ASSET DATA
Asset Condemnation Approval/Rejection
Sr# | BER | RBER | View RBER Document | RBER Remark | Asset# / Barcode# | District | Manufacturer | Department | Facility Name | Asset | Criticality | Remark | Approve | Reject |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
1 | Pending | Approved | View | Equipment completed 9 year, End of life letter from oem received | 1340681 | Kannur | Robonik | LABORATORY | 24X7 PHC CHERUTHAZHAM | Biochemistry Analyzer | CRITICAL |
|
||
2 | Pending | Approved | View | Equipment completed 12+ year reapiring cost of equipment is 66%of asset value, both criteria met | 1345310 | Kannur | Lifeplus | DRESSING ROOM | 24X7 PHC KELAKAM | Sterlizer | NON CRITICAL |
|
||
3 | Pending | Approved | View | Totally Rusted- Verified- Recommended RBER | 0541748 | Kottayam | nill | WARD | 24X7 PHC MARANGATTUPALLY | Sterlizer | NON CRITICAL |
|
||
4 | Pending | Approved | View | Equipment completed 8 year 8 months, eol declared model robonick prietest touch as per tender clause 5.3.14.2 | 1340949 | Kannur | Robonik | LABORATORY | 24X7 PHC UDAYAGIRI | Biochemistry Analyzer | CRITICAL |
|
||
5 | Pending | Approved | View | Equipment completed 8+ year, repairing cost of equipment is more than 100%, both criteria for rber met | 1430055 | KASARGODE | Mindray | LABORATORY | CHC CHERUVATHUR | Haematology Analyser | CRITICAL |
|
||
6 | Pending | Approved | View | Documents verified- Quotation above 100%- Year met- Recommended for RBER | 0431232 | Alappuzha | Prokan | LABORATORY | CHC CHUNAKKARA | Haematology Analyser | CRITICAL |
|
||
7 | Pending | Approved | View | verified | 0530343 | Kottayam | NISCO INDIA | OP | CHC KOODALLOOR | B P Apparatus | NON CRITICAL |
|
||
8 | Pending | Approved | View | Equipment completed 7 year 7 month with an asset cost of rupees 850, quotation not submitted since spares are not available in the market | 0930044 | Palakkad | Ross max | OP | CHC KOZHINJAMPARA | Weighing Machine | NON CRITICAL |
|
||
9 | Pending | Approved | View | EQUIPMENT COMPLETED 10 YEAR 5 MONTH EOL DECLARED MODEL AS PER TENDER CLAUSE 5.3.14.2 RECOMMEND EQUIPMENT FOR CONDEMINATION | 1430390 | KASARGODE | Robonik | LABORATORY | CHC KUMBLA | Biochemistry Analyzer | CRITICAL |
|
||
10 | Pending | Approved | View | Equipment outlived 12 year, repairing cost of equipment is more than 100% with respect to asset value, Both criteria for RBR met | 1330311 | Kannur | REMI | LABORATORY | CHC MATTANNUR | Centrifuge | NON CRITICAL |
|
||
11 | Pending | Approved | View | Equipment completed 5 +year, quotation not submitted since spares are not available in the market | 1030698 | Malappuram | Comfo care | nursing station | CHC NEDUVA | Nebuliser | NON CRITICAL |
|
||
12 | Pending | Approved | View | Equipment completed 12+ year, repairing cost of equipment is 93% of asset value, both criteria for rber met | 1130777 | Kozhikode | REMI | LABORATORY | CHC Orkkatteri | Centrifuge | NON CRITICAL |
|
||
13 | Pending | Approved | View | Vendor not Existing- images and Documents verified | 0610106 | Idukki | jove automation | BLOOD BANK | DH IDUKKI | Blood Collection Monitor | NON CRITICAL |
|
||
14 | Pending | Approved | View | Images and Documents Verified- Year met- Below 10K asset- Recommended for RBEr | 0610631 | Idukki | Anita Industries | ISOLATION WARD | DH IDUKKI | B P Apparatus | NON CRITICAL |
|
||
15 | Pending | Approved | View | Equipment completed 12 year 11 month, Defibrilator philips heart startxl is an obselate model as per tender clause 5.3.14.2 we recommend equipment for condemination | 1410471 | KASARGODE | Philips | OT | DH KANHANGAD | Defibrillator | NON CRITICAL |
|
||
16 | Pending | Approved | View | Equipment completed 12 year,repairing cost of equipment is 84%,both criteria for rber met | 1410437 | KASARGODE | Allengers | OT | DH KANHANGAD | C-ARM | NON CRITICAL |
|
||
17 | Pending | Approved | View | Equipment completed 12 year, repairing cost of equipment is 60% of asset value, both criteria met for RBER | 1310517 | Kannur | Kelvinator | DENTAL DEPARTMENT | DH KANNUR | Pharmacy Refrigerator | NON CRITICAL |
|
||
18 | Pending | Approved | View | Equipment outlived 9+ year, repairing cost of equipment is more than 100%. both criteria met for RBER. | 1011708 | Malappuram | ADONICS | RADIOLOGY | DH NILAMBUR | X Ray Machine | CRITICAL |
|
||
19 | Pending | Approved | View | Equipment completed 17+ year ,repairing cost of equipment is more than 100%,the cost of repairing the equipment surpases its orginal asset value of rupees 9600, meeting the criteria for condemination | 0911233 | Palakkad | Electrolux | LABORATORY | DH PALAKKAD | Pharmacy Refrigerator | NON CRITICAL |
|
||
20 | Pending | Approved | View | Equipment outlived 19+ year,Orginal equipment manufacture provided End of life letter and recommend for decomission of equipment considering age of the equipment, increasing maintanence cost, and new technology standards. | 1010315 | Malappuram | Terumo penpol | BLOOD BANK | DH PERINTHALMANNA | Blood Collection Monitor | NON CRITICAL |
|
||
21 | Pending | Approved | View | Equipment completed 8 year,repairing cost of equipment is 80% of asset value,both critera met for rber | 1010870 | Malappuram | Hamilton | ICU | DH TIRUR | Ventilator | CRITICAL |
|
||
22 | Pending | Approved | View | Equipment completed 12 year,End of life declared model as per tenderclause5.3.14.2 criteria for rber met | 1012135 | Malappuram | Robonik | LABORATORY | DISTRICT TB CENTRE MANJERI | Biochemistry Analyzer | CRITICAL |
|
||
23 | Pending | Approved | View | Equipment outlived 13+year,the spare & components necessary to put the instruments back in working condition are no longer avaiable with oem Remi,model discontinued and letter from oem attached as per tender clause 5.3.14.2 | 1410848 | KASARGODE | Remi | BLOOD BANK | GH KASARAGODE | Blood Bank Refrigerator | NON CRITICAL |
|
||
24 | Pending | Approved | View | Equipment completed 13 years, reparing cost of equipment is 82% of asset value, both criteria met for rber | 1410966 | KASARGODE | Air Liquid | ICU | GH KASARAGODE | Ventilator | CRITICAL |
|
||
25 | Pending | Approved | View | Equipment outlived 13+year, Oem philips india ltd has declared Heartstart XL defibrilator as an end of support device, so as pewr tender clause 5.3.14.2 | 1410927 | KASARGODE | Philips | ICU | GH KASARAGODE | Defibrillator | NON CRITICAL |
|
||
26 | Pending | Approved | View | Spare not available with Vendor- Cyrix Explanation letter attached- Verified- Recommended RBER | 0510202 | Kottayam | Alan Electronic System | OT | GH KOTTAYAM | Cautery | NON CRITICAL |
