the equipment for condemnination under tender clause 5.3.14.2 | 30-08-2024 | 0841449 | Thrissur | nill | OP | PHC POYYA | Oxygen Concentrator | NON CRITICAL |
|
||
746 | Pending | Approved | View | Equipment outlived 4+ year with an asset cost of rupees 1353. Quotation not submitted since spares are not available in the market | 17-04-2024 | 1120643 | Kozhikode | Beurer | casulity | THQH KUTTIADI | Nebuliser | NON CRITICAL |
|
||
747 | Pending | Approved | View | Equipment completed 4 years with an asset cost of rupees 1600.Quotation not submitted since spares are not available in the market | 29-05-2024 | 1030222 | Malappuram | SAMSO | IP | CHC Valavannur | Weighing Machine | NON CRITICAL |
|
||
748 | Pending | Approved | View | VERIFIED | 18-07-2024 | 0420025 | Alappuzha | NIDEK | CASUALTY | THQH CHERTHALA | Nebuliser | NON CRITICAL |
|
||
749 | Pending | Approved | View | VERIFIED | 18-07-2024 | 0420027 | Alappuzha | OPTIMA | CASUALTY | THQH CHERTHALA | Nebuliser | NON CRITICAL |
|
||
750 | Pending | Approved | View | VERIFIED | 17-04-2024 | 0420052 | Alappuzha | PHILIPS | CASUALTY | THQH CHERTHALA | Defibrillator | NON CRITICAL |
|
||
751 | Pending | Approved | View | Equipment completed 6 year 10 months with an asset cost of rupees 1353, Quotation not submitted since spares are not available in the market | 27-11-2024 | 1321665 | Kannur | BPL | OP | THQH KOOTHUPARAMBA | Nebuliser | NON CRITICAL |
|
||
752 | Pending | Approved | View | Equipment completed 4 year 8 months with an asset cost of rupees 1353, Quotation not submitted since spares are not available in the market | 20-09-2024 | 1321666 | Kannur | BPL | OP | THQH KOOTHUPARAMBA | Nebuliser | NON CRITICAL |
|
||
753 | Pending | Approved | View | Equipment completed 6 year 10 months with an asset cost of rupees 2200, Quotation not submitted since spares are not avaialble in the market | 27-12-2024 | 0841546 | Thrissur | Med Vision | OP | PHC KARALAM | B P Apparatus | NON CRITICAL |
|
||
754 | Pending | Approved | View | VERIFIED | 23-08-2024 | 0112914 | Thiruvananthapuram | Lifecare | GERIATRIC WARD | GH NEYYATTINKARA | Nebuliser | NON CRITICAL |
|
||
755 | Pending | Approved | View | VERIFIED | 17-07-2024 | 0541850 | Kottayam | Robonik | LABORATORY | 24X7 PHC ERATTUPETTA | Biochemistry Analyzer | CRITICAL |
|
||
756 | Pending | Approved | View | VERIFIED | 25-04-2024 | 0420064 | Alappuzha | Ishnee | WARD 5 | THQH CHERTHALA | Nebuliser | NON CRITICAL |
|
||
757 | Pending | Approved | View | Flood Damage- Unserviceable- Verified- Recommended RBAR | 30-10-2023 | 0141216 | Thiruvananthapuram | Infi | nursing station | UPHC KANNAMMOOLA | B P Apparatus | NON CRITICAL |
|
||
758 | Pending | Approved | View | Flood Damage- Verified- Recommended RBER | 30-10-2023 | 0141215 | Thiruvananthapuram | OMRON HEALTHCARE CO.LTD. | nursing station | UPHC KANNAMMOOLA | Nebuliser | NON CRITICAL |
|
||
759 | Pending | Approved | View | Flood Damage- Verified- Recommended RBER | 30-10-2023 | 0141214 | Thiruvananthapuram | Safe & Touch | nursing station | UPHC KANNAMMOOLA | Nebuliser | NON CRITICAL |
|
||
760 | Pending | Approved | View | VERIFIED | 13-11-2024 | 0141221 | Thiruvananthapuram | NISCOMED | nursing station | UPHC KANNAMMOOLA | Fetal doppler | NON CRITICAL |
|
||
761 | Pending | Approved | View | VERIFIED | 23-09-2024 | 0141226 | Thiruvananthapuram | Labtronics | LABORATORY | UPHC KANNAMMOOLA | Colorimeter | NON CRITICAL |
|
||
762 | Pending | Approved | View | Equipment completed 4+ year with an asset cost of rupees 1323, Quotation not submitted since spares are not available in the market | 24-01-2024 | 0830500 | Thrissur | nill | OP | CHC MATTATHUR | B P Apparatus | NON CRITICAL |
|
||
763 | Pending | Approved | View | VERIFIED | 21-08-2024 | 0420079 | Alappuzha | SUPREME ENTERPRISES | WARD 5 | THQH CHERTHALA | Suction Apparatus | NON CRITICAL |
|
||
764 | Pending | Approved | View | VERIFIED | 23-08-2024 | 0112083 | Thiruvananthapuram | MONARCH SURGICAL INSUSTRIES | OT | GH THIRUVANANTHAPURAM | Electrical Suction Unit | NON CRITICAL |
|
||
765 | Pending | Approved | View | VERIFIED | 04-04-2024 | 0112020 | Thiruvananthapuram | ASCON | OT | GH THIRUVANANTHAPURAM | Operating Microscope | NON CRITICAL |
|
||
766 | Pending | Approved | View | VERIFIED | 07-12-2024 | 0112943 | Thiruvananthapuram | SOCOMEC | X RAY | GH NEYYATTINKARA | ONLINE UPS AST PBS | NON CRITICAL |
|
||
767 | Pending | Approved | View | verified | 26-11-2024 | 0112969 | Thiruvananthapuram | ASIA POWER | LABORATORY | GH NEYYATTINKARA | Invertor With UPS | NON CRITICAL |
|
||
768 | Pending | Approved | View | Quotation attached- Recommended RBER | 11-12-2023 | 0113001 | Thiruvananthapuram | Transasia Biomedical LTD. | LABORATORY | GH NEYYATTINKARA | ESR Analyser | NON CRITICAL |
|
||
769 | Pending | Approved | View | verified | 21-08-2024 | 0420175 | Alappuzha | Maestros Mediline Systems LTD. | Minor OT | THQH CHERTHALA | Pulse Oximeter | NON CRITICAL |
|
||
770 | Pending | Approved | View | Equipment completed 8+ year ,Oem declared robonick, prietest touch semiauto analyser is an end of life declared | 27-12-2023 | 0841705 | Thrissur | nill | LABORATORY | PHC VETTILAPARA | Biochemistry Analyzer | CRITICAL |
|
||
771 | Pending | Approved | View | VERIFIED | 07-09-2024 | 0420331 | Alappuzha | Cryo Scienific | LABORATORY | THQH CHERTHALA | Pharmacy Refrigerator | NON CRITICAL |
|
||
772 | Pending | Approved | View | Verified Document | 21-12-2023 | 0641719 | Idukki | Lifecare | SUB CENTRE | 24X7 PHC KARIMANNOOR | B P Apparatus | NON CRITICAL |
|
||
773 | Pending | Approved | View | VERIFIED | 11-03-2024 | 0420340 | Alappuzha | BPL | ECG ROOM | THQH CHERTHALA | ECG Machine | CRITICAL |
|
||
774 | Pending | Approved | View | VERIFIED | 04-04-2024 | 0420341 | Alappuzha | BPL | ECG ROOM | THQH CHERTHALA | ECG Machine | CRITICAL |
|
||
775 | Pending | Approved | View | VERIFIED | 17-07-2024 | 0420344 | Alappuzha | CELFROST | LABORATORY | THQH CHERTHALA | Deep Freezer | NON CRITICAL |
|
||
776 | Pending | Approved | View | VERIFIED | 20-03-2024 | 0113157 | Thiruvananthapuram | Rossmax | observation room | GH NEYYATTINKARA | B P Apparatus | NON CRITICAL |
|
||
777 | Pending | Approved | View | VERIFIED | 23-08-2024 | 0113184 | Thiruvananthapuram | Lifecare | opd | GH NEYYATTINKARA | Nebuliser | NON CRITICAL |
|
||
778 | Pending | Approved | View | VERIFIED | 31-12-2024 | 0113186 | Thiruvananthapuram | OMRON HEALTHCARE CO.LTD. | opd | GH NEYYATTINKARA | Nebuliser | NON CRITICAL |
|
||
779 | Pending | Approved | View | VERIFIED | 11-10-2024 | 0113183 | Thiruvananthapuram | ELKO | opd | GH NEYYATTINKARA | B P Apparatus | NON CRITICAL |
|
||
780 | Pending | Approved | View | VERIFIED | 23-08-2024 | 0113222 | Thiruvananthapuram | Cool | C WARD | GH NEYYATTINKARA | Nebuliser | NON CRITICAL |
|
||
781 | Pending | Approved | View | VERIFIED | 23-08-2024 | 0113224 | Thiruvananthapuram | PHILIPS | C WARD | GH NEYYATTINKARA | Nebuliser | NON CRITICAL |
|
||
782 | Pending | Approved | View | Total damage- verified- Recommended RBER | 28-10-2023 | 0113220 | Thiruvananthapuram | PHILIPS | C WARD | GH NEYYATTINKARA | Nebuliser | NON CRITICAL |
|
||
783 | Pending | Approved | View | VERIFIED | 18-11-2024 | 0630659 | Idukki | PROKAN | LABORATORY | CHC MUTTOM | Haematology Analyser | CRITICAL |
|
||
784 | Pending | Approved | View | VERIFIED | 29-11-2024 | 0630670 | Idukki | LABOPEPPETE | LABORATORY | CHC MUTTOM | Micro Pippette | NON CRITICAL |
|
||
785 | Pending | Approved | View | VERIFIED | 29-11-2024 | 0630665 | Idukki | Agappe Diagnostics LTD. | LABORATORY | CHC MUTTOM | Micro Pippette | NON CRITICAL |
|
||
786 | Pending | Approved | View | Equipment completed 6 year 5 month, repairing cost of equipment is more than 67% of asset value, both criteria for rber met, as per tender clause 5.3.14.1 we recommend equipment for condemination | 11-11-2024 | 1343461 | Kannur | REMI ELEKTROTECHNIK LTD. | LABORATORY | 24X7 PHC KEEHPALLY | Centrifuge | NON CRITICAL |
|
||
787 | Pending | Approved | View | VERIFIED | 23-08-2024 | 0113322 | Thiruvananthapuram | EZ Life | WARD 2 | GH NEYYATTINKARA | Nebuliser | NON CRITICAL |
|
||
788 | Pending | Approved | View | VERIFIED | 03-07-2024 | 0113328 | Thiruvananthapuram | Local purchase | WARD2 | GH NEYYATTINKARA | Nebuliser | NON CRITICAL |
|
||
789 | Pending | Approved | View | Equipment completed 13+ year reapiring cost of equipment is more than 100%,Both criteria for rber met | 15-03-2024 | 0831829 | Thrissur | nill | OP | CHC PERINJANAM | Fetal doppler | NON CRITICAL |
|
||
790 | Pending | Approved | View | Equipment outlived 12+year, Authorised service provider for BPL 9613 fetal doppler has state that machiene get old and spare parts are not available so as per tender clause 5.3.14.2 we recommend equipment for condemination | 06-04-2024 | 0831830 | Thrissur | BPL | OP | CHC PERINJANAM | Fetal doppler | NON CRITICAL |
|
||
791 | Pending | Approved | View | Equipment installed on 31-07-2018,and out lived 5+year, repairing cost of equipment is 63.5% of asset value,Both criteria for RBER met | 07-02-2024 | 1330613 | Kannur | DLAB | LABORATORY | CHC IRIKKUR | Centrifuge | NON CRITICAL |
|
||
792 | Pending | Approved | View | VERIFIED | 17-10-2024 | 0742579 | Ernakulam | Sknol | OP | UPHC CHAMBAKKARA | Weighing Machine | NON CRITICAL |
|
||
793 | Pending | Approved | View | Equipment completed 3 year 5 month with an asset cost of rupees 1875, Quotation not submitted since spares are not available in the market | 23-07-2024 | 1121184 | Kozhikode | Safe & Touch | MALE WARD | THQH KOYILANDY | Nebuliser | NON CRITICAL |
|
||
794 | Pending | Approved | View | Equipment completed 6 year 2 month with an asset cost of rupees 1353, Quotation not submitted since spares are not available in the market. | 19-12-2024 | 1121407 | Kozhikode | Lifecare | COVID ICU | THQH NADAPURAM | Nebuliser | NON CRITICAL |
|
||
795 | Pending | Approved | View | Equipment completed 8+ year, Oem robonick provided EOL for the said equipment | 24-07-2024 | 1441177 | KASARGODE | Robonik | LABORATORY | PHC Perla | Semi-Auto Biochemistry Analyzer | NON CRITICAL |
|
||
796 | Pending | Approved | View | VERIFIED | 08-08-2024 | 0731827 | Ernakulam | Lifecare | observation room | CHC MOOTHAKUNNAM | B P Apparatus | NON CRITICAL |
|
||
797 | Pending | Approved | View | Equipment aged upto 4+ year,repairing cost of equipment is 97% with respect to asset value, both criteria met for rber | 12-10-2023 | 1341610 | Kannur | Med Vision | main center | PHC PARIYARAM | B P Apparatus | NON CRITICAL |
|
||
798 | Pending | Approved | View | Equipment completed 7+ year with an asset cost of rupees 850.Quotation not submitted since spares are not available in the market | 01-08-2024 | 1343130 | Kannur | S Cure | OP | PHC MANGATTIDAM | Weighing Machine | NON CRITICAL |
|
||
799 | Pending | Approved | View | VERIFIED | 22-10-2024 | 0141730 | Thiruvananthapuram | Power UPS | LABORATORY | PHC ADAYAMON | UPS | NON CRITICAL |
|
||
800 | Pending | Approved | View | VERIFIED | 18-07-2024 | 0113540 | Thiruvananthapuram | FABWELL ENGINEERS | CSSD | GH NEYYATTINKARA | Sterlizer | NON CRITICAL |
|
||
801 | Pending | Approved | View | Equipment completed 3+ year with an asset cost of rupees 1600,Quotation not submitted since spares are not available in the market | 03-01-2024 | 0831971 | Thrissur | SAMSO | OP | CHC THIRUVILWAMALA | Weighing Machine | NON CRITICAL |
|
||
802 | Pending | Approved | View | Equipment completed 5 year with an asset cost of rupees 2200, Quotation not submitted since spares are not avaialble in the market | 10-10-2024 | 1343730 | Kannur | Infi | OP | PHC PUZHATHY | B P Apparatus | NON CRITICAL |
|
||
803 | Pending | Approved | View | Equipment outlived 7+ years,with an asset cost of rupees10800,have rusted and weakened body structure posing hazards to patients and user,so we recommend equipment for condemination to mitigate risk and patient safety | 08-04-2024 | 1120941 | Kozhikode | NIL | MALE WARD | THQH KUTTIADI | Suction Apparatus | NON CRITICAL |
|
||
804 | Pending | Approved | View | Equipment completed 17+ year quotation not submitted since spares are not available in the market | 14-05-2024 | 0831260 | Thrissur | Godrej | LABORATORY | CHC ALAPAD | Pharmacy Refrigerator | NON CRITICAL |
|
||
805 | Pending | Approved | View | Equipment outlived 13+ year, Repairirng cost of quipment is 68% of asst value, both criteria for rber met | 20-01-2024 | 0831273 | Thrissur | LG | ICTC LAB | CHC ALAPAD | Pharmacy Refrigerator | NON CRITICAL |
|
||
806 | Pending | Approved | View | Equipment completed 6 year 4 month with an asset cost of rupees 2200,Quotation not submitted since spares are not available in the market | 31-08-2024 | 1330253 | Kannur | Beurer | OP | CHC MATTANNUR | B P Apparatus | NON CRITICAL |
|
||
807 | Pending | Approved | View | Equipment completed 12 year 10 months, repairing cost of equipment is 67% of asset value, both criteria for rber met | 28-09-2024 | 1330010 | Kannur | Godrej | OP | CHC KARIVELLUR | Pharmacy Refrigerator | NON CRITICAL |
|
||
808 | Pending | Approved | View | verified | 04-01-2024 | 0741884 | Ernakulam | OMRON | public health | PHC CHERANELLOORE | B P Apparatus | NON CRITICAL |
|
||
809 | Pending | Approved | View | VERIFIED | 28-11-2024 | 0113715 | Thiruvananthapuram | RADICAL | LABORATORY | DISTRICT MODEL HOSPITAL PEROORKADA | Laboratory Microscope | NON CRITICAL |
|
||
810 | Pending | Approved | View | VERIFIED | 10-07-2024 | 0113732 | Thiruvananthapuram | Biorad | Blood storage unit | DISTRICT MODEL HOSPITAL PEROORKADA | Elisa Reader | NON CRITICAL |
|
||
811 | Pending | Approved | View | VERIFIED | 19-07-2024 | 0620132 | Idukki | NIL | OT | THQH ADIMALY | Drill System | NON CRITICAL |
|
||
812 | Pending | Approved | View | VERIFIED | 27-05-2024 | 0420531 | Alappuzha | Truchek | PAY WARD | THQH HARIPPAD | Nebuliser | NON CRITICAL |
|
||
813 | Pending | Approved | View | VERIFIED | 24-06-2024 | 0742404 | Ernakulam | Lifecare | public health | PHC AYYAMPUZHA | B P Apparatus | NON CRITICAL |
|
||
814 | Pending | Approved | View | VERIFIED | 20-09-2024 | 0741705 | Ernakulam | Dr Morepen | public health | PHC MALAYATTOR | Weighing Machine | NON CRITICAL |
|
||
815 | Pending | Approved | View | verified | 26-07-2024 | 0113639 | Thiruvananthapuram | Local purchase | pediatric | DISTRICT MODEL HOSPITAL PEROORKADA | Weighing Machine | NON CRITICAL |
|
||
816 | Pending | Approved | View | VERIFIED | 31-07-2024 | 0741298 | Ernakulam | ECG Mac | observation room | 24X7 PHC EROOR | ECG Machine | CRITICAL |
|
||
817 | Pending | Approved | View | Equipment Completed 8 years, Robonick prietest touch Semi auto analyser is an obselate model as per tender clause 5.3.14.2 we recommend equipment for condemination | 30-05-2024 | 1142032 | Kozhikode | HANDYPET | LABORATORY | UPHC KANNANCHERI | Micro Pippette | NON CRITICAL |
|
||
818 | Pending | Approved | View | Equipment completed 5 year 6 month quotation not submitted since spares are not available in the market | 15-10-2024 | 1121556 | Kozhikode | Lifecare | FEMALE WARD | THQH KOYILANDY | B P Apparatus | NON CRITICAL |
|
||
819 | Pending | Approved | View | Equipment completed 3+year with an asset cost of rupees 2200,Quotation not submitted since spares are not available in the market | 24-06-2024 | 1121559 | Kozhikode | EKCO | FEMALE WARD | THQH KOYILANDY | B P Apparatus | NON CRITICAL |
|
||
820 | Pending | Approved | View | Equipment completed 3+ year with an asset cost of rupees 2200,Quotation not submitted since spares are not available in the market | 24-06-2024 | 1121560 | Kozhikode | EKCO | FEMALE WARD | THQH KOYILANDY | B P Apparatus | NON CRITICAL |
|
||
821 | Pending | Approved | View | upon inspecting the machiene,it was discovered that compressor motor, seivebed and main board faulty.bottom body is melted due to low quality,these spares need replacement,the supplier and installation stock information is unavailable with enduser as the equipment was donated without manufacturing details, due to lack of spare part information we advise condemning the equipment as a special case 5.3.14.2 | 29-02-2024 | 0831749 | Thrissur | nill | PAIN & PALLIATIVE | CHC PERINJANAM | Oxygen Concentrator | NON CRITICAL |
|
||
822 | Pending | Approved | View | VERIFIED | 18-07-2024 | 0421312 | Alappuzha | perfecxa | CASUALTY | THQH THURAVOOR | Nebuliser | NON CRITICAL |
|
||
823 | Pending | Approved | View | VERIFIED | 05-03-2024 | 0530390 | Kottayam | ANITA INDUSTRIES | OP | CHC EDAYAZHAM | B P Apparatus | NON CRITICAL |
|
||
824 | Pending | Approved | View | Equipment completed 5 years 9 months with an asset cost of rupees 2200,Quotation not submitted since spares are not available in the market | 30-08-2024 | 1330285 | Kannur | A&D Medicals | OP | CHC MATTANNUR | B P Apparatus | NON CRITICAL |
|
||