|
||
27 | Pending | Approved | View | VERIFIED | 0510869 | Kottayam | MICRO SURGICAL EQUIPMENTS PVT LTD | WARD 6 | GH KOTTAYAM | B P Apparatus | NON CRITICAL |
|
||
28 | Pending | Approved | View | Verified- Recommended RBER- Machine very Old | 0510712 | Kottayam | Videocon | EYE OT | GH KOTTAYAM | Pharmacy Refrigerator | NON CRITICAL |
|
||
29 | Pending | Approved | View | Quotation and service report attached- Recommended RBER | 0112852 | Thiruvananthapuram | B BRAUN | DIALYSIS UNIT | GH NEYYATTINKARA | Dialysis Machine | CRITICAL |
|
||
30 | Pending | Approved | View | Mercury Spillage- Verified- Recommended RBER | 0511295 | Kottayam | ELKO METER | CASUALTY | GH PALA | B P Apparatus | NON CRITICAL |
|
||
31 | Pending | Approved | View | Equipment outlived 10 year 11 months,repairing cost of equipment is 75% of asset value,both critera for rber met | 1311179 | Kannur | Biorad | BLOOD BANK | GH THALASSERY | Elisa Washer | NON CRITICAL |
|
||
32 | Pending | Approved | View | Rber | 1311433 | Kannur | Videocon | LABORATORY | GH THALASSERY | Pharmacy Refrigerator | NON CRITICAL |
|
||
33 | Pending | Approved | View | Equipment covered 5 years, repairing cost is 75% , with respect to asset value, both criteria for RBER Met | 1311736 | Kannur | supreme | FM WARD | GH THALASSERY | Suction Apparatus | NON CRITICAL |
|
||
34 | Pending | Approved | View | VERIFIED | 0111959 | Thiruvananthapuram | ASCON | OT | GH THIRUVANANTHAPURAM | Operating Microscope | NON CRITICAL |
|
||
35 | Pending | Approved | View | VERIFIED | 0112023 | Thiruvananthapuram | APPASAMY ASSOCIATES | OT | GH THIRUVANANTHAPURAM | Phaco Emulsification Unit | NON CRITICAL |
|
||
36 | Pending | Approved | View | Verified- OEM Letter attached- Recommended RBER | 0111870 | Thiruvananthapuram | KARL STORZ | OT | GH THIRUVANANTHAPURAM | Laser lithotripter with flexible ureteroscope | NON CRITICAL |
|
||
37 | Pending | Approved | View | Documents verified- Both criteria met- Recommended RBER | 0111053 | Thiruvananthapuram | GE | COVID ICU | GH THIRUVANANTHAPURAM | Multipara Monitor | CRITICAL |
|
||
38 | Pending | Approved | View | VERIFIED | 0110234 | Thiruvananthapuram | L& T | MALE WARD | GH THIRUVANANTHAPURAM | Pulse Oximeter | NON CRITICAL |
|
||
39 | Pending | Approved | View | EQUIPMENT COMPLETED 10 YEARS,ASPEN JSB 1200 DISCONTINUED AND OBSELATE MODEL AS PER TENDER CLAUSE 5.3.14.2 RBER CRITERIA MET | 0810484 | Thrissur | Aspen | MICU | GH THRISSUR | Infusuion Pump | CRITICAL |
|
||
40 | Pending | Approved | View | Equipment completed 10 years, ASPEN JSB1200 IS AN DISCONTINUED MODEL AS PER TENDER CLAUSE 5.3.14.2 BOTH CRITERIA FOR RBER MET | 0810493 | Thrissur | Aspen | MICU | GH THRISSUR | Infusuion Pump | CRITICAL |
|
||
41 | Pending | Approved | View | Equipment outlived 16+ year, repairing cost of equipment is 74% with respect to asset value, Both criteria met for RBER as per tender clause 5.3.14.2 | 1111809 | Kozhikode | MCLOARDS | WARD 4 | GMHC KOZHIKODE | Suction Apparatus | NON CRITICAL |
|
||
42 | Pending | Approved | View | VERIFIED | 0341153 | Pathanamthitta | A-Aspen | PP UNIT | PHC ANICADU | ECG Machine | CRITICAL |
|
||
43 | Pending | Approved | View | Equipment completed 10 years,End of life declared model as per tenderclause 5.3.14.2 both criteria for rber met | 1345236 | Kannur | Robonik | LABORATORY | PHC CHITTARIPARAMBA | Biochemistry Analyzer | CRITICAL |
|
||
44 | Pending | Approved | View | Equipment outlived 10+ year, Repairing cost of equipment is more than 100%, Both criteria for RBER Met | 1342623 | Kannur | Aero | OP | PHC KANNAPURAM | Nebuliser | NON CRITICAL |
|
||
45 | Pending | Approved | View | Equipment completed 5 year, quotation not submitted since spares are not available in the market | 0941078 | Palakkad | OMRON | OP | PHC OZHALAPATHY | Weighing Machine | NON CRITICAL |
|
||
46 | Pending | Approved | View | EQUIPMENT COMPLETED 14+ YEAR ,HOSPITAL HAS DECIDED TO CONDEM THE EQUIPMENT, CONDEMINATION REPORT FROM HOSPITAL ATTACHED | 1041019 | Malappuram | LABOMED | LABORATORY | PHC OZHOOR | Laboratory Microscope | NON CRITICAL |
|
||
47 | Pending | Approved | View | Fully Rusted and Dilapidated condition- Images attached- Verified- Recommended for RBER | 0340980 | Pathanamthitta | LG | observation room | PHC PRAMADAM | Pharmacy Refrigerator | NON CRITICAL |
|
||
48 | Pending | Approved | View | Equipment completed 8+ year, End of life letter from OEM received | 1342816 | Kannur | Robonik | LABORATORY | PHC RAMANTHALI | Biochemistry Analyzer | CRITICAL |
|
||
49 | Pending | Approved | View | Very Old machine- Asset value Below 10K- Images attached- Verified- Recommended for RBER | 0543507 | Kottayam | Whirlpool | OP | PHC THALANAD | Pharmacy Refrigerator | NON CRITICAL |
|
||
50 | Pending | Approved | View | Document verified- RBER Approved | 0543495 | Kottayam | Diamond | OP | PHC THALANAD | B P Apparatus | NON CRITICAL |
|
||
51 | Pending | Approved | View | Verified- Recommended for RBER | 0543432 | Kottayam | Infi Neb | OP | PHC THALAYAZAHAM | Nebuliser | NON CRITICAL |
|
||
52 | Pending | Approved | View | Equipment completed 16 years.repairing cost is more than 60% of asset value, Both criteria FOR RBER Met, as per tenderclause 5.3.14.1 | 1241103 | Wayanad | Electrolux | LABORATORY | PHC THONDERNADU | Pharmacy Refrigerator | NON CRITICAL |
|
||
53 | Pending | Approved | View | Equipment completed 8+year, End of life letter provide by oem robonick as per tenderclause 5.3.14.2 | 1041250 | Malappuram | Robonik | LABORATORY | PHC VATTAMKULAM | Biochemistry Analyzer | CRITICAL |
|
||
54 | Pending | Approved | View | Equipment completed 10 year,Eol letter provided by oem attached, as per tender clause 5.3.14.2 | 1241206 | Wayanad | Robonik | LABORATORY | PHC Vellamunda | Biochemistry Analyzer | CRITICAL |
|
||
55 | Pending | Approved | View | RBER | 0843770 | Thrissur | Robonik | LABORATORY | PHC VILVATTOM | Biochemistry Analyzer | CRITICAL |
|
||
56 | Pending | Approved | View | Equipment completed 12 year,repairing cost of equipment is more than 100% both criteria for rber met | 1311996 | Kannur | Cornley | LABORATORY | RPH LAB THALAP KANNUR | Electrolyte Analyser | CRITICAL |
|
||
57 | Pending | Approved | View | Equipment completed 9 year with an asset cost of rupees 1353.Quotation not submitted since spares are not available in the market | 0920917 | Palakkad | NA | Labour Ward | THQH ALATHUR | Nebuliser | NON CRITICAL |
|
||
58 | Pending | Approved | View | VERIFIED | 0420249 | Alappuzha | Aspen | WARD 1 | THQH CHERTHALA | Infusuion Pump | CRITICAL |
|
||
59 | Pending | Approved | View | VERIFIED | 0720734 | Ernakulam | SURGIVAC | OT | THQH FORTKOCHI | Suction Apparatus | NON CRITICAL |
|
||
60 | Pending | Approved | View | VERIFIED | 0420779 | Alappuzha | SKANRAY | pediatric | THQH HARIPPAD | Pulse Oximeter | NON CRITICAL |
|
||
61 | Pending | Approved | View | VERIFIED | 0421040 | Alappuzha | BPL | ECG ROOM | THQH KAYAMKULAM | ECG Machine | CRITICAL |
|
||
62 | Pending | Approved | View | Verified- Recommended RBER | 0722365 | Ernakulam | Philips | OT | THQH KOTHAMANGALAM | OT Light | CRITICAL |
|
||
63 | Pending | Approved | View | Equipment completed 12+ year, repairing cost of equipment is more than 100%, both criteria for rber met | 0920467 | Palakkad | Allengers | OT | THQH MANNARKKAD | C-ARM | NON CRITICAL |
|
||
64 | Pending | Approved | View | Letter attached- Images and Documents Verified- Recommended for RBER | 0620715 | Idukki | Videocon | LABORATORY | THQH NEDUMKANDAM | Pharmacy Refrigerator | NON CRITICAL |
|
||
65 | Pending | Approved | View | Equipment completed 12 years,mindray dp9900 plus is an obselate model as per tender clause 5.3.14.2 Met rber criteria | 0921465 | Palakkad | Mindray | Labour Ward | THQH PATTAMBI | Ultrasound Machine | CRITICAL |
|
||
66 | Pending | Approved | View | Equipment completed 11 year 7 month,repairing cost of equipment is 91% both critera for rber met | 1120071 | Kozhikode | Mindray | LABORATORY | THQH THAMARASSERY | Haematology Analyser | CRITICAL |
|
||
67 | Pending | Approved | View | EQUIPMENT COMPLETED 7 YEAR, FULLY RUSTED, QUOTATION NOT SUBMITTED SINCE SPARES ARE NOT AVAILABLE IN THE MARKET | 1120220 | Kozhikode | Pragathi | NICU | THQH THAMARASSERY | Weighing Machine | NON CRITICAL |
|
||
68 | Pending | Approved | View | Verified- Recommended RBER | 0121261 | Thiruvananthapuram | Philips | OT | THQH VARKALA | Defibrillator | NON CRITICAL |
|
||
69 | Pending | Approved | View | Equipment completed 10+year, repairing cost of equipment is more than 100%,both criteria for RBER met | 0913284 | Palakkad | Cornley | LABORATORY | TRIBAL SPECIALITY HOSPITAL KOTTATHARA | Electrolyte Analyser | CRITICAL |
|
||
70 | Pending | Approved | View | VERIFIED | 0140728 | Thiruvananthapuram | DISHA | OP | UPHC CHAKKAI | B P Apparatus | NON CRITICAL |
|
||
71 | Pending | Approved | View | VERIFIED | 0140722 | Thiruvananthapuram | REMI | LABORATORY | UPHC CHAKKAI | Centrifuge | NON CRITICAL |
|
||
72 | Pending | Approved | View | Equipment completed 7 year 4 month repairing cost of equipment is 83% of asset value, both criteria for rber met | 1111318 | Kozhikode | Chromate | BLOOD BANK | WOMEN & CHILDREN HOSPITAL KOZHIKODE | Elisa Reader | NON CRITICAL |
|
||
73 | Pending | Approved | View | Equipment completed 7 years, repairing cost of equipment is 82% both criteria for rber met | 1331063 | Kannur | Prokan | LABORATORY | CHC MAYYIL | Haematology Analyser | CRITICAL |
|
||
74 | Pending | Approved | View | Equipment completed 5 year 3 months, OEM informed spare parts are not available model obselate as per tender clause 5.3.14.2 | 1311185 | Kannur | Biorad | BLOOD BANK | GH THALASSERY | Elisa Reader | NON CRITICAL |
|
||
75 | Pending | Approved | View | Equipment completed 4+ year quotation not submitted since spares are not available in the market. | 0845467 | Thrissur | ELKO | SUB CENTRE | PHC PERUMBILAVU | B P Apparatus | NON CRITICAL |
|
||
76 | Pending | Approved | View | Year not met- Special case- Images and Documents verified- Recommended for RBER | 0543629 | Kottayam | LIFE CARE EQUIPMENTS | SUB CENTRE | PHC VIZHIKKATHODU | B P Apparatus | NON CRITICAL |
|
||
77 | Pending | Approved | View | Equipment completed 13 years, repairing cost of equipment is 69.5% both criteria for Rber met | 1330525 | Kannur | LG | OP | CHC IRIKKUR | Pharmacy Refrigerator | NON CRITICAL |
|
||
78 | Pending | Approved | View | EQUIPMENT COMPLETED 4 YEAR WITH AN ASSET COST OF RUPEES 715.QUOTATION NOT SUBMITTED SINCE SPARES ARE NOT AVAILABLE IN THE MARKET | 1341991 | Kannur | Samso Electronics | maincenter | PHC MALAPPATTAM | Weighing Machine | NON CRITICAL |
|
||
79 | Pending | Approved | View | Equipment completed 5 year 11 month,with an asset cost of rupees 850.Quotation not submitted since spares are not available in the market | 1340813 | Kannur | healcure | OP | PHC ALAKODE (T) | Weighing Machine | NON CRITICAL |
|
||
80 | Pending | Approved | View | Equipment outlived 6+ year,Repairing cost more than 100%,Both critria Met for RBER | 1330876 | Kannur | OMRON | OP | CHC PAPINISERY | B P Apparatus | NON CRITICAL |
|
||
81 | Pending | Approved | View | Equipment completed 4 year, Repairing cost of equipment is 81% with respect to asset value, Both criteria for RBER Met | 1344755 | Kannur | ELKO | SUB CENTRE | PHC CHELORA | B P Apparatus | NON CRITICAL |
|
||
82 | Pending | Approved | View | Equipment completed 4 year 3 month , spare quote not submitted since spare are not available in the market, | 1341869 | Kannur | Samso Electronics | SUB CENTRE | PHC KADIRUR | Weighing Machine | NON CRITICAL |
|
||
83 | Pending | Approved | View | One more Month Required to met Criteria- Digital BP- Special case- | 0542831 | Kottayam | DIAMOND DELUX | SUB CENTRE | PHC PALLICKATHODU | B P Apparatus | NON CRITICAL |
|
||
84 | Pending | Approved | View | Verified- Recommended RBER | 0541913 | Kottayam | Anita Industries | OP | 24X7 PHC POONJAR | B P Apparatus | NON CRITICAL |
|
||
85 | Pending | Approved | View | Equipment completed 5 year, with an asset cost of rupees 2200, quotation not submitted since spares are not available in the market | 0920802 | Palakkad | DIAMOND | MEDICAL WARD | THQH ALATHUR | B P Apparatus | NON CRITICAL |
|
||
86 | Pending | Approved | View | Equipment completed 4 year 9 month, oem declared merilyzer celquant 3 aspire is discontinued , end of life declared model | 1130114 | Kozhikode | Merilyzer CliniQuant | LABORATORY | CHC Kodanchery | Haematology Analyser | CRITICAL |
|
||
87 | Pending | Approved | View | verified- Recommended RBER | 0611314 | Idukki | ELKO | PAIN & PALLIATIVE | DH THODUPUZHA | B P Apparatus | NON CRITICAL |
|
||
88 | Pending | Approved | View | Documents and Images verified- Recommended for RBER | 0541842 | Kottayam | SAMSO | OP | 24X7 PHC ERATTUPETTA | Weighing Machine | NON CRITICAL |
|
||
89 | Pending | Approved | View | Mercury BP- Documents verified- Recommended RBER | 0120659 | Thiruvananthapuram | ELKEOMETER | CASUALTY | THQH PARASSALA | B P Apparatus | NON CRITICAL |
|
||