825 | Pending | Approved | View | Equipment completed 5 year, repairing cost is more than 100% both criteria for rber met | 25-10-2023 | 1330286 | Kannur | Beurer | OP | CHC MATTANNUR | B P Apparatus | NON CRITICAL |
|
||
826 | Pending | Approved | View | Equipment completed 4 year 1 month with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market | 31-08-2024 | 1330287 | Kannur | OMRON | OP | CHC MATTANNUR | B P Apparatus | NON CRITICAL |
|
||
827 | Pending | Approved | View | Equipment completed 7 +year, reapiring cost of equipment more than 100% ,Both criteria for RBER Met | 04-11-2023 | 1330313 | Kannur | BPE | LABORATORY | CHC MATTANNUR | UPS | NON CRITICAL |
|
||
828 | Pending | Approved | View | Equipment outlived 4+year with an asset cost of rupees 2200,Quotation not submitted since spares are not available in the market | 02-07-2024 | 0940312 | Palakkad | OMRON HEALTHCARE CO.LTD. | OP | PHC KOLLENGODE | B P Apparatus | NON CRITICAL |
|
||
829 | Pending | Approved | View | Equipment completed 8 year 11 months , oem informed that equipment is outdated and obselate model, as per tender clause 5.3.14.2 we recommend equipment for condemination | 07-09-2024 | 1330406 | Kannur | Appasamy Associates PVT.LTD. | EYE OP | CHC IRIVERI | Autorefracto Keratometer | NON CRITICAL |
|
||
830 | Pending | Approved | View | VERIFIED | 28-11-2024 | 0113927 | Thiruvananthapuram | Local purchase | khrw payward | DISTRICT MODEL HOSPITAL PEROORKADA | Nebuliser | NON CRITICAL |
|
||
831 | Pending | Approved | View | Equipment completed 9+ year with an asset cost of rupees 1353.Quotation not submitted since spares are not available in the market | 13-06-2024 | 0831435 | Thrissur | Easy Care | OP | CHC THRIPRAYAR | Nebuliser | NON CRITICAL |
|
||
832 | Pending | Approved | View | VERIFIED | 28-02-2024 | 0113929 | Thiruvananthapuram | READY MIST | OLD PAYWARD | DISTRICT MODEL HOSPITAL PEROORKADA | Nebuliser | NON CRITICAL |
|
||
833 | Pending | Approved | View | VERIFIED | 30-09-2024 | 0114030 | Thiruvananthapuram | SERI | PPS OT | DISTRICT MODEL HOSPITAL PEROORKADA | Sterlizer | NON CRITICAL |
|
||
834 | Pending | Approved | View | VERIFIED | 19-03-2024 | 0320553 | Pathanamthitta | Hygeia | Casuality | THQH KONNI | ECG Machine | CRITICAL |
|
||
835 | Pending | Approved | View | VERIFIED | 13-11-2024 | 0114133 | Thiruvananthapuram | Namat Shree Vinayak Udyog | LABOUR ROOM | DISTRICT MODEL HOSPITAL PEROORKADA | B P Apparatus | NON CRITICAL |
|
||
836 | Pending | Approved | View | Equipment completed 4+ year repairing cost of equipment is 68%,Both criteria for rber mer | 04-11-2023 | 0820028 | Thrissur | onlineups, bpe | LABORATORY | THQH CHALAKUDY | UPS | NON CRITICAL |
|
||
837 | Pending | Approved | View | VERIFIED | 31-08-2024 | 0114177 | Thiruvananthapuram | RMS | LABOUR ROOM | DISTRICT MODEL HOSPITAL PEROORKADA | ECG Machine | CRITICAL |
|
||
838 | Pending | Approved | View | VERIFIED | 17-10-2024 | 0742687 | Ernakulam | Sknol | SUB CENTRE | UPHC CHAMBAKKARA | Weighing Machine | NON CRITICAL |
|
||
839 | Pending | Approved | View | VERIFIED | 07-03-2024 | 0511011 | Kottayam | Alps | AUDIOLOGY | GH KOTTAYAM | Audiometer | NON CRITICAL |
|
||
840 | Pending | Approved | View | Equipment outlived its shelf life of 11+ year, End of life letter provided by oem attached | 04-11-2023 | 1110127 | Kozhikode | Erba | LABORATORY | RPH LAB KOZHIKODE | Fully Auto Biochemistry Analyzer | NON CRITICAL |
|
||
841 | Pending | Approved | View | Equipment outlived 11+ years repairing cost of equipment is more than 100%,Both criteria for RBER Met | 16-05-2024 | 1110142 | Kozhikode | CORNLEY | LABORATORY | RPH LAB KOZHIKODE | Electrolyte Analyser | CRITICAL |
|
||
842 | Pending | Approved | View | VERIFIED | 23-07-2024 | 0221178 | Kollam | BPL | LABOUR ROOM | THQH KUNDARA | NST Machine (CTG) | NON CRITICAL |
|
||
843 | Pending | Approved | View | Equipment completed 4 year 11 months with an asset cost of rupees 850, Quotation not submitted since spares are not available in the market | 08-11-2024 | 1330838 | Kannur | Solter | OP | CHC ODUVALLITHATTU | Weighing Machine | NON CRITICAL |
|
||
844 | Pending | Approved | View | VERIFIED | 23-09-2024 | 0642284 | Idukki | NIL | OP | PHC MARIYAPURAM | B P Apparatus | NON CRITICAL |
|
||
845 | Pending | Approved | View | VERIFIED | 23-09-2024 | 0642299 | Idukki | SAMSO | OP | PHC MARIYAPURAM | Weighing Machine | NON CRITICAL |
|
||
846 | Pending | Approved | View | Equipment completed 3+ year with an asset cost of rupees 2200, Quottaion not submitted since spares are not available in the market | 22-09-2024 | 0820182 | Thrissur | nill | CASUALTY | THQH CHALAKUDY | B P Apparatus | NON CRITICAL |
|
||
847 | Pending | Approved | View | Equipment completed 9+ year, repairing cost of equipment is more than 100%, both criteria for rber met | 22-08-2024 | 1330640 | Kannur | ALLENGERS | X RAY | CHC PANOOR | X Ray Machine | CRITICAL |
|
||
848 | Pending | Approved | View | Equipment completed 4 year 5 month, quotation not submitted since spares are not available in the market.asset cost of equipment is below 2200 rupees | 01-02-2024 | 1344379 | Kannur | OMRON | main center | PHC KANICHAR | B P Apparatus | NON CRITICAL |
|
||
849 | Pending | Approved | View | Equipment completed 8+ year, repairing cost of equipment 93%, both criteria for rber met | 03-02-2024 | 1142195 | Kozhikode | ASIA POWER | LABORATORY | PHC KUNNAMANGALAM | UPS | NON CRITICAL |
|
||
850 | Pending | Approved | View | Equipment outlived 17+ year With an asset cost of rupees 3100.Quotation not submitted since spares are not available in the market | 14-05-2024 | 1030734 | Malappuram | mes | DRESSING ROOM | CHC NEDUVA | Sterlizer | NON CRITICAL |
|
||
851 | Pending | Approved | View | VERIFIED | 23-07-2024 | 0421051 | Alappuzha | ANITA INDUSTRIES | MEDICAL WARD | THQH KAYAMKULAM | B P Apparatus | NON CRITICAL |
|
||
852 | Pending | Approved | View | VERIFIED | 05-04-2024 | 0421057 | Alappuzha | Truchek | MEDICAL WARD | THQH KAYAMKULAM | B P Apparatus | NON CRITICAL |
|
||
853 | Pending | Approved | View | VERIFIED | 22-08-2024 | 0421060 | Alappuzha | TRUCHECK | MEDICAL WARD | THQH KAYAMKULAM | Nebuliser | NON CRITICAL |
|
||
854 | Pending | Approved | View | Equipment aged upto 8 year and 7 month, End of life for Equipmnt declared by Orginal Equipment Manufacture attached for reference | 13-11-2023 | 1130962 | Kozhikode | Robonik | LABORATORY | CHC Ulliyeri | Biochemistry Analyzer | CRITICAL |
|
||
855 | Pending | Approved | View | Equipment completed 4 year 1o months with an asset cost of rupees 1600.Quotation not submitted since spares are not available in the market | 17-12-2024 | 1344166 | Kannur | SAMSO | SUB CENTRE | PHC MEKKUNNU | Weighing Machine | NON CRITICAL |
|
||
856 | Pending | Approved | View | VERIFIED | 15-01-2024 | 0542458 | Kottayam | Robonik | LABORATORY | PHC KURUPPANTHARA | Biochemistry Analyzer | CRITICAL |
|
||
857 | Pending | Approved | View | Equipment completed 13+ year with an asset cost of rupees 3200,Quotation not submitted since spares are not available in the market | 12-06-2024 | 1030719 | Malappuram | nill | nursing station | CHC NEDUVA | Sterlizer | NON CRITICAL |
|
||
858 | Pending | Approved | View | Equipment completed 15+ year with an asset cost of rupees 3200,Equipment is having significant structural damage,including multiple cracks, rust and leaking. | 31-07-2024 | 1030731 | Malappuram | nill | nursing station | CHC NEDUVA | Sterlizer | NON CRITICAL |
|
||
859 | Pending | Approved | View | Equipment completed 9.4 years, repairing cost of equipment is 69% both criteria for rber met | 20-10-2023 | 0920181 | Palakkad | REMI ELEKTROTECHNIK LTD. | LABORATORY | THQH CHITTUR | Centrifuge | NON CRITICAL |
|
||
860 | Pending | Approved | View | Equipment outlived 6+ year, repairing cost of equipment is more than 100%,Both criteria for RBER Met | 03-04-2024 | 1020523 | Malappuram | REMI ELEKTROTECHNIK LTD. | LABORATORY | THQH PONNANI | Centrifuge | NON CRITICAL |
|
||
861 | Pending | Approved | View | Equipment installed on 27-11-2013, and outlived 10 year, repairing cost of equipment is 100%,both criteria for RBER met | 30-03-2024 | 1345245 | Kannur | Emerson | LABORATORY | PHC CHITTARIPARAMBA | UPS | NON CRITICAL |
|
||
862 | Pending | Approved | View | VERIFIED | 12-03-2024 | 0421184 | Alappuzha | Lifecare | WARD | THQH PULINKUNNU | Nebuliser | NON CRITICAL |
|
||
863 | Pending | Approved | View | Verified- Recommended RBER | 02-02-2024 | 0541927 | Kottayam | nill | OP | 24X7 PHC POONJAR | Spot Light | NON CRITICAL |
|
||
864 | Pending | Approved | View | VERIFIED | 14-05-2024 | 0421672 | Alappuzha | Rayto | LABORATORY | THQH KAYAMKULAM | Elisa Reader | NON CRITICAL |
|
||
865 | Pending | Approved | View | VERIFIED | 13-11-2024 | 0114535 | Thiruvananthapuram | MEDITECH | CASUALTY | WOMEN & CHILDREN HOSPITAL THYCAUD | Nebuliser | NON CRITICAL |
|
||
866 | Pending | Approved | View | Verified | 12-06-2024 | 0114587 | Thiruvananthapuram | ANITA INDUSTRIES | WARD 6 | WOMEN & CHILDREN HOSPITAL THYCAUD | B P Apparatus | NON CRITICAL |
|
||
867 | Pending | Approved | View | VERIFIED | 09-08-2024 | 0743572 | Ernakulam | MURPHY | OP | PHC GOTHURUTH | Infrared Lamp | NON CRITICAL |
|
||
868 | Pending | Approved | View | VERIFIED | 01-11-2024 | 0341759 | Pathanamthitta | OMRON | PP UNIT | PHC RANNI PAZHAVANGADI | B P Apparatus | NON CRITICAL |
|
||
869 | Pending | Approved | View | VERIFIED | 01-03-2024 | 0421812 | Alappuzha | Godrej | STORE | THQH KAYAMKULAM | Pharmacy Refrigerator | NON CRITICAL |
|
||
870 | Pending | Approved | View | Verified Document- Year Met- Asset value below 10K- Recommended for RBER | 09-12-2023 | 0642185 | Idukki | APC | EYE OP | PHC KONNATHADY | UPS | NON CRITICAL |
|
||
871 | Pending | Approved | View | Verified- Recommended RBER | 20-09-2023 | 0114937 | Thiruvananthapuram | videocon | STORE | WOMEN & CHILDREN HOSPITAL THYCAUD | Pharmacy Refrigerator | NON CRITICAL |
|
||
872 | Pending | Approved | View | VERIFIED | 25-07-2024 | 0142222 | Thiruvananthapuram | Not available | OP | PHC ARUVIKKARA | Sterlizer | NON CRITICAL |
|
||
873 | Pending | Approved | View | VERIFIED | 31-07-2024 | 0743635 | Ernakulam | Dr Morepen | observation room | 24X7 PHC CHITTATTUKARA | Nebuliser | NON CRITICAL |
|
||
874 | Pending | Approved | View | Equipment completed 4+ year with an asset cost of rupees 1600.Quotation not submitted since spares are not available in the market | 06-07-2024 | 1041231 | Malappuram | SAMSO | nursing station | PHC NANNAMBRA | Weighing Machine | NON CRITICAL |
|
||
875 | Pending | Approved | View | VERIFIED | 03-12-2024 | 0642226 | Idukki | DIAMOND | SUBCENTER | PHC KONNATHADY | B P Apparatus | NON CRITICAL |
|
||
876 | Pending | Approved | View | Equipment outlived 10+ year,Repairing cost of equipment is 75% of asset value, both criteria for rber met as per tender clause 5.3.14.1 we recommend equipment for condemination | 25-06-2024 | 1130852 | Kozhikode | Rotek Instruments | LABORATORY | CHC VALAYAM | Centrifuge | NON CRITICAL |
|
||
877 | Pending | Approved | View | VERIFIED | 30-07-2024 | 0543079 | Kottayam | MICROLIFE | OP | UPHC PERUMBAIKADU | B P Apparatus | NON CRITICAL |
|
||
878 | Pending | Approved | View | VERIFIED | 13-03-2024 | 0142259 | Thiruvananthapuram | DRAGON LAB | LABORATORY | PHC VEMBAYAM | Micro Pippette | NON CRITICAL |
|
||
879 | Pending | Approved | View | Equipment completed 14 year 6 months repairing cost of equipment is 61% of asset value, both criteria for rber met | 25-09-2024 | 1345474 | Kannur | LG | pharmacy | PHC PERINGALAM | Pharmacy Refrigerator | NON CRITICAL |
|
||
880 | Pending | Approved | View | Equipment completed 4+ year with an asset cost of rupees 5000,Quotation not submitted since spares are not available in the market | 16-07-2024 | 0841327 | Thrissur | SAMSO | public health | PHC NALUKETTU | Weighing Machine | NON CRITICAL |
|
||
881 | Pending | Approved | View | Equipment outlived 4+ year,with an asset cost of rupees 850.Quotation not submitted since spares are not available in the market | 07-02-2024 | 0841317 | Thrissur | nill | OP | PHC NALUKETTU | Weighing Machine | NON CRITICAL |
|
||
882 | Pending | Approved | View | VERIFIED | 16-04-2024 | 0732588 | Ernakulam | Rotek Instruments | LABORATORY | CHC VENGOOR | Centrifuge | NON CRITICAL |
|
||
883 | Pending | Approved | View | Equipment completed 5 year 5 months with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market | 27-11-2024 | 0842244 | Thrissur | OMRON | public health | PHC VELLANGALLUR | B P Apparatus | NON CRITICAL |
|
||
884 | Pending | Approved | View | VERIFIED | 27-03-2024 | 0732604 | Ernakulam | nill | Minor OT | CHC VENGOOR | Spot Light | NON CRITICAL |
|
||
885 | Pending | Approved | View | VERIFIED | 05-06-2024 | 0732609 | Ernakulam | Rossmax | OP | CHC VENGOOR | Weighing Machine | NON CRITICAL |
|
||
886 | Pending | Approved | View | Verified- Recommended RBER | 14-11-2023 | 0744082 | Ernakulam | LOCAL | OP | UPHC MANGATTUMUKKU | B P Apparatus | NON CRITICAL |
|
||
887 | Pending | Approved | View | VERIFIED | 18-06-2024 | 0140232 | Thiruvananthapuram | Dragon | LABORATORY | PHC CHEMMARUTHY | Micropipete 100-1000 variable | NON CRITICAL |
|
||
888 | Pending | Approved | View | Equipment completed 6 years with an asset cost of rupeeds 1353, Quotation not submitted since spares are not available in the market | 23-11-2024 | 0920024 | Palakkad | Medi Gold | CASUALTY | THQH CHITTUR | Nebuliser | NON CRITICAL |
|
||
889 | Pending | Approved | View | Equipment completed 6+ year with an asset cost of rupees 1353, Quotation not submitted since spares are not available in the market | 22-11-2024 | 0920019 | Palakkad | Medi Gold | CASUALTY | THQH CHITTUR | Nebuliser | NON CRITICAL |
|
||
890 | Pending | Approved | View | VERIFED | 06-11-2024 | 0342134 | Pathanamthitta | Lifecare | OP | 24X7 PHC THANNITHODE | Nebuliser | NON CRITICAL |
|
||
891 | Pending | Approved | View | Verified- Recommended RBER | 12-01-2024 | 0110881 | Thiruvananthapuram | EPPENDROF | IRL | State TB Centre | Centrifuge | NON CRITICAL |
|
||
892 | Pending | Approved | View | Equipment installed on 12/09/2019 and aged upto 4 year 3 month repairing cost of equipment is 91% of asset value. Both criteria for RBER met | 04-12-2023 | 1141793 | Kozhikode | APC | LABORATORY | PHC AYANCHERI | UPS | NON CRITICAL |
|
||
893 | Pending | Approved | View | Equipment completed 3+ year with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market | 22-09-2024 | 0840838 | Thrissur | Mehar | SUB CENTRE | PHC VELUR | Fetal doppler | NON CRITICAL |
|
||
894 | Pending | Approved | View | Equipment completed 10 years, oem informed robonick prietest touch is an obselate model, as per tender clause 5.3.14.2 we recommend equipment for condemination | 30-09-2024 | 0842248 | Thrissur | NA | PHYSIOTHERAPY | PHC VELLANGALLUR | Stimulator | NON CRITICAL |
|
||
895 | Pending | Approved | View | Equipment completed 5+ year with an asset cost of rupees 1353, Quotation not submitted since spares are not available in the market | 09-08-2024 | 0930066 | Palakkad | Lifecare | OP | CHC KODUVAYUR | Nebuliser | NON CRITICAL |
|
||
896 | Pending | Approved | View | Equipment completed 6 year 2 month with an asset cost of rupees 1353, Quotation not submitted since spares are not available in the market | 24-12-2024 | 0930063 | Palakkad | PHILIPS | OP | CHC KODUVAYUR | Nebuliser | NON CRITICAL |
|
||
897 | Pending | Approved | View | Equipment completed 5+ year with an asset cost of rupees 1353, Quotation not submitted since spares are not available in the market | 09-08-2024 | 0930061 | Palakkad | PHILIPS | OP | CHC KODUVAYUR | Nebuliser | NON CRITICAL |
|
||
898 | Pending | Approved | View | Equipment completed 4+ year,with an asset cost of rupees 5000,Quotation not submitted since spares are not available in the market | 02-07-2024 | 0841377 | Thrissur | SAMSO | public health | PHC KAKKAD | Weighing Machine | NON CRITICAL |
|
||
899 | Pending | Approved | View | Equipment completed 22+ year with an asset cost of rupees 19990, The actual book value of the equipment is 11485 rupees, The repairing cost of equipment is 84% with respect to orginal book value, both criteria for rber met consider the orginal book value | 03-04-2024 | 1343077 | Kannur | Godrej | OP | PHC CHUZHALI | Pharmacy Refrigerator | NON CRITICAL |
|
||
900 | Pending | Approved | View | VERIFIED | 24-07-2024 | 0732665 | Ernakulam | nill | main centre | CHC VENGOOR | B P Apparatus | NON CRITICAL |
|
||
901 | Pending | Approved | View | Equipment completed 8 year 7 months repairing cost of equipment is more than 100% both criteria for rber met | 07-11-2024 | 0930173 | Palakkad | PROKAN | LABORATORY | CHC KUZALMANNAM | Haematology Analyser | CRITICAL |
|
||
902 | Pending | Approved | View | VERIFIED | 22-04-2024 | 0114376 | Thiruvananthapuram | REMI ELEKTROTECHNIK LTD. | LABORATORY | CDH PULAYANARKKOTTA | Centrifuge | NON CRITICAL |
|
||
903 | Pending | Approved | View | VERIFIED | 17-05-2024 | 0744085 | Ernakulam | LOCAL | OP | PHC CHELLANAM | B P Apparatus | NON CRITICAL |
|
||
904 | Pending | Approved | View | VERIFIED | 05-03-2024 | 0222056 | Kollam | Lifecare | FS WARD | THQH PUNALUR | Nebuliser | NON CRITICAL |
|
||
905 | Pending | Approved | View | Equipment completed 5 year with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market | 17-10-2024 | 1344753 | Kannur | EKCO | SUB CENTRE | PHC CHELORA | B P Apparatus | NON CRITICAL |
|
||
906 | Pending | Approved | View | Equipment completed 5 year 1 month with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market | 30-11-2024 | 0940559 | Palakkad | Instapro | SUB CENTRE | PHC PUTHUNAGARAM | B P Apparatus | NON CRITICAL |
|
||
907 | Pending | Approved | View | Equipment completed 9+ year,with an asset cost of rupees 1353, Quotation not submitted since spares are not available in the market | 16-09-2023 | 0820845 | Thrissur | EASY BREATH | WARD | THQH CHAVAKAD | Nebuliser | NON CRITICAL |
|
||
908 | Pending | Approved | View | Equipment completed 4+ year repairing cost more than 100% Both criteria for RBER Met | 09-10-2023 | 1344070 | Kannur | SAMSO | SUB CENTRE | PHC ANGADIKKADAVU | Weighing Machine | NON CRITICAL |
|
||
909 | Pending | Approved | View | Equipment completed 14+ year,Reparing cost of equipment is more than 100%,Both criteria for rber met | 09-03-2024 | 0820799 | Thrissur | Sonomed | EYE OP | THQH CHAVAKAD | Ocular Ultrasound | NON CRITICAL |
|
||
910 | Pending | Approved | View | Equipment completed 8 year 7 Months,repairing cost of equipment is more than 64% of asset value, both criteria for rber met as per tender clause 5.3.14.1 we recommend equipment for condemination | 26-09-2024 | 1020435 | Malappuram | SKANRAY HEALTH CARE PVT LTD. | OP | THQH WANDOOR | Pulse Oximeter | NON CRITICAL |
|
||
911 | Pending | Approved | View | VERIFIED | 23-08-2024 | 0142834 | Thiruvananthapuram | PHILIPS | OP | PHC PERUMATHURA | Nebuliser | NON CRITICAL |
|
||
912 | Pending | Approved | View | VERIFIED | 03-10-2024 | 0440853 | Alappuzha | ECG Mac | INJECTION ROOM | PHC KANJIKUZHY | ECG Machine | CRITICAL |
|
||
913 | Pending | Approved | View | VERIFIED | 16-08-2024 | 0440856 | Alappuzha | EVERLOW | nursing station | PHC KANJIKUZHY | Oxygen Concentrator | NON CRITICAL |
|
||
914 | Pending | Approved | View | VERIFIED | 10-09-2024 | 0340609 | Pathanamthitta | OMRON | main center | PHC PANDALAM | B P Apparatus | NON CRITICAL |
|
||
915 | Pending | Approved | View | VERIFIED | 10-09-2024 | 0340610 | Pathanamthitta | OMRON | main center | PHC PANDALAM | B P Apparatus | NON CRITICAL |
|
||
916 | Pending | Approved | View | Equipment completed 5 years, with an asset cost of rupees 1600. Quotation not submitted because since spares are not available in the market | 03-01-2024 | 0930377 | Palakkad | SAMSO | INJECTION ROOM | CHC VADAKKANCHERY | Weighing Machine | NON CRITICAL |
|
||
917 | Pending | Approved | View | VERIFIED | 08-11-2024 | 0340625 | Pathanamthitta | BIOSYSTEM | LABORATORY | PHC PANDALAM | Micro Pippette | NON CRITICAL |
|
||
918 | Pending | Approved | View | Equipment completed 5 year 6 month with an asset cost of rupees 1800.Quotation not submitted because spares are not available in the market | 24-04-2024 | 1344703 | Kannur | Infi | OP | UPHC MUTHATHI | B P Apparatus | NON CRITICAL |
|
||
919 | Pending | Approved | View | VERIFIED | 18-11-2024 | 0142837 | Thiruvananthapuram | Lifecare | FWC | PHC VETTORE | B P Apparatus | NON CRITICAL |
|
||
920 | Pending | Approved | View | VERIFIED | 18-11-2024 | 0142496 | Thiruvananthapuram | Acon | FWC | PHC VETTORE | B P Apparatus | NON CRITICAL |
|
||
921 | Pending | Approved | View | Equipment completed 8 year 9 months, quotation not submitted since spares are not available in the market | 03-02-2024 | 1344735 | Kannur | OMRON | SUB CENTRE | UPHC MUTHATHI | B P Apparatus | NON CRITICAL |
|
||
922 | Pending | Approved | View | verified | 11-01-2024 | 0743019 | Ernakulam | nill | observation room | PHC AVOLY | B P Apparatus | NON CRITICAL |
|
||
923 | Pending | Approved | View | VERIFIED | 23-08-2024 | 0120098 | Thiruvananthapuram | IBIS MEDICAL | LABOUR ROOM | THQH ATTINGAL | Infant Radiant Warmer | CRITICAL |
|
||
924 | Pending | Approved | View | VERIFIED | 17-08-2024 | 0120121 | Thiruvananthapuram | DELTA | LABOUR ROOM | THQH ATTINGAL | Weighing Machine | NON CRITICAL |
|
||
925 | Pending | Approved | View | Equipment completed 3 year 3 month, quotation not submitted since spares are not available in the market | 02-02-2024 | 1331159 | Kannur | Sknol | OP | CHC PINARAYI | Weighing Machine | NON CRITICAL |
|
||
926 | Pending | Approved | View | VERIFIED | 20-05-2024 | 0120124 | Thiruvananthapuram | Local purchase | FEMALE WARD | THQH ATTINGAL | Oxygen Concentrator | NON CRITICAL |
|
||
927 | Pending | Approved | View | VERIFIED | 22-06-2024 | 0120158 | Thiruvananthapuram | OMRON | MALE WARD | THQH ATTINGAL | B P Apparatus | NON CRITICAL |
|
||
928 | Pending | Approved | View | VERIFIED | 21-03-2024 | 0142479 | Thiruvananthapuram | OMRON | OP | PHC PUTHUKURICHY | B P Apparatus | NON CRITICAL |
|
||
929 | Pending | Approved | View | Equipment completed 11+ years,confident, C70D, Dental xray end of life declared model as per tender clause 5.3.14.2 recommend equipment for condmeination | 13-02-2024 | 1020632 | Malappuram | Confident Dental Equipments Ltd. | X RAY | THQH WANDOOR | Dental X Ray Machine | NON CRITICAL |
|
||
930 | Pending | Approved | View | VERIFIED | 04-07-2024 | 0115028 | Thiruvananthapuram | ANITA INDUSTRIES | OT | WOMEN & CHILDREN HOSPITAL THYCAUD | B P Apparatus | NON CRITICAL |
|
||
931 | Pending | Approved | View | Equipment completed 8 year, robonick prietets touch semi auto analyser is an end of life declared model, Eol letter provided by oem robonick attached, | 27-01-2024 | 0930368 | Palakkad | Robonik | LABORATORY | CHC VADAKKANCHERY | Biochemistry Analyzer | CRITICAL |
|
||
932 | Pending | Approved | View | Equipment completed 7 years, quotation not submitted since spares are not available in the market | 03-01-2024 | 0930360 | Palakkad | OMRON HEALTHCARE CO.LTD. | OP | CHC VADAKKANCHERY | Nebuliser | NON CRITICAL |
|
||
933 | Pending | Approved | View | VERIFIED | 17-07-2024 | 0744354 | Ernakulam | SAMSO | OP | 24X7 PHC KARUMALLOR | Weighing Machine | NON CRITICAL |
|
||
934 | Pending | Approved | View | VERIFIED | 30-10-2024 | 0441054 | Alappuzha | TRUCHEK | OP | PHC BHARANIKAVU | B P Apparatus | NON CRITICAL |
|
||
935 | Pending | Approved | View | VERIFIED | 29-06-2024 | 0441055 | Alappuzha | ELKO | OP | PHC BHARANIKAVU | B P Apparatus | NON CRITICAL |
|
||
936 | Pending | Approved | View | Equipment outlived 9+year, End of life letter from orginal equipment manufacture attached | 28-11-2023 | 1020641 | Malappuram | Robonik | LABORATORY | THQH WANDOOR | Biochemistry Analyzer | CRITICAL |
|
||
937 | Pending | Approved | View | VERIFIED | 02-12-2024 | 0744401 | Ernakulam | Handy pette | LABORATORY | 24X7 PHC KARUMALLOR | Micro Pippette | NON CRITICAL |
|
||
938 | Pending | Approved | View | Equipment completed 7+ year with an asset cost of rupees 4500, Repairing cost of equipment is 65% of asset value, both criteria for rber met,as per tender clause 5.3.14.1 we recommend equipment for condemination | 17-09-2024 | 1041804 | Malappuram | APC | LABORATORY | UPHC Tanur | UPS | NON CRITICAL |
|
||
939 | Pending | Approved | View | VERIFIED | 28-11-2024 | 0541623 | Kottayam | nill | OP | PHC ATHIRAMPUZHA | Sterlizer | NON CRITICAL |
|
||
940 | Pending | Approved | View | VERIFIED | 31-07-2024 | 0744761 | Ernakulam | PRAGATHI MEDICAL SYSTEMS | main centre | 24X7 PHC VALAKOM | Weighing Machine | NON CRITICAL |
|
||
941 | Pending | Approved | View | verified | 14-12-2023 | 0744424 | Ernakulam | SAMSO | public health | PHC KADUNGALLOR | Weighing Machine | NON CRITICAL |
|
||
942 | Pending | Approved | View | VERIFIED | 30-07-2024 | 0340399 | Pathanamthitta | Fisher & Paykel Health Care | ILR Room | PHC VALLICODE | Autoclave Cooker Type | NON CRITICAL |
|
||
943 | Pending | Approved | View | VERIFIED | 17-05-2024 | 0744813 | Ernakulam | Lifecare | OP | 24X7 PHC MANJALLOOR | B P Apparatus | NON CRITICAL |
|
||
944 | Pending | Approved | View | Oem Medequip had discontinued support for the oxymed-owgels 3/5lpm oxygen concentrator, obselate letter attached from oem as per tender clause 5.3.14.2 | 29-11-2023 | 1330121 | Kannur | Lifecare | DIALYSIS UNIT | CHC AZHIKODE | Oxygen Concentrator | NON CRITICAL |
|
||
945 | Pending | Approved | View | RBER | 30-05-2024 | 0230011 | Kollam | APC | LABORATORY | CHC SOORANAD NORTH | UPS | NON CRITICAL |
|
||
946 | Pending | Approved | View | VERIFIED | 20-11-2024 | 0542784 | Kottayam | nill | OP | PHC KALAKETTI | Oxygen Concentrator | NON CRITICAL |
|
||
947 | Pending | Approved | View | VERIFIED | 07-11-2024 | 0745035 | Ernakulam | LIFE LINE | STORE | 24X7 PHC ELOOR | B P Apparatus | NON CRITICAL |
|
||
948 | Pending | Approved | View | Equipment completed 5 year with an asset cost of rupees 1353. Quotation not submitted since spares are not available in the market | 23-02-2024 | 0912737 | Palakkad | NA | CH Ward | WOMEN & CHILDREN HOSPITAL PALAKKAD | Nebuliser | NON CRITICAL |
|
||
949 | Pending | Approved | View | Equipment outlived 5+ year,with an asset cost of rupees 1353,Quotation not submitted,since spares are not available in the market | 03-07-2024 | 0912753 | Palakkad | Medi Gold | CASUALTY | WOMEN & CHILDREN HOSPITAL PALAKKAD | Nebuliser | NON CRITICAL |
|
||
950 | Pending | Approved | View | Equipment 2+ year , found compressor motor, seive bed and main board are faulty, upon inspection found that supplier details are unavailable with end user because the equipment was donated during covide pandemic | 24-02-2024 | 0831028 | Thrissur | NIL | PAIN & PALLIATIVE | CHC VATANAPPALLY | Oxygen Concentrator | NON CRITICAL |
|
||
951 | Pending | Approved | View | VERIFIED | 19-09-2024 | 0642862 | Idukki | NIL | LABORATORY | 24X7 PHC VANNAPURAM | UPS | NON CRITICAL |
|
||
952 | Pending | Approved | View | Equipment completed 3+ year, with an asset cost of rupees 2200,Quoatation not submitted since spares are not available in the market | 19-08-2024 | 0842528 | Thrissur | OMRON HEALTHCARE CO.LTD. | OP | PHC PAMBOOR | B P Apparatus | NON CRITICAL |
|
||
953 | Pending | Approved | View | VERIFIED | 21-10-2024 | 0230030 | Kollam | Whirlpool | IP | CHC SOORANAD NORTH | Pharmacy Refrigerator | NON CRITICAL |
|
||
954 | Pending | Approved | View | VERIFIED | 06-08-2024 | 0120883 | Thiruvananthapuram | ANITA INDUSTRIES | CASUALTY | THQH CHIRAYINKEEZH | B P Apparatus | NON CRITICAL |
|
||
955 | Pending | Approved | View | Equipment completed 15 year,repairing cost more than 100%, both criteria for Rber met | 25-03-2024 | 1220528 | Wayanad | Rotek Instruments | ICTC LAB | THQH VYTHIRI | Centrifuge | NON CRITICAL |
|
||
956 | Pending | Approved | View | VERIFIED | 11-09-2024 | 0440719 | Alappuzha | Lifecare | SUB CENTRE | 24X7 PHC VETTAKKAL | B P Apparatus | NON CRITICAL |
|
||
957 | Pending | Approved | View | Verified- Recommended RBER | 14-11-2023 | 0744941 | Ernakulam | DIAMOND | OP | 24X7 PHC KALLOORKAD | B P Apparatus | NON CRITICAL |
|
||
958 | Pending | Approved | View | Verified- Recommended RBER | 31-10-2023 | 0745118 | Ernakulam | APC | LABORATORY | 24X7 PHC KAKKANAD | UPS | NON CRITICAL |
|
||
959 | Pending | Approved | View | Equipment outlived 3+ year with an asset cost of rupees1353,Quotation not submitted since spares are not available in the market | 28-06-2024 | 0930405 | Palakkad | Lifecare | Labour Ward | CHC KUZALMANNAM | Nebuliser | NON CRITICAL |
|
||
960 | Pending | Approved | View | Document Verified- RBER Approved | 13-11-2023 | 0543498 | Kottayam | LIFE LINE | OP | PHC THALANAD | B P Apparatus | NON CRITICAL |
|
||
961 | Pending | Approved | View | Equipment completed 3+ year, the supplier details is unavailable from end user as the equipment was donated during the covid pandemic | 19-07-2024 | 1130189 | Kozhikode | Shenyang Canta Medical Tech Co. Ltd. | OP | CHC Melady | Oxygen Concentrator | NON CRITICAL |
|
||
962 | Pending | Approved | View | Equipmnt outlived 6 years, repair cost of equipment is 68% of assetvalue, Both criteria for RBER Met | 15-02-2024 | 0821137 | Thrissur | IBALL | LABORATORY | THQH KODUNGALLUR | ONLINE UPS AST PBS | NON CRITICAL |
|
||
963 | Pending | Approved | View | Equipment completed 6+ year with an asset cost of rupees 1800,Quotation not submitted since spares are not available in the market | 07-08-2024 | 0941077 | Palakkad | SAMSO | OP | PHC OZHALAPATHY | Weighing Machine | NON CRITICAL |
|
||
964 | Pending | Approved | View | EQUIPMENT COMPLETED 5 YEAR 6 MONTH WITH AN ASSET COST OF RUPEES 1800,QUOTATION NOT SUBMITTED SINCE SPARES ARE NOT AVAILABLE IN THE MARKET | 24-04-2024 | 0941059 | Palakkad | SAMSO | OP | PHC OZHALAPATHY | Weighing Machine | NON CRITICAL |
|
||
965 | Pending | Approved | View | Equipment outlived 10 year,repairing cost of equipment is 87.32% of asset value, Both criteria for Rber met, so we recommend equipment for condemination | 08-04-2024 | 0920803 | Palakkad | Medi Gold | MEDICAL WARD | THQH ALATHUR | Nebuliser | NON CRITICAL |
|
||
966 | Pending | Approved | View | VERIFIED | 29-06-2024 | 0441586 | Alappuzha | ELKO | OP | PHC ERAMALLIKARA | B P Apparatus | NON CRITICAL |
|
||
967 | Pending | Approved | View | VERIFIED | 08-08-2024 | 0441601 | Alappuzha | S Cure | STORE | PHC ERAMALLIKARA | B P Apparatus | NON CRITICAL |
|
||
968 | Pending | Approved | View | Equipment completed 4 year, quotation not submitted since spares are not available in the market | 09-01-2024 | 1041898 | Malappuram | SAMSO | nurse station | PHC Marakkara | Weighing Machine | NON CRITICAL |
|
||
969 | Pending | Approved | View | Equipment completed 6+ year with an asset cost of rupees 1353,Quotation not submitted since spares are not available in the market | 12-07-2024 | 0821239 | Thrissur | Infi-neb | FEMALE WARD | THQH KODUNGALLUR | Nebuliser | NON CRITICAL |
|
||
970 | Pending | Approved | View | VERIFIED | 13-04-2024 | 0745204 | Ernakulam | EKCO | OP | PHC RAMA VARMA PALACE DISPENSARY THRIPUNITHURA | B P Apparatus | NON CRITICAL |
|
||
971 | Pending | Approved | View | VERIFIED | 05-07-2024 | 0221308 | Kollam | Local purchase | OT | THQH NEDUNGOLAM | OT Table | CRITICAL |
|
||
972 | Pending | Approved | View | VERIFIED | 01-01-2025 | 0430098 | Alappuzha | NA | DRESSING ROOM | CHC KURATHIKAD | Oxygen Concentrator | NON CRITICAL |
|
||
973 | Pending | Approved | View | VERIFIED | 06-04-2024 | 0531360 | Kottayam | LG | OP | CHC RAMAPURAM | Sterlizer | NON CRITICAL |
|
||
974 | Pending | Approved | View | Equipment completed 4 year 3 months with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market. | 16-12-2024 | 1030458 | Malappuram | Rossmax | SUBCENTER | CHC TANUR | B P Apparatus | NON CRITICAL |
|
||
975 | Pending | Approved | View | RBER | 01-04-2024 | 0220983 | Kollam | SHIN-NIPPON | Ophthalmology | THQH KOTTARAKKARA | Auto refractometer | NON CRITICAL |
|
||
976 | Pending | Approved | View | VERIFIED | 23-07-2024 | 0430211 | Alappuzha | NA | OP | CHC CHEMPUMPURAM | Sterlizer | NON CRITICAL |
|
||
977 | Pending | Approved | View | VERIFIED | 29-06-2024 | 0441674 | Alappuzha | HYKON | LABORATORY | PHC PALAMEL | UPS | NON CRITICAL |
|
||
978 | Pending | Approved | View | VERIFIED | 11-04-2024 | 0121137 | Thiruvananthapuram | Confident | DENTAL DEPARTMENT | THQH VARKALA | Dental Chair | NON CRITICAL |
|
||
979 | Pending | Approved | View | Equipment outlived 8 year, Repairing cost of equipment is more than 100% of asset value,Both criteria for RBER Met | 23-05-2024 | 1312304 | Kannur | ALFA | TB CENTER | TB Centre | UPS | NON CRITICAL |
|
||
980 | Pending | Approved | View | VERIFIED | 16-12-2024 | 0130291 | Thiruvananthapuram | Power UPS | LABORATORY | CHC PUTHENTHOPPE | UPS | NON CRITICAL |
|
||
981 | Pending | Approved | View | VERIFIED | 06-11-2024 | 0130288 | Thiruvananthapuram | V-GUARD | LABORATORY | CHC PUTHENTHOPPE | UPS | NON CRITICAL |
|
||
982 | Pending | Approved | View | VERIFIED | 06-11-2024 | 0130289 | Thiruvananthapuram | V-GUARD | LABORATORY | CHC PUTHENTHOPPE | UPS | NON CRITICAL |
|
||
983 | Pending | Approved | View | Equipment completed 2+ year quotation not submitted since spares are not available in the market | 29-01-2024 | 1031123 | Malappuram | ANITA INDUSTRIES | SUBCENTER | CHC PURATHUR | B P Apparatus | NON CRITICAL |
|
||
984 | Pending | Approved | View | VERIFIED | 27-05-2024 | 0121386 | Thiruvananthapuram | PRAGATHI MEDICAL SYSTEMS | LABOUR ROOM | THQH VARKALA | Weighing Machine | NON CRITICAL |
|
||
985 | Pending | Approved | View | Equipment completed 8 year, repairing cost of equipment is 79.4% of asset value, both criteria met | 27-01-2024 | 1220585 | Wayanad | Transasia Biomedical LTD. | LABORATORY | THQH VYTHIRI | ESR Analyser | NON CRITICAL |