90 | Pending | Approved | View | VERIFIED | 0510210 | Kottayam | CHAOTUO | OT | GH KOTTAYAM | Sterlizer | NON CRITICAL |
|
||
91 | Pending | Approved | View | Flood Damage- Documents verified- Recommended RBER | 0141212 | Thiruvananthapuram | Life care | nursing station | UPHC KANNAMMOOLA | B P Apparatus | NON CRITICAL |
|
||
92 | Pending | Approved | View | Equipment completed 5+ year with an asset cost of rupees 850, Quotation not submitted since spares are not available in the market | 0941592 | Palakkad | Unigrip | OP | PHC MUTHALAMADA | Weighing Machine | NON CRITICAL |
|
||
93 | Pending | Approved | View | VERIFIED | 0114845 | Thiruvananthapuram | Electrolux | BLOOD BANK | WOMEN & CHILDREN HOSPITAL THYCAUD | Pharmacy Refrigerator | NON CRITICAL |
|
||
94 | Pending | Approved | View | Equipment completed 3 year with an asset cost of rupees 2200, in fully damaged condition becuase of main board and display with flex cable damaged, Quotation not submitted since spares are not available in the market | 0940835 | Palakkad | Omron | OP | PHC KUMARAMPUTHUR | B P Apparatus | NON CRITICAL |
|
||
95 | Pending | Approved | View | Covid supply- No vendor details- Donated- Spare parts not available in market- Verified- Recommended RBER | 0513233 | Kottayam | nill | MEDICAL WARD | GH KANJIRAPALLY | Oxygen Concentrator | NON CRITICAL |
|
||
96 | Pending | Approved | View | Equipment completed 6+year,repairing cost of equipment is 87% both criteria for Rber met | 1011320 | Malappuram | KEMI | LABORATORY | DH NILAMBUR | Centrifuge | NON CRITICAL |
|
||
97 | Pending | Approved | View | verified | 0541924 | Kottayam | READY UNIT | OP | 24X7 PHC POONJAR | Suction Apparatus | NON CRITICAL |
|
||
98 | Pending | Approved | View | Equipment completed 5 + year, with an asset cost of rupees 2200, quotation not submitted since spares are not available in the market | 0920915 | Palakkad | Life care | LABOUR ROOM | THQH ALATHUR | B P Apparatus | NON CRITICAL |
|
||
99 | Pending | Approved | View | Equipment competleted 5+ year, repairing cost of equipment is more than 71% of asset value,both criteria for rber met | 1311919 | Kannur | enertech | urology | GH THALASSERY | UPS | NON CRITICAL |
|
||
100 | Pending | Approved | View | Year not met- Special case- Document verified- Recommended RBER | 0130153 | Thiruvananthapuram | LIFECARE | OP | CHC ANCHUTHENGU | B P Apparatus | NON CRITICAL |
|
||
101 | Pending | Approved | View | Both criteria met- Recommended RBER | 0141572 | Thiruvananthapuram | ECG Mac | ECG ROOM | PHC KEEZHATTINGAL | ECG Machine | CRITICAL |
|
||
102 | Pending | Approved | View | Eqruipment completed 3+ year with an asset cost of rupees 1875 ,Quotation not submitted since spares are not available in the market | 1010567 | Malappuram | Trucheck | oncology | DH TIRUR | Nebuliser | NON CRITICAL |
|
||
103 | Pending | Approved | View | VERIFIED | 0711498 | Ernakulam | LIFE PLUS | INJECTION ROOM | DH ALUVA | Fetal doppler | NON CRITICAL |
|
||
104 | Pending | Approved | View | VERIFIED | 0411127 | Alappuzha | Meditech | MEDICAL WARD | DH MAVELIKARA | Nebuliser | NON CRITICAL |
|
||
105 | Pending | Approved | View | Unserviceable letter attached- Year met- Recommended for RBER | 0310056 | Pathanamthitta | HYKON | LABORATORY | GH PATHANAMTHITTA | UPS | NON CRITICAL |
|
||
106 | Pending | Approved | View | VERIFIED | 0310110 | Pathanamthitta | SOCOMEC | RADIOLOGY | GH PATHANAMTHITTA | UPS | NON CRITICAL |
|
||
107 | Pending | Approved | View | Recommended for RBER.End of life by OEM. | 0512142 | Kottayam | Alan Electronic System | main ot | GH CHANGANACHERRY | Cautery | NON CRITICAL |
|
||
108 | Pending | Approved | View | Totaly rusted- Verified- Recommended RBER | 0610914 | Idukki | Electrolux | MALE WARD | DH THODUPUZHA | Pharmacy Refrigerator | NON CRITICAL |
|
||
109 | Pending | Approved | View | Mercury Spillage- Verified- Recommended RBER | 0511297 | Kottayam | ELKO METER | CASUALTY | GH PALA | B P Apparatus | NON CRITICAL |
|
||
110 | Pending | Approved | View | Verified- Recommended RBER | 0511301 | Kottayam | Dr Gene | CASUALTY | GH PALA | Weighing Machine | NON CRITICAL |
|
||
111 | Pending | Approved | View | Equipment covered 7+ year, quotation not submitted since spares are not available in the market | 0912440 | Palakkad | CITIZEN SYSTEMS | Antenatal Ward | WOMEN & CHILDREN HOSPITAL PALAKKAD | B P Apparatus | NON CRITICAL |
|
||
112 | Pending | Approved | View | Equipment completed 5+ year, repairing cost of equipment is 89% of asset value,both criteria for RBER Met. | 0912442 | Palakkad | perfecxa | Antenatal Ward | WOMEN & CHILDREN HOSPITAL PALAKKAD | Fetal doppler | NON CRITICAL |
|
||
113 | Pending | Approved | View | Verified- Recommended RBER | 0611250 | Idukki | Alpha Fiberonics | ISOLATION WARD | DH THODUPUZHA | Nebuliser | NON CRITICAL |
|
||
114 | Pending | Approved | View | Equipment completed 4 year with an asset cost of rupees 1353, quotation not submitted since spares are not available in the market | 0912486 | Palakkad | Instapro | PICU | WOMEN & CHILDREN HOSPITAL PALAKKAD | Nebuliser | NON CRITICAL |
|
||
115 | Pending | Approved | View | Equipment surpassed its shelf life of 12 years and 6 months,no stock details available at hospital,attested document by store keeper attached, as per tender clause 5.3.14.2 which stipulates manufacture not existing the equipment should evaluate for condemination we exclate these as a special case to tender invite authority | 1411353 | KASARGODE | MEMMERT GmbH | TB CENTER | DISTRICT TB CENTRE KASARAGOD | Hot Air Oven | NON CRITICAL |
|
||
116 | Pending | Approved | View | Equipment completed 6 year, oem cura is not existing as per tender clause 5.3.14.2 we recommending these as a special case and recommend equipment for condemination | 0812018 | Thrissur | Cura | DIALYSIS UNIT | DH WADAKKANCHERRY | Dialyzer Reprocessing Unit | NON CRITICAL |
|
||
117 | Pending | Approved | View | Equipment completed 6+ year, repairing cost of equipment is 96.74% of asset value,both criteria for rber met | 0912641 | Palakkad | WELCH ALLYN | CASUALTY | WOMEN & CHILDREN HOSPITAL PALAKKAD | Otoscope | NON CRITICAL |
|
||
118 | Pending | Approved | View | Verified- Recommended RBER | 0110569 | Thiruvananthapuram | digital, dr morepen | CARDIOLOGY OP | GH THIRUVANANTHAPURAM | B P Apparatus | NON CRITICAL |
|
||
119 | Pending | Approved | View | Equipment completed 8 year repairing cost of equipment is 79% of asset value, both criteria for rber met | 0811908 | Thrissur | ZEISS | LABORATORY | TB CENTER THRISSUR | Laboratory Microscope | NON CRITICAL |