|
||
986 | Pending | Approved | View | Equipment completed 13+ year , repairing cost of equipment is 82% of asset value, both criteria for rber met as per tender clause 5.3.14.2 we recommend equipment for condemination | 22-08-2024 | 0921030 | Palakkad | MAQUET | CASUALTY | THQH OTTAPALAM | Ventilator | CRITICAL |
|
||
987 | Pending | Approved | View | VERIFIED | 31-12-2024 | 0430322 | Alappuzha | Whirlpool | pharmacy | CHC MUHAMMA | Pharmacy Refrigerator | NON CRITICAL |
|
||
988 | Pending | Approved | View | Equipment outlived 4+year with an asset cost of rupees 5000.Quotation not submitted since spares are not available in the market | 01-12-2023 | 1143172 | Kozhikode | SAMSO | SUB CENTRE | PHC KUNNAMANGALAM | Weighing Machine | NON CRITICAL |
|
||
989 | Pending | Approved | View | Equipment completed 6+year, quotation not submitted since spares are not available in the market | 16-01-2024 | 1143182 | Kozhikode | Lifecare | SUB CENTRE | PHC KUNNAMANGALAM | Nebuliser | NON CRITICAL |
|
||
990 | Pending | Approved | View | Equipment aged upto 12+ year with an asset cost of rupees 7440,Repairing cost of equipment is 64% of asset value, both criteria for rber met | 06-08-2024 | 1031336 | Malappuram | Labtronics | LABORATORY | CHC MANKADA | Colorimeter | NON CRITICAL |
|
||
991 | Pending | Approved | View | Equipment completed 5 year 2 month repairing cost of equipment is 79% of asset value, both criteria for rber met. | 19-09-2024 | 0821222 | Thrissur | MASIMO | NICU | THQH KODUNGALLUR | Pulse Oximeter | NON CRITICAL |
|
||
992 | Pending | Approved | View | Images and Documents verified- Recommended for RBER | 19-10-2023 | 0542964 | Kottayam | nill | OP | PHC PARATHODU | B P Apparatus | NON CRITICAL |
|
||
993 | Pending | Approved | View | VERIFIED | 13-07-2024 | 0542963 | Kottayam | nill | OP | PHC PARATHODU | B P Apparatus | NON CRITICAL |
|
||
994 | Pending | Approved | View | VERIFIED | 02-11-2024 | 0430363 | Alappuzha | SAMSO | OP | CHC EDATHUA | Weighing Machine | NON CRITICAL |
|
||
995 | Pending | Approved | View | Equipment completed 5+ year, repairing cost of equipment is more than 100%,Both criteria for rber met | 08-06-2024 | 1240010 | Wayanad | SAMSO | OP | PHC KURUKKANMOOLA | Weighing Machine | NON CRITICAL |
|
||
996 | Pending | Approved | View | VERIFIED | 04-09-2024 | 0220137 | Kollam | ALLENGERS | X RAY | THQH KADAKKAL | X Ray Machine | CRITICAL |
|
||
997 | Pending | Approved | View | Equipment completed 6+ year, repairing cost of equipment is 100% of asset value,Both criteria for RBER Met | 07-07-2024 | 1240127 | Wayanad | APC | LABORATORY | 24X7 PHC Appappara | UPS | NON CRITICAL |
|
||
998 | Pending | Approved | View | Equipment completed 13+ year , repairing cost of equipment is 76% of asset value, both criteria for rber met, as per tender clause 5.3.14.1 we recommend equipment for condemination | 20-09-2024 | 0921070 | Palakkad | MAQUET | MALE WARD | THQH OTTAPALAM | Ventilator | CRITICAL |
|
||
999 | Pending | Approved | View | VERIFIED | 07-06-2024 | 0130178 | Thiruvananthapuram | Not available | LABORATORY | CHC ANCHUTHENGU | UPS | NON CRITICAL |
|
||
1000 | Pending | Approved | View | Equipment outlived 4+ year, with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market | 15-02-2024 | 1142588 | Kozhikode | Lifecare | SUB CENTRE | PHC Puthiyappa | B P Apparatus | NON CRITICAL |
|
||
1001 | Pending | Approved | View | VERIFIED | 18-07-2024 | 0130192 | Thiruvananthapuram | HYKON | LABORATORY | CHC ANCHUTHENGU | UPS | NON CRITICAL |
|
||
1002 | Pending | Approved | View | VERIFIED | 23-08-2024 | 0130147 | Thiruvananthapuram | Oxymed | OP | CHC ANCHUTHENGU | Oxygen Concentrator | NON CRITICAL |
|
||
1003 | Pending | Approved | View | Equipment completed 11+ year,Repairing cost of equipment is 85% of asset value, both criteria for rber met | 06-05-2024 | 0832396 | Thrissur | Rotek Instruments | LABORATORY | CHC MULLASSERY | Vortex Mixer | NON CRITICAL |
|
||
1004 | Pending | Approved | View | VERIFIED | 08-08-2024 | 0430507 | Alappuzha | DELTA | LABORATORY | CHC VELIYANADU | UPS | NON CRITICAL |
|
||
1005 | Pending | Approved | View | VERIFIED | 20-12-2023 | 0221023 | Kollam | Biotech | PHYSIOTHERAPY | THQH KADAKKAL | Traction Unit | NON CRITICAL |
|
||
1006 | Pending | Approved | View | VERIFIED | 14-10-2024 | 0221031 | Kollam | Supreme Enterprises | Palliative | THQH KADAKKAL | Suction Apparatus | NON CRITICAL |
|
||
1007 | Pending | Approved | View | VERIFIED | 18-07-2024 | 0121536 | Thiruvananthapuram | LONGFIAN | OP | THQH VITHURA | Oxygen Concentrator | NON CRITICAL |
|
||
1008 | Pending | Approved | View | Equipment completed 3 year 5month, with an asset value of rupees 1353, Quotation not submitted since spares are not available in the market | 24-09-2024 | 0832638 | Thrissur | Lifecare | WARD | CHC VADAKKEKAD | Nebuliser | NON CRITICAL |
|
||
1009 | Pending | Approved | View | Equipment outlived 3+ year, repairing cost of equipment is more than 100%, both criteria for RBER Met | 25-03-2024 | 1240302 | Wayanad | V-GUARD | LABORATORY | UPHC MUNDERI | Stabilizer | NON CRITICAL |
|
||
1010 | Pending | Approved | View | VERIFIED | 27-05-2024 | 0121550 | Thiruvananthapuram | MEDITECH | CASUALTY | THQH VITHURA | Nebuliser | NON CRITICAL |
|
||
1011 | Pending | Approved | View | VERIFIED | 02-09-2024 | 0542979 | Kottayam | NIDEK Medical India | SUB CENTRE | PHC PARATHODU | Fetal doppler | NON CRITICAL |
|
||
1012 | Pending | Approved | View | VERIFIED | 30-08-2024 | 0542986 | Kottayam | SAMSO | SUB CENTRE | PHC PARATHODU | Weighing Machine | NON CRITICAL |
|
||
1013 | Pending | Approved | View | Equipment completed 3+ year with an asset cost of rupees 2250, Quotation not submitted since spares are not available in the market | 28-07-2024 | 0832429 | Thrissur | safe Touch | PAIN & PALLIATIVE | CHC MULLASSERY | Airbed | NON CRITICAL |
|
||
1014 | Pending | Approved | View | Equipment completed 9+ year, with an asset cost of rupees 1323, Quotation not submitted since spaeres are not available in the market | 21-08-2024 | 0941661 | Palakkad | ANITA INDUSTRIES | PP UNIT | PHC MELAMURI | B P Apparatus | NON CRITICAL |
|
||
1015 | Pending | Approved | View | Equipment completed 4 years 10 months with an asset cost of rupees 1323, Quotation not submitted as the necessary spare parts are not available in the market | 18-12-2024 | 0832206 | Thrissur | NISCOMED | public health | CHC PAZHANJI | B P Apparatus | NON CRITICAL |
|
||
1016 | Pending | Approved | View | VERIFIED | 24-07-2024 | 0230096 | Kollam | Not available | LABORATORY | CHC MYNAGAPALLI | UPS | NON CRITICAL |
|
||
1017 | Pending | Approved | View | upon inspecting the machiene found compressor motor, seive bed, and main baord are faulty,spare parts required ,unfortunately the supplier and installation details are not available becuase equipment was donated without manufacturing details, due to lack of spare part information we advise condemning the equipment as a special case. | 30-11-2023 | 0843515 | Thrissur | NIL | STORE | PHC CHOWANNUR | Oxygen Concentrator | NON CRITICAL |
|
||
1018 | Pending | Approved | View | Equipment completed 5+ year,repairing cost of equipment is more than 100%,Both criteria for RBER Met | 13-04-2024 | 1240199 | Wayanad | 3C Health | PAIN & PALLIATIVE | PHC KAPPUKUNNU | Oxygen Concentrator | NON CRITICAL |
|
||
1019 | Pending | Approved | View | Verified- Recommended RBER | 22-01-2024 | 0130523 | Thiruvananthapuram | TRUCHECK | observation room | CHC KANYAKULANGARA | Nebuliser | NON CRITICAL |
|
||
1020 | Pending | Approved | View | Recommended RBER | 12-01-2024 | 0130524 | Thiruvananthapuram | Ultra neb | observation room | CHC KANYAKULANGARA | Nebuliser | NON CRITICAL |
|
||
1021 | Pending | Approved | View | VERIFIED | 13-11-2024 | 0230510 | Kollam | MEDITECH | OP | CHC PATHANAPURAM | Nebuliser | NON CRITICAL |
|
||
1022 | Pending | Approved | View | VERIFIED | 10-07-2024 | 0543676 | Kottayam | NIL | OP | PHC PERUVA | Sterlizer | NON CRITICAL |
|
||
1023 | Pending | Approved | View | VERIFIED | 24-06-2024 | 0746580 | Ernakulam | LABOVISION | LABORATORY | 24X7 PHC KUTTAMPUZHA | Laboratory Microscope | NON CRITICAL |
|
||
1024 | Pending | Approved | View | VERIFIED | 24-06-2024 | 0746583 | Ernakulam | LABOMED | LABORATORY | 24X7 PHC KUTTAMPUZHA | Laboratory Microscope | NON CRITICAL |
|
||
1025 | Pending | Approved | View | Equipment completed 19+ year, Repairing cost of equipment is more than 60% both criteria for rber met | 21-03-2024 | 0832342 | Thrissur | Whirlpool | OP | CHC OLLUR | Pharmacy Refrigerator | NON CRITICAL |
|
||
1026 | Pending | Approved | View | VERIFIED | 24-06-2024 | 0746587 | Ernakulam | LABOMED | LABORATORY | 24X7 PHC KUTTAMPUZHA | Laboratory Microscope | NON CRITICAL |
|
||
1027 | Pending | Approved | View | verified | 12-06-2024 | 0114594 | Thiruvananthapuram | Life-Line | INFERTILITY CLINIC | WOMEN & CHILDREN HOSPITAL THYCAUD | B P Apparatus | NON CRITICAL |
|
||
1028 | Pending | Approved | View | VERIFIED | 10-12-2024 | 0230546 | Kollam | Rotek Instruments | LABORATORY | CHC PATHANAPURAM | Centrifuge | NON CRITICAL |
|
||
1029 | Pending | Approved | View | VERIFIED | 07-12-2024 | 0115036 | Thiruvananthapuram | LIFE LINE | WARD 5 | WOMEN & CHILDREN HOSPITAL THYCAUD | B P Apparatus | NON CRITICAL |
|
||
1030 | Pending | Approved | View | Equipment completed 6+ year with an asset cost of rupees 2200,Quotation not submitted since spares are not available in the market | 13-09-2023 | 0820979 | Thrissur | AccuSure | NCD CLINIC | THQH CHAVAKAD | B P Apparatus | NON CRITICAL |
|
||
1031 | Pending | Approved | View | Equipment completed 2 year 8 months,the supplier data is unavailable due to the equipment was donated during covid pandemic | 31-05-2024 | 0822178 | Thrissur | CHINESE | CASUALTY | THQH PUTHUKKAD | Oxygen Concentrator | NON CRITICAL |
|
||
1032 | Pending | Approved | View | Equipment completed 5 year 4 months with an asset cost of rupees 1353, Quotation not submitted since spares are not available in the market | 20-09-2024 | 0920941 | Palakkad | Medi Gold | CH Ward | THQH ALATHUR | Nebuliser | NON CRITICAL |
|
||
1033 | Pending | Approved | View | Equipment completed 9 year 10 month, repairing cost of equipment is 74% of asset value, both criteria for rber met | 04-01-2025 | 0822207 | Thrissur | Goley | LABOUR ROOM | THQH PUTHUKKAD | Fumigator | NON CRITICAL |
|
||
1034 | Pending | Approved | View | Equipment completed 2 year 9 months, supplier details are not available with the end user as the equipment was donated during covid pandemic | 08-06-2024 | 0822214 | Thrissur | CHINESE | PAIN & PALLIATIVE | THQH PUTHUKKAD | Oxygen Concentrator | NON CRITICAL |
|
||
1035 | Pending | Approved | View | Equipment completed 5+ year with an asset cost of rupees 2200,Quotation not submitted since spares are not available in the market | 16-01-2024 | 0920921 | Palakkad | OMRON | Labour Ward | THQH ALATHUR | B P Apparatus | NON CRITICAL |
|
||
1036 | Pending | Approved | View | Equipment completed 4 years, quotation not submitted since spares are not available in the market | 27-01-2024 | 0920919 | Palakkad | NIL | Labour Ward | THQH ALATHUR | B P Apparatus | NON CRITICAL |
|
||
1037 | Pending | Approved | View | Equipment completed 5+ year, unfortunately no supplier details is available with end user, as the equipment was donated without manufacturing details during covid pandamic, due to lack of spare part information we request tender invite authority to condem the equipment | 29-11-2024 | 0832438 | Thrissur | nill | PAIN & PALLIATIVE | CHC OLLUR | Oxygen Concentrator | NON CRITICAL |
|
||
1038 | Pending | Approved | View | Equipment completed 3+ year, found compressor motor, seivebed and main board are faulty, the supplier and installation stock details are not available at hospital becuase the equipment was donated item,due to the lack of spare part information we advise the condemning of equipment | 16-12-2023 | 0820986 | Thrissur | nill | PAIN & PALLIATIVE | THQH CHAVAKAD | Oxygen Concentrator | NON CRITICAL |
|
||
1039 | Pending | Approved | View | VERIFIED | 18-11-2024 | 0531738 | Kottayam | MINDRAY | ICU | CHC ERUMELI | Multipara Monitor | CRITICAL |
|
||
1040 | Pending | Approved | View | VERIFIED | 06-12-2024 | 0531737 | Kottayam | MINDRAY | ICU | CHC ERUMELI | Multipara Monitor | CRITICAL |
|
||
1041 | Pending | Approved | View | VERIFIED | 01-10-2024 | 0120781 | Thiruvananthapuram | Yuwell | CASUALTY | THQH NEMOM | Oxygen Concentrator | NON CRITICAL |
|
||
1042 | Pending | Approved | View | VERIFIED | 05-04-2024 | 0230603 | Kollam | MEDICRAFT | DENTAL DEPARTMENT | CHC MAYANAD | Sterlizer | NON CRITICAL |
|
||
1043 | Pending | Approved | View | VERIFIED | 10-10-2024 | 0121734 | Thiruvananthapuram | Lifecare | STORE | THQH NEMOM | Nebuliser | NON CRITICAL |
|
||
1044 | Pending | Approved | View | VERIFIED | 27-09-2024 | 0121739 | Thiruvananthapuram | NIDEK | STORE | THQH NEMOM | Nebuliser | NON CRITICAL |
|
||
1045 | Pending | Approved | View | Equipment completed 6 year 3 months with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market | 04-01-2025 | 0843095 | Thrissur | SCURE | OP | PHC PUTHUR | B P Apparatus | NON CRITICAL |
|
||
1046 | Pending | Approved | View | VERIFIED | 04-10-2024 | 0121754 | Thiruvananthapuram | PANASONIC | STORE | THQH NEMOM | Pharmacy Refrigerator | NON CRITICAL |
|
||
1047 | Pending | Approved | View | Equipment aged upto 2 year 9 months with an asset cost of rupees 2200.Quotation not submitted since spares are not available in the market | 12-03-2024 | 0930928 | Palakkad | nill | OP | CHC CHERPULASSERY | B P Apparatus | NON CRITICAL |
|
||
1048 | Pending | Approved | View | Equipment completed 9 year 4 months with an asset cost of rupees 1353, Quotation not submitted since spares are not available in the market | 22-11-2024 | 0930934 | Palakkad | OPTIMA | OP | CHC CHERPULASSERY | Nebuliser | NON CRITICAL |
|
||
1049 | Pending | Approved | View | Equipment completed 5 year? 10 months with an asset cost of rupees 1353. Quotation not submitted since spares are not available in the market | 21-09-2024 | 0930937 | Palakkad | OPTIMA | OP | CHC CHERPULASSERY | Nebuliser | NON CRITICAL |
|
||
1050 | Pending | Approved | View | Equipment completed 13+ year, Authorised supplier declared that Ecg machiene Edan se1200 is an obselate model,as per tender clause 5.3.14.2 we recommend equipment for condemination | 14-02-2024 | 0822300 | Thrissur | EDAN | FEMALE WARD | THQH CHELAKKARA | ECG Machine | CRITICAL |
|
||
1051 | Pending | Approved | View | VERIFIED | 05-08-2024 | 0121764 | Thiruvananthapuram | APC | X RAY | THQH NEMOM | UPS | NON CRITICAL |
|
||
1052 | Pending | Approved | View | VERIFIED | 21-07-2024 | 0111306 | Thiruvananthapuram | LOCAL | HISTOPATHOLOGY | STATE PUBLIC HEALTH LABORATORY | Tissue Floatation Bath | NON CRITICAL |
|
||
1053 | Pending | Approved | View | Equipment completed 4+ year with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market | 19-07-2024 | 0941898 | Palakkad | Truchek | OP | PHC MANKARA | B P Apparatus | NON CRITICAL |
|
||
1054 | Pending | Approved | View | VERIFIED | 22-06-2024 | 0143043 | Thiruvananthapuram | PHILIPS | OP | PHC KADAKAMPALLY | Nebuliser | NON CRITICAL |
|
||
1055 | Pending | Approved | View | VERIFIED | 22-06-2024 | 0143046 | Thiruvananthapuram | AccuSure | OP | PHC KADAKAMPALLY | Nebuliser | NON CRITICAL |
|
||
1056 | Pending | Approved | View | VERIFIED | 12-08-2024 | 0340440 | Pathanamthitta | Not available | main center | PHC ERATHU | B P Apparatus | NON CRITICAL |
|
||
1057 | Pending | Approved | View | Equipment completed 4 year 7 months with an asset cost of rupees 1600,Quotation not submitted since spares are not available in the market | 01-08-2024 | 0941765 | Palakkad | SAMSO | SUB CENTRE | PHC THENKURISSI | Weighing Machine | NON CRITICAL |
|
||
1058 | Pending | Approved | View | Equipment completed 4 year 7 months with an asset cost of rupes 2200.Quotation not submitted since spares are not available in the market | 01-08-2024 | 0941763 | Palakkad | Mycare | SUB CENTRE | PHC THENKURISSI | B P Apparatus | NON CRITICAL |
|
||
1059 | Pending | Approved | View | VERIFIED | 01-07-2024 | 0544211 | Kottayam | NISCOMED | OP | PHC MEENADOM | B P Apparatus | NON CRITICAL |
|
||
1060 | Pending | Approved | View | Equipment completed 6 year 8 month with an asset cost of rupees 850, Quotation not submitted since spares are not available in the market | 22-10-2024 | 1240673 | Wayanad | Healcure | OP | PHC VENGAPALLY | Weighing Machine | NON CRITICAL |
|
||
1061 | Pending | Approved | View | Totally Rusted Dilapidated condition- Hospital letter attached- Recommended for RBER | 08-11-2023 | 0340836 | Pathanamthitta | Not available | PAIN & PALLIATIVE | PHC PANDALAM THEKKEKARA | Sterlizer | NON CRITICAL |
|
||
1062 | Pending | Approved | View | VERIFIED | 24-08-2024 | 0130926 | Thiruvananthapuram | DIAMOND | nurse station | CHC KATTAKADA | B P Apparatus | NON CRITICAL |
|
||
1063 | Pending | Approved | View | Equipment outlived 2 year 9 month,with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market | 08-04-2024 | 0942210 | Palakkad | OMRON HEALTHCARE CO.LTD. | OP | PHC KOTTAYI | B P Apparatus | NON CRITICAL |
|
||
1064 | Pending | Approved | View | Equipment completed 5+ year with an asset cost of rupees1875,Quotation not submitted since spares are not available in the market | 01-06-2024 | 0822323 | Thrissur | BPL | STORE | THQH CHELAKKARA | Nebuliser | NON CRITICAL |
|
||
1065 | Pending | Approved | View | Equipment completed 2 year 9 months The supplier details are not available in hospital as the equipment was donated during covid pandemic | 02-07-2024 | 0843273 | Thrissur | nill | PAIN & PALLIATIVE | PHC VALLACHIRA | Oxygen Concentrator | NON CRITICAL |
|
||
1066 | Pending | Approved | View | Equipment completed 6+ year with an asset cost of rupees 2200,Quotation not submitted since spares are not available in the market | 25-07-2024 | 0843747 | Thrissur | OMRON | OP | PHC VILVATTOM | B P Apparatus | NON CRITICAL |
|
||
1067 | Pending | Approved | View | Machine totally Rusted and Dilapidated condition- Images verified- Hospital Letter attached- Recommended for RBer | 08-11-2023 | 0340867 | Pathanamthitta | Not available | LABORATORY | PHC PANDALAM THEKKEKARA | Laboratory Microscope | NON CRITICAL |
|
||
1068 | Pending | Approved | View | Verified Documents- Customer Letter attached- RBER Approved | 16-11-2023 | 0340868 | Pathanamthitta | LABOVISION | LABORATORY | PHC PANDALAM THEKKEKARA | Laboratory Microscope | NON CRITICAL |
|
||
1069 | Pending | Approved | View | Equipment outlived 4+year with an asset cost of rupees 1353, Quotation not submitted since spares are not available in the market | 06-05-2024 | 0941367 | Palakkad | Lifecare | OP | PHC MUTHALAMADA | Nebuliser | NON CRITICAL |
|
||
1070 | Pending | Approved | View | Equipment completed 4+ year with an asset cost of rupees 1600,Quotation not submitted since spares are not available in the market | 24-04-2024 | 0843793 | Thrissur | SAMSO | SUBCENTER | PHC VILVATTOM | Weighing Machine | NON CRITICAL |
|
||
1071 | Pending | Approved | View | VERIFIED | 15-10-2024 | 0430997 | Alappuzha | Lifecare | OP | CHC Thrikunnapuzha | Nebuliser | NON CRITICAL |
|
||
1072 | Pending | Approved | View | Equipment completed 12 years,Repairing cost of equipment is more than 100% both criteria for rber met | 13-04-2024 | 0822712 | Thrissur | Anands | OT | THQH KUNNAMKULAM | Suction Apparatus | NON CRITICAL |
|
||
1073 | Pending | Approved | View | Equipment completed 5 year 8 months with an asset cost of rupees 1353, Quotation not submitted since spares are not available in the market | 25-11-2024 | 0822779 | Thrissur | S Cure | FEMALE WARD | THQH KUNNAMKULAM | Nebuliser | NON CRITICAL |
|
||
1074 | Pending | Approved | View | VERIFIED | 03-06-2024 | 0230162 | Kollam | Infi-neb | IP | CHC CHAVARA | Nebuliser | NON CRITICAL |
|
||
1075 | Pending | Approved | View | VERIFIED | 20-06-2024 | 0241204 | Kollam | Infi | OP | PHC CHAVARA | B P Apparatus | NON CRITICAL |
|
||
1076 | Pending | Approved | View | VERIFIED | 09-07-2024 | 0242453 | Kollam | LIFE LINE | OP | 24X7 PHC CHADAYAMANGALAM | Nebuliser | NON CRITICAL |
|
||
1077 | Pending | Approved | View | Equipment completed 9 year 2 months with an assert cost of rupees 1000, the micropippete has found fail calibration during routine testing the device has exhibited disrespancies in volume dispensing , which are beyond acceptable tolerance limits for accurate and relaible lab work, so we recommedn equipment for condemination | 26-09-2024 | 1240448 | Wayanad | Dr. Pette | LABORATORY | PHC MOOPAINAD | Micro Pippette | NON CRITICAL |
|
||
1078 | Pending | Approved | View | Equipment completed 11 year 2 month, repairing cost of equipment is 97% of asset value, nboth criteria for rber met | 27-11-2024 | 1231034 | Wayanad | Erba | LABORATORY | CHC NALLOORNAD | Biochemistry Analyzer | CRITICAL |
|
||
1079 | Pending | Approved | View | Equipment completed 4 year 11 months with an asset cost of rupees 1600, Quotation not submitted since spares are not available in the market | 18-11-2024 | 0843852 | Thrissur | SAMSO | SUB CENTRE | PHC VARANDARAPILLY | Weighing Machine | NON CRITICAL |
|
||
1080 | Pending | Approved | View | VERIFIED | 20-07-2024 | 0121870 | Thiruvananthapuram | DIAMOND | OP | TH Malayinkeezh | B P Apparatus | NON CRITICAL |
|
||
1081 | Pending | Approved | View | Verified- Recommended RBER | 21-12-2023 | 0121877 | Thiruvananthapuram | Cura | ECG ROOM | TH Malayinkeezh | ECG Machine | CRITICAL |
|
||
1082 | Pending | Approved | View | VERIFIED | 20-03-2024 | 0121883 | Thiruvananthapuram | Dr.Odin | STORE | TH Malayinkeezh | B P Apparatus | NON CRITICAL |
|
||
1083 | Pending | Approved | View | Verified- Recommended RBER | 09-01-2024 | 0121922 | Thiruvananthapuram | PHILIPS | WARD | TH Malayinkeezh | Nebuliser | NON CRITICAL |
|
||
1084 | Pending | Approved | View | Equipment completed 2+ year, Supplier details is unavailable from end user as the equipment was donated without manufacturing details, so we request for condemning equipment as per tender cluase 5.3.14..2 | 07-12-2023 | 0842641 | Thrissur | Lifecare | PAIN & PALLIATIVE | PHC ENGANDIYUR | Oxygen Concentrator | NON CRITICAL |
|
||
1085 | Pending | Approved | View | Images and Documents verified- Recommended for RBER | 03-11-2023 | 0544195 | Kottayam | SAMSO | OP | PHC MEENADOM | Weighing Machine | NON CRITICAL |
|
||
1086 | Pending | Approved | View | VERIFIED | 18-11-2024 | 0241234 | Kollam | KELVINATOR | OP | UPHC VAADI | Pharmacy Refrigerator | NON CRITICAL |
|
||
1087 | Pending | Approved | View | Equipment completed 3 + year with an asset cost of rupees 1353,Quotation not submitted since spares are not available in the market | 18-01-2024 | 0941613 | Palakkad | NA | OP | 24X7 PHC KUNISSERY | Nebuliser | NON CRITICAL |
|
||
1088 | Pending | Approved | View | Equipment completed 3 year 6 months, with an asset cost of rupees 1353, Quotation not submitted since spares are not available in the market | 19-09-2024 | 0843897 | Thrissur | DeVilbiss | OP | PHC ELAVALLY | Nebuliser | NON CRITICAL |
|
||
1089 | Pending | Approved | View | VERIFIED | 02-12-2024 | 0443395 | Alappuzha | videocon | OP | PHC PALLIPPAD | Pharmacy Refrigerator | NON CRITICAL |
|
||
1090 | Pending | Approved | View | Equipment completed 3 year 11 months with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market | 25-10-2024 | 0942002 | Palakkad | Lifecare | Subcentre | PHC PIRAYIRI | B P Apparatus | NON CRITICAL |
|
||
1091 | Pending | Approved | View | Verified | 13-06-2024 | 0431068 | Alappuzha | NA | DRESSING ROOM | CHC MUTHUKULAM | Sterlizer | NON CRITICAL |
|
||
1092 | Pending | Approved | View | Equipment already condemned from hospital as per order NO:88/2024 Dated 13/05/2024 | 15-06-2024 | 0844656 | Thrissur | videocon | STORE | PHC PATTIKAD | Pharmacy Refrigerator | NON CRITICAL |
|
||
1093 | Pending | Approved | View | EQUIPMENT COMPLETED 4 YEAR, REPAIRING COST IS MORE THAN 100% BOTH CRITERIA FOR RBER MET, RECOMMEND FOR CONDEMINATION | 15-02-2024 | 1241274 | Wayanad | SAMSO | OP | PHC Vellamunda | Weighing Machine | NON CRITICAL |
|
||
1094 | Pending | Approved | View | VERIFIED | 04-03-2024 | 0443545 | Alappuzha | SAMSO | LABORATORY | 24X7 PHC ARATTUPUZHA | Weighing Machine | NON CRITICAL |
|
||
1095 | Pending | Approved | View | VERIFIED | 30-10-2024 | 0443548 | Alappuzha | SAMSO | LABORATORY | 24X7 PHC ARATTUPUZHA | Weighing Machine | NON CRITICAL |
|
||
1096 | Pending | Approved | View | VERIFIED | 04-12-2024 | 0230217 | Kollam | HMS | PHYSIOTHERAPY | CHC OCHIRA | Moist Heat Therapy | NON CRITICAL |
|
||
1097 | Pending | Approved | View | VERIFIED | 30-08-2024 | 0442988 | Alappuzha | Microtek | OP | 24X7 PHC EZHUPUNNA | B P Apparatus | NON CRITICAL |
|
||
1098 | Pending | Approved | View | VERIFIED | 29-11-2024 | 0241463 | Kollam | Diamond | OP | PHC K.S PURAM | B P Apparatus | NON CRITICAL |
|
||
1099 | Pending | Approved | View | VERIFIED | 25-03-2024 | 0443718 | Alappuzha | Owgels oxy-med | OP | PHC CHERUTHANA | Nebuliser | NON CRITICAL |
|
||
1100 | Pending | Approved | View | VERIFIED | 27-05-2024 | 0443719 | Alappuzha | OMRON HEALTHCARE CO.LTD. | OP | PHC CHERUTHANA | B P Apparatus | NON CRITICAL |
|
||
1101 | Pending | Approved | View | VERIFIED | 17-07-2024 | 0443235 | Alappuzha | NA | OP | PHC KUPPAPURAM | Oxygen Concentrator | NON CRITICAL |
|
||
1102 | Pending | Approved | View | VERIFIED | 21-08-2024 | 0431223 | Alappuzha | MEDITECH | STORE | CHC CHUNAKKARA | Nebuliser | NON CRITICAL |
|
||
1103 | Pending | Approved | View | VERIFIED | 21-05-2024 | 0431259 | Alappuzha | LABOPETTE | LABORATORY | CHC CHUNAKKARA | Micro Pippette | NON CRITICAL |
|
||
1104 | Pending | Approved | View | Equipment outlived 6+year,With an asset cost of rupees 1323, Quotation not submitted since spares are not availble in the market | 18-05-2024 | 0931064 | Palakkad | Diamond | OP | CHC CHALISSERY | B P Apparatus | NON CRITICAL |
|
||
1105 | Pending | Approved | View | VERIFIED | 18-01-2024 | 0112610 | Thiruvananthapuram | Smiths Medicals | OT | DH NEDUMANGAD | PATIENT WARMER | NON CRITICAL |
|
||
1106 | Pending | Approved | View | VERIFIED | 09-09-2024 | 0443945 | Alappuzha | OMRON | OP | 24X7 PHC THAKAZHY | B P Apparatus | NON CRITICAL |
|
||
1107 | Pending | Approved | View | VERIFIED | 14-08-2024 | 0122114 | Thiruvananthapuram | Dr.Odin | CASUALTY | THQH FORT THIRUVANANTHAPURAM | B P Apparatus | NON CRITICAL |
|
||
1108 | Pending | Approved | View | VERIFIED | 10-07-2024 | 0130644 | Thiruvananthapuram | TRUCHECK | WARD | CHC PALODE | Nebuliser | NON CRITICAL |
|
||
1109 | Pending | Approved | View | VERIFIED | 07-05-2024 | 0431306 | Alappuzha | ANITA INDUSTRIES | OP | CHC AMBALAPUZHA | B P Apparatus | NON CRITICAL |
|
||
1110 | Pending | Approved | View | VERIFIED | 06-11-2024 | 0431356 | Alappuzha | CONTEC MEDICAL SYSTEMS CO.LTD. | OP | CHC AMBALAPUZHA | ECG Machine | CRITICAL |
|
||
1111 | Pending | Approved | View | Equipment completed 7 year 5 months, repairing cost of equipment is 77% of asset value, both criteria for rber met | 14-10-2024 | 1043263 | Malappuram | APC | LABORATORY | PHC EDAPATTA | UPS | NON CRITICAL |
|
||
1112 | Pending | Approved | View | VERIFIED | 19-11-2024 | 0241564 | Kollam | APC | LABORATORY | HCD-Karunagapally | UPS | NON CRITICAL |
|
||
1113 | Pending | Approved | View | Equipment completed 6 hyear 3 months with an asset cost of rupees 850. Quotation not submitted since spares are not available in the market | 23-12-2024 | 0942396 | Palakkad | Unigrip | OP | PHC AYALUR | Weighing Machine | NON CRITICAL |
|
||
1114 | Pending | Approved | View | VERIFIED | 26-10-2024 | 0241582 | Kollam | MASIMO | ECG ROOM | PHC ALAPPATU | Pulse Oximeter | NON CRITICAL |
|
||
1115 | Pending | Approved | View | VERIFIED | 02-09-2024 | 0544469 | Kottayam | NIL | SUB CENTRE | PHC PUTHUPPALLY | Fetal doppler | NON CRITICAL |
|
||
1116 | Pending | Approved | View | VERIFIED | 15-07-2024 | 0442676 | Alappuzha | BPE | LABORATORY | 24X7 PHC VAYALAR | UPS | NON CRITICAL |
|
||
1117 | Pending | Approved | View | Upon inspecting the machiene found compressor motor , seivebed and mainboard are faulty, Equipment installed on 20-01-2024, unfortunately the supplier information is unavailable with end user as the equipment was donated without manufacturing/supplier details.Due to the lack of spare part information we request tender inviting authority to condem the equipment | 21-01-2024 | 0845025 | Thrissur | nill | OP | PHC ELANAD | Oxygen Concentrator | NON CRITICAL |
|
||
1118 | Pending | Approved | View | Equipment completed 4 year, repairing cost of equipment is more than 100% both criteria for rber met, as per tender clause 5.3.14.1 | 31-10-2023 | 1240532 | Wayanad | SAMSO | observation room | PHC CHETHALAYAM | Weighing Machine | NON CRITICAL |
|
||
1119 | Pending | Approved | View | VERIFIED | 19-09-2024 | 0241259 | Kollam | Amtz | OP | PHC SAKTHIKULANGARA | Oxygen Concentrator | NON CRITICAL |
|
||
1120 | Pending | Approved | View | Equipment completed 3 year 11 months with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market | 18-12-2024 | 0844383 | Thrissur | Shenzhen Goldway Industrial Inc. | observation room | PHC EDAVILANGU | B P Apparatus | NON CRITICAL |
|
||
1121 | Pending | Approved | View | Equipment completed 3+ year with an asset cost of rupees 2200,Quotation not submitted since spares are not available in the market | 17-03-2024 | 0844139 | Thrissur | OMRON | public health | PHC KAIPPAMANGALAM | B P Apparatus | NON CRITICAL |
|
||
1122 | Pending | Approved | View | Equipment completed 5 year 9 months repairing cost of equipment is 88% of asset value, both criteria for rber met | 21-11-2024 | 0844091 | Thrissur | Rotek Instruments | LABORATORY | PHC KAIPPAMANGALAM | Centrifuge | NON CRITICAL |
|
||
1123 | Pending | Approved | View | VERIFIED | 26-09-2024 | 0240133 | Kollam | BPL | OP | UPHC KARUNAGAPALLY | ECG Machine | CRITICAL |
|
||
1124 | Pending | Approved | View | Equipment completed 3+ year with an asset cost of rupees 1600.Quotation not submitted since spares are not available in the market | 20-07-2024 | 1240571 | Wayanad | SAMSO | OP | 24X7 PHC VAZHAVATTA | Weighing Machine | NON CRITICAL |
|
||
1125 | Pending | Approved | View | Checked machiene found compressor motor,seive bed, and main board are faulty, need replacement of spare parts to make equipment operational, unfortunately supplier and stock information is unavailable from end user as the equipment was donated without manufacturing and supplier details, due to lack of spare part information we recommend equipment for condemination | 18-03-2024 | 0845164 | Thrissur | nill | PAIN & PALLIATIVE | PHC CHOONDAL | Oxygen Concentrator | NON CRITICAL |
|
||
1126 | Pending | Approved | View | VERIFIED | 06-09-2024 | 0130713 | Thiruvananthapuram | Local purchase | LABORATORY | CHC ARYANADU | Micro Pippette | NON CRITICAL |
|
||
1127 | Pending | Approved | View | Equipment outlived 8 year 11 months, repairing cost of equipment is more than 100%, both criteria for rber met | 22-11-2024 | 1044498 | Malappuram | Agappe Diagnostics LTD. | LABORATORY | 24X7 PHC POTHUKAL | Biochemistry Analyzer | CRITICAL |
|
||
1128 | Pending | Approved | View | verified | 24-08-2024 | 0130755 | Thiruvananthapuram | MASIMO | nurse station | CHC ARYANADU | Pulse Oximeter | NON CRITICAL |
|
||
1129 | Pending | Approved | View | VERIFIED | 08-08-2024 | 0143143 | Thiruvananthapuram | OMRON HEALTHCARE CO.LTD. | observation room | PHC CHETTIVILAKAM | B P Apparatus | NON CRITICAL |
|
||
1130 | Pending | Approved | View | Equipment completed 4+ year repairing cost of equipment is more than 68%,both criteria for rber met | 03-10-2023 | 0942752 | Palakkad | nill | LABORATORY | PHC KUMBIDI | UPS | NON CRITICAL |
|
||
1131 | Pending | Approved | View | VERIFIED | 03-06-2024 | 0240760 | Kollam | Jiangsu Jumao X-Care Medical Equipment Co,Ltd | OP | PHC PORUVAZHY | Oxygen Concentrator | NON CRITICAL |
|
||
1132 | Pending | Approved | View | Equipment completed 5+ year repairing cost of equipment is more than 100% both criteria for rber met | 13-06-2024 | 0844162 | Thrissur | LOCAL | OP | PHC PADIYUR | Oxygen Concentrator | NON CRITICAL |
|
||
1133 | Pending | Approved | View | Equipment completed 3 year 7 months, with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market | 04-11-2024 | 0844914 | Thrissur | EKCO | OP | MCH GURUVAYUR | B P Apparatus | NON CRITICAL |
|
||
1134 | Pending | Approved | View | Equipment completed 4 year 5 months,with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market | 04-11-2024 | 0844918 | Thrissur | S Cure | OP | MCH GURUVAYUR | B P Apparatus | NON CRITICAL |
|
||
1135 | Pending | Approved | View | Equipment completed 5 year 8 months, repairing cost of equipment is more than 100%, both criteria for rber met | 14-10-2024 | 0844173 | Thrissur | LOCAL | OP | PHC PADIYUR | Sterlizer | NON CRITICAL |
|
||
1136 | Pending | Approved | View | Equipment completed 9 year 3 months , repairing cost of equipment is 61% of asset value, both criteria for rber met | 20-09-2024 | 0920504 | Palakkad | Faith Innovations | OT | THQH MANNARKKAD | Fumigator | NON CRITICAL |
|
||
1137 | Pending | Approved | View | VERIFIED | 20-07-2024 | 0130619 | Thiruvananthapuram | Not available | LABORATORY | CHC Vellanad | UPS | NON CRITICAL |
|
||
1138 | Pending | Approved | View | VERIFIED | 24-04-2024 | 0130687 | Thiruvananthapuram | Handy pette | LABORATORY | CHC Vellanad | Micro Pippette | NON CRITICAL |
|
||
1139 | Pending | Approved | View | VERIFIED | 10-06-2024 | 0130658 | Thiruvananthapuram | Easy Care | OP | CHC Vellanad | Nebuliser | NON CRITICAL |