|
||
120 | Pending | Approved | View | Verified- OEM Letter- Recommended RBER | 0110584 | Thiruvananthapuram | Appasamy Associates | EYE OP | GH THIRUVANANTHAPURAM | Slit Lamp | NON CRITICAL |
|
||
121 | Pending | Approved | View | VERIFIED- RECOMMENDED RBER | 0110616 | Thiruvananthapuram | Diamond | GENERAL WARD | GH THIRUVANANTHAPURAM | B P Apparatus | NON CRITICAL |
|
||
122 | Pending | Approved | View | Found crack on the boiler- Service not possible- Image attached- Verified- Recommended RBER | 0110682 | Thiruvananthapuram | Shiva | DENTAL DEPARTMENT | GH THIRUVANANTHAPURAM | Sterlizer | NON CRITICAL |
|
||
123 | Pending | Approved | View | VERIFIED | 0110845 | Thiruvananthapuram | SUPREME ENETERPRISES | OT | GH THIRUVANANTHAPURAM | Suction Apparatus | NON CRITICAL |
|
||
124 | Pending | Approved | View | One more month required to meet Criteria- Documents verified- Recommended RBER | 0140100 | Thiruvananthapuram | ELKO | immunization | UPHC MAMBAZHAKKARA | B P Apparatus | NON CRITICAL |
|
||
125 | Pending | Approved | View | Equipment completed 11+ year repairing cost of equipment is 79% of asset value, both criteria for rber met | 0811896 | Thrissur | ZEISS | LABORATORY | TB CENTER THRISSUR | Laboratory Microscope | NON CRITICAL |
|
||
126 | Pending | Approved | View | EQUIPMENT COMPLETED 10 YEARS,INFUSION PUMP ASPEN JSB1200 HAS BEEN DISCONTINUED AND OBSELATE MODEL AS PER TENDERCLAUSE5.3.14.2. RBER CRITERIA MET | 0840035 | Thrissur | SAMSO | INJECTION ROOM | UFHC Porkkalengad | Weighing Machine | NON CRITICAL |
|
||
127 | Pending | Approved | View | verified | 0722675 | Ernakulam | Beurer | CASUALTY | THQH ANGAMALI | B P Apparatus | NON CRITICAL |
|
||
128 | Pending | Approved | View | VERIFIED | 0540071 | Kottayam | ASIA POWER | LABORATORY | PHC KALLARA | UPS | NON CRITICAL |
|
||
129 | Pending | Approved | View | Manufacture cura healthcare not existing, As per tender clause 5.3.14.2 which stipulates that manufacture not existing the equipment should be evaluated for condemination, we highlighting the tender inviting authority as these as special case | 1340215 | Kannur | NIL | DIALYSIS UNIT | UDC URUVACHAL | Bicarbonate Mixer | NON CRITICAL |
|
||
130 | Pending | Approved | View | Two more months required to meet Criteria- Special case- Images and Documents verified-Recommended RBER | 0540108 | Kottayam | NIL | OP | PHC KALLARA | Sterlizer | NON CRITICAL |
|
||
131 | Pending | Approved | View | Verified- Recommended RBER | 0140841 | Thiruvananthapuram | REMI | LABORATORY | UPHC VATTIYOORKAVU | Centrifuge | NON CRITICAL |
|
||
132 | Pending | Approved | View | verified | 0340038 | Pathanamthitta | Not available | LABORATORY | PHC Chenneerkkara | UPS | NON CRITICAL |
|
||
133 | Pending | Approved | View | Documents Verified- Criteria met- Recommended RBER | 0130075 | Thiruvananthapuram | Labatronics | LABORATORY | CHC POONTHURA | Colorimeter | NON CRITICAL |
|
||
134 | Pending | Approved | View | Verified- Recommended for RBER | 0540462 | Kottayam | SAMSO | OP | PHC THIRUVARPU | Weighing Machine | NON CRITICAL |
|
||
135 | Pending | Approved | View | Verified- Recommended RBER | 0540460 | Kottayam | SAMSO | OP | PHC THIRUVARPU | Weighing Machine | NON CRITICAL |
|
||
136 | Pending | Approved | View | Equipment completed 4 year, repairing cost of equipment is more than 90%with respect to asset value, both criteria for Rber Met | 1320079 | Kannur | ELKO | SUB CENTRE | THQH PERAVOOR | B P Apparatus | NON CRITICAL |
|
||
137 | Pending | Approved | View | VERIFIED | 0730176 | Ernakulam | REMI | LABORATORY | CHC VENGOLA | Centrifuge | NON CRITICAL |
|
||
138 | Pending | Approved | View | Recommended RBER | 0111780 | Thiruvananthapuram | Ready Mist | WARD 9 | GH THIRUVANANTHAPURAM | Nebuliser | NON CRITICAL |
|
||
139 | Pending | Approved | View | RECOMMENDED RBER | 0111805 | Thiruvananthapuram | LIFECARE | WARD 7 | GH THIRUVANANTHAPURAM | Suction Apparatus | NON CRITICAL |
|
||
140 | Pending | Approved | View | VERIFIED | 0111869 | Thiruvananthapuram | MONARCH | OT | GH THIRUVANANTHAPURAM | Suction Apparatus | NON CRITICAL |
|
||
141 | Pending | Approved | View | Equipment completed 4 year 5 month, Quotation not submitted since spares are not available in the market | 1341575 | Kannur | SAMSO | SUBCENTER | PHC KANKOL ALAPADAMBA | Weighing Machine | NON CRITICAL |
|
||
142 | Pending | Approved | View | Verified- Recommended RBER | 0640755 | Idukki | NIL | DUTY ROOM OP | 24X7 PHC KANCHIYAR | B P Apparatus | NON CRITICAL |
|
||
143 | Pending | Approved | View | Equipment outlived 10 year, end of life declared model, as per tender clause 5.3.14.2 we recommend equipment for condemination | 1210668 | Wayanad | Appasamy Associates | EYE OP | DH MANANTHAVADY | OT Table | CRITICAL |
|
||
144 | Pending | Approved | View | Equipment completed 3 year and 3 month, quotation not submitted because spares are not available in the market | 1321331 | Kannur | Lifecare | WARD | THQH IRITTY | Nebuliser | NON CRITICAL |
|
||
145 | Pending | Approved | View | VERIFIED | 0420052 | Alappuzha | Philips | CASUALTY | THQH CHERTHALA | Defibrillator | NON CRITICAL |
|
||
146 | Pending | Approved | View | VERIFIED | 0420064 | Alappuzha | Ishnee | WARD 5 | THQH CHERTHALA | Nebuliser | NON CRITICAL |
|
||
147 | Pending | Approved | View | Flood Damage- Unserviceable- Verified- Recommended RBAR | 0141216 | Thiruvananthapuram | Infi | nursing station | UPHC KANNAMMOOLA | B P Apparatus | NON CRITICAL |
|
||
148 | Pending | Approved | View | Flood Damage- Verified- Recommended RBER | 0141215 | Thiruvananthapuram | Omaron | nursing station | UPHC KANNAMMOOLA | Nebuliser | NON CRITICAL |
|
||
149 | Pending | Approved | View | Flood Damage- Verified- Recommended RBER | 0141214 | Thiruvananthapuram | Safe & Touch | nursing station | UPHC KANNAMMOOLA | Nebuliser | NON CRITICAL |
|
||
150 | Pending | Approved | View | VERIFIED | 0112020 | Thiruvananthapuram | Ascon | OT | GH THIRUVANANTHAPURAM | Operating Microscope | NON CRITICAL |
|
||
151 | Pending | Approved | View | Quotation attached- Recommended RBER | 0113001 | Thiruvananthapuram | Transasia | LABORATORY | GH NEYYATTINKARA | ESR Analyser | NON CRITICAL |
|
||
152 | Pending | Approved | View | Verified Document | 0641719 | Idukki | Life care | SUB CENTRE | 24X7 PHC KARIMANNOOR | B P Apparatus | NON CRITICAL |
|
||
153 | Pending | Approved | View | VERIFIED | 0420341 | Alappuzha | BPL | ECG ROOM | THQH CHERTHALA | ECG Machine | CRITICAL |
|
||