|
||
1140 | Pending | Approved | View | VERIFIED | 09-10-2024 | 0130686 | Thiruvananthapuram | Not available | DRESSING ROOM | CHC Vellanad | Sterlizer | NON CRITICAL |
|
||
1141 | Pending | Approved | View | rber | 18-03-2023 | 1241712 | Wayanad | Carer | Palliative | 24X7 PHC CHEERAL | Oxygen Concentrator | NON CRITICAL |
|
||
1142 | Pending | Approved | View | Equipment outlived 9+ year, repairing cost of equipment is more than 100%,Both criteria for RBER Met | 09-06-2024 | 1230903 | Wayanad | MINDRAY | LABORATORY | CHC PORUNNANNUR | Haematology Analyser | CRITICAL |
|
||
1143 | Pending | Approved | View | Equipment completed 3 years,repairing cost of equipment is 72% of asset value, both criteria for rber met | 03-01-2024 | 0942870 | Palakkad | NIL | pharmacy | PHC PUDUSSERY | B P Apparatus | NON CRITICAL |
|
||
1144 | Pending | Approved | View | Equipment completed 3+ year, with an asset cost of rupees 1353, Quotation not submitted since spares are not available in the market | 20-09-2024 | 1131059 | Kozhikode | Rose Max | OP | CHC VALAYAM | Nebuliser | NON CRITICAL |
|
||
1145 | Pending | Approved | View | Equipment completed 4 year 9 month with an asset cost of rupees 1353, Quotation not submitted since spares are not available in the market | 20-09-2024 | 1131062 | Kozhikode | TRUCHECK | OP | CHC VALAYAM | Nebuliser | NON CRITICAL |
|
||
1146 | Pending | Approved | View | Equipment completed 10+ year, repairing cost of equipment is more than 100%,Both criteria for RBER Met | 04-07-2024 | 0921380 | Palakkad | B.BRAUN | DIALYSIS UNIT | THQH OTTAPALAM | Dialysis Machine | CRITICAL |
|
||
1147 | Pending | Approved | View | VERIFIED | 27-12-2024 | 0142041 | Thiruvananthapuram | Local purchase | nursing station | PHC AMBOORI | Oxygen Concentrator | NON CRITICAL |
|
||
1148 | Pending | Approved | View | Equipment completed 9+ year, reapairing cost of equipment is more than 70%,Both criteria for rber met | 09-08-2024 | 0942852 | Palakkad | REMI ELEKTROTECHNIK LTD. | LABORATORY | PHC PUDUSSERY | Centrifuge | NON CRITICAL |
|
||
1149 | Pending | Approved | View | VERIFIED | 31-12-2024 | 0143192 | Thiruvananthapuram | Lifecare | FWC | PHC CHETTIVILAKAM | B P Apparatus | NON CRITICAL |
|
||
1150 | Pending | Approved | View | VERIFIED | 24-08-2024 | 0142928 | Thiruvananthapuram | Lifecare | FWC | PHC CHEMBOOR | B P Apparatus | NON CRITICAL |
|
||
1151 | Pending | Approved | View | VERIFIED | 12-09-2024 | 0142623 | Thiruvananthapuram | LOCAL | FWC | 24X7 PHC VALIYATHURA | Nebuliser | NON CRITICAL |
|
||
1152 | Pending | Approved | View | VERIFIED | 08-07-2024 | 0142953 | Thiruvananthapuram | TRUCHECK | OP | UPHC SECRTERIATE (RAJAJI NAGAR) | Nebuliser | NON CRITICAL |
|
||
1153 | Pending | Approved | View | VERIFIED | 07-12-2024 | 0142962 | Thiruvananthapuram | NA | OP | UPHC SECRTERIATE (RAJAJI NAGAR) | Nebuliser | NON CRITICAL |
|
||
1154 | Pending | Approved | View | VERIFIED | 12-09-2024 | 0142613 | Thiruvananthapuram | TRUCHECK | PP UNIT | 24X7 PHC VALIYATHURA | Nebuliser | NON CRITICAL |
|
||
1155 | Pending | Approved | View | VERIFIED | 12-09-2024 | 0142615 | Thiruvananthapuram | Ready Matic | PP UNIT | 24X7 PHC VALIYATHURA | B P Apparatus | NON CRITICAL |
|
||
1156 | Pending | Approved | View | VERIFIED | 09-09-2024 | 0142645 | Thiruvananthapuram | Infi | OP | 24X7 PHC VALIYATHURA | B P Apparatus | NON CRITICAL |
|
||
1157 | Pending | Approved | View | Equipment completed 5+ year,repairing cost of equipment is more than 65% of asset vaule, both criteria foir RBER met | 20-06-2024 | 1031479 | Malappuram | NA | LABORATORY | CHC CHUNGATHARA | Haemo Mixer | NON CRITICAL |
|
||
1158 | Pending | Approved | View | VERIFIED | 02-09-2024 | 0531179 | Kottayam | Godrej | OP | CHC ULLANADU | Pharmacy Refrigerator | NON CRITICAL |
|
||
1159 | Pending | Approved | View | Equipment completed 5+ year, repairing cost of equipment is 74.9% of asset value, both criteria for rber met | 01-03-2024 | 1042665 | Malappuram | Meril Diagnostics Pvt,Ltd. | LABORATORY | PHC Kodur | Biochemistry Analyzer | CRITICAL |
|
||
1160 | Pending | Approved | View | VERIFIED | 08-08-2024 | 0144332 | Thiruvananthapuram | AccuSure | immunization | 24X7 PHC POZHIYOOR | B P Apparatus | NON CRITICAL |
|
||
1161 | Pending | Approved | View | VERIFIED | 08-08-2024 | 0144374 | Thiruvananthapuram | Elkometer | FWC | 24X7 PHC POZHIYOOR | B P Apparatus | NON CRITICAL |
|
||
1162 | Pending | Approved | View | VERIFIED | 31-12-2024 | 0144268 | Thiruvananthapuram | Local purchase | nursing station | PHC MARANALLOOR | Oxygen Concentrator | NON CRITICAL |
|
||
1163 | Pending | Approved | View | VERIFIED | 27-03-2024 | 0131196 | Thiruvananthapuram | Lifecare | WARD | CHC POOVAR | Nebuliser | NON CRITICAL |
|
||
1164 | Pending | Approved | View | verified | 12-02-2024 | 0743851 | Ernakulam | Lifecare | observation room | 24X7 PHC KEEZHMAD | Nebuliser | NON CRITICAL |
|
||
1165 | Pending | Approved | View | Verified- Recommender RBER | 16-01-2024 | 0144289 | Thiruvananthapuram | NA | LABORATORY | PHC MARANALLOOR | Centrifuge | NON CRITICAL |
|
||
1166 | Pending | Approved | View | VERIFIED | 10-04-2024 | 0743158 | Ernakulam | nill | observation room | 24X7 PHC CHOWARA | X Ray Film Viewer | NON CRITICAL |
|
||
1167 | Pending | Approved | View | VERIFIED | 08-01-2024 | 0144415 | Thiruvananthapuram | EKCO | nurse station | PHC PARANIYAM | B P Apparatus | NON CRITICAL |
|
||
1168 | Pending | Approved | View | VERIFIED | 16-01-2024 | 0144538 | Thiruvananthapuram | ANITA INDUSTRIES | nurse station | PHC PARANIYAM | B P Apparatus | NON CRITICAL |
|
||
1169 | Pending | Approved | View | VERIFIED | 11-11-2024 | 0144454 | Thiruvananthapuram | Whirlpool | LABORATORY | PHC CHEMBOOR | Pharmacy Refrigerator | NON CRITICAL |
|
||
1170 | Pending | Approved | View | VERIFIED | 24-08-2024 | 0144565 | Thiruvananthapuram | DIAMOND | OP | PHC KULATHOOR | B P Apparatus | NON CRITICAL |
|
||
1171 | Pending | Approved | View | VERIFIED | 10-07-2024 | 0240494 | Kollam | LG | OP | 24X7 PHC PERUMON | Pharmacy Refrigerator | NON CRITICAL |
|
||
1172 | Pending | Approved | View | Equipment completed 6 year 10 months repairing cost of equipment is 81%, both criteria for rber met | 17-10-2024 | 1042931 | Malappuram | Labtronics | LABORATORY | UPHC PANNAKKAD | Colorimeter | NON CRITICAL |
|
||
1173 | Pending | Approved | View | Equipment completed 6 year 2 months with an asset cost of rupees 850, Quotation not submitted since spares are not available in the market | 10-12-2024 | 0944102 | Palakkad | Unigrip | OP | PHC PALLIPURAM | Weighing Machine | NON CRITICAL |
|
||
1174 | Pending | Approved | View | Equipment completed 6+ years, with an asset cost of rupees 2200,Quotation not submitted since spares are not available in the market | 13-05-2024 | 1040230 | Malappuram | OMRON HEALTHCARE CO.LTD. | OP | 24X7 PHC THIRUNNAVAYA | B P Apparatus | NON CRITICAL |
|
||
1175 | Pending | Approved | View | The steriliser installed on 30/06/2000, with an asset cost of rupees 3800.is in critical condition, extensive rusting and leaking compromise its structural integrity,posing risk to safety and reliability during operation | 12-02-2024 | 1031688 | Malappuram | nill | OP | 24X7 PHC PERUVALLUR | Sterlizer | NON CRITICAL |
|
||
1176 | Pending | Approved | View | Equipment completed 6 year 9 months the equipment is non repairable as per authorised dealer | 17-10-2024 | 0921621 | Palakkad | Tanita India Pvt Ltd. | NCD CLINIC | THQH PATTAMBI | Body Composition Analyser | NON CRITICAL |
|
||
1177 | Pending | Approved | View | VERIFIED | 22-05-2024 | 0744574 | Ernakulam | EKCO | OP | PHC MUDAKUZHA | B P Apparatus | NON CRITICAL |
|
||
1178 | Pending | Approved | View | VERIFIED | 27-03-2024 | 0732810 | Ernakulam | Lifecare | STORE | CHC KALADY | Nebuliser | NON CRITICAL |
|
||
1179 | Pending | Approved | View | VERIFIED | 16-04-2024 | 0144606 | Thiruvananthapuram | Local purchase | nurse station | FHC Pallichal | Pharmacy Refrigerator | NON CRITICAL |
|
||
1180 | Pending | Approved | View | Equipment completed 10+ year Repairing cost of equipment is more than 100%,Both criteria for rber met | 01-08-2024 | 0921333 | Palakkad | B.BRAUN | DIALYSIS UNIT | THQH OTTAPALAM | Dialysis Machine | CRITICAL |
|
||
1181 | Pending | Approved | View | Equipment outlived 5 year 9 month, OEM informed that the model is obselate and service not available,End of life letter from Oem received as per tender clause 5.3.14.2 we recommend equipment for condemination | 31-05-2024 | 0921312 | Palakkad | STRIKER | OT | THQH OTTAPALAM | Drill System | NON CRITICAL |
|
||
1182 | Pending | Approved | View | Equipment completed 7+ year, repairing cost of the equipment is more than 100%,both criteria for rber met | 31-05-2024 | 0943853 | Palakkad | Champion | LABORATORY | 24X7 PHC PUDUR | UPS | NON CRITICAL |
|
||
1183 | Pending | Approved | View | VERIFIED | 02-09-2024 | 0520010 | Kottayam | DISHA | CASUALTY | THQH VAIKOM | B P Apparatus | NON CRITICAL |
|
||
1184 | Pending | Approved | View | VERIFIED | 31-08-2024 | 0520011 | Kottayam | DISHA | CASUALTY | THQH VAIKOM | B P Apparatus | NON CRITICAL |
|
||
1185 | Pending | Approved | View | VERIFIED | 02-11-2023 | 0732826 | Ernakulam | Lifecare | EYE OP | CHC MALAYIDAMTHURUTH | B P Apparatus | NON CRITICAL |
|
||
1186 | Pending | Approved | View | Equipment installed on 04/09/2018 andf aged upto 5+ year, with an asset cost of rupees 1600.Quotation not submitted since spares are not available in the market | 16-04-2024 | 0943890 | Palakkad | SAMSO | OP | 24X7 PHC PUDUR | Weighing Machine | NON CRITICAL |
|
||
1187 | Pending | Approved | View | Equipment completed 3 year 9 months with an asst cost of rupees 1600,Quotation not submitted since spares are not available in the market | 29-08-2024 | 0942933 | Palakkad | SAMSO | OP | PHC POOKATUKAVU | Weighing Machine | NON CRITICAL |
|
||
1188 | Pending | Approved | View | Equipment installed on 12-08-2014 and outlived 9+ years, End of life letter provided by oem states that equipment is very old model and spareparts are not available | 06-04-2024 | 0921266 | Palakkad | EDAN | ECG ROOM | THQH OTTAPALAM | ECG Machine | CRITICAL |
|
||
1189 | Pending | Approved | View | Verified- Recommended RBER | 01-12-2023 | 0144529 | Thiruvananthapuram | Misaki | nursing station | PHC KANJIRAMKULAM | Weighing Machine | NON CRITICAL |
|
||
1190 | Pending | Approved | View | Verified- Recommended RBER | 01-12-2023 | 0144530 | Thiruvananthapuram | EKCO | OP | PHC KANJIRAMKULAM | B P Apparatus | NON CRITICAL |
|
||
1191 | Pending | Approved | View | verified | 23-08-2024 | 0520416 | Kottayam | Voltech | RADIOLOGY | THQH PAMPADY | UPS | NON CRITICAL |
|
||
1192 | Pending | Approved | View | Equipment installled on 01-06-2011 and aged upto 13 years,corrosion has created cracks in between the lowerside of chamber and boiler causes water and steam leak which will affect sterilisation process and safety.moreover OEM have notified EOL for the said equipment | 17-01-2024 | 0921252 | Palakkad | Reico | OT | THQH OTTAPALAM | Sterlizer | NON CRITICAL |
|
||
1193 | Pending | Approved | View | Equipment completed 8 year 4 months with an asset cost of rupees 1353, Quotation not submitted since spares are not available in the market | 19-09-2024 | 0930561 | Palakkad | OMRON HEALTHCARE CO.LTD. | observation room | CHC KOPPAM | Nebuliser | NON CRITICAL |
|
||
1194 | Pending | Approved | View | Equipment completed 5+ year, repairing cost of equipment is more than 70%,both criteria for rber met | 21-05-2024 | 0931515 | Palakkad | REMI ELEKTROTECHNIK LTD. | LABORATORY | CHC AGALI | Centrifuge | NON CRITICAL |
|
||
1195 | Pending | Approved | View | VERIFIED | 21-08-2024 | 0145025 | Thiruvananthapuram | Dr. Pette | LABORATORY | PHC UZHAMALAKKAL | Micro Pippette | NON CRITICAL |
|
||
1196 | Pending | Approved | View | VERIFIED | 27-05-2024 | 0143242 | Thiruvananthapuram | C-POULO | LABORATORY | PHC BALARAMAPURAM | Centrifuge | NON CRITICAL |
|
||
1197 | Pending | Approved | View | Equipment installed on 31-03-2011, aged upto 12+ year, Repairing cost of equipment is 61% both criteria for Rber met | 22-03-2024 | 0942574 | Palakkad | Whirlpool | OP | PHC CHALAVARA | Pharmacy Refrigerator | NON CRITICAL |
|
||
1198 | Pending | Approved | View | VERIFIED | 01-07-2024 | 0541269 | Kottayam | SAMSO | SUB CENTRE | PHC KARIKKATTOOR | Weighing Machine | NON CRITICAL |
|
||
1199 | Pending | Approved | View | VERIFIED | 14-06-2024 | 0640808 | Idukki | Infi | OP | PHC EDAVETTY | B P Apparatus | NON CRITICAL |
|
||
1200 | Pending | Approved | View | VERIFIED | 16-11-2024 | 0143340 | Thiruvananthapuram | OMRON | observation room | PHC MANICKAL | B P Apparatus | NON CRITICAL |
|
||
1201 | Pending | Approved | View | VERIFIED | 21-02-2024 | 0543918 | Kottayam | Infi | OP | 24X7 PHC PARAMPUZHA | B P Apparatus | NON CRITICAL |
|
||
1202 | Pending | Approved | View | Verified- Recommended RBER | 09-01-2024 | 0520392 | Kottayam | Godrej | OP | THQH KURAVILANGAD | Pharmacy Refrigerator | NON CRITICAL |
|
||
1203 | Pending | Approved | View | VERIFIED | 22-06-2024 | 0145062 | Thiruvananthapuram | ROMSONS | FWC | PHC THONNAKAL | B P Apparatus | NON CRITICAL |
|
||
1204 | Pending | Approved | View | VERIFIED | 22-06-2024 | 0145063 | Thiruvananthapuram | ANITA INDUSTRIES | FWC | PHC THONNAKAL | B P Apparatus | NON CRITICAL |
|
||
1205 | Pending | Approved | View | VERIFIED | 22-06-2024 | 0145071 | Thiruvananthapuram | SAMSO | FWC | PHC THONNAKAL | Weighing Machine | NON CRITICAL |
|
||
1206 | Pending | Approved | View | Equipment completed 4 year 11 month with an asset cost of rupees 1600, Quotation not submitted since spares are not available in the market. | 06-11-2024 | 0832716 | Thrissur | SAMSO | public health | CHC MALA | Weighing Machine | NON CRITICAL |
|
||
1207 | Pending | Approved | View | Verified- Recommended RBER | 31-10-2023 | 0145863 | Thiruvananthapuram | SAMSO | FWC | 24X7 PHC THIRUVALLOM | Weighing Machine | NON CRITICAL |
|
||
1208 | Pending | Approved | View | VERIFIED | 06-04-2024 | 0145910 | Thiruvananthapuram | DIAMOND | FWC | PHC Mukkola | B P Apparatus | NON CRITICAL |
|
||
1209 | Pending | Approved | View | Equipment completed 7 year 10 month, with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market | 05-11-2024 | 0822849 | Thrissur | OMRON | public health | PHC VARAVOOR | B P Apparatus | NON CRITICAL |
|
||
1210 | Pending | Approved | View | Equipment completed 4 year 8 months with an asset cost of rupees 2200, Quotaion not submitted since spares are not available in the market | 16-09-2024 | 0832307 | Thrissur | Lifecare | public health | CHC VELLANIKKARA | B P Apparatus | NON CRITICAL |
|
||
1211 | Pending | Approved | View | Equipment completed 5+ year with an asset cost of rupees 1600, Quotation not submitted since spares are not available in the market | 09-12-2024 | 0822918 | Thrissur | SAMSO | public health | PHC OLLUKKARA | Weighing Machine | NON CRITICAL |
|
||
1212 | Pending | Approved | View | VERIFIED | 20-03-2024 | 0131474 | Thiruvananthapuram | Lifecare | nurse station | CHC POOVAR | Nebuliser | NON CRITICAL |
|
||
1213 | Pending | Approved | View | Equipment completed 3+ year with an asset cost of rupees 2200,Quotation not submitted since spares are not available in the market | 25-09-2023 | 0845573 | Thrissur | K-Life | OP | PHC AVINISSERY | B P Apparatus | NON CRITICAL |
|
||
1214 | Pending | Approved | View | Equipment completed 9 year,repairing cost morethan 100% both criteria for rber met | 15-04-2024 | 0943965 | Palakkad | DIAMOND | SUB CENTRE | PHC PUDUPPARIYARAM | B P Apparatus | NON CRITICAL |
|
||
1215 | Pending | Approved | View | VERIFIED | 20-07-2024 | 0145158 | Thiruvananthapuram | DIAMOND | FWC | PHC CHENKAL | B P Apparatus | NON CRITICAL |
|
||
1216 | Pending | Approved | View | Equipment completed 4 year 11 months with an asset cost of rupees 1600, Quotation not submitted since spares are not available in the market | 04-12-2024 | 0822978 | Thrissur | SAMSO | public health | PHC MADAKKATHARA | Weighing Machine | NON CRITICAL |
|
||
1217 | Pending | Approved | View | Equipment completed 5 year, quotation not submitted since spares are not available in the market | 03-01-2024 | 0944501 | Palakkad | perfecxa | SUB CENTRE | PHC PUDUSSERY | B P Apparatus | NON CRITICAL |
|
||
1218 | Pending | Approved | View | verified | 19-12-2023 | 0747097 | Ernakulam | Easy Care | SUB CENTRE | UPHC KADAVANTHRA | B P Apparatus | NON CRITICAL |
|
||
1219 | Pending | Approved | View | Equipment completed 4+year, quotation not submitted since spares are not available in the market | 15-01-2024 | 1011192 | Malappuram | Rossmax | post natal ward | WOMEN & CHILDREN HOSPITAL PONNANI | Weighing Machine | NON CRITICAL |
|
||