154 | Pending | Approved | View | Total damage- verified- Recommended RBER | 0113220 | Thiruvananthapuram | Philips | C WARD | GH NEYYATTINKARA | Nebuliser | NON CRITICAL |
|
||
155 | Pending | Approved | View | Equipment aged upto 4+ year,repairing cost of equipment is 97% with respect to asset value, both criteria met for rber | 1341610 | Kannur | Medvision | main center | PHC PARIYARAM | B P Apparatus | NON CRITICAL |
|
||
156 | Pending | Approved | View | verified | 0741884 | Ernakulam | OMRON | public health | PHC CHERANELLOORE | B P Apparatus | NON CRITICAL |
|
||
157 | Pending | Approved | View | VERIFIED | 0530390 | Kottayam | Anita Industries | OP | CHC EDAYAZHAM | B P Apparatus | NON CRITICAL |
|
||
158 | Pending | Approved | View | Equipment completed 5 year, repairing cost is more than 100% both criteria for rber met | 1330286 | Kannur | Beurer | OP | CHC MATTANNUR | B P Apparatus | NON CRITICAL |
|
||
159 | Pending | Approved | View | Equipment completed 7 +year, reapiring cost of equipment more than 100% ,Both criteria for RBER Met | 1330313 | Kannur | BPE | LABORATORY | CHC MATTANNUR | UPS | NON CRITICAL |
|
||
160 | Pending | Approved | View | VERIFIED | 0511011 | Kottayam | Alps | AUDIOLOGY | GH KOTTAYAM | Audiometer | NON CRITICAL |
|
||
161 | Pending | Approved | View | Equipment outlived its shelf life of 11+ year, End of life letter provided by oem attached | 1110127 | Kozhikode | Erba | LABORATORY | RPH LAB KOZHIKODE | Fully Auto Biochemistry Analyzer | NON CRITICAL |
|
||
162 | Pending | Approved | View | Equipment completed 4 year 5 month, quotation not submitted since spares are not available in the market.asset cost of equipment is below 2200 rupees | 1344379 | Kannur | OMRON | main center | PHC KANICHAR | B P Apparatus | NON CRITICAL |
|
||
163 | Pending | Approved | View | Equipment completed 8+ year, repairing cost of equipment 93%, both criteria for rber met | 1142195 | Kozhikode | ASIA POWER | LABORATORY | PHC KUNNAMANGALAM | UPS | NON CRITICAL |
|
||
164 | Pending | Approved | View | VERIFIED | 0421057 | Alappuzha | Truchek | MEDICAL WARD | THQH KAYAMKULAM | B P Apparatus | NON CRITICAL |
|
||
165 | Pending | Approved | View | Equipment aged upto 8 year and 7 month, End of life for Equipmnt declared by Orginal Equipment Manufacture attached for reference | 1130962 | Kozhikode | PRIETEST-ROBONIK | LABORATORY | CHC Ulliyeri | Biochemistry Analyzer | CRITICAL |
|
||
166 | Pending | Approved | View | VERIFIED | 0542458 | Kottayam | Robonik | LABORATORY | PHC KURUPPANTHARA | Biochemistry Analyzer | CRITICAL |
|
||
167 | Pending | Approved | View | Equipment completed 9.4 years, repairing cost of equipment is 69% both criteria for rber met | 0920181 | Palakkad | REMI | LABORATORY | THQH CHITTUR | Centrifuge | NON CRITICAL |
|
||
168 | Pending | Approved | View | Equipment completed 10 years 3 months,repairing cost of equipment is more than 100% both criteria for RBER Met. | 1345245 | Kannur | Emerson | LABORATORY | PHC CHITTARIPARAMBA | UPS | NON CRITICAL |
|
||
169 | Pending | Approved | View | Verified- Recommended RBER | 0541927 | Kottayam | nill | OP | 24X7 PHC POONJAR | Spot Light | NON CRITICAL |
|
||
170 | Pending | Approved | View | VERIFIED | 0421812 | Alappuzha | Godrej | STORE | THQH KAYAMKULAM | Pharmacy Refrigerator | NON CRITICAL |
|
||
171 | Pending | Approved | View | Verified Document- Year Met- Asset value below 10K- Recommended for RBER | 0642185 | Idukki | APC | EYE OP | PHC KONNATHADY | UPS | NON CRITICAL |
|
||
172 | Pending | Approved | View | Verified- Recommended RBER | 0114937 | Thiruvananthapuram | Videocon | STORE | WOMEN & CHILDREN HOSPITAL THYCAUD | Pharmacy Refrigerator | NON CRITICAL |
|
||
173 | Pending | Approved | View | VERIFIED | 0732588 | Ernakulam | Rotek | LABORATORY | CHC VENGOOR | Centrifuge | NON CRITICAL |
|
||
174 | Pending | Approved | View | Verified- Recommended RBER | 0744082 | Ernakulam | LOCAL | OP | UPHC MANGATTUMUKKU | B P Apparatus | NON CRITICAL |
|
||
175 | Pending | Approved | View | Verified- Recommended RBER | 0110881 | Thiruvananthapuram | Eppendrof | IRL | State TB Centre | Centrifuge | NON CRITICAL |
|
||
176 | Pending | Approved | View | Equipment aged upto 4 year 3 month,repairing cost of equipment is 91% of asset cost of the equipment, hence both criteria for RBER Met. | 1141793 | Kozhikode | APC | LABORATORY | PHC AYANCHERI | UPS | NON CRITICAL |
|
||
177 | Pending | Approved | View | Equipment completed 4+ year repairing cost more than 100% Both criteria for RBER Met | 1344070 | Kannur | SAMSO | SUB CENTRE | PHC ANGADIKKADAVU | Weighing Machine | NON CRITICAL |
|
||
178 | Pending | Approved | View | Equipment completed 5 years, with an asset cost of rupees 1600. Quotation not submitted because since spares are not available in the market | 0930377 | Palakkad | SAMSO | INJECTION ROOM | CHC VADAKKANCHERY | Weighing Machine | NON CRITICAL |
|
||
179 | Pending | Approved | View | Equipment completed 5 year 6 month with an asset cost of rupees 1800.Quotation not submitted because spares are not available in the market | 1344703 | Kannur | Infi | OP | UPHC MUTHATHI | B P Apparatus | NON CRITICAL |
|
||
180 | Pending | Approved | View | Equipment completed 8 year 9 months, quotation not submitted since spares are not available in the market | 1344735 | Kannur | OMRON | SUB CENTRE | UPHC MUTHATHI | B P Apparatus | NON CRITICAL |
|
||
181 | Pending | Approved | View | verified | 0743019 | Ernakulam | nill | observation room | PHC AVOLY | B P Apparatus | NON CRITICAL |
|
||
182 | Pending | Approved | View | Equipment completed 3 year 3 month, quotation not submitted since spares are not available in the market | 1331159 | Kannur | SKNOL | OP | CHC PINARAYI | Weighing Machine | NON CRITICAL |
|
||
183 | Pending | Approved | View | Equipment completed 11+ years,confident, C70D, Dental xray end of life declared model as per tender clause 5.3.14.2 recommend equipment for condmeination | 1020632 | Malappuram | CONFIDENT | X RAY | THQH WANDOOR | X Ray Machine | CRITICAL |
|
||
184 | Pending | Approved | View | Equipment completed 8 year, robonick prietets touch semi auto analyser is an end of life declared model, Eol letter provided by oem robonick attached, | 0930368 | Palakkad | Robonik | LABORATORY | CHC VADAKKANCHERY | Biochemistry Analyzer | CRITICAL |
|
||
185 | Pending | Approved | View | Equipment completed 7 years, quotation not submitted since spares are not available in the market | 0930360 | Palakkad | Omaron | OP | CHC VADAKKANCHERY | Nebuliser | NON CRITICAL |
|
||
186 | Pending | Approved | View | Equipment outlived 9+year, End of life letter from orginal equipment manufacture attached | 1020641 | Malappuram | Robonik | LABORATORY | THQH WANDOOR | Biochemistry Analyzer | CRITICAL |
|
||
187 | Pending | Approved | View | verified | 0744424 | Ernakulam | SAMSO | public health | PHC KADUNGALLOR | Weighing Machine | NON CRITICAL |
|
||
188 | Pending | Approved | View | Oem Medequip had discontinued support for the oxymed-owgels 3/5lpm oxygen concentrator, obselate letter attached from oem as per tender clause 5.3.14.2 | 1330121 | Kannur | Life care | DIALYSIS UNIT | CHC AZHIKODE | Oxygen Concentrator | NON CRITICAL |
|
||
189 | Pending | Approved | View | Equipment completed 5 year with an asset cost of rupees 1353. Quotation not submitted since spares are not available in the market | 0912737 | Palakkad | NA | CH Ward | WOMEN & CHILDREN HOSPITAL PALAKKAD | Nebuliser | NON CRITICAL |
|
||
190 | Pending | Approved | View | Verified- Recommended RBER | 0744941 | Ernakulam | Diamond Deluxe | OP | 24X7 PHC KALLOORKAD | B P Apparatus | NON CRITICAL |
|
||
191 | Pending | Approved | View | Verified- Recommended RBER | 0745118 | Ernakulam | APC by schneider electric | LABORATORY | 24X7 PHC KAKKANAD | UPS | NON CRITICAL |
|
||
192 | Pending | Approved | View | Document Verified- RBER Approved | 0543498 | Kottayam | LIFE LINE | OP | PHC THALANAD | B P Apparatus | NON CRITICAL |
|
||
193 | Pending | Approved | View | EQUIPMENT COMPLETED 5 YEAR 6 MONTH WITH AN ASSET COST OF RUPEES 1800,QUOTATION NOT SUBMITTED SINCE SPARES ARE NOT AVAILABLE IN THE MARKET | 0941059 | Palakkad | SAMSO | OP | PHC OZHALAPATHY | Weighing Machine | NON CRITICAL |
|
||
194 | Pending | Approved | View | Equipment completed 4 year, quotation not submitted since spares are not available in the market | 1041898 | Malappuram | SAMSO | nurse station | PHC Marakkara | Weighing Machine | NON CRITICAL |
|
||
195 | Pending | Approved | View | Equipment completed 2+ year quotation not submitted since spares are not available in the market | 1031123 | Malappuram | ELKOMETER | SUBCENTER | CHC PURATHUR | B P Apparatus | NON CRITICAL |
|
||
196 | Pending | Approved | View | Equipment completed 8 year, repairing cost of equipment is 79.4% of asset value, both criteria met | 1220585 | Wayanad | Transasia | LABORATORY | THQH VYTHIRI | ESR Analyser | NON CRITICAL |
|
||
197 | Pending | Approved | View | Equipment outlived 4 year and 1 month, below 10,000 asset value equipment, quotation not submitted, since spares are not available in the market | 1143172 | Kozhikode | SAMSO | SUB CENTRE | PHC KUNNAMANGALAM | Weighing Machine | NON CRITICAL |
|
||
198 | Pending | Approved | View | Equipment completed 6+year, quotation not submitted since spares are not available in the market | 1143182 | Kozhikode | Lifecare | SUB CENTRE | PHC KUNNAMANGALAM | Nebuliser | NON CRITICAL |
|
||
199 | Pending | Approved | View | Images and Documents verified- Recommended for RBER | 0542964 | Kottayam | nill | OP | PHC PARATHODU | B P Apparatus | NON CRITICAL |
|
||
200 | Pending | Approved | View | The digital bp apparatus installed on 10/10/2019 with an asset cost of rupees 2200,completed 4 year and 4 months of working.quotation not submitted since spares are not available in th market | 1142588 | Kozhikode | Life care | SUB CENTRE | PHC Puthiyappa | B P Apparatus | NON CRITICAL |
|
||
201 | Pending | Approved | View | more than 100% quotation, Documents verified- Recommended for RBER | 0221023 | Kollam | Biotech | PHYSIOTHERAPY | THQH KADAKKAL | Traction Unit | NON CRITICAL |
|
||
202 | Pending | Approved | View | Verified- Recommended RBER | 0130523 | Thiruvananthapuram | Trucheck | observation room | CHC KANYAKULANGARA | Nebuliser | NON CRITICAL |
|
||
203 | Pending | Approved | View | Recommended RBER | 0130524 | Thiruvananthapuram | Ultra Neb | observation room | CHC KANYAKULANGARA | Nebuliser | NON CRITICAL |
|
||
204 | Pending | Approved | View | Equipment completed 5+ year with an asset cost of rupees 2200,Quotation not submitted since spares are not available in the market | 0920921 | Palakkad | OMRON | Labour Ward | THQH ALATHUR | B P Apparatus | NON CRITICAL |
|
||
205 | Pending | Approved | View | Equipment completed 4 years, quotation not submitted since spares are not available in the market | 0920919 | Palakkad | NIL | Labour Ward | THQH ALATHUR | B P Apparatus | NON CRITICAL |
|
||
206 | Pending | Approved | View | Totally Rusted Dilapidated condition- Hospital letter attached- Recommended for RBER | 0340836 | Pathanamthitta | Not available | PAIN & PALLIATIVE | PHC PANDALAM THEKKEKARA | Sterlizer | NON CRITICAL |
|
||
207 | Pending | Approved | View | Machine totally Rusted and Dilapidated condition- Images verified- Hospital Letter attached- Recommended for RBer | 0340867 | Pathanamthitta | Not available | LABORATORY | PHC PANDALAM THEKKEKARA | Laboratory Microscope | NON CRITICAL |
|
||
208 | Pending | Approved | View | Verified Documents- Customer Letter attached- RBER Approved | 0340868 | Pathanamthitta | LABOVISION | LABORATORY | PHC PANDALAM THEKKEKARA | Laboratory Microscope | NON CRITICAL |
|
||
209 | Pending | Approved | View | Verified- Recommended RBER | 0121877 | Thiruvananthapuram | Cura | ECG ROOM | TH Malayinkeezh | ECG Machine | CRITICAL |
|
||
210 | Pending | Approved | View | VERIFIED | 0121883 | Thiruvananthapuram | Dr. Odin | STORE | TH Malayinkeezh | B P Apparatus | NON CRITICAL |
|
||
211 | Pending | Approved | View | Verified- Recommended RBER | 0121922 | Thiruvananthapuram | Philips | WARD | TH Malayinkeezh | Nebuliser | NON CRITICAL |
|
||
212 | Pending | Approved | View | Images and Documents verified- Recommended for RBER | 0544195 | Kottayam | SAMSO | OP | PHC MEENADOM | Weighing Machine | NON CRITICAL |
|
||
213 | Pending | Approved | View | Equipment completed 3 + year with an asset cost of rupees 1353,Quotation not submitted since spares are not available in the market | 0941613 | Palakkad | NA | OP | 24X7 PHC KUNISSERY | Nebuliser | NON CRITICAL |
|
||
214 | Pending | Approved | View | EQUIPMENT COMPLETED 4 YEAR, REPAIRING COST IS MORE THAN 100% BOTH CRITERIA FOR RBER MET, RECOMMEND FOR CONDEMINATION | 1241274 | Wayanad | SAMSO | OP | PHC Vellamunda | Weighing Machine | NON CRITICAL |
|
||
215 | Pending | Approved | View | VERIFIED | 0443545 | Alappuzha | SAMSO | LABORATORY | 24X7 PHC ARATTUPUZHA | Weighing Machine | NON CRITICAL |
|
||
216 | Pending | Approved | View | VERIFIED | 0443718 | Alappuzha | Owgels Oxy-Med | OP | PHC CHERUTHANA | Nebuliser | NON CRITICAL |
|
||