1220 | Pending | Approved | View | Equipment completed 10+ year , Enquired oem for service support and found that oem Not existing, arranged local vendor for service found that repairing cost of equipment is 87% of asset value both criteria for rber met | 24-06-2024 | 1112120 | Kozhikode | NIL | LABORATORY | CDST Lab MCH Kozhikode | Deep Freezer | NON CRITICAL |
|
||
1221 | Pending | Approved | View | VERIFIED | 15-07-2024 | 0133664 | Thiruvananthapuram | Infi | nurse station | CHC VAKKOM | B P Apparatus | NON CRITICAL |
|
||
1222 | Pending | Approved | View | Equipment completed 11+ year,Repairing cost of equipment is 69% of asset value,both criteria for rber met | 23-04-2024 | 0812768 | Thrissur | Radiant Enterprise | eye ot | GH IRINJALAKUDA | Fumigator | NON CRITICAL |
|
||
1223 | Pending | Approved | View | Equipment completed 5+ year repairing cost of equipment is 69% of asset value, both criteriua for rber met | 16-07-2024 | 0845892 | Thrissur | DIAMOND | SUBCENTER | PHC NALUKETTU | B P Apparatus | NON CRITICAL |
|
||
1224 | Pending | Approved | View | VERIFIED | 08-07-2024 | 0131676 | Thiruvananthapuram | SAMSO | FWC | CHC PALODE | Weighing Machine | NON CRITICAL |
|
||
1225 | Pending | Approved | View | Equipment completed 5+ year with an asset cost of rupees 6000.actual book value of equipment is 2590, Quotation not submitted since spares are not available in the market | 24-12-2024 | 0944978 | Palakkad | SAMSO | SUB CENTRE | PHC KANNAMBRA | Weighing Machine | NON CRITICAL |
|
||
1226 | Pending | Approved | View | VERIFIED | 29-06-2024 | 0421468 | Alappuzha | Neona | RADIOLOGY | THQH CHERTHALA | UPS | NON CRITICAL |
|
||
1227 | Pending | Approved | View | Equipment outlived 5+ year repairing cost of equipment is more than 100%, both criteria for RBER Met | 25-03-2024 | 1240847 | Wayanad | SAMSO | SUB CENTRE | PHC VENGAPALLY | Weighing Machine | NON CRITICAL |
|
||
1228 | Pending | Approved | View | Equipment completed 6 year 8 months repairing cost of equipment is 86% of asset value, both criteria for rber met | 28-09-2024 | 0921812 | Palakkad | SOCOMEC | X RAY | THQH MANNARKKAD | UPS | NON CRITICAL |
|
||
1229 | Pending | Approved | View | Equipment completed 4+ year, with an asset cost of rupees 1353, Quotation not submitted since spares are not available in the market | 03-07-2024 | 0912777 | Palakkad | K life | PICU | WOMEN & CHILDREN HOSPITAL PALAKKAD | Nebuliser | NON CRITICAL |
|
||
1230 | Pending | Approved | View | Equipment completed 5 year 10 month , repairing cost of equipment is 74.8% of asset value, both criteria for rber met as per tender clause 5.3.14.1 we recommend equipment for condemination | 31-12-2024 | 0912926 | Palakkad | AKAS MEDICAL | CATH ICU | DH PALAKKAD | Syringe Pump | CRITICAL |
|
||
1231 | Pending | Approved | View | Documents verified- Recommended RBER | 29-11-2023 | 0115858 | Thiruvananthapuram | DIAMOND | WARD 8 | GH THIRUVANANTHAPURAM | B P Apparatus | NON CRITICAL |
|
||
1232 | Pending | Approved | View | VERIFIED EOL | 23-04-2024 | 0412702 | Alappuzha | OPTOPOL TECHNOLOGY | EYE OP | GH ALAPPUZHA | Automatic Field Analyzer | NON CRITICAL |
|
||
1233 | Pending | Approved | View | VERIFIED | 16-10-2024 | 0115873 | Thiruvananthapuram | Safe An B Touch | SURGICAL WARD | GH THIRUVANANTHAPURAM | Nebuliser | NON CRITICAL |
|
||
1234 | Pending | Approved | View | VERIFIED | 22-04-2024 | 0115816 | Thiruvananthapuram | Lifecare | SURGICAL WARD | GH THIRUVANANTHAPURAM | B P Apparatus | NON CRITICAL |
|
||
1235 | Pending | Approved | View | Equipment outlived 12 years, repairing cost of equipment is more than 100%,Both criteria for RBER Met | 24-04-2024 | 1312665 | Kannur | MINDRAY | LABORATORY | DH KANNUR | Haematology Analyser | CRITICAL |
|
||
1236 | Pending | Approved | View | Equipment completed 5+ year oem informed that powered instruments , stryker instruments core is an obselate model as per tender clause 5.3.14.2 we recommend equipment condition for condemination | 04-07-2024 | 1112353 | Kozhikode | STRYKER | OT | GH KOZHIKODE | Powered Instrument Drive | NON CRITICAL |
|
||
1237 | Pending | Approved | View | VERIFIED | 09-08-2024 | 0722944 | Ernakulam | BIO PLUS | OT | THQH THRIPUNITHURA | Nebuliser | NON CRITICAL |
|
||
1238 | Pending | Approved | View | VERIFIED | 24-07-2024 | 0722967 | Ernakulam | S Cure | CASUALTY | THQH NORTH PARAVOOR | Nebuliser | NON CRITICAL |
|
||
1239 | Pending | Approved | View | VERIFIED | 24-07-2024 | 0722969 | Ernakulam | LIFE LINE | CASUALTY | THQH NORTH PARAVOOR | Nebuliser | NON CRITICAL |
|
||
1240 | Pending | Approved | View | VERIFIED | 24-07-2024 | 0723055 | Ernakulam | AccuSure | CASUALTY | THQH KARUVELIPPADY | Nebuliser | NON CRITICAL |
|
||
1241 | Pending | Approved | View | Equipment completed 3 year 6 month,Micropipette piston rod damaged and volume is not acheiving properly, Quotation not submitted since spares are not available in the market | 25-03-2024 | 1212069 | Wayanad | EPPENDROF | LABORATORY | Public Health Laboratory Sultan Bathery | Pippette | NON CRITICAL |
|
||
1242 | Pending | Approved | View | Equipment completed 3 year 6 month, quotation not submitted since spares are not available in the market | 25-03-2024 | 1212070 | Wayanad | EPPENDROF | LABORATORY | Public Health Laboratory Sultan Bathery | Pippette | NON CRITICAL |
|
||
1243 | Pending | Approved | View | Equipment installed on 09-06-2016 and aged upto 7 year 9 months, the equipment get by donation and there is no supplier details /oem details available so as per tender clause 5.3.14.2 we recommend equipment for condemination | 25-03-2024 | 1212072 | Wayanad | TARSONS | LABORATORY | Public Health Laboratory Sultan Bathery | Centrifuge | NON CRITICAL |
|
||
1244 | Pending | Approved | View | verified | 13-11-2023 | 0743180 | Ernakulam | OMRON | public health | 24X7 PHC CHOWARA | B P Apparatus | NON CRITICAL |
|
||
1245 | Pending | Approved | View | VERIFIED | 04-04-2024 | 0743211 | Ernakulam | DIAMOND | PALIATIVE | 24X7 PHC CHOWARA | B P Apparatus | NON CRITICAL |
|
||
1246 | Pending | Approved | View | VERIFIED | 09-08-2024 | 0722216 | Ernakulam | nill | CASUALTY | THQH KOTHAMANGALAM | Suction Apparatus | NON CRITICAL |
|
||
1247 | Pending | Approved | View | VERIFIED | 24-07-2024 | 0740809 | Ernakulam | SAMSO | OP | 24X7 PHC PANANGAD | Weighing Machine | NON CRITICAL |
|
||
1248 | Pending | Approved | View | Equipment installed on 27-03-2019 outlived 4 year 11 month, repairing cost more than 100%.Both criteria for RBER Met | 15-03-2024 | 1130263 | Kozhikode | SCIENTIFIC MEDICAL SYSTEMS | PHYSIOTHERAPY | CHC OLAVANNA | Digital Stimulator (IFT+MST+TENS) | NON CRITICAL |
|
||
1249 | Pending | Approved | View | Equipment outlived 13+ year with an asset cost of rupees 1300, Quotation not submitted since spares are not available in the market. | 06-12-2024 | 0845902 | Thrissur | Misaki | SUB CENTRE | PHC KAKKAD | Weighing Machine | NON CRITICAL |
|
||
1250 | Pending | Approved | View | VERIFIED | 17-08-2024 | 0145396 | Thiruvananthapuram | Yuwell | OP | PHC KILIMANOOR | Oxygen Concentrator | NON CRITICAL |
|
||
1251 | Pending | Approved | View | VERIFIED | 10-06-2024 | 0131739 | Thiruvananthapuram | Local purchase | OT | CHC KESAVAPURAM | OT Table | CRITICAL |
|
||
1252 | Pending | Approved | View | VERIFIED | 10-06-2024 | 0131743 | Thiruvananthapuram | Local purchase | OT | CHC KESAVAPURAM | Infant Radiant Warmer with Phototherapy Unit | CRITICAL |
|
||
1253 | Pending | Approved | View | Equipment completed 5 year with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market | 20-11-2024 | 1441587 | KASARGODE | Not available | SUB CENTRE | PHC KUMBADAJE | B P Apparatus | NON CRITICAL |
|
||
1254 | Pending | Approved | View | VERIFIED | 03-09-2024 | 0530942 | Kottayam | S Cure | SUB CENTRE | CHC PAIKA | B P Apparatus | NON CRITICAL |
|
||
1255 | Pending | Approved | View | Verified- Recommended RBER | 17-09-2023 | 0114684 | Thiruvananthapuram | ELKO | WARD2 | GH THIRUVANANTHAPURAM | B P Apparatus | NON CRITICAL |
|
||
1256 | Pending | Approved | View | VERIFIED | 20-03-2024 | 0110931 | Thiruvananthapuram | EKCO | COVID ICU | GH THIRUVANANTHAPURAM | B P Apparatus | NON CRITICAL |
|
||
1257 | Pending | Approved | View | VERIFIED | 22-04-2024 | 0110982 | Thiruvananthapuram | HLL Lifecare Limited | FSW | GH THIRUVANANTHAPURAM | B P Apparatus | NON CRITICAL |
|
||
1258 | Pending | Approved | View | VERIFIED | 22-02-2024 | 0310189 | Pathanamthitta | HEMOPHARM | BLOOD BANK | GH PATHANMTHITTA | BLOOD BAG TUBE SEALER | NON CRITICAL |
|
||
1259 | Pending | Approved | View | Equipment aged upto 17+ year , while source replacement parts, found manufacture PICKS Ceased its operations and manufacture not exisiting, the equipment have no type approval condisering the circumstance manufacture doesnot exist and potential safety risk and legal consequences we strongly recommend equipment for condemination | 24-09-2024 | 0810835 | Thrissur | Picks Medical Systems PVT.LTD. | OT | GH THRISSUR | C-ARM | NON CRITICAL |
|
||
1260 | Pending | Approved | View | Equipment outlived 13+ years, Repairing cost of equipment is 72% of asset value, both criteria for Rber met | 18-04-2024 | 0831056 | Thrissur | REMI ELEKTROTECHNIK LTD. | LABORATORY | CHC VALAPAD | Centrifuge | NON CRITICAL |
|
||
1261 | Pending | Approved | View | Equipment completed 5+year, reapairing cost of equipment is 60% of asset value,Both criteria for RBER met | 23-05-2024 | 0931191 | Palakkad | NA | LABORATORY | CHC ALANALLUR | Centrifuge | NON CRITICAL |
|
||
1262 | Pending | Approved | View | Equipment completed 17+ year, the model has no longer service available and recommend for condemination, as per tender clause 5.3.14.2 end of life declared model | 16-07-2024 | 0812861 | Thrissur | premir medical system | OT | GH THRISSUR | Ventilator | CRITICAL |
|
||
1263 | Pending | Approved | View | VERIFIED | 18-10-2024 | 0114085 | Thiruvananthapuram | Sri Balaji Surgicals | eye ot | DISTRICT MODEL HOSPITAL PEROORKADA | Cautery | NON CRITICAL |
|
||
1264 | Pending | Approved | View | Equipment completed 7+ year, Repairing cost of equipment is more than 68%,Both criteria for rber met | 07-10-2023 | 0850948 | Thrissur | Beurer | OP | PHC ARIMBUR | Digital BP Apparatus | NON CRITICAL |
|
||
1265 | Pending | Approved | View | VERIFIED | 10-07-2024 | 0250517 | Kollam | Godrej | PAY WARD | DH KOLLAM | Pharmacy Refrigerator | NON CRITICAL |
|
||
1266 | Pending | Approved | View | Equipment outlived 8+ year,semi auto analyser Robonick prietest touch is an obselate model as per tender clause 5.3.14.2 we recommend equipment for condemination | 20-06-2024 | 1141133 | Kozhikode | Robonik | LABORATORY | UPHC KALLUNIRA | Biochemistry Analyzer | CRITICAL |
|
||
1267 | Pending | Approved | View | VERIFIED | 04-03-2024 | 0450354 | Alappuzha | Handy Neb | CH Ward | DH MAVELIKARA | Nebuliser | NON CRITICAL |
|
||
1268 | Pending | Approved | View | VERIFIED | 04-03-2024 | 0450353 | Alappuzha | Handy Neb | CH Ward | DH MAVELIKARA | Nebuliser | NON CRITICAL |
|
||
1269 | Pending | Approved | View | Equipment completed 7 year 7 months, repairing cost of equipment is more than 100%, both criteria for rber met | 02-08-2024 | 1450171 | KASARGODE | APC | LABORATORY | THQH MANGALPADY | UPS 3KVA | NON CRITICAL |
|
||
1270 | Pending | Approved | View | Equipment completed 6 year with an sset cost of rupees 1600, Quotation not submitted since spares are not available in the market | 22-11-2024 | 0950357 | Palakkad | AMKAY | FEMALE WARD | CHC CHERPULASSERY | Nebuliser | NON CRITICAL |
|
||
1271 | Pending | Approved | View | Equipment completed 6+ year with an asset cost of rupees 1353, Quotation not submitted since spares are not available in the market | 16-08-2024 | 0950035 | Palakkad | TRUCHEK | MM WARD | CHC NEMMARA | Nebuliser | NON CRITICAL |
|
||
1272 | Pending | Approved | View | Equipment completed 4 year 9 months with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market | 25-10-2024 | 0950047 | Palakkad | Med Vision | FM WARD | CHC NEMMARA | BP Apparatus | NON CRITICAL |
|
||
1273 | Pending | Approved | View | Equipment completed 7 year 9 month, repairing cost of equipment is more than 100% , both criteria met for rber | 04-01-2024 | 1350002 | Kannur | HYKON | ot 3 | DH KANNUR | UPS | NON CRITICAL |
|
||
1274 | Pending | Approved | View | Equipment completed 1 year 3 months with an asset cost of rupees 1568, Quotation not submitted since spares are not available in the market | 19-07-2024 | 1251156 | Wayanad | Sargen | OP | UPHC Vengoor | Nebuliser | NON CRITICAL |
|
||
1275 | Pending | Approved | View | Equipment installed on 20-06-2015 aged upto 8 years, Robonick prietest touch semi auto analyser is an obselate model, as per tender clause 5.3.14.2 we recommend equipment for condemination | 15-05-2024 | 1250331 | Wayanad | Robonik | LABORATORY | PHC KAPPUKUNNU | Semi-Auto Biochemistry Analyzer | NON CRITICAL |
|
||
1276 | Pending | Approved | View | VERIFIED | 24-07-2024 | 0750735 | Ernakulam | OMRON HEALTHCARE CO.LTD. | MM WARD | THQH FORTKOCHI | Nebuliser | NON CRITICAL |
|
||
1277 | Pending | Approved | View | VERIFIED | 09-10-2024 | 0651072 | Idukki | OPTIMA | OP | UPHC Kattappana | B P Apparatus | NON CRITICAL |
|
||
1278 | Pending | Approved | View | Equipment completed 5+ year with an asset cost of rupees 2117, Quotation not submitted since spares are not available in the market | 06-01-2024 | 0850150 | Thrissur | EKCO | public health | PHC THONURKARA | Digital BP Apparatus | NON CRITICAL |
|
||
1279 | Pending | Approved | View | VERIFIED | 24-10-2024 | 0514015 | Kottayam | York Scientific Industries PVT.LTD. | LABORATORY | GH KANJIRAPALLY | Hot Air Oven | NON CRITICAL |
|
||
1280 | Pending | Approved | View | VERIFIED | 27-11-2024 | 0514009 | Kottayam | Yuwell | PALIATIVE | GH KANJIRAPALLY | Oxygen Concentrator | NON CRITICAL |
|
||
1281 | Pending | Approved | View | verified | 12-11-2024 | 0532289 | Kottayam | Beurer | OP | CHC KOOTTICKAL | B P Apparatus | NON CRITICAL |
|
||
1282 | Pending | Approved | View | Donated item- No vendor available- Documents verified- Recommended RBER | 08-02-2024 | 0521303 | Kottayam | NA | WARD4 | THQH VAIKOM | Oxygen Concentrator | NON CRITICAL |
|
||
1283 | Pending | Approved | View | VERIFIED | 08-02-2024 | 0521282 | Kottayam | SAMSO | WARD 4 | THQH VAIKOM | Weighing Machine | NON CRITICAL |
|
||
1284 | Pending | Approved | View | VERIFIED | 28-02-2024 | 0521275 | Kottayam | AEROSOL | WARD 1 | THQH VAIKOM | Nebuliser | NON CRITICAL |
|
||
1285 | Pending | Approved | View | VERIFIED | 05-05-2024 | 0521267 | Kottayam | APC | LABORATORY | THQH VAIKOM | UPS | NON CRITICAL |
|
||
1286 | Pending | Approved | View | VERIFIED | 21-02-2024 | 0520889 | Kottayam | SOCOMEC | RADIOLOGY | THQH KURAVILANGAD | UPS | NON CRITICAL |
|
||
1287 | Pending | Approved | View | VERIFIED | 27-11-2024 | 0512690 | Kottayam | BPL | CH WARD | GH PALA | Pulse Oximeter | NON CRITICAL |
|
||
1288 | Pending | Approved | View | VERIFIED | 14-03-2024 | 0512731 | Kottayam | CONTEC MEDICAL SYSTEMS CO.LTD. | ECG ROOM | GH CHANGANACHERRY | ECG Machine | CRITICAL |
|
||
1289 | Pending | Approved | View | verified | 22-03-2024 | 0511111 | Kottayam | vitalograph | COPD clinic | GH KOTTAYAM | Vibash | NON CRITICAL |
|
||
1290 | Pending | Approved | View | VERIFIED | 22-11-2024 | 0513376 | Kottayam | Tokyo BOEKI medical system | LABORATORY | GH KANJIRAPALLY | Biochemistry Analyzer | CRITICAL |
|
||
1291 | Pending | Approved | View | Equipment outlived 3+years,repairing cost of equipment is more than 100%,both criteria for rber met | 06-07-2024 | 1250152 | Wayanad | FUJIFILM Corporation | RADIOLOGY | DH MANANTHAVADY | UPS 3KVA | NON CRITICAL |
|
||
1292 | Pending | Approved | View | Verified- Recommended RBER | 08-01-2024 | 0121473 | Thiruvananthapuram | Datex Ohmeda Inc. | OT | THQH CHIRAYINKEEZH | Anesthasia Machine | CRITICAL |
|
||
1293 | Pending | Approved | View | VERIFIED | 31-12-2024 | 0117047 | Thiruvananthapuram | Transasia Biomedical LTD. | LABORATORY | GMHC PEROORKADA THIRVANANTHAPURAM | Haemo Mixer | NON CRITICAL |
|
||
1294 | Pending | Approved | View | VERIFIED | 26-03-2024 | 0151136 | Thiruvananthapuram | Lifecare | PAIN & PALLIATIVE | PHC KANJIRAMKULAM | B P Apparatus | NON CRITICAL |
|
||
1295 | Pending | Approved | View | VERIFIED | 08-08-2024 | 0150661 | Thiruvananthapuram | OLIVE | OP | CHC MANAMBOOR | Oxygen Concentrator | NON CRITICAL |
|
||
1296 | Pending | Approved | View | VERIFIED | 18-06-2024 | 0151421 | Thiruvananthapuram | Lifecare | OP | THQH VARKALA | Nebulizer | NON CRITICAL |
|
||
1297 | Pending | Approved | View | VERIFIED | 10-07-2024 | 0151165 | Thiruvananthapuram | Infi | OP | PHC KOTTUKAL | B P Apparatus | NON CRITICAL |
|
||
1298 | Pending | Approved | View | verified | 24-04-2024 | 0150004 | Thiruvananthapuram | ANITA INDUSTRIES | OP | PHC CHETTIVILAKAM | BP Apparatus | NON CRITICAL |
|
||
1299 | Pending | Approved | View | VERIFIED | 27-12-2024 | 0151072 | Thiruvananthapuram | SAMSO | OP | PHC KALLIYOORE | Weighing Machine | NON CRITICAL |
|
||