217 | Pending | Approved | View | VERIFIED | 0112610 | Thiruvananthapuram | Smiths Medicals | OT | DH NEDUMANGAD | PATIENT WARMER | NON CRITICAL |
|
||
218 | Pending | Approved | View | Equipment completed 4 year, repairing cost of equipment is more than 100% both criteria for rber met, as per tender clause 5.3.14.1 | 1240532 | Wayanad | SAMSO | observation room | PHC CHETHALAYAM | Weighing Machine | NON CRITICAL |
|
||
219 | Pending | Approved | View | rber | 1241712 | Wayanad | Carer | Palliative | 24X7 PHC CHEERAL | Oxygen Concentrator | NON CRITICAL |
|
||
220 | Pending | Approved | View | Equipment completed 3 years,repairing cost of equipment is 72% of asset value, both criteria for rber met | 0942870 | Palakkad | NIL | pharmacy | PHC PUDUSSERY | B P Apparatus | NON CRITICAL |
|
||
221 | Pending | Approved | View | Equipment completed 5+ year, repairing cost of equipment is 74.9% of asset value, both criteria for rber met | 1042665 | Malappuram | Meril Diagnostics | LABORATORY | PHC Kodur | Biochemistry Analyzer | CRITICAL |
|
||
222 | Pending | Approved | View | Verified- Recommender RBER | 0144289 | Thiruvananthapuram | NA | LABORATORY | PHC MARANALLOOR | Centrifuge | NON CRITICAL |
|
||
223 | Pending | Approved | View | VERIFIED | 0144415 | Thiruvananthapuram | ELKO | nurse station | PHC PARANIYAM | B P Apparatus | NON CRITICAL |
|
||
224 | Pending | Approved | View | VERIFIED | 0144538 | Thiruvananthapuram | ELKOMETER | nurse station | PHC PARANIYAM | B P Apparatus | NON CRITICAL |
|
||
225 | Pending | Approved | View | The steriliser installed on 30/06/2000, with an asset cost of rupees 3800.is in critical condition, extensive rusting and leaking compromise its structural integrity,posing risk to safety and reliability during operation | 1031688 | Malappuram | nill | OP | 24X7 PHC PERUVALLUR | Sterlizer | NON CRITICAL |
|
||
226 | Pending | Approved | View | VERIFIED | 0732826 | Ernakulam | Life care | EYE OP | CHC MALAYIDAMTHURUTH | B P Apparatus | NON CRITICAL |
|
||
227 | Pending | Approved | View | Verified- Recommended RBER | 0144529 | Thiruvananthapuram | MISAKI | nursing station | PHC KANJIRAMKULAM | Weighing Machine | NON CRITICAL |
|
||
228 | Pending | Approved | View | Verified- Recommended RBER | 0144530 | Thiruvananthapuram | ELKO | OP | PHC KANJIRAMKULAM | B P Apparatus | NON CRITICAL |
|
||
229 | Pending | Approved | View | VERIFIED | 0543918 | Kottayam | INFI CHECK | OP | 24X7 PHC PARAMPUZHA | B P Apparatus | NON CRITICAL |
|
||
230 | Pending | Approved | View | Verified- Recommended RBER | 0520392 | Kottayam | Godrej | OP | THQH KURAVILANGAD | Pharmacy Refrigerator | NON CRITICAL |
|
||
231 | Pending | Approved | View | Verified- Recommended RBER | 0145863 | Thiruvananthapuram | SAMSO | FWC | 24X7 PHC THIRUVALLOM | Weighing Machine | NON CRITICAL |
|
||
232 | Pending | Approved | View | Equipment completed 5 year, quotation not submitted since spares are not available in the market | 0944501 | Palakkad | perfecxa | SUB CENTRE | PHC PUDUSSERY | B P Apparatus | NON CRITICAL |
|
||
233 | Pending | Approved | View | verified | 0747097 | Ernakulam | Easy care | SUB CENTRE | UPHC KADAVANTHRA | B P Apparatus | NON CRITICAL |
|
||
234 | Pending | Approved | View | Equipment completed 4+year, quotation not submitted since spares are not available in the market | 1011192 | Malappuram | ROSSMAX | post natal ward | WOMEN & CHILDREN HOSPITAL PONNANI | Weighing Machine | NON CRITICAL |
|
||
235 | Pending | Approved | View | Documents verified- Recommended RBER | 0115858 | Thiruvananthapuram | DIAMOND | WARD 8 | GH THIRUVANANTHAPURAM | B P Apparatus | NON CRITICAL |
|
||
236 | Pending | Approved | View | verified | 0743180 | Ernakulam | OMRON | public health | 24X7 PHC CHOWARA | B P Apparatus | NON CRITICAL |
|
||
237 | Pending | Approved | View | Verified- Recommended RBER | 0114684 | Thiruvananthapuram | ELKO | WARD2 | GH THIRUVANANTHAPURAM | B P Apparatus | NON CRITICAL |
|
||
238 | Pending | Approved | View | VERIFIED | 0110931 | Thiruvananthapuram | ELKO | COVID ICU | GH THIRUVANANTHAPURAM | B P Apparatus | NON CRITICAL |
|
||
239 | Pending | Approved | View | VERIFIED | 0310189 | Pathanamthitta | HEMOPHARM | BLOOD BANK | GH PATHANMTHITTA | BLOOD BAG TUBE SEALER | NON CRITICAL |
|
||
240 | Pending | Approved | View | VERIFIED | 0450354 | Alappuzha | Handy Neb | CH Ward | DH MAVELIKARA | Nebuliser | NON CRITICAL |
|
||
241 | Pending | Approved | View | VERIFIED | 0450353 | Alappuzha | Handy Neb | CH Ward | DH MAVELIKARA | Nebuliser | NON CRITICAL |
|
||
242 | Pending | Approved | View | Equipment completed 7 year 9 month, repairing cost of equipment is more than 100% , both criteria met for rber | 1350002 | Kannur | HYKON | ot 3 | DH KANNUR | UPS | NON CRITICAL |
|
||
243 | Pending | Approved | View | Donated item- No vendor available- Documents verified- Recommended RBER | 0521303 | Kottayam | NA | WARD4 | THQH VAIKOM | Oxygen Concentrator | NON CRITICAL |
|
||
244 | Pending | Approved | View | VERIFIED | 0521282 | Kottayam | SAMSO | WARD 4 | THQH VAIKOM | Weighing Machine | NON CRITICAL |
|
||
245 | Pending | Approved | View | VERIFIED | 0521275 | Kottayam | AEROSOL | WARD 1 | THQH VAIKOM | Nebuliser | NON CRITICAL |
|
||
246 | Pending | Approved | View | Verified- Recommended RBER | 0121473 | Thiruvananthapuram | Sreegokulam | OT | THQH CHIRAYINKEEZH | Anesthasia Machine | CRITICAL |
|
||
247 | Pending | Approved | View | VERIFIED | 0144531 | Thiruvananthapuram | LG | pharmacy | PHC KANJIRAMKULAM | Pharmacy Refrigerator | NON CRITICAL |
|
||
248 | Pending | Approved | View | VERIFIED | 0150011 | Thiruvananthapuram | ECG Mac | ECG ROOM | PHC PARANIYAM | ECG Machine 3 Channel | CRITICAL |
|
||
249 | Pending | Approved | View | Verified- Recommended for RBER | 0546279 | Kottayam | IBALL | LABORATORY | PHC MUTHOLY | UPS OF Hematology analyzer | NON CRITICAL |
|
||
250 | Pending | Approved | View | Equipment completed 3+ year, quotation not submitted since spares are not available in the market | 1150841 | Kozhikode | SAMSO | CASUALTY | THQH KOYILANDY | Digital Weighing Machine | NON CRITICAL |
|
||
251 | Pending | Approved | View | VERIFIED | 0511991 | Kottayam | LIFECARE | WARD | K.R. NARAYANAN MEMORIAL SUPER SPECIALTY HOSPITAL UZHAVOOR | Nebulizer | NON CRITICAL |
|
||
252 | Pending | Approved | View | VERIFIED | 0546439 | Kottayam | Infinity Mediquip India | WARD | 24X7 PHC MARANGATTUPALLY | Nebuliser | NON CRITICAL |
|
||
253 | Pending | Approved | View | VERIFIED | 0543737 | Kottayam | Infi | OP | PHC VELIYANNOOR | Digital BP Apparatus | NON CRITICAL |
|