1300 | Pending | Approved | View | verified | 10-07-2024 | 0117055 | Thiruvananthapuram | ELKO | WARD 7 | WOMEN & CHILDREN HOSPITAL THYCAUD | B P Apparatus | NON CRITICAL |
|
||
1301 | Pending | Approved | View | VERIFIED | 16-01-2024 | 0144531 | Thiruvananthapuram | LG | pharmacy | PHC KANJIRAMKULAM | Pharmacy Refrigerator | NON CRITICAL |
|
||
1302 | Pending | Approved | View | VERIFIED | 20-07-2024 | 0112464 | Thiruvananthapuram | SUPREME ENTERPRISES | CASUALTY | DH NEDUMANGAD | Suction Apparatus | NON CRITICAL |
|
||
1303 | Pending | Approved | View | verified | 23-08-2024 | 0115075 | Thiruvananthapuram | comet | WARD 1 | WOMEN & CHILDREN HOSPITAL THYCAUD | Pulse Oximeter | NON CRITICAL |
|
||
1304 | Pending | Approved | View | Equipment completed 6 year 1 month with an asset value of 1800, Quotation not submitted since spares are not available in the market | 21-11-2024 | 0950181 | Palakkad | Eagle | Physiotherapy | CHC KODUVAYUR | Weighing Machine | NON CRITICAL |
|
||
1305 | Pending | Approved | View | verified | 14-04-2024 | 0115893 | Thiruvananthapuram | MEDe VAC | WARD 2 | GH THIRUVANANTHAPURAM | Suction Apparatus | NON CRITICAL |
|
||
1306 | Pending | Approved | View | VERIFIED | 20-03-2024 | 0115760 | Thiruvananthapuram | LIFELINE | PMR | GH THIRUVANANTHAPURAM | B P Apparatus | NON CRITICAL |
|
||
1307 | Pending | Approved | View | VERIFIED | 15-07-2024 | 0115745 | Thiruvananthapuram | LOCAL | ENT OP | GH THIRUVANANTHAPURAM | ENT Head Light | NON CRITICAL |
|
||
1308 | Pending | Approved | View | VERIFIED | 22-01-2024 | 0150011 | Thiruvananthapuram | ECG Mac | ECG ROOM | PHC PARANIYAM | ECG Machine 3 Channel | CRITICAL |
|
||
1309 | Pending | Approved | View | Equipment completed 5+ year, oem hykon informed that the ups model HT1101L has determined to be non serviceable,because the model are no longer available on the market, so as per tender clause 5.3.14.2 we recommend equipment for condemination | 16-08-2024 | 1012599 | Malappuram | HYKON | LABORATORY | PH LAB MALAPPURAM | UPS | NON CRITICAL |
|
||
1310 | Pending | Approved | View | VERIFIED | 09-08-2024 | 0731979 | Ernakulam | Dr.Morepen | observation room | CHC MOOTHAKUNNAM | B P Apparatus | NON CRITICAL |
|
||
1311 | Pending | Approved | View | VERIFIED | 31-07-2024 | 0747126 | Ernakulam | OMRON HEALTHCARE CO.LTD. | OP | PHC CHOORNIKKARA | B P Apparatus | NON CRITICAL |
|
||
1312 | Pending | Approved | View | Verified- Recommended for RBER | 16-01-2024 | 0546279 | Kottayam | IBALL | LABORATORY | PHC MUTHOLY | UPS OF Hematology analyzer | NON CRITICAL |
|
||
1313 | Pending | Approved | View | Equipment completed 1 year 6 month with an asset cost of rupees zero.the supplier details are not available in the hospital because machiene received on covid pandemic. | 29-04-2024 | 1330481 | Kannur | Dynarex | PAIN & PALLIATIVE | CHC IRIVERI | Oxygen Concentrator | NON CRITICAL |
|
||
1314 | Pending | Approved | View | VERIFIED | 22-07-2024 | 0351230 | Pathanamthitta | Lifecare | CASUALTY | THQH RANNI | Nebuliser | NON CRITICAL |
|
||
1315 | Pending | Approved | View | Equipment completed 8 year 10 months, repairing cost of equipment is 81% of asset value, both criteria for rber met | 24-09-2024 | 1150439 | Kozhikode | Not available | OP | CHC Melady | X Ray Film Viewer | NON CRITICAL |
|
||
1316 | Pending | Approved | View | VERIFIED | 05-06-2024 | 0514057 | Kottayam | MediTech Industries | CASUALTY | GH PALA | BP Apparatus | NON CRITICAL |
|
||
1317 | Pending | Approved | View | Equipment completed 5 year 11 months, repairing cost of equipment is more than 100%, both criteria for rber met | 28-10-2024 | 1150127 | Kozhikode | ECG Mac | ECG ROOM | GH KOZHIKODE | ECG Machine | CRITICAL |
|
||
1318 | Pending | Approved | View | Equipment completed 5 year 8 months repaiirng cost of equipment is 83.67% of asset value,both criteria for rber met | 18-12-2024 | 1110769 | Kozhikode | Biotech | PMR | GH KOZHIKODE | Ultrasound Therapy Unit | NON CRITICAL |
|
||
1319 | Pending | Approved | View | Equipment completed 3+ year, quotation not submitted since spares are not available in the market | 16-01-2024 | 1150841 | Kozhikode | SAMSO | CASUALTY | THQH KOYILANDY | Digital Weighing Machine | NON CRITICAL |
|
||
1320 | Pending | Approved | View | VERIFIED | 07-03-2024 | 0511991 | Kottayam | Lifecare | WARD | K.R. NARAYANAN MEMORIAL SUPER SPECIALTY HOSPITAL UZHAVOOR | Nebuliser | NON CRITICAL |
|
||
1321 | Pending | Approved | View | VERIFIED | 21-01-2024 | 0546439 | Kottayam | Infi | WARD | 24X7 PHC MARANGATTUPALLY | Nebuliser | NON CRITICAL |
|
||
1322 | Pending | Approved | View | Equipment completed 8+ year repairing cost of equipment is more than 100%, both criteria for rber met | 24-06-2024 | 1130780 | Kozhikode | RAPID | LABORATORY | CHC Orkkatteri | Haematology Analyzer - 3 part | CRITICAL |
|
||
1323 | Pending | Approved | View | VERIFIED | 08-02-2024 | 0543737 | Kottayam | Infi | OP | PHC VELIYANNOOR | Digital BP Apparatus | NON CRITICAL |
|
||
1324 | Pending | Approved | View | Equipment completed 3 years and 1 month,with an asset cost of rupees 850, Quotation not submitted since spare parts are not available in the market | 16-12-2024 | 1350438 | Kannur | Beurer | OP | CHC PINARAYI | Weighing Machine | NON CRITICAL |
|
||
1325 | Pending | Approved | View | VERIFIED | 18-06-2024 | 0310227 | Pathanamthitta | Godrej | SURGICAL WARD | GH PATHANAMTHITTA | Refrigerator | NON CRITICAL |
|
||
1326 | Pending | Approved | View | VERIFIED | 31-07-2024 | 0731955 | Ernakulam | EUR SG Technology Germany | OP | CHC EDAPPALLY | Nebuliser | NON CRITICAL |
|
||
1327 | Pending | Approved | View | Equipment completed 1 year 10 month with an asset cost of rupees 1353,Quotation not submitted since spares are not available in the market | 16-09-2024 | 1412126 | KASARGODE | Point of Care | FM WARD | GH KASARAGODE | Nebuliser | NON CRITICAL |
|
||
1328 | Pending | Approved | View | Equipment completed 3+ year, with an asset cost of rupees zero,Quotation not submitted since spares are not available in the market | 17-07-2024 | 1051128 | Malappuram | Dr Morepen | OP | PHC THUVVUR | Digital BP Apparatus | NON CRITICAL |
|
||
1329 | Pending | Approved | View | VERIFIED | 28-09-2024 | 0740637 | Ernakulam | Lifecare | OP | 24X7 PHC KADAVOOR | Digital BP Apparatus | NON CRITICAL |
|
||
1330 | Pending | Approved | View | Equipment outlived 1 year 8 months, with an asset cost of rupees 5000, Quotation not submitted since spares are not available in the market | 19-11-2024 | 1051901 | Malappuram | Sknol | OP | PHC NANNAMBRA | Digital Weighing Machine | NON CRITICAL |
|
||
1331 | Pending | Approved | View | VERIFIED | 04-10-2024 | 0150842 | Thiruvananthapuram | Goley | WARD | THQH NEMOM | Suction Apparatus | NON CRITICAL |
|
||
1332 | Pending | Approved | View | VERIFIED | 16-10-2024 | 0150843 | Thiruvananthapuram | Universal Surgical Co. | WARD | THQH NEMOM | Suction Apparatus | NON CRITICAL |
|
||
1333 | Pending | Approved | View | Equipment completed 17+ year, repairing cost of equipment is 73% of asset value, both criteria for rber met as per tender clause 5.3.14.1 we recommend equipment for condemination | 29-11-2024 | 0850466 | Thrissur | ALMICRO | LABORATORY | CHC CHERPU | Laboratory Microscope | NON CRITICAL |
|
||
1334 | Pending | Approved | View | VERIFIED | 15-07-2024 | 0115329 | Thiruvananthapuram | LOCAL | OP | CDH PULAYANARKKOTTA | X Ray Film Viewer | NON CRITICAL |
|
||
1335 | Pending | Approved | View | VERIFIED | 26-06-2024 | 0514088 | Kottayam | MONARCH SURGICAL INSUSTRIES | ENT OP | GH KANJIRAPALLY | Suction Apparatus | NON CRITICAL |
|
||
1336 | Pending | Approved | View | VERIFIED | 16-04-2024 | 0751460 | Ernakulam | Hosco | Palliative care | PHC MOOKKANNOR | Oxygen Concentrator | NON CRITICAL |
|
||
1337 | Pending | Approved | View | VERIFIED | 07-10-2024 | 0514091 | Kottayam | Infi | CASUALTY | GH KANJIRAPALLY | Digital BP Apparatus | NON CRITICAL |
|
||
1338 | Pending | Approved | View | VERIFIED | 20-06-2024 | 0250987 | Kollam | BPL | OP | PHC TRIKOVILVATTOM | ECG Machine | CRITICAL |
|
||
1339 | Pending | Approved | View | VERIFIED | 10-10-2024 | 0550095 | Kottayam | Jiangsu Yuyue Medical Equipment & Supply Co.Ltd. | public health | PHC MEENACHIL | Oxygen Concentrator | NON CRITICAL |
|
||
1340 | Pending | Approved | View | VERIFIED | 23-08-2024 | 0250078 | Kollam | SOCOMEC | CASUALTY | THQH PUNALUR | UPS | NON CRITICAL |
|
||
1341 | Pending | Approved | View | Equipment completed 8 year, repairing cost of equipment is 81% of asset value, both criteria for rber met | 24-09-2024 | 1150349 | Kozhikode | LABTECH MEDICO (P) LTD | LABORATORY | GH KOZHIKODE | Micropipette 500uL | NON CRITICAL |
|
||
1342 | Pending | Approved | View | Equipment completed 16+ year Oem schiller issued obselate letter for defibrilator schiller -cardioplus, as per tender clause 5.3.14.2 we request tender inviting authority to make equipment condem | 13-12-2024 | 0850452 | Thrissur | SCHILLER | COPD | GH THRISSUR | Defibrillator | NON CRITICAL |
|
||
1343 | Pending | Approved | View | VERIFIED | 12-06-2024 | 0150483 | Thiruvananthapuram | ANITA INDUSTRIES | CASUALTY | GH THIRUVANANTHAPURAM | B P Apparatus | NON CRITICAL |
|
||
1344 | Pending | Approved | View | VERIFIED | 05-08-2024 | 0150924 | Thiruvananthapuram | Lucky Surgical | DIALYSIS UNIT | GH THIRUVANANTHAPURAM | Suction Apparatus | NON CRITICAL |
|
||
1345 | Pending | Approved | View | The jaundice meter MBJ-20 Installed on 30-07-2022,has shown significantly result variation and instability , compromising neonatal care,authorised supplier m/s biomedix solution identified a faulty main pcb with the microcontroller, and the oem repair cost are 83% of asset value making repair impractical, replacement part have not resolved the accuracy issues due to poor quality, we urge tender inviting authority to decommision and discard the mbj-20 immediately to ensure neonatal safety | 04-04-2024 | 0852873 | Thrissur | Shanghai International Holding Corp GMBH | LABOUR ROOM | THQH CHAVAKAD | Bilirubinometer | NON CRITICAL |
|
||
1346 | Pending | Approved | View | VERIFIED | 12-06-2024 | 0151011 | Thiruvananthapuram | EMCO MEDITEK PVT LTD. | ICU | GH THIRUVANANTHAPURAM | Syringe Pump | CRITICAL |
|
||
1347 | Pending | Approved | View | VERIFRIED | 06-09-2024 | 0151024 | Thiruvananthapuram | Diamond | SUBCENTER | PHC ANAD | B P Apparatus | NON CRITICAL |
|
||
1348 | Pending | Approved | View | VERIFIED | 23-08-2024 | 0153090 | Thiruvananthapuram | OMRON | High Dependency Unit(HDU) | DH NEDUMANGAD | Nebuliser | NON CRITICAL |
|
||
1349 | Pending | Approved | View | Equipment outlived 3+ year with an asset cost of rupees 2200, Quotation not submitted since spares are not available in the market | 10-05-2024 | 0951196 | Palakkad | LOCAL | OP | CHC KOPPAM | Digital BP Apparatus | NON CRITICAL |
|
||
1350 | Pending | Approved | View | VERIFIED | 27-09-2024 | 0151025 | Thiruvananthapuram | S.CURE | SUBCENTER | PHC ANAD | Weighing Machine | NON CRITICAL |
|
||
1351 | Pending | Approved | View | VERIFIED | 06-09-2024 | 0150758 | Thiruvananthapuram | OMRON HEALTHCARE CO.LTD. | FWC | CHC VENPAKAL | Digital BP Apparatus | NON CRITICAL |
|
||
1352 | Pending | Approved | View | Equipment completed 4+ year, supplier union chemsit informed that there are no available spare from manufacturer becuase the model is obselate,considering the circumstance and adherance to tender clause 5.3.14.2 we recommend tender inviting authority to consider this equipment for condemination | 23-08-2024 | 1453313 | KASARGODE | Guilin Woodpecker Medical Instrument Co,Ltd. | DENTAL DEPARTMENT | GH KASARAGODE | EndoMotor | NON CRITICAL |
|
||
1353 | Pending | Approved | View | VERIFIED | 10-07-2024 | 0151978 | Thiruvananthapuram | Local purchase | MALE WARD | THQH ATTINGAL | Oxygen Concentrator | NON CRITICAL |
|
||
1354 | Pending | Approved | View | Equipment installed on 24/04/2012 and aged upto 12+ year,repairing cost of equipment is more than 100%, both criteria for Rber met | 23-05-2024 | 0951823 | Palakkad | Guilin Woodpecker Instruments co.ltd | DENTAL DEPARTMENT | CHC KUZALMANNAM | Ultrasonic Scalar | NON CRITICAL |
|
||
1355 | Pending | Approved | View | Equipment completed 2+ year the supplier details is unavailable from end user as the equipment was donated without manufacturing as per the tender clause 5.3.1.4.2 we request for recommending equipment for condemination | 31-05-2024 | 0852928 | Thrissur | TATA | pharmacy | PHC ELAVALLY | Oxygen Concentrator | NON CRITICAL |
|
||
1356 | Pending | Approved | View | VERIFIED | 24-09-2024 | 0153925 | Thiruvananthapuram | LIFE LINE MEDICAL DEVICES | FEMALE WARD | THQH ATTINGAL | Nebulizer | NON CRITICAL |
|
||
1357 | Pending | Approved | View | Equipment completed 4 year 7 months with an asset cost of rupees 1562, Quotation not submitted since spares are not available in the market | 13-11-2024 | 0951635 | Palakkad | OMRON HEALTHCARE CO.LTD. | CASUALTY | CHC NEMMARA | Nebuliser | NON CRITICAL |
|
||
1358 | Pending | Approved | View | Equipment completed 6 year 9 months with an asset cost of rupees 2500, Quotation not submitted since spares are not available in the market | 04-12-2024 | 0852746 | Thrissur | OMRON | public health | CHC MULLASSERY | Digital BP Apparatus | NON CRITICAL |
|
||
1359 | Pending | Approved | View | Equipment completed 4+ year with an reference asset cost of 2200 rupees, quotation not submitted since spares are not available in the market | 21-08-2024 | 1352833 | Kannur | nill | casulity | THQH PERAVOOR | Nebuliser | NON CRITICAL |
|
||
1360 | Pending | Approved | View | VERIFIED | 04-09-2024 | 0251710 | Kollam | LG | pharmacy | UPHC KARUNAGAPALLY | Refrigerator | NON CRITICAL |
|
||
1361 | Pending | Approved | View | VERIFIED | 18-06-2024 | 0730400 | Ernakulam | Hemodiaz Life Sciences Pvt,Ltd. | Palliative | CHC KALADY | Oxygen Concentrator | NON CRITICAL |
|
||
1362 | Pending | Approved | View | VERIFIED | 18-10-2024 | 0150781 | Thiruvananthapuram | LOCAL | OP | TH Malayinkeezh | Nebuliser | NON CRITICAL |
|
||
1363 | Pending | Approved | View | VERIFIED | 09-10-2024 | 0150779 | Thiruvananthapuram | Infi | OP | TH Malayinkeezh | Nebuliser | NON CRITICAL |
|
||
1364 | Pending | Approved | View | Equipment completed 8 year 7 month, repairing cost of equipment is 76% of asset value, both criteria for rber met | 21-11-2024 | 0950191 | Palakkad | ANITA INDUSTRIES | causality | THQH PATTAMBI | BP Apparatus | NON CRITICAL |
|
||
1365 | Pending | Approved | View | Equipment completed 4 year 11 months with an asset cost of rupees 3199.Quotation not submitted since spares are not available in the market | 23-08-2024 | 1251619 | Wayanad | SAMSO | public health | CHC PULPALLY | Baby Weighing Machine-Digital | NON CRITICAL |
|
||
1366 | Pending | Approved | View | VERIFIED | 26-09-2024 | 0252087 | Kollam | S.CURE | SUBCENTER | UPHC VAADI | Weighing Machine Adult (Digital) | NON CRITICAL |
|
||
1367 | Pending | Approved | View | VERIFIED | 02-11-2024 | 0210006 | Kollam | MINDRAY | LABORATORY | DH KOLLAM | Haematology Analyser | CRITICAL |
|
||
1368 | Pending | Approved | View | VERIFIED | 27-08-2024 | 0543402 | Kottayam | MIC Electronics Ltd | PALLEATIVE UNIT | PHC T V Puram | Oxygen Concentrator | NON CRITICAL |
|
||
1369 | Pending | Approved | View | VERIFIED | 07-08-2024 | 0112643 | Thiruvananthapuram | TITAN | SNCU | SNCU IN SAT HOSPITAL | Oto Acoustic Emission Screener (OAE) | NON CRITICAL |
|
||
1370 | Pending | Approved | View | VERIFIED | 17-08-2024 | 0153418 | Thiruvananthapuram | OMRON HEALTHCARE CO.LTD. | WARD | CDH PULAYANARKKOTTA | Nebuliser | NON CRITICAL |
|
||
1371 | Pending | Approved | View | Equipment completed 1 year 5 month with an asset cost of rupees 1700. Quotation not submitted since spares are not available in the market | 05-12-2024 | 1250619 | Wayanad | Beurer | OP | Poothikkad - UHWC - Wayanad | Digital BP Apparatus | NON CRITICAL |
|
||
1372 | Pending | Approved | View | VERIFIED | 08-10-2024 | 0353557 | Pathanamthitta | AEROSOL | public health | NUHM PATHANAMTHITTA | Nebuliser | NON CRITICAL |
|
||
1373 | Pending | Approved | View | VERIFIED | 24-09-2024 | 0152034 | Thiruvananthapuram | HOSCO PRIVATE LIMITED | OP | PHC THONNAKAL | Oxygen Concentrator | NON CRITICAL |
|
||
1374 | Pending | Approved | View | VERIFIED | 18-09-2024 | 0152038 | Thiruvananthapuram | HOSCO PRIVATE LIMITED | PALIATIVE | CHC PALLIKKAL | Oxygen Concentrator | NON CRITICAL |
|
||
1375 | Pending | Approved | View | VERIFIED | 27-12-2024 | 0532849 | Kottayam | Yuwell | PAIN & PALLIATIVE | CHC KADAPLAMATTOM | Oxygen Concentrator | NON CRITICAL |
|
||
1376 | Pending | Approved | View | Equipment completed 4 year7 month with an asset cost of rupees 1935, Quotation not submitted as the necessary spares are not available in the market | 27-12-2024 | 0854769 | Thrissur | Sknol | OT | GH THRISSUR | Digital Weighing Machine | NON CRITICAL |
|